Stryker
Theatre, Surgical and Examination Lights
SIDNE Visum LED and StrykeCam 2 In-Light Camera Package User Guide Rev C Feb 2016
User Guide
396 Pages
Preview
Page 1
SIDNE® Stryker Visum™ LED Surgical Light & StrykeCam™ 2 In-Light Camera Package 0240020832
User Guide
Contents Warnings and Cautions... 7 Product Description... 9 Intended Use...9 Package Contents...9
Setup and Interconnection... 10 Quick Startup Guide...10 Installing Driver Upgrade...10 Connecting the Visum™ LED Light and Camera ® to SIDNE . ...12
Operation. ... 13 Operating Room Lights...15 StrykeCam™ 2 In-Light Camera...19 Perilite...25
Troubleshooting ... 27
Warnings and Cautions Please read this manual and follow its instructions carefully. The words warning, caution, and note carry special meanings and should be carefully reviewed: Warning
Indicates risks to the safety of the patient or user. Failure to follow warnings may result in injury to the patient or user.
Caution
Indicates risks to the equipment. Failure to follow cautions may result in product damage.
Note
Provides special information to clarify instructions or present additional useful information. An exclamation mark within a triangle is intended to alert the user to the presence of important operating and maintenance instructions in the manual. A lightning bolt within a triangle is intended to warn of the presence of hazardous voltage. Refer all service to authorized personnel.
Warning
1. 2.
3. 4.
To avoid potential serious injury to the user and the patient and/or damage to this device, please note the following warnings:
Read this manual thoroughly and be familiar with its contents prior to using this equipment. Use of the Visum LED Surgical Light and Camera is restricted to a qualified physician or technician familiar with the device and the equipment to which it connects. Misusing the Visum™ LED Surgical Light and Camera may cause injury to the patient and/or damage to system components. Unpack the package carefully and check if any damage occurred during shipment. If damage is detected, DO NOT use the equipment. Be aware of the position of the table at all times in relation to the patient and the surroundings.
7
Warning
1. 2.
8
To avoid potential serious injury to the user and the patient and/or damage to this device, please note the following electrical warnings:
Always power off the Visum™ LED Surgical Light and Camera when connecting or disconnecting accessories. Before and after each use, inspect the cord insulation for cracks, nicks, cuts, dents, or depressions. Any irregularity will decrease the effectiveness of the insulation and may result in shocks to the patient or operating room personnel.
Product Description The SIDNE® Stryker Visum™ LED Surgical Light and StrykeCam™ 2 In-Light Camera package (Visum LED Surgical Light and Camera package) is a software interface between the SIDNE device control system and the Visum™ LED Surgical Light and Camera that provides voice-activated or touchscreen control of the Visum LED Light and Camera. Note: To purchase the SIDNE-compatible version of the Visum LED Surgical Light and Camera package, contact your local Stryker representative.
Intended Use The Visum LED Light and Camera package is intended to allow SIDNE users to control the Visum LED Light and Camera through voice and touchscreen commands. It is intended for use in any surgery requiring overhead lights and camera. It can be used in any operating room setting where both the Stryker Visum LED Light and Camera and the SIDNE device control system are installed.
Package Contents The Visum LED Light and Camera package includes the following components: 1. 50-foot USB Cable (SIDNE interface cable) 2. SIDNE Visum™ LED Surgical Light and Camera Driver 1.0 USB dongle
9
Setup and Interconnection Quick Startup Guide This section outlines the startup instructions for experienced users. Refer to the relevant section(s) of this manual for further details. 1. Power on the SIDNE® console. 2. Install the Visum™ LED Surgical Light and Camera Driver 1.0 upgrade. SIDNE will power off. 3. Power on SIDNE. Wait for SIDNE to say, “SIDNE ready” before connecting the Visum LED Light and Camera. • Connect the device adapter cable to the Visum LED Light and Camera. • Connect the SIDNE interface cable to SIDNE. 4. Power on the Visum LED Light and Camera. SIDNE should say, “O.R. Lights connected.”
Installing Driver Upgrade The SIDNE Visum™ LED Surgical Light and Camera Driver Upgrade 1.0 allows the SIDNE console to communicate with the Visum LED Light and Camera. Caution
Read the Operational Notes at the end of this procedure before using the SIDNE console to control the Visum LED Light and Camera.
Caution
Do not attempt to install any other software driver until the installation of the Visum™ LED Surgical Light and Camera Driver is complete and the SIDNE console has restarted.
1. 2.
10
Power on the SIDNE console and wait until SIDNE says, “SIDNE ready.” Insert the dongle into the device-driver dongle port on the front SIDNE panel (see below).
3.
Follow the voice instructions given by SIDNE. To avoid equipment damage, please do not remove the dongle card before SIDNE® says, “Please remove card.” After installation is complete, SIDNE® will power off.
When SIDNE says this...
Do this...
“Software card detected. Please stand by.”
Wait for the next instruction. Do not remove the dongle until SIDNE says, “Please remove card.”
“Software installation started. Please wait.”
Wait for the next instruction.
“Programming started.” “10% complete, 20% complete...”
Wait for the next instruction.
“Programming failed.”
SIDNE will announce this message only if the programming failed. If this occurs, follow the steps below: 1. Remove the dongle, restart SIDNE, and insert the dongle. 2. If the same problem occurs again, contact Stryker Technical Support.
“Programming complete. Please wait.”
Wait for the next instruction.
“Software installation complete.”
Wait for the next instruction.
“Please remove card.”
Remove the dongle.
4.
Power SIDNE on. When SIDNE says, “SIDNE ready,” the driver is installed and ready for use.
11
Connecting the Visum™ LED Light and Camera to SIDNE® SIDNE can control up to four Visum LED Surgical Lights. To do this, a second power control box is necessary. Use the Expansion port on the first power supply box to control the additional two lights. To enable SIDNE to control four surgical lights, do the following: 1. Power on the lights and camera. 2. Power on SIDNE. Wait for SIDNE to say, “SIDNE ready.” 3. Connect the USB A – to – B (male) cable from an available SIDNE device port to the SIDNE port on the first power supply box.
4. 5. 6.
12
Insert a second Stryker USB A – to – B cable into the expansion port. Connect the 50-foot cable (leading from the expansion port) to the SIDNE® port on the second power supply box (see diagram). SIDNE should say, “O.R. Lights connected” within 20 seconds after the Visum LED Light and Camera and SIDNE have been connected. If this does not occur, refer to the Troubleshooting section of this manual for instructions.
Operation Status Bar The status bar, shown below, appears at the bottom of the monitor screen.
1 1. 2. 3.
2
3
Indicates the brightness level of the last light accessed. Indicates the focus setting of the in-light camera. Indicates the the shutter setting of the in-light camera.
13
O.R. Lights Touchscreen
14
Operating Room Lights Touchscreen Commands The following table describes the tablet buttons and level meters and their function. Button Group
Buttons/Levels
Description
Light–specific controls
Power
Powers on/off a particular light. Green indicates the light is on. Blue indicates the light is off.
Brightness level
Increases or decreases the light intensity. The range is from 1 through 8. If this button has a gray background, the light is at the maximum intensity.
High
Sets a light to the maximum intensity. If this button has a gray background, the light is at the maximum intensity.
Low
Sets a light to the minimum intensity. If this button has a gray background, the light is at the minimum intensity.
+
Increases the intensity of a light by one unit. If this button has a gray background, the light is at the maximum intensity.
-
Decreases the intensity of a light by one unit. If this button has a gray background, the light is at the minimum intensity.
On
Powers on all the lights. If this button has a gray background, at least one light is on.
Off
Powers off all the lights. If this button has a gray background, all lights are off.
High
Sets all lights to the maximum intensity.
Low
Sets all lights to the minimum intensity.
Brighter
Increases the intensity of all lights by one unit. If this button has a gray background, the light is at the maximum intensity.
Darker
Decreases the intensity of all lights by one unit. If this button has a gray background, the light is at the minimum intensity or is off.
All light controls
15
LED Lights Menu
16
Voice Commands The following table shows the SIDNE® voice commands for controlling the Visum™ LED Surgical Lights. Top Level
Device Selection
SIDNE
O. R. Lights
Function Selection Light 1, Light 2, Light 3, or Light 4
All lights
Description
Activate
Powers on the light.
Brighter
Increase the light intensity by one unit.
Darker
Decreases the light intensity by one unit.
High
Sets the light intensity to maximum.
Low
Sets the light intensity to minimum.
Off
Powers off the light.
Activate
Powers on all the connected lights.
Brighter
Increases the light intensity of all connected lights by one unit.
Darker
Decreases the light intensity of all connected lights by one unit.
High
Sets the light intensity of all the connected lights to maximum.
Low
Sets the light intensity of all the connected lights to minimum.
Off
Powers off all the connected lights.
17
SIDNE® Announcements Voice Message
Playback Condition
Connected
Visum™ LED Surgical Lights are powered on and connected to SIDNE.
Disconnected
Visum LED Surgical Lights are disconnected or powered off.
Light error
An error has been reported on one or more of the connected Visum LED Surgical Lights. Servicing may be necessary.
O. R. lights …
Added as a prefix added to a feedback message when not in the O. R. Lights voice command menu.
Operating Notes 1. 2.
18
Use the activate or the brightness commands to turn on any light when the darker and off buttons are gray. A status bar will be displayed on the monitor when using voice control. The far left section shows the brightness level of the last individual light accessed using the voice menu or, by default, the lowest numbered light head connected. Say “Light” with the number of the desired light in the “O.R. Lights” menu and the status bar will update to show the brightness of that light. For example, say “Light 2” to see the brightness lights.
StrykeCam™ 2 In-Light Camera Camera Touchscreen Menu
19
Touchscreen Commands The following table describes the tablet commands and their functions. Button Group
Buttons/Levels
Description
Power
Power
Powers on/off a particular light. Green indicates the light is on. Blue indicates the light is off.
Shutter
Automatic
Sets the camera to automatic shutter control. Orange indicates the camera is set to automatic shutter control.
Manual
Sets the camera to manual shutter control. Orange indicates the camera is set to manual shutter control.
Brighter
Increases the shutter speed by one unit. Gray indicates the camera is at maximum shutter speed. Using this button will set the camera to manual shutter mode.
Darker
Decreases the shutter speed by one unit. Gray indicates the camera is at minimum iris. Using this button will set the camera to manual iris mode.
Automatic
Sets the camera to automatic focus control. Orange indicates the camera is set to automatic focus control.
Manual
Sets the camera to manual focus control. Orange indicates the camera is set to manual focus control.
Focus out
Begins to focus out of the current view. Releasing the button will stop the focus out. A gray background indicates the camera is at the maximum focus. Using this button will set the camera to manual focus mode.
Focus in
Begins to focus into the current view. Releasing the button will stop the focus in. A gray background indicates the camera is at the minimum focus. Using this button will set the camera to manual focus mode.
Focus
20
Zoom controls Zoom out
Begins to zoom out to a wide-angle view. Releasing the button will stop the zoom. A gray background indicates the camera is at the minimum zoom.
Zoom in
Begins to zoom in to a telephoto view. Releasing the button will stop the zoom. A gray background indicates the camera is at the maximum zoom.
StrykeCam 2 In-Light Camera Voice Menu
21
Voice Commands The following table shows the SIDNE® voice commands for controlling the StrykeCam 2 In-Light Camera. Top Level
Device
SIDNE O.R. lights
Function Selection
Description
Overhead Activate camera Zoom out
Powers on the camera.
Zoom in
Continuously zooms in to a telephoto view.
Stop
Stops the zoom command.
Continuously zooms out to a wide-angle view.
Shutter Automatic Sets Iris/shutter to automatically control brightness.
22
Manual
Sets Iris/shutter to manually control brightness.
Brighter
Continuously increases the light level until the stop command or other camera command is given.
Darker
Continuously decreases the light level until the stop command or other camera command is given.
Stop
Stops the shutter adjustment.
Focus
Off
Automatic Sets the camera to automatic focus mode. Manual
Sets the camera to manual focus mode.
Out
Continuously focuses the camera out until the stop command or other camera command is given.
In
Continuously focuses the camera in until the stop command or other command is given.
Stop
Stops the camera from focusing. Powers off the camera.
Note: Continuously means until “Stop” is said or pressed by the user, or within 5 seconds, whichever occurs first.
23
Operating Notes 1.
2. 3.
4.
24
While the camera is executing any of the following voice commands initiated by the SIDNE® interface-Zoom in, Zoom out, Focus in, Focus out, Shutter brighter, Shutter darker-the command will continue executing until • the stop command or another camera command has been issued • it has reached its maximum or minimum limit • the command times out. Issuing a camera command other than stop will stop the current execution and start the execution of the new command issued. Please note that there is no indication on the SIDNE® Interface (Voice and Tablet) regarding the maximum and minimum limits for the following commands: • Brightness/Darkness • Focus In/Out • Zoom In/Out The only indication when these quantities have reached their limits will be when the appropriate buttons/commands are grayed out. A status bar will be displayed on the monitor when using voice control. The center section shows the status of the camera focus and camera shutter modes, either automatic, manual, or off.