Synthes Small Battery Drive User’s Manual Rev G April 2010
User’s Manual
40 Pages
Preview
Page 1
Small Battery Drive. For orthopaedic trauma applications.
User’s Manual
Table of Contents
Introduction
Operating Instructions
Product Information
Small Battery Drive
2
12 V Battery
3
14.4 V Battery
4
Universal Battery Charger
5
Universal Battery Charger II
6
Attachments for Small Battery Drive
7
Universal Battery Charger Universal Battery Charger II
10
Battery Pack
11
Drive Unit
12
Attachments
13
Attachments for use with 12 V and 14.4 V Battery
14
Attachments for use with 14.4 V Battery
20
Light Adaptor for Small Battery Drive
24
Graphic Case
26
Sterilization
28
Troubleshooting
29
Set Lists
31
Cutting Tools
34
Synthes
Small Battery Drive (532.010)
– Variable speed: 0 – 3,840 rpm (max. speed varies by attachment) – Maximum torque: 0 – 6.0 Nm (max. torque varies by attachment) – Weight, including 12 V battery: 810 grams, including 14.4 V battery: 916 grams – Cannulation: 3.2 mm Attachment release button Allows easy exchange of attachments
Attachment coupling Accepts a wide variety of attachments Instant-reverse or oscillation trigger Reverses rotation to counterclockwise direction or activates oscillation mode Variable-speed trigger Controls rotational speed Release buttons for battery casing Allow easy intraoperative battery exchange
Battery casing Holds battery in sterile environment
This device complies with the following standards: UL 2601-1 EN 60601 (IEC 601-1) EN 60601-1-1 (VDE 0750 part 1 and 2) EN 55011 (VDE 0875 part 11) Limiting values of class B, group 1
2
Synthes Small Battery Drive User’s Manual
Mode switch turns drive off for safe attachment exchange and selects “on” or “oscillation” mode
12 V Battery (532.003)
– Voltage: 12 VDC – Capacity: 0.5 Amp hours – Type: NiCd – Working temperature: 0°C to 50°C (32°F to 122°F) – Charging time: 30 minutes maximum – Discharge protection: prevents total battery discharge, significantly decreases memory effect, maintains battery capacity, and increases battery life – Indications include drilling, sawing, and inserting K-wires Important: Do not sterilize batteries! Batteries will no longer function.
Drilling
Sawing
K-wire insertion
Synthes
3
14.4 V Battery (532.033)
– Voltage: 14.4 VDC – Capacity: 0.5 Amp hours – Type: NiCd – Working temperature: 0°C to 50°C (32°F to 122°F) – Charging time: 40 minutes maximum – Discharge protection: prevents total battery discharge, significantly decreases memory effect, maintains battery capacity, and increases battery life – Indications include DHS, DCS, and intramedullary reaming Important: Do not sterilize batteries! Batteries will no longer function.
Reaming
4
Synthes Small Battery Drive User’s Manual
Universal Battery Charger (530.601)
– Four charging bays accept Small Battery Drive, Power Drive and Battery Power Line batteries – One (1) battery can be placed in each charging bay allowing a maximum of four (4) batteries to be charged at the same time
Status indicator panels
Power switch Fuse: 2 x 5 AT 250 V (SPT 5 x 20 mm) Socket for power cord
Complies with standards: UL 60950 IEC 60950 EN 55022 class B EN 61000-3-2, 3 EN 61000-4-2, 3, 4, 5, 6, 11
Synthes
5
Universal Battery Charger II (05.001.204)
The Universal Battery Charger II includes four independent charging bays.
Front view
1 2 3 4
5
1 2 3 4 5
Charging bays (4) Symbols for battery type ON / OFF display Control display for each charging bay Ventilation holes
3
Rear view 3 Small Battery Drive Battery (532.003 or 532.033) 6
Complies with standards: UL 60601-1, 1st Edition IEC 60601-1, 2nd Edition
10PB
Universal Battery Charger II With respect to electrical shock, fire and mechanical hazards only in accordance with UL 60601-1/CAN/CSA C22.2 No. 601.1
IEC 60601-1-2 IEC 60601-1-4
6
Synthes Small Battery Drive User’s Manual
7 8 9
6 7 8 9
Ventilation holes Power switch Fuses: 225 AT Power cord connection
Attachments for Small Battery Drive
Drilling speed attachments 532.011
Mini Quick Coupling
532.012
Stryker J-latch Coupling With 12 V battery: Maximum speed: 3,200 rpm Maximum torque: 0.5 Nm With 14.4 V battery: Maximum speed: 3,840 rpm Maximum torque: 0.6 Nm Accepts mini quick coupling or Stryker J-latch drill bits.
532.022
Quick Coupling for K-Wires Cannulation: 3.2 mm (fully open) With 12 V battery: Maximum speed: 800 rpm Maximum torque: 3.0 Nm With 14.4 V battery: Maximum speed: 900 rpm Maximum torque: 3.2 Nm To insert / remove Kirschner wires and guide pins 0.6 mm –3.2 mm in diameter (any length)
The following drilling speed attachments are geared for a maximum drive speed of 1,200 rpm with the 12 V battery and 1,440 rpm with the 14.4 V battery. The maximum torque is 1.3 Nm with the 12 V battery and 1.6 Nm with the 14.4 V battery.* 05.001.250
AO Quick Coupling for Drill Bits Cannulation: 2.1 mm Accepts cutting tools and instruments with the AO quick coupling fitting.
05.001.252
Small Chuck with Key Cannulation: 3.3 mm Accepts round and triangular shafts: up to 4.5 mm
* Performance specifications on file at Synthes.
Synthes
7
Attachments for Small Battery Drive continued
05.001.253
Large Chuck with Key Cannulation: 3.3 mm Accepts round and triangular shafts: up to 7.3 mm
Saw attachments 532.021
Sagittal Saw Attachment Maximum speed: With 12 V battery= 15,000 oscillations per minute With 14.4 V battery= 18,000 oscillations per minute
532.023
Oscillating Saw Attachment II Maximum speed: With 12 V battery= 15,000 oscillations per minute With 14.4 V battery= 18,000 oscillations per minute
532.026
Oscillating Saw Attachment, large Maximum speed: With 12 V battery= 15,000 oscillations per minute With 14.4 V battery= 18,000 oscillations per minute
Reaming speed attachments The following reaming speed attachments are geared for a maximum speed of 384 rpm and a maximum torque of 6.0 Nm with the 14.4 V battery.* All reduction drive units allow intramedullary reaming with most IM nail systems. Note: The reaming speed attachments should only be used with the 14.4 V battery for optimal performance. 05.001.251
AO Screw Attachment Cannulation: 2.0 mm Accepts AO coupling screwdriver
* Performance specification on file at Synthes.
8
Synthes Small Battery Drive User’s Manual
Reaming speed attachments continued 05.001.254
Chuck with Key Cannulation: 3.2 mm Accepts round and triangular shafts: up to 7.3 mm
532.015
Large Quick Coupling Cannulation: 3.2 mm Accepts DHS and DCS triple reamers, screwdriver shafts and cannulated drill bits with large quick coupling fitting
532.017
AO Reduction Drive Unit Cannulation: 3.2 mm Accepts cutting tools and instruments with AO reaming fitting
532.018
Hudson Reduction Drive Unit Cannulation: 3.2 mm Accepts cutting tools and instruments with Hudson fitting
532.019
Trinkle Reduction Drive Unit Cannulation: 3.2 mm Accepts cutting tools and instruments with Trinkle fitting
532.020
Modified Trinkle Reduction Drive Unit Cannulation: 3.2 mm Accepts cutting tools and instruments with modified Trinkle fitting
Synthes
9
Operating Instructions
Universal Battery Charger (530.601) Universal Battery Charger II (05.001.204) Note: For instructions for the Universal Battery Charger II, please refer to the user’s manual. Connect the charger to the electrical cord, and plug the cord into a 110/120 VAC outlet. Turn on the charger by pressing the power switch to ON. All LEDs on the status indicator panels will illuminate (red, yellow and green), when the charger is empty and ready to charge (Figure 1). Insert the battery into one of the appropriate charging bays. Only the yellow LED should remain illuminated, indicating that the battery is charging (Figure 2).
Figure 1
When the green LED illuminates, the battery is fully charged and operational (Figure 3). The battery should be left in the charging bay until needed. The battery can be removed during the charging process; however, the battery may not be fully charged and operational duration may be shorter. Important: Always leave the charger switched on when a battery is inserted in a charging bay. Leaving a battery in the charger when it is switched off or unplugged may permanently discharge the battery.
Figure 2
Never attempt to insert more than one battery in the same charging bay. Note: If LEDs do not follow this sequence, refer to the troubleshooting section, page 29.
Figure 3
10
Synthes Small Battery Drive User’s Manual
Operating Instructions - Battery Pack
The battery pack assembly requires a scrubbed person and a circulating person to insert the battery and preserve the sterile field.
1
2
Scrubbed person
Scrubbed person
Scrubbed person 1 Hold the sterilized battery casing and open the lid. 2 Place the sterilized battery insertion shield on the battery casing. Circulating person 3 Carefully insert the nonsterile battery into the sterile battery casing. 4 Press firmly until it engages. 5 Remove the battery insertion shield.
3
4
Circulating person
Scrubbed person 6 Close and lock the lid.
Scrubbed person
5
Circulating person
Scrubbed person
Circulating person
Scrubbed person
Synthes
11
Operating Instructions - Drive Unit
Speed control For forward/reverse operation, position the mode switch to “ON”. For forward (clockwise) rotation, press the variable-speed trigger of the drive unit. For reverse (counterclockwise) rotation, press both triggers. For forward/oscillate operation, position the mode switch to “OSC DRILL”. For forward rotation, press the variable-speed trigger of the drive unit. For oscillating, press both triggers. Important: Mode switch must be in the “ON” or “OSC DRILL” position to operate drive.
Attachment release button
Attachment coupling
Instant-reverse or oscillation trigger
Variable-speed trigger
Battery casing
12
Synthes Small Battery Drive User’s Manual
Release buttons for battery casing
Operating Instructions - Attachments
Important: To prevent injuries, the mode switch should always be in the “OFF” position when inserting or removing instruments.
Install attachment Insert the attachment into the attachment coupling. It is not necessary to press the release buttons for installation. Check the correct engagement of the attachment by pulling it slightly. Remove attachment Simultaneously press the attachment release buttons and remove the attachment from the coupling.
Synthes
13
Operating Instructions - Attachments for use with 12 V and 14.4 V Batteries
Quick Coupling for K-Wires (532.022) – For insertion of Kirschner wires from 0.6 mm to 3.2 mm diameters (see page 12 for installation and removal of attachments). Important: To prevent injuries, the mode switch should always be in the “OFF” position when inserting or removing instruments.
Insert K-wire into Quick Coupling for K-Wires Select and set the adjustment range for the diameter of the K-wire according to the inscription. • = 0.6 mm – 1.2 mm •• = 1.2 mm – 2.2 mm ••• = 2.2 mm – 3.2 mm Insert the K-wire from the front opening to the desired working depth. Note: The K-wire is “presqueezed” upon insertion into the coupling to prevent slippage during handling. Pulling back on the lever grasps the K-wire securely for insertion into bone.
Insert K-wire into bone Grasp the K-wire by pulling the lever toward the drive unit, and press the lower (forward) trigger. Remove K-wire from bone Slide the drive unit and coupling over the K-wire. Grasp the wire by squeezing the lever toward the drive unit, and press both triggers (reverse) simultaneously to remove the wire from the bone. Note: The mode switch must be in the “ON” position to enable reverse.
14
Synthes Small Battery Drive User’s Manual
Mini Quick Coupling (532.011) Stryker J-latch Coupling (532.012) – Accept mini quick coupling or Stryker J-latch drill bits To insert an instrument into the Mini Quick Coupling or Stryker J-latch Coupling, retract the collar of the coupling and insert the instrument shaft completely with a slight rotation to align the keyway. When the shaft is fully inserted, release the collar. Pull on the instrument slightly to ensure that it is held securely.
AO Quick Coupling for Drill Bits (05.001.250) – Accepts quick coupling drill bits – Allows use of Radiolucent Drive (511.30) with Adaptor (532.031) See page 12 for installation and removal of attachments. Important: To prevent injuries, the mode switch should always be in the “OFF” position when inserting or removing instruments.
Insert an instrument into the quick coupling Insert the instrument shaft completely with a slight rotation to align the keyway. Pull on the instrument slightly to ensure that it is held securely. Note: Collar does not need to be retracted to insert instrument.
05.001.250
Remove the instrument Retract the collar of the quick coupling to release the instrument.
Synthes
15
Operating Instructions - Attachments for use with 12 V and 14.4 V Batteries continued
Large Chuck with Key (05.001.253) – Accepts round shafts from 1.0 mm to 6.0 mm in diameter, or triangular shafts up to 7.3 mm
Small Chuck with Key (05.001.252) – Accepts round shafts from 0.5 mm to 4.0 mm in diameter, or triangular shafts up to 4.5 mm See page 12 for installation and removal of attachments. Important: To prevent injuries, the mode switch should always be in the “OFF” position when inserting or removing instruments. Always remove the key from the chuck before operating the drive unit.
Insert an instrument Open the chuck jaws by turning the key counterclockwise, or by manually turning the collar. Insert the instrument shaft into the opened chuck. Close the chuck manually by rotating the collar, keeping the shaft centered in the jaws. Tighten the chuck by turning the key clockwise. Remove an instrument Turn the key counterclockwise to open the jaws. Remove the instrument.
16
Synthes Small Battery Drive User’s Manual
Sagittal Saw Attachment (532.021) – Accepts small saw blades
Saw blade mount Locking knob
Important: To prevent injuries, the mode switch should always be in the “OFF” position when inserting or removing instruments. The saw blade should always be handled near the hub. Note: To minimize heat generation and provide optimal sawing results, always use new Synthes saw blades. Used blades cannot be resharpened.
Position the saw attachment Once mounted, the saw attachment can be locked in eight different positions in 45° increments. To reposition the saw attachment, push the adjustment sleeve and turn the attachment to the desired position (Figure 1). Insert saw blade Open the blade coupling by pulling the locking knob and turning it counterclockwise (Figure 2). Insert the new saw blade and adjust it to the desired position (Figure 3). The blade position can be locked in 45° increments. Squeeze the blade coupling closed and turn the locking knob clockwise to tighten (Figure 4).
Figure 1
Figure 2
Figure 3
Figure 4
Remove saw blade Open the blade coupling. Remove the blade. Important: Mode switch should always be in the “ON” position while using the Oscillating Saw Attachment.
Synthes
17
Operating Instructions - Attachments for use with 12 V and 14.4 V Batteries continued
Oscillating Saw Attachment II (532.023) – Accepts crescentic saw blades and intraoral saw blades (shown with dimensions on page 33). Large Oscillating Saw Attachment (532.026) – Please refer to Operating Instructions in package insert.
Important: To prevent injuries, the mode switch should always be in the “OFF” position when inserting or removing instruments. The saw blade should always be handled near the hub. Synthes recommends using a new saw blade for each surgery, as used saw blades may increase the risk of necrosis and increase the cutting time. All oscillating and intraoral saw blades are sold sterile. Insert saw blade Pull back the collar of the attachment.
Crescent blades for halux valgus 03.000.311S 03.000.312S 03.000.313S 03.000.316S
Saw blades, 105° angled for intraoral
Insert a saw blade into the saw blade coupling and move it into the desired position.
03.000.343S
Release the collar. Pull lightly on the saw blade to confirm that it is secure.
03.000.341S
03.000.342S
03.000.340S Remove saw blade Pull back on the collar of the attachment, and remove the saw blade.
18
Synthes Small Battery Drive User’s Manual
Adaptor for Radiolucent Drive (532.031) – Allows use of Radiolucent Drive (511.30) with Small Battery Drive Important: To prevent injuries, the mode switch should always be in the “OFF” position when inserting or removing instruments.
Insert adaptor – Insert the Quick Coupling for Drill Bits (05.001.250) into the Small Battery Drive – Insert the Adaptor (532.031) into the quick coupling – Slide the Radiolucent Drive (511.30) over the adaptor and the quick coupling and rotate until the Radiolucent Drive engages in the adaptor Remove adaptor Pull the Radiolucent Drive off the Small Battery Drive and its attachments. Remove the adaptor from the quick coupling.
Small Battery Drive with Quick Coupling for Drill Bits and Adaptor for Radiolucent Drive
See page 12 for installation and removal of attachments.
Small Battery Drive and Radiolucent Drive assembled
Radiolucent Drive (511.30)
Adaptor (532.031)
Quick Coupling for Drill Bits (05.001.250)