Synthes
Synthes Ultrasonic Surgical Systems
Electric Pen Drive 05.001.010 Instructions for Use Feb 2017
Instructions for Use
64 Pages
Preview
Page 1
ELECTRIC PEN DRIVE Compact drive with specific attachments for a wide spectrum of applications
This publication is not intended for distribution in the USA.
INSTRUCTIONS FOR USE
TABLE OF CONTENTS
INTRODUCTION
CONSOLES
General Information
3
Explanation of Symbols
5
Standard Consoles
6
Basic Console
7
Set-up of Consoles
8
Speed Regulation
10
Irrigation11
ELECTRIC PEN DRIVE SYSTEM
ATTACHMENTS
System Electric Pen Drive 60,000 rpm (05.001.010)
13
Hand Switch (05.001.012)
14
Foot Switch, 1 Pedal (05.001.016)
15
Foot Switch, 2 Pedals (05.001.017)
16
General Information
17
Drill Attachments
18
Screw Attachments
20
Kirschner Wire Attachment
21
Saw Attachments
22
Burr Attachments
24
Adapter for Intra Coupling
26
Perforators27 Craniotome Attachment
Electric Pen Drive
Instructions for Use
29
DePuy Synthes
1
Table of Contents
CUTTING TOOLS
General Information
30
CARE AND MAINTENANCE
General Information
31
Preparation Prior to Cleaning
32
Manual Cleaning Instruction
33
Mechanical / Automated Cleaning
35
Instruction with Manual Pre-cleaning
35
Maintenance and Lubrication
40
Function Control
43
Packaging, Sterilization and Storage
44
Repairs and Technical Service
45
Disposal46
TROUBLESHOOTING47
SYSTEM SPECIFICATIONS
49
ELECTROMAGNETIC COMPATIBILITY
55
ORDERING INFORMATION
59
2
DePuy Synthes Electric Pen Drive
Instructions for Use
Introduction
GENERAL INFORMATION
Intended Use The Electric Pen Drive is an electrically-powered system to be used for treatment in general traumatology, as well as for surgery in the hand, foot, spine and maxillofacial regions and neurosurgery. Safety Instructions The surgeon has to evaluate if the machine is suitable for an application, based on power limitation of the machine, attachment and cutting tool regarding bone strength/anatomical situation as well as handling of the machine, attachment and cutting tool regarding bone size. In addition, the contraindications of the implant have to be respected. Please refer to the corresponding “Surgical Techniques” of the implant system used.
To prevent overheating, always respect the duty cycles for each attachment listed on page 56. For the tool to function properly, Synthes recommends that it is cleaned and serviced after each use in accordance with the process recommended in the chapter “Care and Maintenance”. Compliance with these specifications can considerably extend the service life of the tool. Only use Synthes oil to lubricate the tool. Efficiently working cutting tools are the basis for successful surgery. Therefore, it is mandatory to check used cutting tools after every use for wear and / or damage and to replace them if necessary. We recommend using new Synthes cutting tools for every surgery.
The Electric Pen Drive System is only to be used for patient treatment after careful consultation of the instructions for use. It is recommended that an alternative system is available to use during application, as technical problems can never be completely ruled out.
Cutting tools must be cooled with irrigation liquid to prevent heat necrosis.
The Electric Pen Drive System is designed for use by physicians and trained medical personnel.
If the Electric Pen Drive System is used in conjunction with an implant system make sure to consult the corresponding “Surgical Technique”.
The user of the product is responsible for proper use of the equipment during surgery.
DO NOT use any component if damage is apparent. DO NOT use this equipment in presence of oxygen, nitrous oxide or a mixture consisting of flammable anesthetic and air. Never place the Electric Pen Drive in a magnetic environment since the machine might start unintentionally. To ensure proper operation of the tool, use only Synthes original accessories. Before first and every use, power tools and their accessories / attachments have to run through the complete reprocessing procedure. Protective covers and foils must be fully removed before sterilization.
For important information regarding electromagnetic compatibility (EMC) please refer to the chapter “System Specifications” in this manual. The tool is classified as type B against electrical shock and leakage current. The tool is suitable for use on patients in accordance with IEC 60601-1. This system requires regular maintenance service, at least once a year, in order to maintain its functionality. This service has to be performed by the original manufacturer or an authorized site.
Check instruments for correct adjustment and functioning prior to every use. Always wear personal protective equipment (PPE) including safety goggles when handling with the Electric Pen Drive.
Electric Pen Drive
Instructions for Use
DePuy Synthes
3
Introduction General Information
Unusual Transmissible Pathogens Surgical patients identified as at-risk for CreutzfeldtJakob disease (CJD) and related infections should be treated with single-use instruments. Dispose of instruments used or suspected of use on a patient with CJD after surgery and / or follow current national recommendations. Precautions: • To avoid injuries, the locking mechanism of the tool has to be activated before every manipulation and before placing the tool back down, i.e. the mode switch has to be in the LOCK position ( ). • Should the machine drop on the floor and have visible defects, do not use it anymore and send it to the Synthes service center. • If a product drops on the floor, fragments may split off. This represents a danger for the patient and user as: • these fragments may be sharp. • unsterile fragments may enter the sterile field or hit the patient. Accessories / Scope of delivery The main components of the Electric Pen Drive (EPD) System are the handpiece, hand switch, foot switch, console, electric cables as well as attachments and accessories. An overview of all components belonging to the Electric Pen Drive System can be found in the chapter “Ordering Information”. For using the EPD System the following components are a must: • Electric Pen Drive (05.001.010) • Hand Switch (05.001.012) or Foot Switch (05.001.016) with the Cable Foot Switch-Console (05.001.022) • Console (05.001.006 or 05.001.002) • Cable Electric Pen Drive – Console (05.001.021 or 05.001.025) • At least one attachment belonging to the system and a cutting tool fitting to the attachment. For an optimal function of the system only Synthes cutting tools shall be used.
Synthes recommends the use of the specifically designed Synthes Vario Cases and of the specifically designed Washing Basket (68.001.800) to sterilize and store the system. For care and maintenance special tools are available, such as cleaning brushes, Synthes Maintenance Oil for EPD and APD (05.001.095), Maintenance Spray (05.001.098) and a Maintenance Unit (05.001.099). No oils from other manufacturers may be used. Only Synthes oil may be used. Lubricants with other compositions can cause jamming, can have a toxic effect or can have a negative impact on the sterilization results. Only lubricate the power tool and the attachments when clean. Locating of the instrument or fragments of instruments Synthes instruments are designed and manufactured to perform within the scope of their intended use. However, if a Power Tool or accessory/attachment breaks during use, a visual inspection or a medical imagine device (e.g. CT, Radiation Devices, etc.) can aid in locating the fragments and/or components of the instrument. Storage and Transport Only use the original packaging for dispatch and transport. If the packing material is no longer available, please contact the Synthes office. For storage and transport environmental conditions, see page 55. Warranty / Liability The warranty for the tools and accessories does not cover damage of any kind resulting from wear, improper use, improper reprocessing and maintenance, damaged seal, use of non Synthes cutting tools and lubricants or improper storage and transport. The manufacturer excludes liability for damage resulting from improper use, neglected or unauthorized mainte nance or servicing of the tool. For further information on the warranty please contact your local Synthes office.
4
DePuy Synthes Electric Pen Drive
Instructions for Use
Introduction
EXPLANATION OF SYMBOLS
Do not reuse Products intended for single use must not be reused.
Caution Read the provided instructions for use before operating the device.
Reuse or reprocessing (e.g. cleaning and resterilization) may compromise the structural integrity of the device and / or lead to device failure which may result in patient injury, illness or death. Furthermore, reuse or reprocessing of single use devices may create a risk of contamination e.g. due to the transmission of infectious material from one patient to another. This could result in injury or death of the patient or user.
Indicates that the device complies with the classification type B against electrical shock and electrical leakage current. The device is suitable for use on patients as defined by IEC 60601-1. Do not immerse device in liquids.
calibrate The marked device shall be re-calibrated. annually
10PB
0123
Synthes does not recommend reprocessing contaminated products. Any Synthes product that has been contaminated by blood, tissue and / or bodily fluids / matter should never be used again and should be handled according to hospital protocol. Even though they may appear undamaged, the products may have small defects and internal stress patterns that may cause material fatigue.
Product is UL classified to the requirements of both the United States and Canada The device meets the requirements of directive 93/42/EEC for medical devices. It is authorized by an independent notified body for which it bears the CE symbol. The European directive 2002/96/EC on waste electrical and electronic equipment (WEEE) to this device. This device contains CdappliesMH materials that should be disposed of in accordance with environment protection requirements. Please observe national and local regulations.
IP X4
Locked symbol. Drive Unit is off for safety. Manufacturer Date of manufacture
Range of relative humidity non sterile
Range of atmospheric pressure
Do not use if package is damaged.
Ingress protection rating according to IEC 60529.
Non sterile Non sterile Range of temperature
Electric Pen Drive
Instructions for Use
DePuy Synthes
1
Consoles
STANDARD CONSOLES
The Electric Pen Drive (05.001.010) can be supplied with the Standard Console (05.001.006) or with the Basic Console without integrated irrigation system (05.001.002). Standard Console with irrigation, without torque limiting function (05.001.006) 1 Slide control for adjusting the maximum speed for 1 2 Slide control for adjusting the maximum speed for 2 4 Selection switch for irrigation 5 Connection for Adapter for Colibri * 6 Connection for Electric Pen Drive and Small Electric Drive 1 7 Connection for Electric Pen Drive and Small Electric Drive 2 8 Connection for Foot Switch 9 Twist lock for irrigation pump 10 Adjustment knob for irrigation flow rate 11 LED Irrigation Ready (Activation with Foot Switch)
* T his connection is not of use anymore since the Adaptor for Colibri is no longer available.
6
DePuy Synthes Electric Pen Drive
Instructions for Use
5
1
2
6
7
11
8
4
9
10
Consoles
BASIC CONSOLE
Basic Console (05.001.002) 1 Slide control for adjustment of the maximum speed for 1 5 Connection for Adapter for Colibri * 6 Connection for Electric Pen Drive and Small Electric Drive 1 8 Connection for Foot Switch
1
Note: For more details on the Small Electric Drive please refer to the separate manual and contact your Synthes representative for more information.
5
6
8
* This connection is not of use anymore since the Adaptor for Colibri is no longer available.
Electric Pen Drive
Instructions for Use
DePuy Synthes
7
Consoles
SET-UP OF CONSOLES
Color Coding Set (60038602) A color coding set is supplied with every console. Each set contains 3 text stickers, 8 silicone rings in 3 colors (red, blue and white) and 8 stickers of the colored rings. The stickers and silicone rings can be used to indicate on the console and the cables how the system has to be assembled.
Content of Set No. 60038602
Set-up Before the initial operation of the device, make sure that the power switch (12) is set to position 0. Only connect the console to the power supply with the cable provided via the power plug socket (13) and set the power switch to position 1 (ON). The LED marked with on the front of the console signals the proper operation of the console. If the LED flashes, the console must be sent in for maintenance. A potential equilization connector (14) for an equipotential earth plug is integrated in the console. Existing equipotential earth plugs can be connected there. (15) Fuses: 223AF/250V, Breaking Capacity 1500A
8
DePuy Synthes Electric Pen Drive
Instructions for Use
14 12 15 13
Mounting cables on consoles To mount the respective cables on the consoles, position the nose on the plugs flush with the groove on the plug sockets and insert the plugs. For improved orientation, red dots are provided on the plug and socket that must face upward when connecting the cable. Removing cables To remove the plug, grasp the release sleeve, pull back and remove. The connections 5–8 (page 6) are used to connect the following devices: (5): Connection for a Colibri / Small Battery Drive handpiece (6 and 7): Connection for an Electric Pen Drive and Small Electric Drive (8): C onnection for the Foot Switch Connections not used can be sealed off with the protective caps provided. Precautions: • Do not lay textiles or objects under the console. These can be sucked up and block the ventilation inlet. • Do not block the ventilation opening on the back of the console with any object. • Only place consoles on smooth and flat surfaces. • Do not place the console in the sterile field. • Do not hang the irrigation liquid directly above the console to prevent liquid from dripping on the console. • Do not pull the cable! Always activate the release sleeve. • The use of HF (= high frequency) equipment for tissue coagulation can cause electromagnetic interferences – in this case cables should be separated as far as possible. • Ensure that the power cord can always be disconnected immediately from mains supply.
Electric Pen Drive
Instructions for Use
DePuy Synthes
9
Consoles
SPEED REGULATION
Functions of the Standard Console (05.001.006) and the Basic Console (05.001.002) Adjustment of the maximum speed (1, 2) Speed is automatically optimized and adjusted for each attachment; for certain attachments it is however recommended that you reduce the maximum speed of the connected handpiece. Adjustment can be performed in increments of 25 % with the slide controls for adjustment of maximum speed. It is recommended using this feature for high speed burrs. A letter code on each burr indicates the maximum speed specified by Synthes. Marking
Console Setting Pen 60k
A
100%
B
100%
C
75%
D
50%
E
25%
This is also explained by symbols on the supplied adhesive foil. The adhesive foil can be applied to the console as a reference aid.
11
DePuy Synthes Electric Pen Drive
Instructions for Use
Consoles
IRRIGATION
Function of the Standard Console (05.001.006) The function of the Standard Consoles described in the following is not available on the Basic Console. Irrigation (4) With the integrated irrigation system, tools can be cooled to prevent tissue necrosis affected by excess heat. Irrigation nozzles for the attachments and sterile irrigation tubes are available for the irrigation system. In addition to the OFF position (position 4.1), there are two operating positions for Pen 1 and Pen 2: constant and variable irrigation (refer to Figure 4). If constant irrigation for Pen 1/Pen 2 is chosen, a constant quantity of irrigation liquid will be released for Pen 1/Pen 2. A constant flow rate between 10 and 100 ml/min (0.34 and 3.4 oz/min) can be adjusted with the adjustment knob (10) for the irrigation flow rate (position 4.3/4.5). If variable irrigation for Pen 1/Pen 2 is chosen, the flow rate will be directly proportional to the speed selected on the Hand or Foot Switch, i.e. the higher the selected speed, the higher is the irrigation flow rate. The maximum flow rate can be set between 10 and 100 ml/min (0.34 and 3.4 oz/min) with the adjustment knob (10) for the irrigation flow rate (position 4.2/4.4).
4.2 4.3
4.1
4.5
4.4
4.1 Irrigation OFF 4.2 Variable irrigation for Pen 1 VAR 4.3 Constant irrigation for Pen 1 CON 4.4 Variable irrigation for Pen 2 VAR 4.5 Constant irrigation for Pen 2 CON
11
The LED lights up if the irrigation is activated (11) on the foot switch. Please refer to the chapter of the Foot Switch for a detailed explanation on how to activate or deactivate the irrigation. A Bottle Holder for Irrigation Fluid (05.001.007) is available for the standard console (05.001.006). The Bottle Holder allows to hang a bottle with irrigation fluid near the console.
Bottle Holder
Electric Pen Drive
Instructions for Use
DePuy Synthes
11
Consoles Irrigation
Irrigation Tube Set (05.001.178.01S) and Clips for Irrigation Tube Set (05.001.179.05S), Inserting Irrigation Tube 1. Remove the sterile Irrigation Tube (Fig. 1) from the sterile package. 2. Secure the attachment specific Irrigation Nozzle on the attachment (Fig. 2) used. 3. Fit the irrigation tube on the Irrigation Nozzle used first and then fix it with the clips (Fig 3) on the cable. 4. Route the Irrigation Tube end into the non-sterile area and open the twist lock for the irrigation pump in the direction of the arrow. 5. Insert the Irrigation Tube in the pump in accordance with the marking (see Fig. 4) and close the twist lock. 6. Remove the protective cap from the cannula and connect cannula to irrigation bag. When doing so, make sure that the connection nipple of the cannula is not touched by non-sterile persons while securing irrigation nozzle to the attachment. The vent of the spike has to be open when the irrigation is used. Push the Irrigation Nozzles for the Burr Attachments and Reciprocating Saw Attachment as far as necessary over the attachment from the front.
Fig. 1
Fig. 2
Fig. 3
Push the Irrigation Nozzle for the Sagittal Saw Attachment onto the attachment from the rear (from the attachment coupling side) before mounting the attachment on the handpiece.
Fig. 4
11
DePuy Synthes Electric Pen Drive
Instructions for Use
Electric Pen Drive System
SYSTEM ELECTRIC PEN DRIVE 60,000 RPM (05.001.010)
1 Adjustment sleeve 2 Release sleeve for attachment 3 Cable IN / OUT position 4 LOCK position 5 Forward position (clockwise) 6 Reverse position (anti-clockwise) 7 Lock slide for adjustment sleeve Turning the adjustment sleeve To avoid an unintentional change of the operating mode the lock slide for adjustment sleeve (7) locks the adjustment sleeve automatically. To be able to move the adjustment sleeve the lock slide has to be pushed in the direction indicated by the arrow. After having reached the desired position release the lock slide (7) and the adjustment sleeve (1) is locked in the desired position. Mounting cable on handpiece Move the adjustment sleeve (1) on the handpiece into the position cable IN / OUT (3). Align the groove on the plug with the notch in the sleeve and insert the plug. Turn the adjustment sleeve into the LOCK position (4). The cable is now firmly locked to the handpiece and the handpiece is in locked mode. To remove it again, move the adjustment sleeve (1) into the cable IN / OUT position (3) and remove the cable.
5
2 3
6
1
4
7
FWD / REV switchover By turning the adjustment sleeve into the forward position (5), the handpiece can be switched into a clockwise operating direction. In the reverse position (6) the handpiece can be operated in an anti-clockwise direction. Precautions: In addition to locking the cable, the LOCK position (4) • Do not place the handpiece on magnetic covers is used for safety switch-off when changing attachments or in the immediate vicinity of other magnetic objects. This can activate the handpiece. and tools prevent accidental start-up of the unit. • When two Electric Pen Drive handpieces are connected and the speed is controlled with the For instructions on mounting the attachments see Foot Switch, one handpiece must be switched to page 21. LOCK .Otherwise both handpieces will be blocked for safety reasons. Either a Hand Switch (page 18) or a Foot Switch (pages • In all other cases, the first activated device func19/20) can be used for speed control. tions. As long as this device is activated, all others are deactivated. • HF (= high frequency) equipment for tissue coagulation can cause electromagnetic interferences and inadvertently activate the Electric Pen Drive – cables should be separated as far as possible. Electric Pen Drive
Instructions for Use
DePuy Synthes
11
Electric Pen Drive System
HAND SWITCH (05.001.012)
1 Positioning arrow 2 Pull-out finger rest 3 Positioning arrow 4 Guide groove 5 Locking switch
5 1
Mounting the Hand Switch on the handpiece (05.001.010) Position the Hand Switch on the handpiece so that both positioning arrows (1) of the Hand Switch cover the positioning arrows (3) over the guide grooves (4) of the pen. Then press downward vertically until the Hand Switch clicks into position. Removing To remove the Hand Switch, grasp the lever and pull it away upwards. Operation The length of the Hand Switch can be individually adjusted with the pull-out finger rest (2). The speed can be continuously adjusted by operating the Hand Switch. The Hand Switch can be deactivated (LOCK position) or activated (ON position) with the locking switch (5).
11
DePuy Synthes Electric Pen Drive
Instructions for Use
3
4
2
Electric Pen Drive System
FOOT SWITCH, 1 PEDAL (05.001.016)
Connecting the Foot Switch to a console The Foot Switch can be connected to the console with the Foot Switch Cable (05.001.022). To insert the plugs, align the red dots on the plugs with those on the sockets and insert the plug. A second foot switch can be connected with the second socket in the foot pedal. Use the cable (05.001.026) for this purpose. However, this is not possible in combination with the basic console.
1
The second socket is covered with a protective cap on delivery; this can be removed if necessary. Disconnecting a Foot Switch Grasp the respective plug by the release sleeve, pull back and remove. Operation The speed can be continuously adjusted with the pedal (2).
2
1 Irrigation ON / OFF button 2 Pedal
Briefly pressing the irrigation button (1) connects or disconnects the irrigation. In the ON position the setting preselected with the irrigation selection switch on the console is activated. The LED lights up if the irrigation is activated on the foot switch. If the user holds this button down, the irrigation quantity set with the irrigation flowrate adjustment knob (page 6) is supplied until the button is released. This function is independent of the position selected on the irrigation selection switch (page 6) and of the activation of the pedal (2) or Hand Switch (page 19). Precaution: When the cable is not connected properly to the foot switch, it is possible that the handpiece is activated without pressing the foot switch.
Electric Pen Drive
Instructions for Use
DePuy Synthes
11
Electric Pen Drive System
FOOT SWITCH, 2 PEDALS (05.001.017)
Connecting the Foot Switch to a console The Foot Switch can be connected to the console with the Foot Switch Cable (05.001.022). To insert the plugs, align the red dots on the plugs with those on the sockets and insert the plug. A second foot switch can be connected with the second socket in the foot pedal. Use the cable (05.001.026) for this purpose. However, this is not possible in combination with the basic console.
3 2
The second socket is covered with a protective cap on delivery; this can be removed if necessary.
1
Disconnecting the Foot Switch Grasp the respective plug by the release sleeve, pull back and remove. Operation Operating mode is forward FWD (1) when activating the right pedal and reverse REV (2) for the left pedal. The speed can be continuously adjusted with the pedals (1 and 2). Briefly pressing the irrigation button (3) connects or disconnects the irrigation. In the ON position the setting preselected with the irrigation selection switch on the console is activated. The LED lights up if the irrigation is activated on the foot switch. If the user holds this button down, the irrigation quantity set with the irrigation flowrate adjustment knob (page 6) is supplied until the button is released. This function is independent of the position selected on the irrigation selection switch (page 6) and of the activation of the pedal (1 and 2). Precautions: • When operating the Electric Pen Drive 60 000 rpm (05.001.010) with the Foot Switch (05.001.017) the operating mode is defined by the pedal used on the Foot Switch (05.001.017), i. e. right pedal for forward mode and left pedal for reverse mode regardless of the operating mode (FWD or REV) defined by the adjustment sleeve of the handpiece. • When the cable is not connected properly to the foot switch, it is possible that the handpiece is activated without pressing the foot switch.
11
DePuy Synthes Electric Pen Drive
Instructions for Use
1 FWD 2 REV 3 Irrigation ON / OFF button
Attachments
GENERAL INFORMATION
Mounting the attachments on the handpieces (05.001.010) The attachments can be connected in 8 different positions (45° increments). To mount, turn the release sleeve for the attachments clockwise (see arrow on the release sleeve) until it engages. The release sleeve protrudes slightly from the black section of the handpiece towards the front. Insert the attachment into the attachment coupling from the front and press it lightly against the handpiece. The attachment automatically engages. If the release sleeve accidentally closes, turn the attachment clockwise while applying slight pressure against the handpiece until it engages without holding the release sleeve in place, or repeat the entire connection process. Check secure holding of the attachment on the handpiece by pulling on the attachment.
Release sleeve
Removing the attachments from the handpiece Turn the release sleeve for attachments clockwise until it disengages. Hold the attachment upward while doing so. Then remove the attachment.
Attachments and accessories For easier change of burrs, without having the Burr Attachment or Craniotome Attachment connected to the handpiece, the Handle for change of instruments (05.001.074) can be used.
Warranty: Only use Synthes saw blades, burrs and rasps while working with the Electric Pen Drive attachments. The use of other tools voids the device warranty.
Electric Pen Drive
Instructions for Use
DePuy Synthes
11
Attachments
DRILL ATTACHMENTS
Drill Attachments (05.001.030–05.001.032, 05.001.042–05.001.044, 05.001.110) Speed drill attachments: approx. 1,800 rpm
1
The system includes straight and 45° Drill Attachments with Mini Quick Coupling, J-Latch Coupling and AO / ASIF Coupling, as well as a straight Drill Attachment with Hexagonal Coupling.
1
The 45° Drill Attachment with AO / ASIF Coupling (05.001.044) has a cannulation of 1.6 mm, which permits the use of this attachment for drilling and reaming over Kirschner Wire (e.g. for cannulated screws and for cup and cone technique).
1
Mounting and removing tools Lock unit. Pull back the release sleeve and insert / remove the tool. 1 Release sleeve
Drill Attachment 45°, cannulated, with Jacobs Chuck (05.001.120) Speed drill attachments: approx. 1,800 rpm Clamping range: 0.5 mm–4.7 mm The cannulation of 1.6 mm permits the use of this attachment for drilling and reaming over Kirschner Wire (e.g. for cannulated screws and for cup and cone technique). Mounting and removing tools Lock unit. Open the chuck with the key provided (310.932) or by hand by turning the two moving parts clockwise with respect to each other. Insert / remove the tool. Close the chuck by turning the moving parts counterclockwise and tighten it by turning the key clockwise.
11
DePuy Synthes Electric Pen Drive
Instructions for Use
1 Release sleeve
1
1
Drill Attachment 90°, short (05.001.035) and long (05.001.036) with Mini Quick Coupling Speed: approx. 1,800 rpm Due to their very small angled head the 90° Drill Attachments enable good visibility during operations with narrow access (e.g. intra-oral, shoulder etc.). Mounting and removing tools Lock unit. Move the slide (1) to the side following the arrow on the slide and insert / remove the tool. To secure the tool, push the slide back again.
1
1 Slide
Oscillating Drill Attachment (05.001.033) Frequency: approx. 3,200 osc/min The oscillating drilling movement of the Oscillating Drill Attachment prevents tissue and nerves from wrapping around the drill. This can considerably improve the operating results. Mounting and removing tools Tools with a Mini Quick Coupling can be clamped with the Oscillating Drill Attachment. To do this, lock the unit, pull back the release sleeve and insert / remove the tool. Precaution: The handpiece must be in FWD tion to use the oscillating drill attachment.
1
1 Release sleeve
posi-
Electric Pen Drive
Instructions for Use
DePuy Synthes
11