Teleflex Medical

NextStep® Retrograde Chronic Hemodialysis Catheter

6 Pages

NextStep® Retrograde Chronic Hemodialysis Catheter Rx only.  Product Description:  3. All chronic dialysis catheters should be used only as bridge devices and are not intended for extended use unless no other hemodialysis access options exist.  The Arrow® NextStep® Retrograde Catheter consists of a catheter with a separate hub connection assembly. This design allows catheter tip to be precisely positioned. After the tip has been positioned, the proximal end of the catheter is tunneled retrograde to the exit site. The separate hub connection assembly is then fastened to the proximal end of the catheter using a compression sleeve and threaded compression cap (refer to Figure 1).  4. Air embolism can occur if air is allowed to enter a vascular access device or vein. Do not leave open needles or uncapped, unclamped catheters in venous puncture site. Use only securely tightened LuerLock connections with any vascular access device to guard against inadvertent disconnection.  The NextStep Retrograde Catheter is manufactured from radiopaque polyurethane. Luer Connections  5. Do not apply excessive force in placing or removing catheter or guidewire. Excessive force can cause component damage or breakage. If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained and further consultation requested.  Extension Lines Extension Line Clamps  Hub Connection Assembly  Arterial Port  Double D Cannula Compression Sleeve  6. If guidewire becomes damaged, catheter and guidewire must be removed together.  Threaded Compression Cap  7. Clinicians must be aware of potential entrapment of the guidewire by any implanted device in circulatory system. It is recommended that if patient has a circulatory system implant, catheter procedure be done under direct visualization to reduce risk of guidewire entrapment. Venous Port  Catheter Body  8. Passage of guidewire into the right heart can cause dysrhythmias, right bundle branch block, and a perforation of vessel, atrial or ventricular wall.  Cuff  9. Do not use excessive force when introducing guidewire or tissue dilator as this can lead to vessel perforation, bleeding, or component damage. 10. Patients requiring ventilator support are at an increased risk of pneumothorax during subclavian vein cannulation.  Figure 1  11. Extended use of subclavian vein may be associated with subclavian vein stenosis.  The Arrow NextStep Retrograde Catheter is indicated for use in attaining long-term vascular access for hemodialysis and apheresis. The NextStep Retrograde Catheter is inserted percutaneously and is preferentially placed into the internal jugular (IJ) vein. Alternately, this catheter may be inserted into the subclavian vein although the jugular vein is the preferred site. Catheters greater than 40 cm are intended for femoral vein insertion. The Arrow NextStep Retrograde Catheter is intended for use in adult patients.  12. Clinicians must be aware of complications/undesirable side-effects associated with central venous catheters including, but not limited to:  Indications for use:  • air embolus		 • bacteremia		 • brachial plexus injury		 • caval perforation		 w/bronchial communication • cardiac arrhythmia		 • cardiac tamponade 		 • central venous thrombosis		 • endocarditis		 • exit site infection		 • exsanguination		 • femoral nerve damage		 • hematoma		 • hemorrhage		 • hemothorax		  Contraindications: The Arrow NextStep Retrograde Catheter is intended for long-term hemodialysis vascular access only and should not be used for any other purpose other than those indicated herein. Do not use this catheter in patients with thrombosed vessels.  Clinical Benefits to be Expected: • • • •  Permits long term vascular access for hemodialysis and apheresis. Permits vascular access percutaneously. Permits vascular access by way of the internal jugular, subclavian veins. Catheters greater than 40 cm permit vascular access by way of the femoral veins.  MRI Safety Information: Non-clinical testing demonstrated that the Arrow NextStep Retrograde Chronic Hemodialysis Catheter (CHDC) is MR Conditional. A patient with this device can be safely scanned in an MR system meeting the following conditions: • Static magnetic field of 1.5 Tesla and 3.0 Tesla only • Maximum spatial gradient magnetic field of 4,000 Gauss/cm (40 T/m) or less • Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 4 W/kg (First Level Controlled Operating Mode)  • laceration of vessel • lumen thrombosis • mediastinal injury • puncture/perforation of vessel • pleural injury • pneumothorax • retroperitoneal bleed • right atrial puncture • septicemia • thoracic duct laceration • tunnel infection • vascular thrombosis • venous stenosis • subcutaneous hematoma  Precautions: 1. Do not alter any kit/set component during insertions, use or removal (except as instructed). 2. Procedure must be performed by trained personnel well versed in anatomical landmarks, safe technique and potential complications. 3. Use standard precautions and follow institutional policies for all procedures including safe disposal of devices.  Under the scan conditions defined above, the Arrow NextStep Retrograde CHDC is not expected to produce any MRI-related heating. In non-clinical testing, the image artifact caused by the Arrow NextStep Retrograde CHDC extends approximately 5 mm from the device when imaged with a gradient echo pulse sequence in a 3.0 T MRI system.  4. Examine catheter and extension lines before and after each treatment for any signs of damage. 5. Do not use acetone with this catheter or catheter degradation may occur.  General Warnings and Precautions  6. Do not use sharp instruments near extension lines or catheter.  Warnings:  7. Do not use clamps other than those that are provided. Using other clamps may damage the catheter.  1. Sterile, Single use: Do not reuse, reprocess or resterilize. Reuse of device creates a potential risk of serious injury and/or infection which may lead to death. Reprocessing of medical devices intended for single use only may result in degraded performance or a loss of functionality.  8. Do not clamp tubing repeatedly in the same place. Doing so may weaken the tubing. 9. Avoid clamping near Luer-Lock fittings. 10. Repeated over tightening of bloodlines, syringes and caps will reduce connector life and could lead to potential connector failure.  2. Read all package insert warnings, precautions and instructions prior to use. Failure to do so may result in severe patient injury or death. 1
File Type: PDF
File Size: 846 KB
File Name: Teleflex Medical - Q15232109B0 - NextStep® Retrograde Chronic Hemodialysis Catheter - 2022-04 - 01.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions