Quick Operating Instructions
2 Pages
Preview
Page 1
TERUMO TE - 331/2 SYRINGE PUMP
Operating Instructions
Operation Panel
To Operate
Body weight mode (332)
1) Turn Power On 2) Load Syringe
• Press [POWER] switch for 1 second to turn pump on • Lift clamp and place syringe with flanges in groove • Lower clamp onto syringe • While pressing clutch push the driving arm firmly against syringe plunger
3) Set Flow Rate
• While [RATE] indicator is displayed • Rotate dial at least 1/2 a turn to activate • Rotate clockwise to increase, rotate anticlockwise to decrease • Hold down [STOP/SILENCE] switch while rotating dial To increase in 100mL increments
4) Start 5) Stop 6) Power Off
• Press [START] switch • Press [STOP] switch • hold down [ON/OFF] switch and allow pump to count to 3
To Purge • Press [STOP/SILENCE] switch • Press and hold [PURGE] switch down until desired volume is infused To Change Flow Rate • Press [STOP/SILENCE] switch • Rotate dial to set new infusion rate • Press [START] switch To Check Volume Infused (∑mL) • Press [SELECT] switch • Display shows volume infused To Zero Volume Infused (c∑mL) • Press [STOP/SILENCE] switch • Press [SELECT] switch to display ∑mL • Press & hold [CLEAR (c∑mL)] switch for 3 seconds to zero ∑mL To Change Occlusion Setting • Press & hold both [STOP/SILENCE] and [SELECT] switches for 3 seconds is shown on display • • Keep the [STOP/SILENCE] switch depressed and press the [SELECT] switch to move between L.M.H until desired setting is attained To Change Alarm Volume • Press & hold both [STOP/SILENCE] and [CLEAR (c∑mL)] switches for 3 seconds • is shown on display • Keep the [STOP/SILENCE] switch depressed and press the [CLEAR (c∑mL)] switch to move between settings (1,2,3) until the desired setting is attained To Use Body Weight Mode (TE-332 only) • Hold down [ug/kg/min] button for 3 secs • To set individual patient parameters press [rate] button and use dial to set each value • Once all parameters are set the rate is automatically calculated. Press [START] button
h
FOR FURTHER OPERATING INSTRUCTIONS PLEASE REFER TO THE TERUMO TE-331/2 OPERATING MANUAL ® Registered Trademark
Terumo Australia Pty Limited Level 4, Building B, 11 Talavera Road Macquarie Park NSW 2113 T: +61 2 9878 5122 F: +61 2 9878 5085 T: 1800 837 866 F: 1800 334 190 E: [email protected]
GHP130108D421V1
PO Box 137-059 Parnell Auckland 1151, New Zealand T: 0800 66 77 57 F: 0800 66 99 88