TOSHIBA
Aplio Ultrasound Systems
Aplio iXXX TUS-AIxxx series Acoustic Power and Surface Temperature Data Guide Sept 2016
Guide
231 Pages
Preview
Page 1
ACOUSTIC POWER AND SURFACE TEMPERATURE DATA 2B771-144EN*H
DIAGNOSTIC ULTRASOUND SYSTEM
TUS−AI900
TUS−AI800
TUS−AI700
No. 2B771-144EN*H
ACOUSTIC POWER AND SURFACE TEMPERATURE DATA FOR DIAGNOSTIC ULTRASOUND SYSTEM
MODELS TUS-AI900 / TUS-AI800 / TUS-AI700 (2B771-144EN*H)
IMPORTANT! Read and understand this manual before operating the equipment. After reading, keep this manual in an easily accessible place.
TOSHIBA MEDICAL SYSTEMS CORPORATION 2016 ALL RIGHTS RESERVED
Issued : 2016-09
No. 2B771-144EN*H
IMPORTANT! 1. No part of this manual may be copied or reprinted, in whole or in part, without written permission. 2. The contents of this manual are subject to change without prior notice and without our legal obligation.
No. 2B771-144EN*H
Table of Contents How to read the ACOUSTIC POWER TABLE -------------------------------------------------- 5 Transducer maximum surface temperature ---------------------------------------------------- 7 ACOUSTIC POWER TABLE • PSI-30BX ------------------------------------------------------------------------------------------------------------------------ 8 • PSI-30VX ----------------------------------------------------------------------------------------------------------------------- 16 • PSI-70BT ----------------------------------------------------------------------------------------------------------------------- 24 • PST-25BT ---------------------------------------------------------------------------------------------------------------------- 32 • PST-30BT ---------------------------------------------------------------------------------------------------------------------- 39 • PST-50BT ---------------------------------------------------------------------------------------------------------------------- 47 • PVI-475BT --------------------------------------------------------------------------------------------------------------------- 54 • PVI-475BX --------------------------------------------------------------------------------------------------------------------- 62 • PVT-350BTP ------------------------------------------------------------------------------------------------------------------ 70 • PVT-375BT -------------------------------------------------------------------------------------------------------------------- 77 • PVT-375SC -------------------------------------------------------------------------------------------------------------------- 85 • PVT-382BT -------------------------------------------------------------------------------------------------------------------- 93 • PVT-674BT ------------------------------------------------------------------------------------------------------------------ 100 • PVT-675MVL ---------------------------------------------------------------------------------------------------------------- 108 • PVT-681MVL ---------------------------------------------------------------------------------------------------------------- 113 • PVT-712BT ------------------------------------------------------------------------------------------------------------------ 120 • PVT-745BTF / PVT-745BTH -------------------------------------------------------------------------------------------- 125 • PVT-745BTV ---------------------------------------------------------------------------------------------------------------- 131 • PVT-781VT / PVT-781VTE ---------------------------------------------------------------------------------------------- 136 • PVL-715RST_Convex ---------------------------------------------------------------------------------------------------- 144 • PVL-715RST_Linear ------------------------------------------------------------------------------------------------------ 150 • PLI-1205BX ------------------------------------------------------------------------------------------------------------------ 155 • PLI-2004BX ------------------------------------------------------------------------------------------------------------------ 163 • PLT-308BTP ---------------------------------------------------------------------------------------------------------------- 168
-3-
No. 2B771-144EN*H
• PLT-704SBT ---------------------------------------------------------------------------------------------------------------- 173 • PLT-705BT ------------------------------------------------------------------------------------------------------------------ 180 • PLT-1005BT ----------------------------------------------------------------------------------------------------------------- 187 • PLT-1204BT ----------------------------------------------------------------------------------------------------------------- 195 • PEI-512VX ------------------------------------------------------------------------------------------------------------------- 202 • PET-508MA------------------------------------------------------------------------------------------------------------------ 208 • PET-512MD ------------------------------------------------------------------------------------------------------------------ 215 • PET-805LA ------------------------------------------------------------------------------------------------------------------ 222 • PC-20M ----------------------------------------------------------------------------------------------------------------------- 228 • PC-50M ----------------------------------------------------------------------------------------------------------------------- 229
-4-
No. 2B771-144EN*H
How to read the ACOUSTIC POWER TABLE Symbols used in the table are described below. MI
:
Index of the effect of mechanical shocks by ultrasound that affects the tissue
TIS scan
:
Index of the effect of heating by ultrasound that affects the soft tissue in scanning mode (such as 2D and CDI modes)
TIS nonscan
:
Index of the effect of heating by ultrasound that affects the soft tissue in non-scanning mode (such as M and PW modes)
TIB
:
Index of the effect of heating by ultrasound that affects fetuses and the heads of newborns
TIC
:
Index of the effect of heating by ultrasound that affects the cranial bones of children and adults
Izpta,α
:
Value of the temporal-average intensity at the point where the temporal average intensity calculated considering the tissue attenuation (α = 0.3 dB/cm-MHz) is a 2 maximum or local maximum within a specified region (mW/cm )
Aaprt
:
Transducer aperture area (cm )
Dim. of Aaprt
:
Active aperture dimension (cm) (X: array direction, Y: lens direction)
Pr,α
:
Peak-rarefactional acoustic pressure of pulse wave in water calculated considering the tissue attenuation (α = 0.3 dB/cm-MHz) (MPa)
P
:
Amount of ultrasonic power per unit of time emitted from the transducer (mW) For scanning mode, the amount of ultrasonic power measured with a 1-cm acoustic window
Pα(zs)
:
P calculated considering the tissue attenuation (α = 0.3 dB/cm-MHz) at the depth zs (cm) (mW)
Ita,α(zs)
:
Temporal-average acoustic intensity calculated considering the tissue attenuation 2 (α = 0.3 dB/cm-MHz) at the depth zs (cm) (mW/cm )
zs
:
Depth where min. of [Pα(zs), Ita,α(zs)*1 cm ] is maximized in the depth range z ≥ zbp (cm)
zbp
:
zbp = 1.69*
zb
:
Depth considering the tissue attenuation (0.6 dB/cm-MHz) (cm)
z at max. Ipi,α
:
Depth where the pulse-intensity integral is maximized when calculation considering the tissue attenuation (α = 0.3 dB/cm-MHz) is performed (cm)
deq(zb)
:
Beam width when the beam area at the depth zb undergoes circular approximation (cm)
fawf
:
Acoustic working frequency (MHz)
td
:
Pulse duration (ns)
prr
:
Pulse repetition rate (Hz)
pr at max. Ipi
:
Peak-rarefactional acoustic pressure under conditions where the pulse-intensity integral is maximized in water (MPa)
2
2
(Aaprt)
-5-
No. 2B771-144EN*H
deq at max. Ipi
:
Beam width when the beam area (under conditions where the pulse-intensity integral is maximized in water) undergoes circular approximation (cm)
FL
:
Focal length, or azimuthal and elevational lengths (mm)
Ipa,α at max. MI :
:
α
NOTE:
Tissue conversion value of pulse-average acoustic intensity under conditions where the MI is maximized when the MI is calculated considering the tissue attenuation 2 (α = 0.3 dB/cm-MHz) (W/cm ) Tissue attenuation coefficient (α = 0.3 dB/cm-MHz)
1.
For details related to the acoustic power, refer to the operation manual (fundamentals volume).
2.
The maximum values under fixed conditions in each mode are shown for the MI/TI values given on the following pages. These values may be smaller depending on the operating conditions.
3.
Regarding to the CHI Dual display, the maximum value is included in the MI/TI values given on the following CHI pages.
-6-
No. 2B771-144EN*H
Transducer maximum surface temperature The maximum values of surface temperature test results as specified in IEC 60601-2-37 are shown below. The transducers with a maximum surface temperature of 41°C or more calculated considering the expanded uncertainty are shown below. Transducer name
Maximum surface temperature (°C)
PSI-30VX
47±1
*2
PSI-70BT
41±1
*3
PST-50BT
40±1
*3
PVT-674BT
41±1
*3
PVT-681MVL
41±1
*3
PVT-712BT
41±1
*3
PVT-745BTF / PVT-745BTH
41±1
*3
PVT-745BTV
41±1
*3
PVT-781VT / PVT-781VTE
41±1
*3
PVL-715RST_Convex
41±1
*3
PVL-715RST_Linear
41±1
*3
PLI-1205BX
41±1
*3
PLI-2004BX
41±1
*3
PLT-308BTP
40±1
*3
PLT-704SBT
41±1
*3
PLT-705BT
41±1
*3
PLT-1005BT
41±1
*3
PLT-1204BT
41±1
*3
PEI-512VX
42±1
*3
PET-508MA
40±1
*3
PET-512MD
40±1
*3
PET-805LA
41±1
*3
*1
*1: Values following the "±" mark indicate the expanded uncertainty with a confidence level of 95%, k = 2. *2: Result of still air test *3: Result of simulated use test
-7-
No. 2B771-144EN*H Acoustic Power Table for Track3 ACOUSTIC OUTPUT REPORTING TABLE for TRACK3 Transducer Model: PSI-30BX
Operating Mode: 2D-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
TIB
TIC
0.7
2.7
Non- scan A aprt≦1cm
2
A aprt>1cm
2
1.6
0.7
(MPa) (mW)
1.9 59.4
1.6 66.7
N/A N/A
N/A N/A
1.6 66.7
2.3 61.1
(mW)
48.8
50.2
N/A
N/A
50.2
46.4
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
2.0 2.4 2.8 2.8 0.3 1.3 2.9 1.2 1843 4810 2.1 0.2 29 44 81.5
1.8 2.5 4.7 7.4 0.9 2.3 10.0 1.2 724 2466 2.8 0.3 101 101 126.5
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.8 2.5 4.7 7.4 0.9 2.3 10.0 1.2 724 2466 2.8 0.3 101 101 126.5
2.5 2.5 4.4 4.6 0.6 1.6 10.0 1.2 1155 2461 2.9 0.4 105 100 151.4
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 301 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm)
-8-
No. 2B771-144EN*H Acoustic Power Table for Track3 ACOUSTIC OUTPUT REPORTING TABLE for TRACK3 Transducer Model: PSI-30BX
Operating Mode: 2D/M-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
A aprt≦1cm 0.5
1.6
TIB
TIC
1.3
2.5
Non- scan 2
A aprt>1cm
2
(MPa) (mW)
1.6 6.5
1.4 33.4
N/A N/A
2.3 60.8
1.9 62.3
2.2 44.4
(mW)
5.3
25.1
N/A
45.7
52.3
33.7
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
1.8 2.5 2.4 2.4 0.3 1.6 5.3 1.2 1155 475 1.9 0.2 53 60 73.9
1.8 2.5 4.7 7.4 0.9 2.3 10.0 1.2 724 1584 2.6 0.3 101 101 97.3
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.8 2.5 4.7 7.4 0.9 2.3 10.0 1.2 724 475 4.2 0.3 101 101 339.1
1.9 2.5 3.8 3.9 1.2 1.3 10.0 1.2 1938 475 2.2 0.9 157 135 102.5
2.5 2.5 4.4 4.6 0.6 1.6 10.0 1.2 1155 1901 2.8 0.4 105 100 142.5
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 307 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm) Transducer Model: PSI-30BX
Operating Mode: BDF-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
TIB
TIC
1.3
2.5
Non- scan A aprt≦1cm
2
A aprt>1cm
2
1.6
1.3
(MPa) (mW)
1.9 7.9
1.7 69.7
N/A N/A
N/A N/A
1.7 69.7
1.7 68.3
(mW)
6.2
41.2
N/A
N/A
41.2
40.3
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
2.5 2.5 4.2 4.5 0.4 1.3 6.0 1.2 1901 327 2.3 0.3 60 90 93.7
2.0 2.5 0.5 0.8 1.9 3.8 1.9 1.2 728 1618 1.9 1.5 560 560 83.5
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
2.0 2.5 0.5 0.8 1.9 3.8 1.9 1.2 728 1618 1.9 1.5 560 560 83.5
2.0 2.5 0.5 0.8 1.9 3.8 1.9 1.2 728 1593 1.9 1.5 559 559 83.2
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 536 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm)
-9-
No. 2B771-144EN*H Acoustic Power Table for Track3 ACOUSTIC OUTPUT REPORTING TABLE for TRACK3 Transducer Model: PSI-30BX
Operating Mode: MDF-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
A aprt≦1cm
1.6
TIB
TIC
2.8
2.0
Non- scan 2
A aprt>1cm 1.3
2
(MPa) (mW)
1.9 3.6
0 0
N/A N/A
2.3 117.7
2.1 82.9
1.5 96.4
(mW)
2.8
0
N/A
71.3
60.3
70.1
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
2.5 2.5 4.2 4.5 0.4 1.3 6.0 1.2 1901 87 2.3 0.3 60 90 93.7
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.9 2.5 0.5 0.7 1.9 3.8 1.9 1.2 731 726 2.5 1.5 160 160 149.0
2.5 2.5 5.5 5.9 0.5 1.8 7.5 1.2 752 2564 3.0 0.3 75 100 161.7
2.5 2.5 4.0 6.8 1.1 1.8 10.0 1.2 766 2564 2.4 0.4 105 120 90.4
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 480 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm)
- 10 -
No. 2B771-144EN*H Acoustic Power Table for Track3 ACOUSTIC OUTPUT REPORTING TABLE for TRACK3 Transducer Model: PSI-30BX
Operating Mode: PWD-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
A aprt≦1cm N/A
1.6
TIB
TIC
3.2
2.0
Non- scan 2
A aprt>1cm 1.4
2
(MPa) (mW)
2.1 91.1
0 0
N/A N/A
2.5 129.8
1.6 82.7
2.8 130.7
(mW)
69.2
0
N/A
78.7
65.0
76.1
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
2.2 2.2 3.1 4.5 0.8 1.8 1.5 1.2 1525 1599 2.8 0.4 76 76 152.5
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.9 2.5 0.5 5.0 1.8 3.8 1.9 1.2 792 500 4.8 0.8 132 132 234.0
1.9 2.0 3.1 3.6 0.6 1.8 1.2 1.2 1520 4006 2.0 0.3 62 62 80.4
2.0 2.5 0.5 0.8 1.9 3.8 1.9 1.2 728 500 3.1 1.5 279 279 232.0
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 545 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm) Transducer Model: PSI-30BX
Operating Mode: 2D/PWD-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
A aprt≦1cm 1.1
1.6
TIB
TIC
3.1
2.5
Non- scan 2
A aprt>1cm
2
(MPa) (mW)
2.2 11.8
2.4 10.9
2.6 54.5
N/A N/A
1.2 56.3
2.3 52.6
(mW)
9.6
8.3
39.8
N/A
48.1
40.0
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
1.8 2.5 2.4 2.4 0.3 1.6 10.0 1.2 1155 751 2.5 0.2 53 60 138.2
2.5 2.5 4.4 4.6 0.6 1.6 10.0 1.2 1155 377 3.2 0.4 105 100 177.4
1.2 1.6 3.3 3.4 0.2 3.8 0.8 1.2 727 1599 4.1 0.2 40 40 356.9
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.3 1.7 1.8 1.9 0.4 1.8 0.8 1.2 1545 9006 1.3 0.2 42 42 40.9
2.5 2.5 4.4 4.6 0.6 1.6 10.0 1.2 1155 2000 3.0 0.4 105 100 160.4
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 544 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm)
- 11 -
No. 2B771-144EN*H Acoustic Power Table for Track3 ACOUSTIC OUTPUT REPORTING TABLE for TRACK3 Transducer Model: PSI-30BX
Operating Mode: BDF/PWD-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
A aprt≦1cm 1.2
1.6
TIB
TIC
1.6
2.2
Non- scan 2
A aprt>1cm
2
(MPa) (mW)
2.5 20.1
1.5 35.0
2.3 31.1
N/A N/A
2.0 53.6
1.7 44.8
(mW)
14.7
20.8
22.1
N/A
40.2
26.4
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
1.9 2.0 3.6 3.9 0.5 2.4 1.2 1.2 595 1074 3.4 0.3 60 60 229.3
2.0 2.5 0.5 0.8 1.9 3.8 1.9 1.2 728 1031 1.7 1.5 560 560 67.3
1.3 1.7 3.4 3.5 0.2 3.8 0.8 1.2 733 1200 3.7 0.2 42 42 268.0
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
2.3 2.3 3.1 4.7 0.9 1.8 1.6 1.2 1529 1000 2.6 0.4 82 82 127.0
2.0 2.5 0.5 0.8 1.9 3.8 1.9 1.2 728 1031 1.9 1.5 560 560 84.9
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 408 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm) Transducer Model: PSI-30BX
Operating Mode: B-TDI-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
TIB
TIC
0.6
1.7
Non- scan A aprt≦1cm
2
A aprt>1cm
2
1.6
0.6
(MPa) (mW)
1.9 22.0
2.0 39.3
N/A N/A
N/A N/A
2.0 39.3
2.4 38.5
(mW)
17.9
29.2
N/A
N/A
29.2
28.6
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
2.1 2.4 2.9 2.9 0.3 1.3 3.0 1.2 1847 1748 2.1 0.2 30 45 82.1
1.8 2.5 0.5 4.2 1.8 2.4 1.9 1.2 673 1246 2.9 1.0 170 170 131.8
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.8 2.5 0.5 4.2 1.8 2.4 1.9 1.2 673 1246 2.9 1.0 170 170 131.8
1.8 2.5 0.5 4.2 1.8 2.4 1.9 1.2 673 825 3.4 1.0 280 280 189.4
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 96 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm)
- 12 -
No. 2B771-144EN*H Acoustic Power Table for Track3 ACOUSTIC OUTPUT REPORTING TABLE for TRACK3 Transducer Model: PSI-30BX
Operating Mode: M-TDI-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
A aprt≦1cm
1.6
TIB
TIC
2.6
1.9
Non- scan 2
A aprt>1cm 0.9
2
(MPa) (mW)
1.9 10.1
0 0
N/A N/A
1.9 73.2
2.0 45.1
2.0 45.1
(mW)
8.3
0
N/A
54.3
34.2
34.2
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
2.1 2.4 2.9 2.9 0.3 1.3 3.0 1.2 1847 606 2.1 0.2 30 45 82.1
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.8 2.5 0.5 4.2 1.8 2.4 1.9 1.2 673 1371 2.7 1.0 275 275 117.3
2.5 2.5 4.4 4.6 0.6 1.6 10.0 1.2 1155 1097 2.6 0.4 105 100 117.9
2.5 2.5 4.4 4.6 0.6 1.6 10.0 1.2 1155 1097 2.6 0.4 105 100 117.9
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 326 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm) Transducer Model: PSI-30BX
Operating Mode: CWD-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
A aprt≦1cm 1.2
0.1
TIB
TIC
4.2
3.2
Non- scan 2
A aprt>1cm N/A
2
(MPa) (mW)
0.1 62.5
0 0
0.1 123.3
N/A N/A
0.1 99.0
0.1 112.0
(mW)
57.5
0
105.8
N/A
87.2
97.9
0.6 0.8 0.5 0.5 0.2 2.0 0.2 1.2 ####### 1395 0.1 0.1 20 20 0.5
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
1.1 1.5 1.5 2.2 0.5 2.0 0.6 1.2 1000000000 1395 0.1 0.2 280 280 0.3
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm) (W/cm2)
0.9 1.0 1.2 1.3 1.0 1.3 1.3 2.1 0.2 0.2 2.0 2.0 0.4 0.5 1.2 1.2 ####### ####### 1395 4006 0.1 0.1 0.2 0.3 42 51 42 51 0.5 0.5
The global maximum values for I zpta,α and MI are I zpta,α = 545 mW/cm2 and MI = 0.1 for the above mode. α = 0.3 (dB/MHz/cm)
- 13 -
No. 2B771-144EN*H Acoustic Power Table for Track3 ACOUSTIC OUTPUT REPORTING TABLE for TRACK3 Transducer Model: PSI-30BX
Operating Mode: TDI-PWD-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
A aprt≦1cm N/A
1.6
TIB
TIC
3.4
2.0
Non- scan 2
A aprt>1cm 1.4
2
(MPa) (mW)
2.1 91.1
0 0
N/A N/A
2.5 129.8
1.6 74.1
2.8 130.7
(mW)
69.2
0
N/A
78.7
60.4
76.1
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
2.2 2.2 3.1 4.5 0.8 1.8 1.5 1.2 1525 1599 2.8 0.4 76 76 152.5
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.9 2.5 0.5 5.0 1.8 3.8 1.9 1.2 792 500 4.8 0.8 132 132 234.0
1.6 1.8 2.5 2.7 0.5 1.8 1.0 1.2 1533 4999 1.9 0.3 51 51 76.5
2.0 2.5 0.5 0.8 1.9 3.8 1.9 1.2 728 500 3.1 1.5 279 279 232.0
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 545 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm) Transducer Model: PSI-30BX
Operating Mode: B-TDI/PWD-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
A aprt≦1cm 0.9
1.6
TIB
TIC
1.6
2.3
Non- scan 2
A aprt>1cm
2
(MPa) (mW)
1.9 13.4
2.5 15.2
2.4 33.3
N/A N/A
2.1 30.8
2.1 26.0
(mW)
11.0
11.3
23.6
N/A
26.2
22.7
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
2.0 2.4 2.8 2.8 0.3 1.3 2.9 1.2 1843 1081 2.1 0.2 29 44 81.5
1.8 2.5 0.5 4.2 1.8 2.4 1.9 1.2 673 343 3.5 1.0 280 280 199.6
1.3 1.7 3.4 3.5 0.2 3.8 0.8 1.2 733 1200 3.8 0.2 42 42 285.5
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.3 1.7 1.9 2.0 0.4 1.8 0.8 1.2 1544 1400 2.4 0.2 42 42 135.3
1.1 1.0 1.1 1.3 0.3 1.8 0.6 0.6 1551 1400 2.3 0.2 31 31 136.8
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 336 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm)
- 14 -
No. 2B771-144EN*H Acoustic Power Table for Track3 ACOUSTIC OUTPUT REPORTING TABLE for TRACK3 Transducer Model: PSI-30BX
Operating Mode: CHI-2D-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
TIB
TIC
0.9
2.4
Non- scan A aprt≦1cm
2
A aprt>1cm
2
1.6
0.9
(MPa) (mW)
2.0 30.4
2.1 70.5
N/A N/A
N/A N/A
2.1 70.5
2.1 70.5
(mW)
23.4
52.3
N/A
N/A
52.3
52.3
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
2.3 2.5 3.3 3.3 0.2 1.6 3.4 1.2 920 4810 2.5 0.2 34 47 124.5
1.8 2.5 0.5 4.2 1.8 2.4 1.9 1.2 673 2471 2.9 1.0 280 280 135.9
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.8 2.5 0.5 4.2 1.8 2.4 1.9 1.2 673 2471 2.9 1.0 280 280 135.9
1.8 2.5 0.5 4.2 1.8 2.4 1.9 1.2 673 2471 2.9 1.0 280 280 135.9
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 271 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm)
- 15 -
No. 2B771-144EN*H Acoustic Power Table for Track3 ACOUSTIC OUTPUT REPORTING TABLE for TRACK3 Transducer Model: PSI-30VX
Operating Mode: 2D-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
TIB
TIC
0.8
2.4
Non- scan A aprt≦1cm
2
A aprt>1cm
2
1.6
0.8
(MPa) (mW)
2.1 56.9
2.2 81.4
N/A N/A
N/A N/A
2.2 81.4
2.1 83.0
(mW)
45.5
63.1
N/A
N/A
63.1
65.2
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
1.8 1.5 2.6 2.7 0.3 1.8 0.9 0.9 751 2612 2.5 0.2 42 42 158.6
1.8 1.3 2.6 2.7 0.3 2.0 0.8 0.8 660 5569 2.7 0.2 35 35 197.2
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.8 1.3 2.6 2.7 0.3 2.0 0.8 0.8 660 5569 2.7 0.2 35 35 197.2
1.8 1.3 2.5 2.6 0.3 1.9 0.8 0.8 716 5569 2.5 0.2 35 35 163.9
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 136 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm)
- 16 -
No. 2B771-144EN*H Acoustic Power Table for Track3 ACOUSTIC OUTPUT REPORTING TABLE for TRACK3 Transducer Model: PSI-30VX
Operating Mode: 2D/M-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
A aprt≦1cm 0.5
1.6
TIB
TIC
0.6
1.3
Non- scan 2
A aprt>1cm
2
(MPa) (mW)
2.1 18.5
2.4 29.2
2.4 7.3
N/A N/A
2.2 14.7
2.2 34.4
(mW)
15.2
20.8
5.2
N/A
11.5
27.3
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
1.6 1.5 1.6 1.6 0.2 1.8 0.9 0.9 749 2572 2.4 0.2 20 20 138.0
1.8 1.4 3.0 3.1 0.2 2.7 0.8 0.8 463 2572 3.2 0.2 36 36 292.9
1.8 1.4 3.0 3.1 0.2 2.7 0.8 0.8 463 643 3.2 0.2 36 36 292.9
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.8 1.6 2.7 2.9 0.3 1.9 0.9 0.9 719 643 2.7 0.3 43 43 192.7
1.7 1.2 2.0 2.0 0.2 1.9 0.7 0.7 717 3216 2.5 0.2 26 26 168.7
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 228 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm) Transducer Model: PSI-30VX
Operating Mode: BDF-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
TIB
TIC
1.0
2.3
Non- scan A aprt≦1cm
2
A aprt>1cm
2
1.6
1.0
(MPa) (mW)
2.4 39.6
1.5 76.6
N/A N/A
N/A N/A
1.5 76.6
2.0 84.9
(mW)
29.7
59.4
N/A
N/A
59.4
67.5
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
1.8 1.8 2.9 3.0 0.3 2.3 1.1 1.1 613 2127 3.1 0.2 40 40 259.9
1.8 1.8 3.8 4.0 0.5 2.1 1.1 1.1 1101 1714 2.0 0.4 320 320 91.3
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.8 1.8 3.8 4.0 0.5 2.1 1.1 1.1 1101 1714 2.0 0.4 320 320 91.3
1.8 1.8 3.3 3.5 0.4 1.8 1.1 1.1 1287 1272 2.5 0.3 100 100 157.2
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 238 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm)
- 17 -
No. 2B771-144EN*H Acoustic Power Table for Track3 ACOUSTIC OUTPUT REPORTING TABLE for TRACK3 Transducer Model: PSI-30VX
Operating Mode: MDF-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
A aprt≦1cm
1.6
TIB
TIC
1.9
0.9
Non- scan 2
A aprt>1cm 0.3
2
(MPa) (mW)
2.4 28.9
0 0
N/A N/A
2.0 38.0
2.0 35.7
2.0 38.8
(mW)
21.7
0
N/A
29.5
27.7
30.8
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
1.8 1.8 2.9 3.0 0.3 2.3 1.1 1.1 613 1457 3.1 0.2 40 40 259.9
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.8 1.8 3.2 3.4 0.4 2.1 1.1 1.1 1115 730 2.6 0.3 80 80 178.3
1.8 1.8 2.7 2.9 0.3 2.1 1.1 1.1 1098 1516 2.5 0.2 40 40 177.0
1.8 1.8 3.1 3.3 0.4 1.8 1.1 1.1 1278 724 2.4 0.3 80 80 146.8
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 331 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm)
- 18 -
No. 2B771-144EN*H Acoustic Power Table for Track3 ACOUSTIC OUTPUT REPORTING TABLE for TRACK3 Transducer Model: PSI-30VX
Operating Mode: PWD-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
A aprt≦1cm N/A
1.6
TIB
TIC
2.5
1.1
Non- scan 2
A aprt>1cm 0.4
2
(MPa) (mW)
2.2 38.3
0 0
N/A N/A
1.1 43.5
2.0 46.4
1.1 43.5
(mW)
30.5
0
N/A
34.6
36.9
34.6
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
1.8 1.8 3.1 3.3 0.4 1.8 1.1 1.1 1276 600 2.7 0.3 75 75 183.7
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.8 1.8 3.5 3.7 0.5 1.8 1.1 1.1 1305 1801 1.4 0.4 279 279 39.4
1.8 1.8 2.3 2.4 0.3 1.8 1.1 1.1 1255 2500 2.3 0.2 30 30 144.8
1.8 1.8 3.5 3.7 0.5 1.8 1.1 1.1 1305 1801 1.4 0.4 279 279 39.4
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 545 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm) Transducer Model: PSI-30VX
Operating Mode: 2D/PWD-mode MI
Index label
TIS Scan
Maximum Index Value
Assoc. Acoustic Param.
Other Info
P r,α P Min of [P α(Z s),I ta,α(Z s)] Zs Z bp Zb Z at max. I pi,α d eq(Z b) f awf Dim. of A aprt td prr P r at max. I pi d eq at max. I pi Focal Length I pa,α at max. MI
A aprt≦1cm 0.4
1.6
TIB
TIC
2.5
1.2
Non- scan 2
A aprt>1cm
2
(MPa) (mW)
2.2 3.2
2.0 40.1
2.4 2.3
N/A N/A
2.1 29.1
2.0 42.1
(mW)
2.6
31.5
1.8
N/A
23.8
33.1
(cm) (cm) (cm) (cm) (cm) (MHz) X (cm) Y (cm) (nsec) (Hz) (MPa) (cm) FLx (mm) FLy (mm)
1.6 1.5 1.6 1.6 0.2 1.8 0.9 0.9 1280 300 2.4 0.2 20 20 149.5
1.8 1.4 2.5 2.6 0.3 1.9 0.8 0.8 716 3000 2.3 0.2 36 36 143.0
1.6 1.6 1.8 1.8 0.2 2.3 0.9 0.9 611 300 2.8 0.2 22 22 209.7
N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A
1.6 1.5 1.6 1.6 0.2 1.8 0.9 0.9 1280 3002 2.3 0.2 20 20 137.0
1.8 1.4 2.5 2.6 0.3 1.9 0.8 0.8 716 3000 2.4 0.2 36 36 150.1
(W/cm2)
The global maximum values for I zpta,α and MI are I zpta,α = 543 mW/cm2 and MI = 1.6 for the above mode. α = 0.3 (dB/MHz/cm)
- 19 -