Small Joint Reaming System Cleaning and Sterilization Protocol

1 Page

CLEANING AND STERILIZATION PROTOCOL Small Joint Reaming System CLEANING  Non-sterile products must be carefully cleaned prior to sterilization. Trained personnel must perform cleaning and mechanical inspection prior to sterilization. Compliance is required with the equipment manufacturer’s user instructions (manual and/or machine cleaning, ultrasound treatment, etc.) and recommendations for chemical detergents. Trilliant Surgical recommends the following cleaning and sterilization instructions:  Recommended Manual Cleaning Instructions:  1. Rinse with tap water to remove gross soil 2. Inject water (60mL) into cannulation to remove gross soil 3. Prepare enzymatic detergent (Enzol®) at manufacturer recommendation (1oz/gal) using lukewarm tap water and fully immerse parts 4. Use a soft bristled brush (Spectrum M16 or equivalent) and appropriately sized lumen brush to brush all surfaces 5. Use syringe to inject detergent (60mL) into cannulation 6. Allow articles to dwell in detergent bath for 1 minute 7. Remove parts from bath and rinse using reverse Recommended Automatic Cleaning Instructions: osmosis/deionized (RO/DI) water 1. Rinse with tap water to remove gross soil 8. Fill syringe with RO/DI water (60mL) and flush part 2. Inject water (60mL) into cannulation to remove gross soil cannulation, where applicable 3. Prepare enzymatic detergent (Enzol®) at manufacturer recommendation (1oz/gal) using lukewarm tap water 9. Allow parts to air dry. A clean lint-free cloth may be used to aid in drying and fully immerse parts 10. Visibly inspect for remaining soil on part 4. Use a soft bristled brush (Spectrum M16 or equivalent) and appropriately sized lumen brush to brush all surfaces NON-STERILE PRODUCT STERILIZATION 5. Use syringe to inject detergent (60mL) into cannulation Trilliant Surgical Implant Systems can be packaged non6. Allow articles to dwell in detergent bath for 1 minute 7. Remove parts from bath and rinse using reverse sterile and therefore must be sterilized prior to surgical use. Use of the sterilizer shall comply with the manufacturer’s osmosis/deionized (RO/DI) water 8. Fill syringe with RO/DI water (60mL) and flush part user instructions. The user facility must clean and disinfect instruments prior to sterilization per standard hospital cannulation, where applicable 9. Transfer parts into automated washer (STERIS® procedures. Non-sterile devices are sterilizable by steam Reliance Genfore) for processing using the following sterilization (autoclaving). The following parameters should be followed: parameters: Detergent Type and Concentration  Sterilization Method  Pre-Vacuum Steam  Gravity Steam  Condition  Wrapped*  Wrapped*  Cold tap water  N/A  Temperature  270°F (132°C)  270°F (132°C)  Hot tap water  Time  4 minutes  30 minutes  Enzol® 1 oz/gal  Dry Time  Recommended 50 minutes**  Recommended 40 minutes**  02:00  65.5°C  Prolystica® 2X Neutral 1/8 oz/gal  Rinse 1  01:00  Hot tap water  N/A  Drying  15:00  90°C  N/A  Phase  Recirculation Temperature Time  Pre-wash 1  02:00  Enzyme Wash  02:00  Wash 1  *The system shall be packaged for sterilization by double wrapping in standard central supply wrap (i.e. Bio-Shield® Sterilization Wrap). **Trilliant Surgical has validated the recommended sterilization cycle and dry time for trays. The dry time varies due to load configuration, wrapping method, and material.  10. Visibly inspect for remaining soil on part  This document is controlled by Trilliant Surgical. When downloaded, printed, and/or copied, this document becomes UNCONTROLLED and users should always check Trilliant Surgical’s website, www.trilliantsurgical.com, to ensure they have the latest version.  Trilliant Surgical 727 North Shepherd Drive Suite 100 | Houston, TX 77007 Phone: 800-495-2919 | Fax: 877-778-3864 | www.trilliantsurgical.com  Page 1 of 1 900-06-011 Rev E
File Type: PDF
File Size: 134 KB
File Name: Trilliant Surgical - DJO Surgical - 900-06-011 - Small Joint Reaming System Cleaning and Sterilization Protocol - Rev E.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions