Valleylab
Argon Gas Delivery Unit II Reference Card
Reference Card
2 Pages
Preview
Page 1
Description of Controls, Indicators, and Receptacles Overpressure Indicator - illuminates and an intermittent alarm sounds as the peritoneal cavity pressure approaches the overpressure setpoint; remains illuminated and a continuous alarm sounds when the peritoneal cavity pressure meets or exceeds the setpoint. Peritoneal Pressure Display when the pressure is below 1 mmHg.
Gas Flow Range Selector Low Gas Flow Indicator 1
Tank Indicator, Tank 1
2
Tank Indicator, Tank 2
1
Tank Select Button, Tank 1
2
Tank Select Button, Tank 2 % Argon in Tank 1 or Tank 2
Overpressure Setpoint Display Overpressure Control Knob Handset Tubing or Peritoneal Pressure Tubing Obstruction Indicator - illuminates and an alarm sounds when a gas tubing obstruction occurs. The Gas Flow indicator or the Overpressure Gas Line Purge Button Gas On Indicator
Valleylab™ Argon Gas Delivery Unit II System
Low Pressure Alarm, Tank 1 or illuminates when the tank pressure is below 10% of full tank brief alarm sounds when all available tanks are low.
unit.
supplies power to the
Peritoneal Pressure Tubing Receptacle
Gas Line Receptacle
Refer to the Valleylab Argon Gas Delivery Unit II User’s Guide for further information including warnings and cautions. 1 Prepare the Argon Gas Delivery Unit II for use (refer to the illustration to the right). 2 Prepare the Valleylab Argon Handset for surgery (refer to the illustrations below). 3 Turn on the Argon Gas Delivery Unit II. • Press the power switch to the on position. • If the handset was replaced, press the Gas Line Purge button. 4 Select the electrosurgical generator power settings. • Turn on the electrosurgical generator. • Set the generator power levels.
ESU Interface Cable
Argon Gas Delivery Unit II
5 6
7
8 Gas Flow Indicator
Force FX™ Electrosurgical Generator
adjust the overpressure setting using the Overpressure Control knob. Valleylab Argon Handset • Press the •
button.
Force FX™ Electrosurgical Generator
generator (or if applicable
RF Input Receptacle Flow Rate Display Flow Control Knob
Handset Receptacle
Standard Gas Flow Indicator
Covidien llc, 15 Hampshire Street, MA 02408 USA.
Made in USA Printed in USA. © 2011 Covidien. 1012525
Argon Gas Delivery Unit II
Condition
Possible Cause
Possible Cause
Solution
1. Install gas tanks in the unit. * 2. Open tank valves. 3. Replace tanks in the unit. *
Solution
Possible Cause
Possible Cause
3. Calibration error.
2. Handset particle filter is clogged.
* Refer to the Valleylab Argon Gas Delivery Unit II User’s Guide for further information including warnings and cautions.
3. Argon Gas Delivery Unit II malfunction.
1. Turn the unit off and then on. If the condition is not corrected, refer to Biomedical Engineer or call Covidien Service Department. 2. Turn the unit off and then on. If the condition is not corrected, refer to Biomedical Engineer or call Covidien Service Department. 3. Turn the unit off and then on. If the condition is not corrected, refer to Biomedical Engineer or call Covidien Service Department.
Solution
1. Verify that the handset particle filter is connected to the Gas Line receptacle. Press the Gas Line Purge button. If the condition is not corrected, replace the handset. 2. Verify that the handset tubing is not obstructed. Press the Gas Line Purge button. If the condition is not corrected, replace the handset. 3. Verify that gas line calibration is performed in open air, not in the peritoneal cavity. Press the Gas Line Purge button. If the condition is not corrected, refer to Biomedical Engineer or call Covidien Service Department.
Solution
1. The designated peritoneal 1. Reduce the peritoneal cavity pressure is exceeded. cavity pressure by venting the cannula.
The Peritoneal Pressure display 1. Power up malfunction. shows ER, the Flow Rate display shows an error code number, all indicators flash, front panel controls will not function, and a continuous alarm sounds during a procedure. 2. P ower drop.
Condition
Solution
1. Check length of tubing for pinched section. 2. Tubing kinked. 2. Straighten the tubing. 3. Nozzle on handset plugged. 3. Replace the handset. 4. Damaged handset. 4. Replace the handset.
Possible Cause 1. Compressed tubing.
The Peritoneal Pressure display 1. Handset particle filter is shows ER, the Flow Rate punctured. display shows an error code number, and a brief alarm sounds during the unit self-test or during surgery
Condition
The Overpressure indicator illuminates and an alarm sounds during a procedure.
Solution
1. Check length of tubing for pinched section. 2. Tubing kinked. 2. Straighten the tubing. 3. Nozzle on handset plugged. 3. Replace the handset. 4. Damaged handset 4. Replace the handset.
Possible Cause 1. Compressed tubing.
Low Pressure Alarm - The 1. No gas tanks installed in tank 1 and tank 2 Low Pressure the unit. 2. Gas tank valves not open. indicators flash and a brief 3. Gas tanks are empty. alarm sounds when tank pressure is low in all tanks.
Condition
Peritoneal Pressure Tubing Obstruction Alarm - When a peritoneal pressure tubing obstruction occurs, the Obstruction indicator illuminates, an alarm sounds, the Overpressure indicator flashes and gas flow stops.
Condition
Gas Line Obstruction Alarm When a gas line obstruction occurs, the Obstruction indicator illuminates, an alarm sounds, the Gas Flow indicator flashes and gas flow stops.
Condition
Operating Room Troubleshooting – Alarm Conditions
Tank bar graphs read zero (0) percent (0%).
Condition
Undesirable argon performance.
Condition
No gas flow at the handset
Condition
No RF output at the Argon Handset
Condition
On/Off switch is on, displays not illuminated.
Condition
Solution
Solution
1. Use the handset or Gas On/ Gas Off switch to turn the gas on. 2. Reduce the peritoneal cavity pressure by venting the cannula.
Solution
7. Clear REM alarm. (Refer to generator instruction manual.)
5. Place the electrosurgical generator in ready mode. 6. Increase power levels gradually.*
1. Connect RF link or RF jumper cable to the generator. 2. Plug the three-pin connector into the Argon Gas Delivery Unit II front panel. 3. Check for proper connections at both ends. 4. Turn on the electrosurgical generator or the Argon Gas Delivery Unit II.
Solution
Solution 1. Open gas tank valves. 2. Replace gas tanks.* 3. Attach the tank transducer interface cable to the tank transducer receptacle on the rear panel. 4. Internal cables/gas lines not 4. Refer to Biomedical properly connected. Engineer or call Covidien Service Department. 5. Argon Gas Delivery Unit II 5. Refer to Biomedical malfunction. Engineer or call Covidien Service Department.
1. Both gas tank valves are closed. 2. Both gas tanks are empty. 3. Tank transducer interface cable is not attached to the rear panel.
Possible Cause
1. Place generator in spray or standard coag.* 2. Increase power levels gradually.* 3. Change flow setting.* 4. Straighten handset gas tubing. 5. Reconnect the handset to the Argon Gas Delivery Unit II. 6. Malfunctioning handset . 6. Replace the handset. 7. Eschar on the electrode. 7. Clean electrode. 8. Gas in tank is contaminated 8. Replace with 99.998% pure or not argon. or better argon gas tank. 9. Incompatible generator. 9. Use only Covidien generators specified in the Valleylab Argon Gas Delivery Unit II User’s Guide.
1. Generator set to incorrect coag mode. 2. Coag or cut power set too low. 3. Inappropriate flow setting. 4. Handset gas tubing obstruction. 5. Handset not properly connected.
Possible Cause
2. The peritoneal cavity pressure exceeds the overpressure setpoint. (If the Argon Gas Delivery Unit II has been adjusted to stop gas flow when the overpressure setpoint is exceeded.) 3. Handset gas tubing not 3. Connect handset gas connected to the Argon Gas tubing Delivery Unit II front panel. 4. Argon Gas Delivery Unit II internal cables/gas lines not 4. Refer to Biomedical properly connected. Engineer or call Covidien Service Department.
1. The gas is not turned on.
Possible Cause
2. Handset three-pin connector is plugged into the electrosurgical generator. 3. ESU interface cable not connected. 4. The electrosurgical generator or the Argon Gas Delivery Unit II is not turned on. 5. Electrosurgical generator is in standby mode. 6. Power levels on the electrosurgical generator set too low. 7. REM alarm.
1. RF link or RF jumper cable not connected.
Possible Cause
1. Power entry module fuse(s) 1. Refer to the Valleylab blown. Argon Gas Delivery Unit II Service Manual for fuse replacement instructions. 2. AC line plug not connected 2. Plug AC line into rear of to the Argon Gas Delivery unit. Unit II. 3. Line plug not connected to 3. Plug AC line plug into wall power receptacle. proper AC receptacle.
Possible Cause
Operating Room Troubleshooting – Functional Problems