BVI 6400
QUICK REFERENCE FOR BLADDER VOLUME
1.
2.
If the device is in sleep mode, press any button. The instrument turns on.
Top button
Press the top button in order to enable or disable female mode. Use the female option only for women who have not undergone a hysterectomy.
3.
Have the patient lie in the supine position with abdominal muscles relaxed. 3 cm
4.
Palpate the patient’s symphysis pubis (pubic bone).
5.
Place an ample amount of gel midline on the patient’s abdomen, approximately 1 inch (3 cm) above the symphysis pubis.
6.
Place the instrument on the gel on the patient’s abdomen.
7.
Aim the instrument slightly downward toward the patient’s coccyx (tail bone).
8.
Press and release the Scan button. A scanning symbol
appears in the upper right-hand corner of the LCD screen during the scan.
Scan button
9.
Hold the probe steady until the scan is finished. When you hear the end‑scan tone, the scan is complete.
10.
11.
Using the Results screen, verify that the aim was accurate. NO ARROW
SOLID ARROW
FLASHING ARROW
If no arrow is displayed, then the bladder was completely contained within the scanning cone. The measurement is accurate.
If a solid arrow is displayed, then the bladder was mostly inside the scanning cone. The results are satisfactory, but in order to ensure accuracy, Verathon recommends that you re‑aim in the direction of the arrow and perform the scan again.
If a flashing arrow is displayed, then the scan is off-target, and the bladder was not enclosed within the scanning cone. Re‑aim the probe in the direction indicated by the flashing arrow and perform the scan again.
If you want to save the scan, record a voice annotation as follows: • Press and hold the top button until you hear a tone. • Hold the microphone approximately 6 inches (15 cm) from your mouth and speak clearly. • When you have finished recording, release the top button and listen to your annotation play back. • If you are satisfied with the annotation, do not press any buttons. After several seconds, the BVI 6400 automatically accepts and stores the annotation with the exam.
800.331.2313 | verathon.com
0123
verathon.com https://my.scanpoint.com
BladderScan, the BladderScan symbol, Verathon, and the Verathon Torch symbol are trademarks of Verathon Inc. © 2007, 2010, 2015 Verathon Inc. 0900-0956-11-60
Corporate Headquarters: Verathon Inc. 20001 North Creek Parkway Bothell, WA 98011, USA Tel: 800.331.2313 (USA & Canada only) Tel: 425.867.1348 Fax: 425.883.2896
EC Representative: Verathon Medical (Europe) B.V. Willem Fenengastraat 13 1096 BL Amsterdam, The Netherlands Tel: +31 (0) 20 210 30 91 Fax : +31 (0) 20 210 30 92