BLADDERSCAN
BVI 3000
OPERATIONS & MAINTENANCE MANUAL
BLADDERSCAN
BVI 3000
Operations & Maintenance Manual Effective: March 6, 2015
Caution: Federal (United States) law restricts this device to sale by or on the order of a physician.
CONTACT INFORMATION To obtain additional information regarding your BladderScan system, please contact Verathon® Customer Care or visit verathon.com/contact-us.
Corporate Headquarters: 20001 North Creek Parkway Bothell, WA 98011 U.S.A. 800.331.2313 (US and Canada only) 425.867.1348 Fax: 425.883.2896 verathon.com
Verathon® Medical (Europe) B.V. Willem Fenengastraat 13 1096 BL Amsterdam The Netherlands +31 (0) 20 210 30 91 Fax: +31 (0) 20 210 30 92 verathon.com
0123 Copyright© 2014, 2015 by Verathon Inc. All rights reserved. No part of this manual may be copied or transmitted by any method without the express written consent of Verathon Inc. Verathon, the Verathon torch symbol, BladderScan, and the BladderScan symbol are either trademarks or registered trademarks of Verathon Inc. All other brand and product names are trademarks or registered trademarks of their respective owners. Information in this manual may change at any time without notice. For the most up‑to‑date information, contact Verathon Customer Care or your local representative.
TABLE OF CONTENTS IMPORTANT INFORMATION... 1 OVERVIEW... 1 Product Description... 1 Statement of Intended Use... 1 Essential Performance... 1 Statement of Prescription... 1 Notice to All Users... 1 SAFETY INFORMATION... 2 Biological Safety... 2 Contraindications... 2 Precautions & Warnings... 2 INTRODUCTION... 6 COMPONENTS & FEATURES... 6 Control Unit... 7 Scanhead... 8 Batteries & Battery Charger... 8 SYSTEM COMPONENTS & ACCESSORIES... 9 BUTTONS, ICONS, & CONNECTIONS... 9 Battery Icon... 9 Buttons... 9 SETTING UP...10 ASSEMBLING THE INSTRUMENT...10 Procedure 1. Charge the Batteries...10 Procedure 2. Insert or Remove a Battery...12 Procedure 3. Connect the Scanhead to the Control Unit... 13 Procedure 4. Load The Thermal Paper Roll...14
Operations & Maintenance Manual: Table of Contents
i
CONFIGURING USER SETTINGS...15 Procedure 1. Program the Name...15 Procedure 2. Set the Date...17 Procedure 3. Set the Time...19 Procedure 4. Adjust User Preferences...21 USING THE INSTRUMENT...24 SCANNING A PATIENT...25 Procedure 1. Measure Bladder Volume...25 Procedure 2. Verify Aim & Accuracy...27 Procedure 3. Confirm the Scan Results...29 MANAGING SCAN RESULTS...31 Procedure 1. Add a Patient ID Number (Optional)...31 Procedure 2. Add Notes (Optional)...32 Procedure 3. Print the Scan Result (Optional)...34 Procedure 4. Print a Histogram of Cost Savings (Optional)... 36 CLEANING & DISINFECTING...38 Procedure 1. Clean the Instrument...38 Procedure 2. Disinfect the Instrument...39 MAINTENANCE & TROUBLESHOOTING...40 ANNUAL CERTIFICATION OF CALIBRATION...40 PERIODIC INSPECTIONS...40 Weekly Inspections...40 Monthly Accuracy Check...40 DEVICE REPAIR...40
ii
TROUBLESHOOTING...41 Procedure 1. Run a Self Test...41 Procedure 2. Troubleshoot Power Issues...42 Procedure 3. Troubleshoot an Error Message...42 Procedure 4. Clear a Paper Jam...44 Procedure 5. Test the Printer...45 Procedure 6. Verify Instrument Accuracy...47 DEVICE DISPOSAL...47 WARRANTY...48 Disclaimer of Additional Warranties...48 PRODUCT SPECIFICATIONS...49 SAFETY & PERFORMANCE SUMMARY...49 ACCURACY RANGE...49 COMPONENT SPECIFICATIONS...50 Control Unit & Scanhead Specifications...50 Battery Charger Specifications...52 Battery Specifications...53 ELECTROMAGNETIC COMPATIBILITY...54 Electromagnetic Emissions...54 Electromagnetic Immunity...55 Recommended Separation Distances...57 Accessory Conformance to Standards...57 SYMBOL DIRECTORY...58 GLOSSARY...60
Operations & Maintenance Manual: Table of Contents
iii
IMPORTANT INFORMATION OVERVIEW PRODUCT DESCRIPTION The BladderScan BVI 3000 is a B-mode ultrasonic instrument, portable and battery-operated, intended for the noninvasive measurement of urinary bladder volume. A mechanical sector-scanning transducer provides cross-sectional images of the bladder from twelve scan planes. Based on these images, the BladderScan instrument automatically calculates the estimated bladder volume in milliliters (ml) and displays it on a screen.
STATEMENT OF INTENDED USE The BladderScan BVI 3000 projects ultrasound energy through the lower abdomen of the nonpregnant patient to obtain an image of the bladder, which is used to determine bladder volume noninvasively.
ESSENTIAL PERFORMANCE Essential performance is the system performance necessary to achieve freedom from unacceptable risk. The essential performance of the BladderScan BVI 3000 system is to produce ultrasonic output energy and display numerical values for bladder volume. The system has a passively temperature-controlled transducer assembly.
STATEMENT OF PRESCRIPTION Caution: Federal (United States) law restricts this device to sale by or on the order of a physician.
NOTICE TO ALL USERS The BladderScan BVI 3000 should be used only by individuals who have been trained and authorized by a physician or the institution providing patient care. All operators should read this manual prior to using the instrument. Failure to comply with these instructions may compromise the performance of the instrument.
Operations & Maintenance Manual: Important Information
1
SAFETY INFORMATION BIOLOGICAL SAFETY To date, exposure to pulsed diagnostic ultrasound has not been shown to produce adverse effects. However, ultrasound should be used only by medical professionals when clinically indicated, using the lowest exposure times possible commensurate with clinical utility. The ultrasonic output power of the BladderScan BVI 3000 is not user-adjustable and is limited to the minimum level necessary for effective performance. Data on acoustic output levels can be found in the chapter Product Specifications on page 49.
CONTRAINDICATIONS The BladderScan BVI 3000 instrument is not intended for fetal use or for use on pregnant patients.
PRECAUTIONS & WARNINGS Warnings indicate that injury, death, or other serious adverse reactions may result from use or misuse of the device. Cautions indicate that use or misuse of the device may cause a problem, such as a malfunction, failure, or damage to the product. Throughout the manual, pay attention to sections labeled Important, as these contain reminders or summaries of the following cautions as they apply to a specific component or use situation. To ensure safe and reliable operation for the user and patient, please heed the following warnings and cautions. PRECAUTIONS CAUTION The BladderScan BVI 3000 and related devices may contain mineral oils, batteries, and other environmentally hazardous materials. When the instrument and/or accessories have reached the end of their useful service life, see the section Device Disposal on page 47. CAUTION To avoid paper jams, never fold the end of the paper roll or cut it diagonally or to a point.
CAUTION Failure to adhere to the following when cleaning the control unit or the scanhead may result in permanent equipment damage and void the instrument warranty: • Ensure that you do not immerse either component in water or a cleaning or disinfecting agent. This may permanently damage the instrument. • Do not use Cidex® Plus, as it is not recommended for use with Lexan polycarbonate. • Do not subject any part of the BVI 3000 system to steam sterilization or ethylene oxide sterilization.
2
CAUTION Medical electrical equipment requires special precautions regarding electromagnetic compatibility (EMC) and must be installed and operated according to the instructions in this manual. For more information, see the Electromagnetic Compatibility section on page 54. To maintain electromagnetic interference (EMI) within certified limits, the BladderScan BVI 3000 system must be used with the cables, components, and accessories specified or supplied by Verathon®. For additional information, see the System Components & Accessories and Component Specifications sections. The use of accessories and/or cables other than those specified or supplied may result in increased emissions and/or decreased immunity of the system. The BladderScan BVI 3000 system should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the system should be observed to verify normal operation in the configuration in which it will be used. This device can radiate radio frequency energy and is very unlikely to cause harmful interference with other devices in the vicinity. There is no guarantee that interference will not occur in a particular installation. Evidence of interference may include degradation of performance in this device or other devices when operated simultaneously. If this occurs, try to correct the interference by using the following measures: • Turn devices on and off in the vicinity to determine the source of interference • Reorient or relocate this device or other devices • Increase the separation between devices • Connect the device to an outlet on a circuit different than the other device(s) • Eliminate or reduce EMI with technical solutions (such as shielding) • Purchase medical devices that comply with IEC 60601-1-2 EMC Standards Be aware that portable and mobile radio frequency communications equipment (cellular phones, etc.) may affect medical electrical equipment; take appropriate precautions during operation. WARNINGS WARNING Do not use the BladderScan BVI 3000 instrument on: • A patient who has open skin or wounds in the suprapubic area. • A patient with ascites. • A pregnant patient. WARNING Risk of explosion. If you use the BladderScan BVI 3000 instrument in the presence of flammable anesthetics, the hazard of potential explosion exists.
Operations & Maintenance Manual: Important Information
3
WARNING When using the instrument, be aware of the following conditions that can affect ultrasound transmission and decrease the accuracy of exam results: • Abdominal Surgery- Scar tissue, surgical incisions, sutures, and staples can affect ultrasound transmission and accuracy. Use care when scanning patients who have had abdominal surgery • Catheterization-A catheter in the patient’s bladder may affect the accuracy of the bladder volume measurement; however, the volume measurement may still be clinically useful if it is large (detecting a blocked catheter, for example). • Obesity-Obesity may affect bladder volume measurements. Lift as much abdominal adipose tissue out of the way of the instrument as possible. Apply more pressure to the scanhead to reduce the amount of adipose tissue through which the ultrasound must pass. Accuracy is compromised if you do not obtain an optimal, repeatable image. WARNING Electric shock hazard. Do not attempt to open the system components. This may cause serious injury to the operator or damage to the instrument and will void the warranty. Contact Verathon Customer Care for all servicing needs. WARNING Risk of electric shock or burns. Do not use the BladderScan instrument in conjunction with high‑frequency surgical equipment. WARNING This product may only be cleaned and disinfected by using the approved processes provided in this manual. Cleaning and disinfection methods listed are recommended by Verathon® based on compatibility with component materials. WARNING Availability of cleaning and disinfection products varies by country, and Verathon is unable to test products in every market. For more information, please contact Verathon Customer Care at 1.800.331.2313 or your local representative. For additional contact information, visit verathon.com/contact-us. WARNING Ensure that you follow the manufacturer’s instructions for handling and disposing of the cleaning and disinfection solutions provided in this manual.
4
WARNING Cleaning is critical to ensuring the component is ready for disinfection. Failure to properly clean the device could result in a contaminated instrument after completing the disinfection procedure. WARNING In order to maintain electrical safety, use only the provided medical-approved power supply, battery charger, and batteries. WARNING To reduce the risk of electrical shock, use only the accessories and peripherals recommended by Verathon®. WARNING No modification of this equipment is allowed.
WARNING The battery charger, power supply, and power cables are not intended for patient contact. Ensure six feet (two meters) is maintained between the patient and these components. WARNING Risk of explosion, fire, or serious injury. The BladderScan BVI 3000 system is battery-powered. Failure to note the following when handling the battery may result in serious injury or damage: • Never short-circuit the battery by bringing the battery terminals into contact with any other conductive object. • Never expose the battery to abnormal shock, vibration, or pressure. • Do not disassemble, heat above 60°C (140°F), or incinerate the battery. • Keep battery out of reach of children and in original package until ready to use. • If the battery is leaking or its case is cracked, put on protective gloves to handle it, and discard it immediately. Always dispose of used batteries in compliance with all applicable laws and regulations. • Put insulating tape, such as cellophane tape, on the electrodes during transportation in order to avoid a possible short circuit, fire, or electrical shock.
Operations & Maintenance Manual: Important Information
5
INTRODUCTION COMPONENTS & FEATURES The BladderScan BVI 3000 consists of four main components: the scanhead, the control unit, rechargeable batteries, and the battery charger. In addition, you may purchase optional accessories and additional supplies such as ultrasound gel, thermal paper for the printer, a mobile cart, or a carrying case. For more information about available accessories, see System Components & Accessories on page 9 or contact Verathon®. This section describes the main components and their parts and features. Figure 1.
6
BladderScan BVI 3000 Control Unit and Scanhead
CONTROL UNIT The control unit provides all operating controls for the scanning procedure by means of six buttons. The measured bladder volume and a target-shaped aiming icon are clearly displayed on the LCD screen, helping the operator to achieve accurate measurement results. Figure 2.
Control Unit Components
Battery-eject button Light sensor Battery
Function buttons (5) Power button
Printhead release lever and manual paper-feed wheel
Paper well
Operations & Maintenance Manual: Introduction
Scanhead cable port
7
SCANHEAD The scanhead transmits and receives ultrasound, automatically moving its internal transducer 360º to scan twelve planes and produce a three-dimensional image of the urinary bladder. The scanhead is connected to the control unit by means of a cable. Figure 3.
Scanhead Components
Cable Cable connector
Scan button Patient orientation icon Dome
BATTERIES & BATTERY CHARGER Two NiMH rechargeable batteries are included with the BladderScan BVI 3000. One battery can be recharged in the battery charger while the other is being used to power the BVI 3000. This ensures that there is no instrument downtime due to battery charge. Figure 4.
Battery Charger
Figure 5.
Rechargeable Batteries (2) Battery contacts
Battery charger
Battery contacts
8
Power adapter
SYSTEM COMPONENTS & ACCESSORIES WARNING To reduce the risk of electrical shock, use only the accessories and peripherals recommended by Verathon®. Table 1.
Components & Accessories COMPONENTS
BladderScan BVI 3000 control unit BladderScan BVI 3000 scanhead BladderScan BVI 3000 battery charger ACCESSORIES
Battery, 7.2 volt (V), 2 provided Printer paper BladderScan BVI 3000 In-Service CD, containing the operations & maintenance manual and a video tutorial Acoustic coupling gel Carrying case (Optional) Mobile cart (Optional)
BUTTONS, ICONS, & CONNECTIONS BATTERY ICON The battery icon, located in the upper-right corner of the LCD, indicates the power status of the battery currently installed in your instrument. Table 2.
Battery Icons
ICON STATUS
PERCENT REMAINING
DESCRIPTION
100%
A fully darkened battery icon indicates that the battery is fully charged and ready for use.
75%
A battery icon that is almost full indicates a partially discharged battery.
50%
When the battery icon is half-darkened, the battery is partially discharged. This is the most common display, shown during the majority of the battery’s charge life.
25%
An almost empty battery icon indicates that the battery is nearly discharged. While a few more scans can still be performed, the battery should be replaced at this point.
0%
When the battery is fully discharged, the battery icon is completely clear and the BVI 3000 does not work. The discharged battery must be replaced with a charged one.
BUTTONS The five buttons below the LCD screen change functions based on the content currently displayed on the screen. The button marked with the power symbol turns the BVI 3000 control unit on or off.
Operations & Maintenance Manual: Introduction
9
SETTING UP ASSEMBLING THE INSTRUMENT Procedure 1. CHARGE THE BATTERIES WARNING The battery charger, power supply, and power cables are not intended for patient contact. Ensure six feet (two meters) is maintained between the patient and these components. WARNING Risk of explosion, fire, or serious injury. The BladderScan BVI 3000 system is battery-powered. Failure to note the following when handling the battery may result in serious injury or damage: • Never short-circuit the battery by bringing the battery terminals into contact with any other conductive object. • Never expose the battery to abnormal shock, vibration, or pressure. • Do not disassemble, heat above 60°C (140°F), or incinerate the battery. • Keep battery out of reach of children and in original package until ready to use. • If the battery is leaking or its case is cracked, put on protective gloves to handle it, and discard it immediately. Always dispose of used batteries in compliance with all applicable laws and regulations. • Put insulating tape, such as cellophane tape, on the electrodes during transportation in order to avoid a possible short circuit, fire, or electrical shock. WARNING In order to maintain electrical safety, use only the provided battery charger and batteries. The BladderScan BVI 3000 draws very little power when it is turned off; however, if you do not plan to use your instrument for several weeks, you should remove the battery to prevent it from discharging. The battery that is not in use should be stored in the charger so that it remains fully charged. There is no danger of overcharging the battery. 1. To maintain accessibility, plug in the charger only where it can be easily unplugged.
10
2. Align the battery as shown in the following figure, and then slide the battery into the recess on the top of the charger.
3. Monitor the color indicator lights on top of the battery charger in order to determine the battery’s charging status: • Solid Green-When the battery is low on charge, charging begins in the fast-charge mode. During fast-charge mode, the green light is solid. For a fully discharged battery, fast-charge mode lasts about two to three hours. • Quickly Flashing Green-When the battery reaches 80% of its charge level, the charger starts to “top off” the charge, and the green light flashes quickly. At this point, you may use the battery in the BVI 3000 instrument. • Slowly Flashing Green-If the green light flashes slowly upon inserting a battery, then the battery level is too low for fast-charge mode. The charger slowly charges the battery until the power level is high enough to begin fast-charge mode. When the battery power level is high enough, fast-charge mode begins automatically. • Solid Amber-The amber light indicates that the battery temperature is stabilizing before recharging can begin. This may occur when the battery is taken from a very cold or warm environment, or if the battery is defective. If the light remains amber for over an hour, the battery is defective and must be replaced.
Operations & Maintenance Manual: Setting Up
11
Procedure 2. INSERT OR REMOVE A BATTERY INSERT A BATTERY 1. Ensure the battery has been properly charged according to the instructions in the procedure Charge the Batteries on page 10. 2. If a battery is already in the control unit, remove it according to the instructions in the following section Remove a Battery. 3. Align the charged battery with the battery well as shown in the following figure.
Battery contacts facing up
4. Slide the battery into the well and press until the battery clicks into place. REMOVE A BATTERY 1. Next to the battery well, press the battery-eject button. The battery is released from the control unit.
Battery-eject button
2. Slide the battery out of the control unit. 3. Recharge or store the removed battery in the charger according to the instructions in the procedure Charge the Batteries on page 10.
12
Procedure 3. CONNECT THE SCANHEAD TO THE CONTROL UNIT 1. Align the scanhead cable connector so that the arrow is facing up.
Arrow on scanhead cable connector
2. Press the connector directly into the port until you hear a click.
3. If you want to disconnect the scanhead from the control unit, grip the black plastic ring on the cable connector, and then pull straight out. Do not twist the cable or connector.
Operations & Maintenance Manual: Setting Up
13
Procedure 4. LOAD THE THERMAL PAPER ROLL If the instrument is experiencing a paper jam, see the procedure Clear a Paper Jam on page 44. The BladderScan BVI 3000 has an automatic paper-loading mechanism. When you insert the paper roll correctly, the instrument automatically feeds the paper through the instrument and prepares for printing. 1. On the top of the control unit, open the paper well door.
2. If there is an empty paper roll in the paper well, remove it. 3. If you are loading a new paper roll, cut off the first inch of the new paper. 4. Ensure that the end of the paper roll is cut straight. Do not fold the end of the paper roll, cut it diagonally, or cut it to a point. 5. Turn the BVI 3000 on by pressing the power
button.
6. Insert the end of a new paper roll, with the thermal side down, into the paper input slot. The BVI 3000 senses the presence of the paper and automatically feeds the paper through the instrument. Note: To verify that you are loading the paper with the thermal side down, flick your nail over the paper. If a black mark appears, this is the thermal side.
7. Place the paper roll in the paper well. 8. Close the paper well door. The instrument is ready to print.
14