NeuroTrac™ PelviTone Operation Manual DUAL CHANNEL STIM UNIT
NeuroTrac™ PelviTone
Operators Manual Visit our website: www.VerityMedical.co.uk for detailed application protocols
1
NeuroTrac™ PelviTone Operation Manual
Warnings *
This unit must be used with the guidance of a Physiotherapist or Doctor
*
Type BF equipment
*
Do not insert lead wires into a mains power supply
*
Do not immerse unit into water or any other substance
*
Do not use the NeuroTrac™ PelviTone STIM unit in the presence of a flammable anaesthetic gas mixture and air or with Oxygen or Nitrous Oxide
*
If using rechargeable 9 volt PP3 Nickel Metal Hydride or Ni-Cad batteries be sure to use a CE approved battery charger
*
Never connect the NeuroTrac™ PelviTone STIM device directly to a battery charger or any other mains powered equipment
*
Patient Electrodes are for single patient use only
*
Keep out of reach of children
*
Do not use this stimulator on your face unless under the guidance of a clinician
Revised Issue Date: 10/09/2002 Document Number ETS200-OM002-0 2
NeuroTrac™ PelviTone Operation Manual
Contents Contents
Page
Warnings What is STIM? Contra Indications & Precautions Description of STIM Unit & Functions Quick Start Instructions Lock Button Continence Treatment Programmes Electrode Types & Tips Care & Maintenance Applications Specifications Warranty Clinical References
2 4 5 6 8 10 11 17 18 19 20 21 22
Please contact us about our Vaginal Probe
3
NeuroTrac™ PelviTone Operation Manual
What is STIM? Neuromuscular Stimulation has been used for many years to stimulate muscle and nerve fibres to treat a number of muscle and nerve related conditions. Over the last 30 years numerous clinical trials and papers have been written. The NeuroTrac™ PelviTone is one of a new breed of modern Neuromuscular stimulators which Verity Medical have developed with the Therapist and Patient in mind. Our principle aim is to design products that have high levels of functional use, are sensibly priced, compact and user friendly. The NeuroTrac™ PelviTone is a dual channel device combining several treatment programmes into one unit. Neuromuscular Stimulation is increasingly understood by Therapists and Doctors. There is a better understanding of the mechanisms which exist between nerves and muscles that make it possible to stimulate the neuromuscular system with precise electrical signals. The NeuroTrac™ PelviTone offers precision giving full control of Pulse Widths, Rates, Ramp up times, Work / Rest cycles as well as alternating or synchronous application if two channels are being applied. Customer Care We welcome constructive comments regarding our equipment particularly those that might help us to improve existing features, add new ones or develop new products for the future.
4
NeuroTrac™ PelviTone Operation Manual
Contra Indications & Precautions Before using this equipment you must first seek the advice of your Physiotherapist or Doctor Read this operating manual before using the STIM unit STIM should not be used: *
By patients fitted with a demand style cardiac pacemakers unless so advised by their Doctor
*
During pregnancy [unless medically advised]
*
By patients with undiagnosed pain conditions
*
With patients who have diminished mental capacity or physical competence who cannot handle the device properly
*
On anaesthetised or desensitised skin
*
When driving a vehicle or operating potentially dangerous equipment
*
Do not place electrodes: * Over carotid sinus nerves * Over larynx or trachea * Inside mouth * Over the area of the heart unless so advised by your Doctor * On your facial area unless under strict guidance from a qualified Clinician
*
The patient should use the unit only as prescribed
*
Do not immerse the unit in water or any other liquid
*
Keep unit out of reach of children
*
Only use CE approved vaginal or rectal probes
5
NeuroTrac™ PelviTone Operation Manual
Description of STIM Unit & Functions
Channel A
Connect Probe to Channel A or B
Channel B
CHB
CHA
LCD Display
Lead Wire and Pin Select Programme
Lock
SET
NeuroTrac™ PelviTone
Front
Rear
* PRG button
Selects the desired set programme from P01 - P11 or customised programme PC1 - PC3. Pauses (reducing the intensity (MA) to zero) and escapes from a running programme.
* SET button
Reduces the intensity (MA) to zero and pauses the programme (if a programme is running) and moves the phase one step forward. Displays the menu for programmes PC1 - PC3 and allows the parameters for Time, Work, Rest, Ramp up time, CH.A / CH.B Synchronous or Alternating and delay to be set.
6
NeuroTrac™ PelviTone Operation Manual
Lock symbol
Frequency & Current phase / pulse width of phase phase indicator 3 Hz 250 uS
Programme Total programme time
Low battery indicator
P08
•Work •Rest •Alt •Cont •Mod
Number of phases
3 : 2
Time of current phase
Min
Min
Phase mode
0
0
Channel A milliamps
Channel B milliamps
MA
MA
Example of preset programme Phase being set 3 Hz 250 uS Programme
PC1
•Work •Rest •Alt •Cont •Mod
5 : WK Total programme time adds cumulatively as time added to phases
Min
4
Phase mode
Number of phases Current setting being altered (work in this example)
SEC
Set Work / Rest / Ramp / Delay seconds Example of custom programme 7
NeuroTrac™ PelviTone Operation Manual
Quick Start Instructions 1.
2. 3. 4.
5. 6. 7.
Insert a 9 volt PP3 Alkaline battery. Alternatively insert a rechargeable Nickel Metal Hydride battery {which has a much longer life than the Ni-Cad rechargeable batteries} into the battery compartment. Insert lead wire/s in to Channel A and B, if you are using two channels. Switch on the unit by pressing the on/off button on the front of the unit. Press the PRG button to select one of the pre-set programmes P01 P11 outlined in the tables on page 11 to 16 or PC1 - PC3 for the customised programmes (see page 9 for setting customised programmes). You can press the set button to change to the next phase of the current programme if required. To start the programme, press channel A + and / or B + if you are using both channels, then increase the mA intensity to the desired level. To stop the programme press the on/off button which will turn the unit off or alternatively press the PRG button twice to return to the home screen.
Low Battery Indicator When the battery power is low, the low battery indicator will appear on the screen (shown in the diagram on page 7). When the battery indicator shows one bar, replace the battery. Electrode Disconnection Indicator When an electrode becomes disconnected or when an electrode no longer conducts the electrical current or if the lead wires are faulty, the milliamp level will return to zero and the effected channel will flash on and off. Setting up the Customised Programmes PC1, PC2 or PC3 First, if a programme is running, press the PRG button twice to return to the home screen. Refer to the example of custom programme diagram on page 7.
8
NeuroTrac™ PelviTone Operation Manual
1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14.
Press the PRG button until PC1, PC2 or PC3 is selected. Press and hold the SET button for three seconds, the phase indicator arrow and Hz symbol will flash on and off. Press CH.B +/- to set the fequency between 2 Hz and 100 Hz. Press CH.A +, the µS symbol will flash, press CH.B +/- to set the pulse duration between 50 µS and 450 µS. Press CH.A +, the MIN symbol will flash, press CH.B +/- to set the length of the phase time between 1 and 99 minutes. Set the time to zero to end the programme on this phase. Press CH.A +, the WORK / REST or the CONT symbols will flash, Press the CH.B +/- to select WORK / REST or CONT (continuous). Note: if continuous is selected, the menu will loop back to step 2. Press CH.A +, WK will appear and flash, press CH.B +/- to set the work seconds between 2 and 99 seconds. Press CH.A +, RT will appear and flash, press CH.B +/- to set the rest seconds between 2 and 99 seconds. Press CH.A +, RP will appear and flash, press CH.B +/- to set the ramp seconds between 0.1 and 9.9 seconds. Press CH.A +, AL or SY will appear and flash, press CH.B +/- to select alternating or synchronous current. Note: if alternating is selected, the menu will loop back to step 2. Press CH.A +, DY will appear and flash, press CH.B +/- to set the delay of channel B starting between 0 and 4 seconds after channel A. The menu will now loop back to step 2 and the Hz symbol will flash. To set the next phase, press the set button. The phase symbol will flash over the next phase, continue with step 2 to set this phase. When finished setting the phases, press the PRG button to save the settings and return to the home screen. The programme will be saved permanently.
Setting the phase time of phase 2,3,4 or 5 to zero will cause the programme to end at that phase. Following procedures 1 to 12 can reprogramme a customised programme. If for example there are 5 pre-set phases in one overall programme and only 4 phases are now required, input 0 (zero) into the phase time that is no longer required and press the PRG key to store the new information.
9
NeuroTrac™ PelviTone Operation Manual
Lock Button Frequency & pulse width of phase
Lock symbol (flashes when being unlocked)
10 Hz 200 uS
P08 Programme Hours and minutes unit was used in the locked mode
•Work •Rest •Alt •Cont •Mod
Phase mode
Number of phases
3 : 2 Min
Min
30
20
MA
Average millamps CH.A
Minutes
MA
Average milliamps for CH.B
A "concealed" lock button is included in the NeuroTrac™ PelviTone which allows the clinician to accurately monitor "Home Compliance" of the patent between appointments. It records the time in use and the average intensity (MA). It also locks the customised programmes, stopping them from being altered. To Lock the Unit 1. Select the pre-set or customised programme required. In the case of a customised programme, make sure that the pulse width, frequency, time etc. are set-up correctly. 2. Remove the battery cover and, using a thin rod gently press on the lock button as shown in the diagram on page 6 until you hear a double beep. The unit is now “locked” and cannot be altered until “unlocked”. Note: The lock symbol will appear on the LCD when the unit is “locked”. To Unlock the Unit Remove the battery cover and press the concealed switch with a thin rod until a single beep is heard. Now the LCD will display the average mA used on each channel and the total hours and minutes the unit has been in use as shown in the diagram. To return to normal "unlocked" operation, simply press SET.
10
NeuroTrac™ PelviTone Operation Manual
Continence Treatment Programmes Programme : P01
Pe lvic Floor Pain
Phase 1
Phase time
min
20
Mode Frequency work
Phase 2
Phase 3
Phase 4
Phase 5
Phase 4
Phase 5
Cont Hz
3
Pulse duration
µS
150
Ramp up time
secs
1.0
Ramp down time
secs
0
Work time
secs
Cont
Rest time
secs
0
Alternating Synchronous O verall time
Programme: P02 Phase time
* 20 min
Urge Incontine nce
Phase 1
min
20
Mode
W/R
Frequency work
Hz
10
Pulse duration
µS
250
Ramp up time
secs
1.0
Ramp down time
secs
0
Work time
secs
5
Rest time
secs
5
Alternating Synchronous O verall time
* 20 min
11
Phase 2
Phase 3
NeuroTrac™ PelviTone Operation Manual
Programme: P03 Phase time
Stre s s Incontine nce 1
Phase 1
min
20
Mode Frequency work
Phase 2
Phase Phase Phase 3 4 5
W/R Hz
40
Pulse duration
µS
200
Ramp up time
secs
1.0
Ramp down time
secs
0
Work time
secs
6
Rest time
secs
15
Alternating Synchronous O verall time
Programme: P04 Phase time
* 20 min
Stre s s Phase Incontine nce 2 1 min
Mode
20 W/R
Frequency work
Hz
30
Pulse duration
µS
2 00
Ramp up time
secs
0 .8
Ramp down time
secs
0
Work time
secs
5
Rest time
secs
8
Alternating Synchronous Overall time
* 20 min
12
Phase Phase 2 3
Phase 4
Phase 5
NeuroTrac™ PelviTone Operation Manual
Programme: P05 Phase time
Fre que ncy / Urge 1
Phase 1
min
20
Mode Frequency work
Phase 2
Phase 3
Phase 4
Phase 5
Phase 2
Phase 3
Phase 4
Phase 5
W/R Hz
10
Pulse duration
µS
2 00
Ramp up time
secs
1.0
Ramp down time
secs
0
Work time
secs
5
Rest time
secs
5
Alternating Synchronous O verall time
Programme: P06 Phase time
* 20 min
Fre que ncy / Urge 2
Phase 1
min
15
Mode
Cont
Frequency work
Hz
10
Pulse duration
µS
20 0
Ramp up time
secs
1.0
Ramp down time
secs
0
Work time
secs
Cont
Rest time
secs
0
Alternating Synchronous O verall time
* 15 min
13
NeuroTrac™ PelviTone Operation Manual
Programme: P07 Phase time
Fre que ncy / Urge 3 min
Mode
Phase 1
Phase 2
Phase 3
Phase 4
Phase 5
20 Cont
Frequency work
Hz
10
Pulse duration
µS
2 20
Ramp up time
secs
Ramp down time
secs
Work time
secs
Rest time
secs
Alternating Synchronous O verall time
* 20 min
This programme works continuously with no rest period. It is used in some countries where they have found continuous stimulation can work effectively.
Programme: P08 Phase time
Lack of Se ns itivity min
Mode Frequency work
Hz
Frequency rest
Hz
Pulse duration
µS
Modulation time
secs
Phase 1
Phase 2
Phase 3
Phase 4
Phase 5
3
10
5
4
3
W/R
W/R
W/R
W/R
W/R
3
10
20
30
40
25 0
2 50
2 50
20 0
2 00
Ramp up time
secs
0. 8
0 .8
0 .8
0. 7
0 .7
Ramp down time
secs
0
0
0
0
0
Work time
secs
4
4
4
4
4
Rest time
secs
4
4
4
6
6
*
*
*
*
*
Alternating Synchronous O verall time
25 min
14
NeuroTrac™ PelviTone Operation Manual
Programme: P09
Phase time
Pe lvic Floor Work Out
Phase 1
Phase 2
Phase 3
Phase 4
Phase 5
min
4
15
8
8
10
W/R
W/R
W/R
W/R
W/R
Hz
20
10
20
30
10
Mode Frequency work Pulse duration
µS
25 0
250
2 50
200
2 50
Ramp up time
secs
0. 8
0.8
0 .8
0.6
0 .8
Ramp down time
secs
0
0
0
0
0
Work time
secs
5
5
5
5
5
Rest time
secs
5
5
5
6
7
*
*
*
*
*
Building up Endurance
Phase 1
Phase 2
min
20
Alternating Synchronous O verall time
Programme: P10
Phase time
45 min
Mode
W/R
Frequency work
Hz
20
Pulse duration
µS
25 0
Ramp up time
secs
0. 8
Ramp down time
secs
0
Work time
secs
5
Rest time
secs
5
Alternating Synchronous O verall time
* 20 min
15
Phase 3
Phase 4
Phase 5
NeuroTrac™ PelviTone Operation Manual
Programme: P11
Phase time
R e laxing the Pe lvic M us cle
Phase 1
min
20
Mode
Phase 2
Phase 3
Phase 4
Phase 5
W/R
Frequency work
Hz
2
Pulse duration
µS
2 20
Ramp up time
secs
1.2
Ramp down time
secs
1.2
Work time
secs
6
Rest time
secs
10
Alternating Synchronous O verall time
* 20 min
This programme is to help relax the pelvic muscle. It may be used where the EMG readings are high, in the region of 8 microvolts or more or when the pelvic muscle has been working hard and some fatigue may have resulted. The very low 2 Hz frequency will help to relax the muscle.
W/R = INTERMITTENT WORK/REST CONT = CONTINUOUS
16
NeuroTrac™ PelviTone Operation Manual
Electrode Types & Tips Types Available: *
Normally vaginal or rectal probes are used with this device to achieve effective results. If self-Adhesive Hypoallergenic electrodes are used instead of the internal probes the results tend to be less effective. The typical life span of surface electrodes (if looked after) is 4/6 weeks. We recommend cleaning the skin before use. After use place the electrodes back onto the plasticfilm then enclose them back into the ziptag plastic pouch and store in a cool environment. Types available: VP.001 Internal Vaginal Probe, VP.002 Internal Vaginal Probe, VS.4040 40 x 40 mm (square), VS.5050 50 x 50 mm (square) (recommended for general use), VS.9050 90 x 50 mm (rectangular) VS.10050 100 x 50 mm (rectangular) (recommended for large muscles), VS.13050 130 x 50 mm (rectangular), VS.30 30 mm Diameter (round), VS.50 50 mm Diameter (round), VS.62 62 mm Diameter (round), VS.75 75mm Diameter (round).
A Few Good Tips [Self- Adhesive Electrodes] *
If you find the electrodes will not stick due to oily skin, cleanse the skin with soap and water, then rinse and dry the area around the electrode site. If this does not work, try cleansing the skin with a swab impregnated with alcohol.
*
Clip away hairy skin using scissors; don’t use a razor to remove the hairs!
The electrodes conductive material is water based. If it becomes saturated (e.g. from perspiration), it will loose its adhesive qualities. After use leave the electrodes face up overnight to dry out. At some point the electrodes will become dry. Moisten the adhesive surface with a few drops of water, and apply back onto the plastic film overnight. This procedure will give you a few more days of electrode life.
17
NeuroTrac™ PelviTone Operation Manual
Care & Maintenance Control Unit: * *
Wipe the surface once a week with a damp cloth or antiseptic wipe Do not use cleaning sprays or alcohol based cleaning solutions
Battery: * * * *
Check periodically for any discharge from the battery Remove battery completely from unit if not in use for any extended period of time (typically one week) Low battery indicator of 6.9 volts shown on LCD display. When flashing change battery for a new one Preferably use a heavy duty or PP3 alkaline battery
Lead Wires * * * *
The lead wires should be handled carefully and never stretched, as this can cause the stimulation to function below normal standards or not at all Examine lead wires before each treatment for loose connections or damage Avoid stretching and twisting the lead wires Store the lead wires carefully after each use
Self-Adhesive Electrodes * *
Check the short connectors have not become separated from the electrodes Replace electrodes onto plastic film after use. If they drop onto the floor debris will adhere to conductive gel rendering the electrodes ineffective
Vaginal / Rectal Probes *
Check the connectors have not become separated from the probe Caution: Static electricity may damage this product
NOTE:
Only Verity Medical Ltd or appointed distributors /importers are approved to undertake servicing.
18
NeuroTrac™ PelviTone Operation Manual
Applications *
Promotes continence
*
Increases muscle strength
*
Maintains or improves range of movement
*
Increases and improves the blood supply to the muscle in cases of intermittent caudication
*
Reduces pain
19
NeuroTrac™ PelviTone Operation Manual
Specifications STIM 1.
Dual channel: individually isolated circuits.
2.
Amplitude: 0-90 mA; indication only; actual mA will tend to be less than indicated due to electrode impedance.
3.
Type: Constant Current.
4.
Waveform: Symmetrical, rectangular bi-phasic with zero DC current.
5.
Selectable pulse width: 50µS – 300µS [2% accuracy].
6.
Pulse Rate selection: in the continuous mode 2 – 100 Hz [2% accuracy].
7.
Time duration of the treatment selectable: 1 minute to 90 minutes.
8.
Low Battery Indicator: If the battery goes below 6.9 volts +/- 0.2 volts the battery symbol will flash on/off once every second.
9.
Open Electrode Detect: If an open circuit is detected at the output of channel A or B the output current will be reset at zero.
10.
Ramp up time 0.1 - 9.9 seconds.
11.
If the battery voltage is below 6.6 (+/- 0.2) volts the unit will not turn on.
12.
Physical dimensions: 134 x 69 x 29.7 mm.
13.
Weight: 0.18 KG with battery.
14.
Environmental conditions for storage & transport: -10 to +50 degrees Centigrade 0-90% Humidity.
20