VIMAP TECHNOLOGIES

VIMAP VMX-1010A Check List Rev R2 March 2012

Check List

4 Pages

Número:  F-00047  Fecha Creación:  Check list  Título: Revisión:  12/03/2012  Página:  R2  Rev. Fecha:  Performance Check list :  1 de 4 16/03/2012  remark  Power switch  The front panel green led must be blinking in standby mode and lit continuously in working mode (when switch is on I position)  Standby switch  When the green led is on continuously, the display screen is displaying information. When the green led is blinking, the exhaust valve connector is open. (release gas output)  CO2 bottle connector  No leak with a bottle connected  Startup test  Without gas connected (or bottle closed) -  Screen will show the sequence 1 (see appendix1) A beep is audible (buzzer test) for a short time The warning led is lit for a short time (Led test)  Run /stop Button  Press the Run or stop button: nothing occurs until calibration is done (gas is present)  Gas input: provide gas to the device  Calibration message appears  Run button  Press the run button without tubing connected: warning message: pipe missing must appear (tubing detection test). Connect tubing (only gas patient output) and press run: same message Connect tubing (gas release output) and press run: insufflation must start. Note: with a pressure bottle below 10 bars, the run is disable.  Stop button  Press Stop : insufflation must stop, , the exhaust valve connector is open. (release gas output)