Instructions for use
84 Pages
Preview
Page 1
5380
5380 K
Instructions for use Models 5380, 5380 K
Help Hyperlinks
Full-text search for a term
Other interactive functions
Hyperlinks are cross-references in the text. They are marked with the symbol. Entries in the table of contents also have hyperlink functionality.
•
The Adobe Reader and all Adobe Acrobat program versions offer the experienced user a whole raft of highly useful functions for interactive working with documents. For more information, see the relevant user manuals or online help.
•
Clicking takes you to the relevant page.
•
Clicking in the Acrobat navigation bar takes you back to the start page.
Clicking window.
opens the "Search"
You can search for any term in this Acrobat document or optionally search all of the Acrobat files stored on your computer or those that are accessible.
Display table of contents
Contents
Jump to Help page (this page)
2 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Help V1.0 (06.03.2009)
Version, Imprint, Type plate Instructions for use G67 Version 1.0 (06.03.2009) for Völker healthcare bed models 5380, 5380 K built after April 2005 © by Völker AG 2009 Völker AG Wullener Feld 79 58454 Witten GERMANY Tel.: +49 2302 96096-0 Fax: +49 2302 96096-16 E-mail: [email protected] Internet: www.voelker.de All rights reserved. Reproduction - even in part - is not permitted. We reserve the right to make changes to reflect technical advances. The contents of this document are subject to change without prior notification.
Type plate
The type plate is located on the inside of the head panel. Raise the rear section to read the type plate. For further information on the type plate, see Appendix 66.
Customers are advised to contact the responsible area sales manager before placing an order.
Contents
3 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Help V1.0 (06.03.2009)
Table of contents 1/2 Help 2 Version, Imprint, Type plate 3 Table of contents 4 Foreword 6 Notes 7 General information 8 Designated purpose 9 General regulations and user training / instruction 10 General safety notes 11
Activation 29 General operating instructions 30 Preparation 31 Electrical activation 32 Using the battery pack (option) 3 3 Taking out of service 34 Functional check 35
Operation 36 Key lock-out 37 Brake 38 Side rails 39 Functional description 15 General safety notes 39 Overview 16 Side rails 40 Model 5380 16 Back rest section 42 Model 5380 K 17 Upper leg section 43 Hand control with hook 18 Lower leg section 44 Transverse hand control (optional) 9 Lying surface height 45 Personnel keypad with complete Anti-Trendelenburg and lock-out (option) 20 Trendelenburg positioning 46 Personnel keypad with individual Comfort seating position 47 lock-out (option) 22 Cleaning and disinfection 48 Trapeze bar and accessory holders, Cleaning 49 accessory rail 24 Wipe and spray disinfection 49 Versions and options 25 Accessories 27
Contents
4 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Maintenance 51 Staff training 52 Safety notes 52 Maintenance schedule 53 Technical checks 54 Visual inspection 55 Functional check of the side rails 55 Functional check of the brakes 56 Functional check of the motors 56 Mains connection cable 56 Cabling 56 Housing 56 Mechanical check 56 Changing the battery 56 Measurement as per VDE 0751-1 57 Checking the grab handle 57 Other accessories 57 Troubleshooting 58 Table of faults 59 Service points 61
Help V1.0 (06.03.2009)
Table of contents 2/2 Appendix 62 Symbols used 63 Technical data 64 Classification 65 Type plate 66 Service life / disposal 68 Manufacturer's declarations, forms, electronic instructions for use 69
Contents
5 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Help V1.0 (06.03.2009)
Foreword We are delighted that you have chosen Völker healthcare beds. We are also grateful for the trust you have placed in our company and our products. This step is undoubtedly the result of extensive considerations and examinations of the requirements that you want to make on new healthcare beds based on your previous experiences. You clearly had good reasons to choose Völker healthcare beds. We promise you that Völker healthcare beds will not disappoint you.
It is not without good reason, after all, Völker healthcare beds are now regarded worldwide as extremely innovative medical products. This relates not just to the design principle, which has been completely redeveloped by Völker, but also to the range of product advantages that have been checked and improved time and time again based on their practicality in use. These advantages now benefit the patient in terms of comfort, but also make the everyday work of care personnel easier. Now, every healthcare bed has product features that are of practical benefit to their users. However, as far as we know, none of them offers the range of advantages that a Völker healthcare bed does.
Völker healthcare beds not only look great, but they also offer functions that can be controlled or adjusted mechanically, although most of them are controlled or adjusted using electric motors or electronic components. If you purchase these beds, the responsibility for their correct and appropriate use is transferred to you. Consequently, we strongly advise that you consult the enclosed instructions for use to learn about the technical features, handling and use of all the functions. We wish you every success with Völker healthcare beds
Heinrich Völker Chairman, Völker Aktiengesellschaft
Contents
6 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Help V1.0 (06.03.2009)
Notes The Notes section contains information on the designated purpose of the product, as well as general safety notes.
Contents General information 8 Designated purpose 9 General regulations and training / instruction of users 10 General safety notes 11
Contents
7 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Help V1.0 (06.03.2009)
Notes | General information General notes
Standard design
Warranty and liability
You have purchased a healthcare bed from Völker AG. This healthcare bed has been built in accordance with the applicable national and international standards and the regulations reflecting the current state of the technological art.
The standard design of the bed can be supplied with various options. A description of these options can be found in the section entitled Versions and options 25.
Völker AG is liable for any faults or failures, not including further claims arising in the context of the warranty obligations detailed in the main agreement. Claims for compensation, for whatever legal reason such claims may be raised, will not be entertained.
Völker healthcare beds satisfy all the requirements in terms of safety and functionality. They are tested in accordance with international standards and bear the CE mark, which documents the beds' compliance with essential requirements for medical products.
Copyright protection These instructions for use may only be transferred to third parties with the written consent of Völker AG. All documents are protected under copyright.
We accept no liability for damage and operational faults caused as a result of misuse and/or non-observance of these instructions for use.
Please read the general safety notes 11. Please also note (with particular attention to any warranty claims) the further notes on the following pages.
Contents
We reserve the right to make technical modifications as part of the further development of the healthcare beds which form the subject of these instructions for use.
The portrayal of the accessories does not necessarily match the actual product.
8 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Help V1.0 (06.03.2009)
Notes | Designated purpose Correct use
Inappropriate use
•
The Völker healthcare bed models 5380 and 5380 K are intended for the laying down and care of patients in care institutions and hospitals.
Inappropriate use can cause danger. These include, but are not limited to:
Any attempt to move the bed while it is in braked position;
•
Use of the bed to transport people;
•
Incorrect actuation of electrical functions and uncontrolled positioning;
•
Use of the bed for transport with a vehicle;
•
Operation of the healthcare bed by the occupant without having received prior instruction in how to do so;
•
Overloading of the bed beyond the specified safe working load.
•
Use of electrical equipment on the bed that is not intended for such use (subject to the operator's obligation to exercise care);
•
Pulling on cables to move the bed;
•
Loosening electrical plug connections by pulling on the cable;
•
Use of the bed on a sloping base with more than ten degrees of inclination (the bed's brakes are designed for an angle of inclination of no more than ten degrees);
The bed is intended for use by people over the age of 12 or who are taller than 146 cm. The bed's safe working load is 170 kg. The maximum permissible weight of the individual is derived by deducting the weight of the mattress, the trapeze handle and other accessories from this. Any use of the Völker healthcare bed other than for the purpose intended excludes the company from any possible liability.
Contents
9 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
CAUTION If, in an emergency situation, it is impossible to avoid putting children under the age of 12 or people who are less than 146 cm tall in the bed, protective covers must be placed on the side rails. This also applies to the use of the bed by weak or confused patients. The use of the bed for children under the age of 8 is absolutely forbidden.
Help V1.0 (06.03.2009)
Notes | General regulations and user training / instruction General regulations
User training
Other requirements
The healthcare bed must only be operated and used in accordance with its designated purpose, in line with the conditions of the Medical Products Directive (MPG) and approved legislation pursuant to this, the generally acknowledged rules of technology and the stipulations of occupational safety and accident protection guidelines. The healthcare bed must not be operated in a faulty state that could endanger its occupant, care personnel or third parties.
The healthcare bed may only be operated by individuals whose training or understanding and experience offer surety for correct handling (medical products directive).
Whoever is in charge of the activation, operation or preparation of the bed must have been given a copy of these instructions for use (in printed or electronic form) and have read them.
User instruction
To avoid operating errors and to safeguard the smooth operation of the bed, care personnel must always have access to the safety notes below.
The thorough induction of care personnel in the operation of the bed can be provided by Völker or its representative at the customer's request. Attendance of such training by care personnel can be certified and confirmed by Völker using the form provided for this purpose, specifying the name, date and signature. Patients must be instructed in the use of the bed before care personnel hand over the hand control to them.
Contents
10 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Help V1.0 (06.03.2009)
Notes | General safety notes 1/4 Warning symbols Information marked with this symbol must be read and its content strictly observed.
Before first activation
DANGER represents an immediate threat of danger that can cause serious physical injury or death.
Before the bed is activated for the first time, care personnel must be instructed in the handling of the bed using the instructions for use. The potential dangers that can arise despite correct operation of the bed must also be pointed out in full.
WARNING represents potentially dangerous situations that can lead to serious physical injury or death. CAUTION represents potentially dangerous situations that can cause slight physical injuries. NOTE warns of potential damage to objects or property.
Contents
Before the healthcare bed is put into action for the first time, care personnel must read these instructions for use in full and with care.
Before and during use Before each use of the bed, the user must be sure that the bed is in a good, safe condition and will ensure safe use (Functional check 35).
11 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Position of the care bed CAUTION To avoid injuries caused by falling, we recommend (except while care is being given) that the bed is generally set at its lowest position when the castors are raised. While the bed is in its lowest position, it is automatically braked. Transporting the bed CAUTION The healthcare bed is not designed for transporting people. When transporting the bed, it must always be ensured that the mains connection cable is not touching the floor, the lying surface is in the horizontal position and that the lying surface height is at least 35 cm. The bed must only be moved over solid ground. Never attempt to push it over obstacles of over 1 cm in height. The maximum
Help V1.0 (06.03.2009)
Notes | General safety notes 2/4 angle of inclination of the floor must not exceed 10°. Securing the bed CAUTION "Risk of accident" If the bed is not being transported, the brake must always be applied, since the bed may be required as a support for when the occupant stands or lies down. If the brake is not applied and the bed rolls away, the patient may fall. Once the brake has been applied, check that the bed is actually properly parked. The bed can be in a non-fullybraked status even after first activation or reactivation, and consequently it must be checked that the castors are correctly raised.
Contents
One-sided load on the bed NOTE In order to prevent onesided loads on the bed, it must not be used as a seat for persons other than the occupant (i.e. visitors must not sit on the edge of the bed).
•
Ensure that the side rails (or parts thereof) are either fully raised and locked in position or completely lowered.
•
It must be ensured that the occupant, when the electrical lying surface adjustment is actuated, does not come into contact with the side rail elements. It is also important to ensure that no part of the body is sticking out through the side rails.
•
If the side rails are used with a person whose psychological condition would deem their use necessary, then it must be ensured that the hand control is kept out of their reach or its functions are blocked. It is strongly recommended that side rail covers are used.
Side rails WARNING "Risk of entrapment" In the case of occupants whose physical or mental condition would deem it necessary to use side rails to protect them from falling out of bed, the following safety measures must be observed: •
The legal permissibility of using side rails must be ascertained.
•
The side rails may only be operated by trained care personnel.
12 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Help V1.0 (06.03.2009)
Notes | General safety notes 3/4 •
Height adjustment If these safety measures are not observed by care personDANGER "Risk of entrapment nel, injuries can be caused to between the lower frame and/ hands, knees, fingers, feet, legs or floor and the bed frame when and hips, along with haematolowering the bed" mas and other injuries as a It must be ensured that, during result of entrapment. In children adjustment procedures, no peoor people who are less than ple, limbs, pets, bed linen or other 146 cm tall, non-observance objects are trapped between the of these guidelines can lead to bed frame and the lower frame death! and/or floor.
WARNING "Danger of injury" If the side rails are damaged, the bed must not be used and must be repaired.
Contents
DANGER "Risk of movement" If any movement of the bed could represent a danger for the occupant, all functions must be blocked.
13 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Accessories WARNING "Risk of injury" Only original Völker accessories should be used! Third-party accessories must be subjected to testing before use. Using lifting devices WARNING "Risk of injury" No lifting device must be fastened directly to the bed (patient transport, repair). Patient lifters must also not be used, as a result of the design. The lifting devices specified are devices that can be attached to the bed for transport purposes.
Help V1.0 (06.03.2009)
Notes | General safety notes 4/4 Using oxygen equipment DANGER "Risk of fire" Do not use any oxygen equipment other than that which is administered via nasal prongs or masks. Do not use the bed in a room with a potential explosion risk. (Provided it has been excluded (e.g. based on information in the instructions for use of the equipment being used) that the use of the equipment may cause the O2 concentration to rise to such a degree that there is an explosion risk (even when there is no fault), then the equipment can be used). Rail spacers When using the rail spacer, please read the separate instructions for use for this accessory. During technical checks, the rail spacers should also be checked to ensure they are suitable for the size of side rail used.
Contents
Cleaning and disinfection In order to maintain consistent functioning, the healthcare bed should be cleaned, disinfected and tested at the soonest possible opportunity following each use, so that it can be reused immediately without risk. Dangers can arise from the incorrect cleaning / disinfection 48 of the bed. Maintenance and repair Anyone responsible for carrying out maintenance and repair work must at least have read the safety notes and the service manual and be qualified in accordance with MPBetreibV §§ 4 and 6.
The technical check must be carried out at least once a year and after every lengthy period of non-use. Any discernible damage, such as signs of wear and tear, loose screws or breaks/fractures must be eliminated immediately. Electromagnetic / static interference The healthcare beds in model series 5380 and 5380 K satisfy the EMC requirements in accordance with the law on medical products (MPG). The basis for testing is standard EN 60601-1-2.
Once maintenance work or repairs have been carried out, a technical check 54 must be carried out. During this check, it must be determined that the bed can be used in accordance with the specifications without risk to the occupant, user or third party.
14 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Help V1.0 (06.03.2009)
Functional description The features of the Völker healthcare bed and its function are set out in the section Functional description.
Contents Overview 16 Model 5380 16 Model 5380 K 17 Hand control with hook 18 Transverse hand control (optional) 19 Personnel keypad with complete lock-out 20 Personnel keypad with individual lock-out 22 Trapeze bar and accessory holders, accessory rail 24 Versions and options 25 Accessories 27
Contents
15 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Help V1.0 (06.03.2009)
Functional description | Overview | Model 5380 Hand control, transverse
Adjustable side rails for back section, both sides
Adjustable back section
Adjustable lower leg section Head section Foot section
Seat section, fixed
Brake lever
Contents
Adjustable side rails for lower leg section, both sides
Height adjustment
Adjustable upper leg section
16 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Accessories rail for holding accessories, on both sides
Help V1.0 (06.03.2009)
Functional description | Overview | Model 5380 K Hand control, transverse
Adjustable back section
Adjustable side rails for back section, both sides
Adjustable lower leg section
Model 5380 K Foot section (can be quickly removed, see Fig. on right)
Head section (can be quickly removed, see Fig. on right)
Seat section, fixed
Brake lever
Contents
Adjustable side rails for lower leg section, on both sides
Height adjustment
Adjustable upper leg section
17 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Accessories rail for holding accessories, on both sides
Help V1.0 (06.03.2009)
Functional description | Hand control with hook Reverse: Back rest section up
Back rest section down
Upper leg section up
Upper leg section down
Anti-Trendelenburg positioning 1
Lying surface up
Auto-Contour 2 (optional)
Hand control unlocked
Lying surface down
Hand control locked.
1
Head end raised
2
Back and upper leg section raised
Contents
WARNING When actuating motorised adjustments with the side rails raised, it must be ensured that the occupant does not have any contact with the side rails, and that no parts of his or her body or that of other people are sticking through the side rails or are trapped between the lying surface and the lower frame and/or floor! 18 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Help V1.0 (06.03.2009)
Functional description | Transverse hand control (option) Lying surface Anti-Trendelenburg positioning 1 up
Upper leg section up
Back rest section up
Reverse:
Hand control unlocked
Lying surface down
Auto-Contour 2 (optional)
Upper leg Back rest section down section down Hand control locked.
1
Head end raised
2
Back and upper leg section raised
Contents
WARNING When actuating motorised adjustments with the side rails raised, it must be ensured that the occupant does not have any contact with the side rails, and that no parts of his or her body or that of other people are sticking through the side rails or are trapped between the lying surface and the lower frame and/or floor! 19 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
Help V1.0 (06.03.2009)
Functional description | Personnel keypad with complete lock-out 1/2 (option) Anti-Trendelenburg positioning 1
Lying surface up3
Back rest section up
Upper leg section up
Reverse:
Keypad unlocked
Trendelenburg positioning 2
1
Head end raised
2
Head end lowered
3
The personnel keyboard can be equipped with an automatic function (double-click) on a country-by-country basis (see next page).
Contents
Lying surface down3
Back rest section down
Upper leg section down
WARNING When actuating motorised adjustments with the side rails raised, it must be ensured that the occupant does not have any contact with the side rails, and that no parts of his or her body or that of other people are sticking through the side rails or are trapped between the lying surface and the lower frame and/or floor! 20 of 84
Völker care bed models 5380, 5380 K built after April 2005 – Instructions for use G67
In this position, the keypad and hand control are completely locked.