Diagnostic Instructions
PocketScopes™
Otoscope
Ophthalmoscope
Operating Instruction Manual Retinoscope
Transilluminator
1300232Cfinal
6/28/00
10:07 AM
Page 8
PocketScope Retinoscope
PocketScope Otoscope
Retinoscope Features • External focusing sleeve allows for easy adjustment regardless of what size hand or how instrument is held. Continuous 360˚ rotation. • Dust-free optics to keep instrument clean longer. • Fixation cards that easily attach for use during dynamic retinoscopy. • Collapsible head for compact size.
Throat Illuminator (21111 only) To convert otoscope to throat illuminator, unscrew otoscope head in counterclockwise direction. NOTE: Illuminator section can become hot. Avoid more than momentary direct skin contact. To turn instruments on and off, and vary intensity rotate switch at base of instrument head. As intensity increases, green band gets thicker.
Patents: US 5,189,449 and 5,459,535
NOTE: The retinoscope must be telescoped out by pulling in an outward direction. The retinoscope can only be used if this operation is performed. Practitioner Side
Patient Side Rubber brow rest Dust-free optics
Throat Illuminator
Focusing sleeve
6
On/Off Rheostat Switch 7
1300232Cfinal
6/28/00
10:07 AM
Page 10
PocketScope Ophthalmoscope 1. Rubber Brow Rest. 2. Aperture selection dial with continuous rotation: Allows access to microspot, small, large, fixation, slit and cobalt blue. 3. Lens Selection Dial - When used with multiplier lever, allows single diopter steps through 48 lens combinations (from +22 to –25 diopters). 4. Multiplier Lever - Allows fast movement through the range via 16 diopter shifts. When used with lens selection dial, allows single diopter steps through 48 lens combinations. 5. Two-position filter switch: Engage either open or red-free filter. 6. Larger spot size for wider field of view. 7. Sealed optics. 8. Diopter Display Window - Displays direct reading of diopter values (positive values displayed in green; negative values displayed in red).
To move through the range on the negative (–) side: Rotate lens selection dial in negative (counterclockwise) direction. At –9 diopters move lens dial counterclockwise again. Then move multiplier lever to “–” position. Continue moving lens dial in same counterclockwise direction to access –10 to –25 diopters. Practitioner Side Rubber brow rest Lens selection dial
Lens multiplier
Diopter display window
Patient Side
Filter
Lens Instructions: To move through the range on the plus (+) side: Rotate lens selection dial in positive (clockwise) direction. At +6 diopters move lens dial clockwise again. Then move multiplier lever to “+” position. Continue moving lens dial in same clockwise direction to access +7 thru +22 diopters. 8
Red-Free Filter position Two-position Filter switch
Open spot position
Aperture selection dial
9
1300232Cfinal
6/28/00
10:07 AM
Page 12
PocketScope Transilluminator Transilluminator Features • Fiber Optic transmission for bright cool light transmission. • Curved design to fit into tight areas. Transilluminator Attachment Remove the retinoscope by turning the instrument in a counterclockwise direction. This will disconnect the retinoscope. Connect the transilluminator by turning the instrument in a clockwise direction. Light pipe can be rotated to locate optimum orientation.
Lamp Replacement Retinoscope Lamp Replacement 1. Remove (unscrew) retinoscope head. 2. Grasp end of lamp and pull out (use nail file or similar object if necessary). 3. Insert replacement lamp by carefully aligning pin on lamp shaft with slot in retinoscope. Push lamp in until firmly seated.
Instructions for attaching the transilluminator
10
11
1300232Cfinal
6/28/00
10:07 AM
Page 14
Otoscope Lamp Replacement (21111 only) 1. Remove (unscrew) otoscope head. 2. Unscrew lamp holder in counterclockwise direction and remove from handle. 3. Grasp end of lamp and pull out of holder. 4. Insert new lamp by reversing above process. NOTE: If lamp surface is touched or becomes smudged, wipe clean with a soft cloth. This will assure longer lamp life and higher output.
Ophthalmoscope Lamp Replacement 1. Unscrew ophthalmoscope head in counterclockwise direction. 2. Grasp end of lamp and pull out (use nail file or similar object if necessary). 3. Insert replacement lamp by carefully aligning pin on lamp shaft with slot in ophthalmoscope head. Push lamp in until firmly seated. NOTE: If lamp surface is touched or becomes smudged, wipe clean with a soft cloth. This will assure longer lamp life and higher output.
Align pin to slot in ophthalmoscope head 12
13
1300232Cfinal
6/28/00
10:07 AM
Page 16
Transilluminator Lamp Replacement 1. Place thumb on knurled ring at base of instrument. Unscrew transilluminator in a counterclockwise direction. 2. Grasp end of lamp and pull out (use nail file or similar object if needed). 3. Insert replacement lamp (No. 03400). Push lamp in firmly.
3. Locate and level prepared back panel within 6 feet of an electrical outlet. 4. Mark the location of the two holes and drill holes to suit anchors provided: Hard materials (concrete) require larger holes (1/2 inch). Soft materials (plaster board) require smaller holes (3/16 inch). 5. Set anchors flush with the wall surface. Solid wood walls require no anchor. Thin panel walls require an anchor designed specifically for these materials. 6. Attach back panel to wall. Slide charger down in place until unit snaps into final locked position. 7. Plug transformer into outlet. CAUTION: NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETICS.
Desk/Wall Charger Installation For Desk 1. Select a smooth level surface within 6 feet of a grounded electrical outlet. 2. Plug transformer into outlet. For Wall 1. Turn charger upside down. Push back panel out slightly; then slide upward from charger base. 2. Punch or drill out two round knockouts. Panel may now be used as template. 14
15
1300232Cfinal
6/28/00
10:07 AM
Page 18
PocketScope Handles AA PocketScope Handles (72830) 1. This handle accepts any brand of AA batter ies. Use alkaline batteries for best perfmance. 2. The AA handle can be upgraded to a convertible handle, which allows the handle to be used with AA batteries, Welch Allyn rechargeable batteries and the PocketScope Desk Charger. The upgrade kit part number 72837 includes the rechargeable bottom cap (#72810) plus the rechargeable battery (#72600). The upgrade kit part number 72850 or 72855 includes the rechargeable bottom cap, rechargeable battery and the 792 Series Double Handle Desk Charger.
Convertible PocketScope Handles (72800) For use with AA and Welch Allyn rechargeable batteries This handle can be used with either AA batteries or with Welch Allyn’s PocketScope rechargeable battery (#72600) and Desk Charger. NOTE: Only use the Welch Allyn 792 Series Double-handled Desk Charger with the Welch Allyn Convertible PocketScope handle and rechargeable battery (#72600).
16
17
1300232Cfinal
6/28/00
10:07 AM
Page 20
Batteries Recharging Batteries (72800 only) 1. Plug charger into outlet: Product No. 79280 in the U.S. (120V, 50-60 Hz) Product No. 79285 outside the U.S. (specify country and voltage) 2. To insure full charge and long lamp life, turn instrument(s) to “off” position. 3. Insert handle(s) into charger. Green indicator lights will illuminate when handles are charging properly. NOTE: AA batteries will not charge and will not cause indicator lights to illuminate. NOTE: Only Welch Allyn rechargeable battery (#72600) is approved for use with these handles.
#72600 Nickel Cadmium Rechargeable Battery
Changing Batteries 1. Remove (unscrew) bottom cap. 2. Tap handles on palm of hand until battery slides out. 3. Insert new battery: Rechargeable: Insert in either direction.
18
AA cells: Use alkaline cells for best performance. Insert positive (+) end first for both cells.
+ –
–
+
“AA” Cells
Recycling Batteries Welch Allyn now employs the services of an agency which can disassemble and recycle all components of nickel-cadmium batteries so that nothing gets land-filled or incinerated. Should you wish to dispose of your expended Welch Allyn rechargeable battery via recycling (in the U.S. and Canada only), please send to: Welch Allyn RECYCLE BATTERY 4341 State Street Road Skaneateles Falls, NY 13153-0220
Welch Allyn Canada RECYCLE BATTERY 160 Matheson Blvd. East Mississauga, Ontario CANADA L4Z 1V4
For locations outside the U.S. and Canada, consult your local recycling authority for correct disposal of batteries.
19
1300232Cfinal
6/28/00
10:07 AM
Page 22
Cleaning Recommendations Ophthalmoscopes – Cleaning (and repair) should be performed at Welch Allyn. DO NOT apply alcohol to mirror or lenses as this may degrade optical quality. Otoscope/Retinoscope/Transilluminator – Do not use abrasive cleaners, alcohol or other chemicals. Lens may be cleaned with warm water and soap. Handle – The PocketScope handle wipes clean with a damp cloth. Desk/Wall Charger – DO NOT pour water or any cleaning solution into charging wells. DO NOT autoclave Desk/Wall Charger, instruments or handles.
Replacement Parts To order replacement parts, use these numbers: No. 06200 Retinoscope Lamp No. 03900 Ophthalmoscope Lamp No. 03400 Otoscope/Transilluminator Lamp No. 13000 Ophthalmoscope without handle No. 16210 Retinoscope without handle No. 42000 Transilluminator without handle No. 21111 Otoscope/Throat Illuminator without handle No. 21130 Otoscope without handle 20
No. 211024-501 No. 211005 No. 21504 No. 72800
No. 72801 No. 72830 No. 72831 No. 72837
No. 72850 (Domestic)
No. 72855 (International, specify country) No. 72600
No. 24400 No. 24420 No. 52432 No. 52434
Lamp and shroud assembly Otoscope Lens Insufflator PocketScope Convertible Handle with #72600 Rechargeable Battery PocketScope Convertible Handle (without battery) PocketScope AA Handle with Alkaline Batteries PocketScope AA Handle (without batteries) Upgrade kit to convertible handle (bottom cap & rechargeable battery) Upgrade kit to convertible handle (bottom cap, rechargeable battery & desk charger) Upgrade kit to convertible handle (bottom cap, rechargeable battery & desk charger) Replacement nickel cadmium 2.5v rechargeable battery Set of 4 reusable polypropylene specula (2, 3, 4, 5mm) Set of 3 reusable SofSpec Specula (3, 5, 7mm) Disposable specula – box of 1000 – 2.5mm pediatric size Disposable specula – box of 1000 - 4mm regular size 21
1300232Cfinal
6/28/00
10:07 AM
Page 24
Technical Specifications
Approvals
PocketScope Desk/Wall Charger
Symbols Attention: consult manual
Model 79280 79282 79284 79286
Input Class 2 Country 110-120v~ .05A U.S./Canada/ 50-60 Hz Japan 230v~50 Hz .012A Europe 230v~50 Hz .012A United Kingdom 240v~50 Hz .015A Australia
Output – Model 72800 Rechargeable Handle used with Instrument Heads (charging rate: 45 ma) 9.0v .150a Leakage Current – less than 10 microamps from any exposed metal part when proper ground is used. Weight – 1 lb. 6 oz. Approx. Dimensions – Charger H = 21/2", D = 31/2", W = 41/2" Transport/Storage Conditions -40˚F (-40˚C) to 122˚F (+50˚C) 95% RH max.
22
Type B Type BF Double Insulated Class 2 The mark on this device indicates it has been tested to and conforms with the provisions noted within the 93/42/EEC Medical Device Directive Authorized European Representative European Regulatory Manager Welch Allyn, LTD Kells Road, Navan County Meath Republic of Ireland Tel: 353 46 28122 Fax: 353 46 28536
23
1300232Cfinal
6/28/00
10:07 AM
Page 26
Maintenance
NOTES: 1. All resistors 1/4W., 5% carbon-film unless otherwise marked. 2. All LEDs are T-1 green. 3. All capacitors dipped monolithic ceramic. 2.4V NOM, 500 mA. Hr
Periodic electrical inspection should be performed by qualified personnel at least every six months or more often if used under adverse conditions.
NICKEL CADMIUM BATTERY (2 PLCS) BATTERY WELL 45 MA 2.4VDC
Leakage Current Test: The leakage current from any exposed metal part when a good ground connection is used should be less than 10 microamps. This is equivalent to 9.5 MV in the following circuit:
BATTERY WELL 45 MA 2.4VDC
10.2Ω
.15mfd.
+
+
1000Ω
R1
36 R2
R6 68η
CR3
R3
36 R5
68η HIGH IMPEDANCE VTVM
CR4
27η
R4
27η
For repairs, send unit to: 2 3 TL317 CLP U1
3
2 TL317 CLP U2 1
1 + (WHITE) TRANSFORMER
9V DC @ NOM. C1 INPUT
1µF
Welch Allyn, Inc. Service Department 4341 State Street Road Skaneateles Falls, NY 13153-0220
I -501 110V AC. 50-60 Hz. NOM. (shown) -502 220V AC. 50-60 Hz. NOM. (not shown)
24
Circuit Diagram 25