WelchAllyn
PocketJunior Ophthalmic and Diagnostic Set Bulb Replacement Guide
Replacement Guide
5 Pages
Preview
Page 1
Diagnostic Instruments
PocketJunior The operating and maintenance instructions for the Welch Allyn Halogen PocketJunior set found in this manual are for the Ophthalmic and Diagnostic model sets. Follow these instructions to ensure many years of reliable service. Before using this product, please read the warnings and instructions. Service Please contact your local Welch Allyn authorized service center listed on the bottom of page 2. Warranty The Welch Allyn products listed on this document are guaranteed from 1 year of purchase by Welch Allyn against all manufacturing defects. Welch Allyn will repair or replace, free of charge, any parts of its own manufacture proven to be defective through cause other than misuse, neglect, damage in shipment, or normal wear. Environment Transport/Storage: Operating:
-20°C - +49°C, 95% R.H. Max, 500hPa - 1060hPa Altitude 10°C - 40°C, 95% R.H. Max, 500hPa - 1060hPa Altitude
Symbols: Conforms to: IEC60601-1, IEC 60601-1-2 US
Type BF
IPXØ
ATTENTION: Consult Manual
EMC Framework of Australia
Equipment not protected against the ingress of water.
Safety and Warnings • This product complies with current required standards for electro-magnetic interference and should not present problems to other equipment nor is it affected by other devices. As a precaution, avoid using this device in close proximity to other equipment. • Do not sterilize the Power Handle. • Only authorized Welch Allyn Service Centers can perform service. • POSSIBLE EXPLOSION HAZARD. Equipment not suitable for use in the presence of a flammable anesthetic. • Use of accessories or materials not indicated in the Operator’s Manual can degrade the minimum safety of the equipment. • Do not disassemble or modify any battery. • Remove dead batteries from the handle immediately.
1 of 5
• Slight heating of the handle is normal during extended use of the product. • Remove the batteries from the handle if the instrument is not in use for an extended period of time. Operating Instructions: Retinoscope 1. External focusing sleeve (number 1, figure 1) allows for easy adjustment with continuous 360 degree rotation. 2. Use the retinoscope in a telescoped position only. To telescope the retinoscope, grip head and pull in an outward direction.
Ophthalmoscope 1. Rubber Brow Rest (1). 2. Aperture selection dial (2) with continous rotation: Allows access to microspot, small, large, fixation, slit and cobalt blue. 3. Lens Selection Dial (8) - When used with multiplier lever, allows single diopter steps through 48 lens combinations (from +22 to -25 diopters). 4. Multiplier Lever (7) - Allows fast movement through the range via 16 diopter shifts. 5. Two-position filter switch (4): Engage either open (3) or red-free (5) filter. 6. Diopter Display Window (9) - Displays direct reading of diopter values (positive in green, negative in red).
Lens Instructions To adjust through positive (+) range: • Rotate lens selection dial (8) in positive (clockwise) direction to +6 diopters. Move lens selection dial clockwise again, then move multiplier lever (7) to "+" position. Continue moving lens dial in same clockwise direction to access +7 thru +22 diopters. To adjust through negative (-) range: • Rotate lens selection dial (8) in negative (counterclockwise) direction to -9 diopters. Move lens selection dial counterclockwise again, move multiplier lever (7) to "-" position. Continue moving lens dial in same counterclockwise direction to access -10 thru -25 diopters.
Handle Operation 1. Grip Handle Tube (3) in hand. 2. Using thumb, rotate switch (1) from OFF to ON position as shown in figure 4. 3. To turn unit OFF reverse step 2. Maintenance Lamp Replacement Retinoscope/Otoscope/Ophthalmoscope 1. Remove (unscrew) head. 2. Grasp end of lamp (see fig. 3) and pull out (use nail file or similar object if necessary.) 3. Insert proper replacement lamp by carefully aligning pin on lamp shaft with slot in head. Push lamp in until firmly seated.
2 of 5
NOTE: If lamp surface is touched or becomes smudged, wipe clean with a soft cloth. This will assure longer lamp life and higher light output. CAUTION: Lamp may be hot. Lamp should be allowed to cool before removal.
Battery Replacement 1. 2. 3. 4. 5.
Grip Handle Tube (3) and Top Cap (2) firmly in hand. Unscrew Bottom Cap (4). Gently tap handle in hand to help batteries slide out. Insert new batteries, positive (+) terminal facing Top Cap (see fig. 4 for orientation). Screw in Bottom Cap.
Cleaning Recomendations Wipe all external surfaces with a cloth dampened with a mild detergent and water solution, a solution of 70% isopropyl alcohol and water solution, or a solution of 10% Clorox and water. DO NOT immerse. Wring out the cloth to prevent excess moisture from entering the assembly. Clean instrument windows with a cotton swab dipped in alcohol. NOTE: Excess solution entering the assembly could damage internal components. Use caution to ensure cloth is not saturated with solution. Retinoscope Ophthalmoscope Figure 1 Figure 2
2 Patient Side
1
6 8 5 9 4 2
1 Practitioner Side
7
Lamp Replacement Figure 3
3 Patient Side
Battery/Actuator Orientation Figure 4 ON / ON (marche) / EIN / Encendido (ON) / ON (acceso) / LIGADO / TIL / ON (aan) /PÅ
1 +
- +
3 of 5
-
2
3 OFF / OFF (arrêt) / AUS / Apagado (OFF) / OFF (spento) / DESLIGADO / FRA / OFF (uit) /AV
4
Specifications
Handle & Otoscope Dimensions: 6.5” / 16.5 cm Long Weight:: .290 lb / 132 g (including battery) Voltage: 3V, Class I, Continuous Operation
Handle & Ophthalmoscope Dimensions: 6.38” / 16.1 cm Long Weight:: .285 lb / 129.5 g (including battery) Voltage: 3V, Class I, Continuous Operation Welch Allyn Authorized Service Centers
North America: Welch Allyn, Inc. 4341 State Street Road Skaneateles Falls, NY 13153 Tel: 315-685-4560 or 800-535-6663 Fax:315-685-3361
Welch Allyn Canada, Ltd. 160 Matheson Blvd. E., Unit #2 Mississauga, Ontario, Canada L4Z 1V4 Tel: 905-890-0004 or 800-561-8797 Fax:905-890-0008
Europe: Welch Allyn, GMBH Zollerstrasse 2-4 72417 Jungingen, Germany Tel: 011 49 7477 927186 Fax:011 49 7477 927193
Welch Allyn, France SARL 814 rue Charles de Gaulle 77100 Mareuil les Meaux France Tel:011 33 1 6009 3366 Fax:011 33 1 60 09 67 97
Latin America: MD International 11300 N.W. 41st Street Miami, FL 33178 USA Tel: 305-669-9003 Fax:305-669-1971
4 of 5
Asia Pacific: Welch Allyn Australia Pty The Metro Centre Unit 5 38-46 South Street Rydalmere NSW 2116, Australia Tel: 011 612 9638-3000 Fax:011 612 9638-3500
Welch Allyn Singapore Ltd 300 Beach Road, #25-08 The Concourse Sinapore 199589, Singapore Tel:011 65-291-0882 Fax:011 65-291-5780
Welch Allyn Japan K.K. Bon Marusan 8F, 3-5-1 Kanda Jinbo-Cho, Chiyoda-Ku Tokyo 101-0051 Japan Tel: 011 813-5212-7391 Fax:011 813-3261-7372
Welch Allyn China Room 708 Central Plaza No. 277 Huang Pi Bei Rd. Huang Pi District Shanghai 200003 China Tel:011 86-21-63279631 Fax:011 86-21-63279632
The CE mark on this device indicates it has been tested to and conforms with the provisions noted within the 93/42/EEC Medical Device Directive.
Authorized European Representative: European Regulatory Manager Welch Allyn, LTD., Navan, County Meath, Republic of Ireland Tel: 353 46 67700 FAX: 353 46 28536
Welch Allyn, Inc. 4341 State Street Road PO Box 220 Skaneateles Falls, New York 13153-0220 U.S.A.
Part No. 728207FAX
5 of 5