Welch Allyn Spot Vital Signs LXi
BP SYS
127 73
4 WEST
DIA
mmHg
TEMP
SpO2
985 %
BP SYS
127 73
4 WEST
mmHg
mmHg
AM Reading
01
TEMP
F PULSE
DIA
/min
985
SpO2
%
62
Patient ID 0123456789 Weight Pain Clear Height Respiration Send/Next Reading
AM Reading
01
mmHg
Patient ID 0123456789
Weight Height Respiration
F PULSE
/min
62 Pain Clear Send/Next Reading
Enter
Enter
BRAUn Spot Vital Signs LXi
Directions for Use
Spot Vital Signs LXi
Welch Allyn Spot Vital Signs LXi
Directions for Use
ii
Welch Allyn Spot Vital Signs LXi
Copyright 2005 Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate, in any form, this manual or any part thereof without permission from Welch Allyn. Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual. Welch Allyn®, Spot Vital Signs®, and SureTemp® are registered trademarks of Welch Allyn. SureBP™ Technology is a trademark of Welch Allyn. Braun ThermoScan® is a registered trademark of the Braun Company. LNCS™ is a trademark of, and SET®, LNOP®, and Masimo® are registered trademarks of, Masimo Corporation. Possession or purchase of a Masimo SpO2-equipped device does not convey any express or implied license to use the device with unauthorized sensors or cables which would, alone or in combination with this device, fall within the scope of one or more of the patents relating to this device. Nellcor® and Oxi-Max® are registered trademarks of Nellcor Puritan Bennett Inc. Software in this product is Copyright 2005 Welch Allyn or its vendors. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled, or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title, and ownership of the software remain with Welch Allyn or its vendors. For information about any Welch Allyn product, call Welch Allyn Technical Support: USA +1 800 535 6663 + 1 315 685 4560
Australia +61 2 9638 3000 +800 074 793
Canada +1 800 561 8797
China +86 21 6327 9631
European Call Center +353 46 90 67790
France +33 1 55 69 58 49
Germany +49 7477 9271 70
Japan +81 3 3219 0071
Latin America +1 305 669 9003
Netherlands +31 157 505 000
Singapore +65 6419 8100
South Africa +27 11 777 7555
United Kingdom +44 207 365 6780
Sweden +46 85 853 6551
Manual Material No. 705310 Rev. D Welch Allyn 4341 State Street Road Skaneateles Falls, NY 13153 USA Printed in USA
0297
Directions for Use
iii
Initial Configuration Before using Spot LXi for the first time, you must program an initial configuration screen. See page 23 for more details.
Change Local Defaults Menu Model No.: 45NTO Serial No. 2005040004 Language: BP Units: Temperature Units: Height Units: Weight Units: Date Format: Date: Time Format: Time:
Select
v
Contents Initial Configuration... iii
1 - Introduction... 1 Intended Use... 1 Symbols... 2 Safety Symbols... 2 Button Symbols... 2 Connection Symbols... 3 Agency Symbols... 3 Related Publications... 3 Warnings and Cautions... 4 General Warnings... 4 Blood Pressure Warnings... 5 Temperature Warnings... 7 SpO2 Warnings... 8 General Cautions... 9 Blood Pressure Cautions... 10 Temperature Cautions... 10 SpO2 Cautions... 10 Mises en gardes et avertissements... 11 Mises en garde générales... 11 Mises en garde relatives à la pression artérielle... 12 Mises en garde relatives à la température... 14 Mises en garde relatives au SpO2... 15 Avertissements généraux... 16 Avertissements relatifs à la pression artérielle... 17 Avertissements relatifs à la température... 17 Avertissements relatifs au SpO2... 18 Contents Checklist... 19 Possible Attachments... 19
2 - Controls, Display Window, and Connections... 21 Controls... 21 Display Window... 23 Connections... 25 Braun ThermoScan PRO 4000 Lock... 26 Blood Pressure Hose and Cuff... 26 Thermometer... 26 SpO2 Sensor... 27
vi
Contents
Welch Allyn Spot Vital Signs LXi
Quick Reference Card... 27 AC Power Transformer... 28 Battery... 28 Power On/Off... 28 Standby Mode... 29
3 - Internal Configuration... 31 4 - Blood Pressure Operation... 35 Blood Pressure Cuff Selection... 35 Blood Pressure Measurement... 36
5 - Temperature Operation... 37 Temperature Operation Mode Selection... 37 Normal Mode... 37 Monitor Mode... 40 Temperature Measurement Range Indicators... 40 Ear Temperatures... 41
6 - Pulse Oximetry Operation... 43 7 - Manual Entries and External Device Operation... 45 Manual Entries... 45 Weight, Height, Respiration, and Pain Level... 45 Body Mass Index... 45 Memory Recall... 45 External Devices... 46 Printer... 46 Weight Scale... 46 Barcode Scanner... 47 Wireless Radio... 47
8 - Troubleshooting... 49 Error Codes... 49 Event Causes and Corrective Actions... 51
9 - Specifications... 55 Performance... 55 Blood Pressure Accuracy... 55 Temperature Specifications... 56 SpO2 Specifications... 57 Masimo Sensor Accuracy Guide... 57 Masimo Patents... 57 Nellcor® Sensor Accuracy Guide... 58 Nellcor Patents... 58 Mechanical... 59 Electrical... 59
Directions for Use
Contents
vii
Battery... 59 Environmental... 59 Wireless Radio... 59 Patents... 60 Guidance and Manufacturer’s Declaration... 60 Emissions and Immunity Information... 60
10 - Maintenance and Service... 63 Cleaning... 63 Spot Vital Signs LXi... 63 Blood Pressure Cuff... 63 Blood Pressure Hose and Cable... 63 SureTemp Plus Thermometer... 63 Braun ThermoScan PRO 4000 Thermometer... 64 SpO2 Sensors... 64 Printer... 64 Battery Replacement... 65 Spot Vital Signs LXi... 65 Braun ThermoScan PRO 4000... 66 Printer... 66 Calibration... 67 Blood Pressure Calibration Check... 67 Temperature Calibration Check... 68 Masimo SpO2 Calibration Check... 68 Nellcor SpO2 Functional Check... 68 Service... 68 Technical Assistance... 68 Service Manual/Spare Parts... 69 Service Loaners... 69
11 - Supplies and Accessories... 71 Blood Pressure... 71 Temperature... 72 Pulse Oximetry... 73 Masimo Accessories... 73 Nellcor Accessories... 74 Miscellaneous... 75 Service Contracts... 75
Warranty... 77 Spot LXi... 77 Accessories... 77
1
1 Introduction This Directions for Use manual is a comprehensive guide designed to help you understand the capabilities and operation of your Spot Vital Signs LXi. The information in this manual includes all options available with Spot LXi (e.g., pulse oximetry, barcode scanner, printer, mobile stand, and wall mount). The applicability of some sections of this manual depends on the configuration of your particular device. Read this manual thoroughly before attempting to use the device. Table 1. Available Versions of Spot Vital Signs LXi REF
Description
450T0
SureBP™ Technology with SureTemp Plus Thermometer
450E0
SureBP™ Technology with Braun ThermoScan PRO 4000 Thermometer
45MT0
SureBP™ Technology with Masimo SpO2 and SureTemp Plus Thermometer
45ME0
SureBP™ Technology with Masimo SpO2 and Braun ThermoScan PRO 4000 Thermometer
45NT0
SureBP™ Technology with Nellcor SpO2 and SureTemp Plus Thermometer
45NE0
SureBP™ Technology with Nellcor SpO2 and Braun ThermoScan PRO 4000 Thermometer
Intended Use The Spot Vital Signs LXi automatically measures systolic and diastolic pressure (excluding neonates), pulse rate, temperature (oral, adult axillary, pediatric axillary, rectal, and ear), and pulse oximetry (SpO2) as well as calculates Mean Arterial Pressure (MAP). Furthermore, Spot Vital Signs LXi allows the entry of height, weight, respiration rate, and pain level. Spot Vital Signs LXi also calculates Body Mass Index (BMI) following height and weight entry. The device is intended to be used by clinicians and medically qualified personnel. It is available for sale only upon the order of a physician or licensed health care provider.
2
Introduction
Welch Allyn Spot Vital Signs LXi
Symbols The following symbols are associated with the Spot Vital Signs LXi.
Safety Symbols Identifies information within the manual to avoid injury. Caution: consult accompanying documents
Identifies information within the manual to avoid equipment failure. Pb
Internally Powered, Lead Acid Battery
Handle with Care
Transport Temperature
Storage Humidity
Recycle
Class II Equipment
IPXØ
Equipment is not protected against the ingress of liquid.
Type BF Equipment
On/Off
Do not dispose of this product as unsorted municipal waste. Prepare this product for reuse or separate collection as specified by Directive 2002/96/EC of the European Parliament and the Council of the European Union on Waste Electronic and Electrical Equipment (WEEE). If this product is contaminated, this directive does not apply.
Non-ionizing radiation (RF transmitter)
For more specific disposal information, see www.welchallyn.com/weee, or contact Welch Allyn Customer Service at +44 207 365 6780. Mode of Operation: Continuous
Button Symbols Navigation Buttons
Blood Pressure
Power On/Off
Select
MEM
Memory
Directions for Use
Introduction
Connection Symbols USB Connection
Serial Port Connection
Agency Symbols CERTIFIED TO: CAN/CSA STD C22.2 NO. 601.1 C
US
166292
CONFORMS TO: UL STD 60601-1 IEC 60601-1 EMC Framework of Australia
The CE mark on this product indicates that it has been tested to and conforms with the provisions noted within the 93/42/EEC Medical Device Directive.
0297 EC
REP
European Regulatory Manager Welch Allyn Ltd. Navan Business Park Dublin Road Navan, County Meath, Republic of Ireland Tel.: 353 46 90 67700 Fax: 353 46 90 67756
Related Publications Braun ThermoScan PRO 4000 User’s Guide - for models 450E0, 45NE0, 45ME0. Masimo Directions for Use - for models 45MT0, 45ME0. Nellcor Directions for Use - for models 45NT0, 45NE0.
3
4
Introduction
Welch Allyn Spot Vital Signs LXi
Warnings and Cautions Familiarize all operating personnel with the general safety information in this summary. Specific warnings and cautions are also found throughout this manual.
General Warnings A warning statement in this manual identifies a condition or practice, which if not corrected or discontinued immediately, could lead to patient injury, illness, or death. These warnings pertain to the entire Spot Vital Signs LXi device. WARNING The information in this manual is a comprehensive guide to the operation of Spot LXi. For best results, read this manual thoroughly before using the device. WARNING Spot LXi is designed for medical clinician use. Although this manual may illustrate medical spot-check techniques, only a trained clinician who knows how to take and interpret a patient’s vital signs should use this device. WARNING Spot LXi is not intended for use in environments that are without health care practitioner supervision. WARNING Spot LXi is not intended for continuous monitoring. Do not leave the device unattended while taking measurements on a patient. WARNING To ensure data integrity, save readings and clear the Spot LXi display between patients. WARNING The Spot LXi is not defibrillator proof. WARNING Spot LXi is not intended for use during patient transport. WARNING This device is not suitable for use in the presence of a flammable anesthetic mixture with air or oxygen or nitrous oxide. An explosion may result. WARNING To ensure patient safety, use only accessories and supplies (i.e., cuffs, hoses, temperature probes, SpO2 sensors, etc.) recommended for or supplied with Spot LXi. Using unapproved accessories with Spot LXi can affect patient and/or operator safety. WARNING Take care to prevent water or other fluid from entering any connectors on the device. Should this occur, dry the connectors with warm air. Check the accuracy of all operating functions. WARNING Every three months, inspect the blood pressure cuff, SpO2 cable, and other accessories for fraying or other damage. Replace as necessary. WARNING Do not use Spot LXi on patients who are on heart/lung machines. WARNING Electric shock hazard. There are no user-serviceable parts inside Spot LXi other than battery replacement (see “Battery Replacement” on page 65). An operator may only perform maintenance procedures specifically described in this manual. For service, refer the device to an Authorized Service Center. WARNING This device is not intended for hand-held use during operation.
Directions for Use
Introduction
5
WARNING Do not autoclave. WARNING This device complies with current required standards for electromagnetic interference and should not present problems to other equipment or be affected by other devices. As a precaution, avoid using this device in close proximity to other equipment. WARNING Welch Allyn is not responsible for the integrity of any mounting installation. Welch Allyn recommends that the customer contact their Biomedical Engineering Department or maintenance service to ensure professional installation for safety and reliability of any mounting accessory. WARNING The Spot LXi consists of high-quality precision parts. Protect it from severe impact and shock. A qualified service technician must check any Spot LXi that is dropped or damaged for proper operation prior to further use. Do not use the Spot LXi if you notice any signs of damage. Contact the Welch Allyn Customer Service Department for assistance. WARNING Do not use an SpO2 finger clip sensor and a blood pressure cuff simultaneously on the same limb. Doing so may result in inaccurate pulse rate and perfusion readings. WARNING The USB port is for service only. Do not use this connection for any external equipment connections.
Blood Pressure Warnings These warnings pertain to the Spot LXi blood pressure feature. WARNING Spot LXi is not intended to measure BLOOD PRESSURE on neonatal patients. The AAMI SP10:2002 standard defines neonates as children 28 days or less of age if born at term (37 weeks gestation or more); otherwise up to 44 gestational weeks. WARNING To ensure pediatric blood pressure accuracy and safety, the Child Reusable Two-Piece Blood Pressure Cuff (4500-01), Infant Durable One-Piece Cuff (5082-82-4MQ), and the Infant Disposable One-Piece Cuff (5082-92-4MQ) are the smallest cuffs approved for use with young children and infants. The child’s arm must fit within the range markings on the cuff. WARNING Avoid compression of the blood pressure hose or cuff tubing of Spot LXi. This may cause system errors to occur in the device. WARNING Patients who are experiencing moderate to severe arrhythmias may give inaccurate blood pressure measurements. WARNING Spot LXi does not operate effectively on patients who are experiencing convulsions or tremors. WARNING Use only Welch Allyn blood pressure cuffs and/or hoses. Using other manufacturers’ blood pressure cuffs and/or hoses may produce inaccurate blood pressure readings. WARNING When several blood pressure measurements are taken on the same patient, regularly check the cuff site and extremity for possible ischemia, purpura, and/or neuropathy.
6
Introduction
Welch Allyn Spot Vital Signs LXi
WARNING Do not allow a blood pressure cuff to remain on the patient more than 10 minutes when inflated above 10 mmHg. This may cause patient distress, disturb blood circulation, and contribute to injury of peripheral nerves. WARNING Do not place the cuff on any extremity that is used for intravenous infusions or any area where circulation is compromised. WARNING Excessive cuff tightness may cause venous congestion and discoloration of the limb. WARNING Wrapping the cuff too loosely (preventing proper inflation) may result in errors. WARNING Do not change the connector(s) on the blood pressure cuff tubing of this device to luer type. Luer type connectors are commonly used in intravenous infusion systems. Using the luer connectors on blood pressure cuff tubing creates the risk that the blood pressure tubing could be mistakenly connected to a patient's intravenous line, resulting in the introduction of air into the patient's circulatory system.
Directions for Use
Introduction
7
Temperature Warnings These warnings pertain to the Spot LXi temperature feature.
SureTemp® Plus These warnings are specific to the SureTemp Plus thermometer option. WARNING Use only Welch Allyn probe covers. Using other manufacturers’ probe covers or no probe cover may produce temperature measurement errors and/or inaccuracy. WARNING Long-term continuous monitoring beyond three to five minutes is not recommended in any mode. WARNING Oral/axillary probes (blue ejection button at top of probe) and blue removable probe wells are used for taking oral and axillary temperatures only. Rectal probes (red ejection button) and red removable probe wells are used for taking rectal temperatures only. Use of the probe at the wrong site will result in temperature errors. Use of the incorrect removable probe well could result in patient cross-contamination. WARNING The thermometer connectors and probe are not waterproof. Do not immerse or drip fluids on these items. Should this occur, dry the connectors and probe with warm air. Check all functions for proper operation and accuracy. WARNING Do not take an axillary temperature through patient’s clothing. Direct probe cover to skin contact is required. WARNING Do not autoclave. WARNING Use Welch Allyn single-use disposable probe covers to limit patient cross-contamination. WARNING Incorrect insertion of probe can cause bowel perforation. WARNING Washing hands greatly reduces the risk of cross-contamination and nosocomial infection. WARNING To ensure optimal accuracy, always confirm that the correct mode is selected.
8
Introduction
Welch Allyn Spot Vital Signs LXi
Braun Thermoscan PRO 4000 These warnings are specific to the Braun ThermoScan PRO 4000 thermometer option. WARNING Keep the probe window clean, dry, and undamaged at all times to ensure accurate measurements. To protect the probe window, always keep the thermometer in the storage cover while transporting or when not in use. WARNING Only use Braun ThermoScan probe covers with this thermometer. Using other manufacturer’s probe covers or no probe cover may produce temperature measurement errors and/or inaccuracies. If the thermometer is used without a probe cover attached, clean the lens (see “Braun ThermoScan PRO 4000 Thermometer” on page 64). WARNING Do not autoclave. WARNING The thermometer is not waterproof. Do not immerse or drip fluids on it. Should this occur, dry the thermometer with warm air. Check for proper operation and accuracy.
SpO2 Warnings These warnings pertain to the Spot LXi SpO2 feature. WARNING Only use Spot LXi with Masimo or Nellcor SpO2 option with Masimo or Nellcor brand sensors and accessories, respectively. Using the wrong or unapproved sensors or cables may cause improper performance. WARNING The SpO2 sensor and extension cables are intended for use only for pulse oximetry measurements. Do not attempt to connect these cables to a PC or any similar device. WARNING Before using, carefully read the sensor Directions for Use, including all warnings, cautions, and instructions. WARNING Do not use a damaged sensor or pulse oximetry cable or a sensor with exposed optical components. WARNING Incorrect application or a long duration of use of an SpO2 sensor may cause tissue damage. Inspect the sensor site periodically as directed in the sensor’s Directions for Use. WARNING Certain ambient environmental conditions, sensor application errors, and certain patient conditions may affect SpO2 readings and pulse signal. WARNING Do not immerse the sensor or patient cables in water, solvents, or cleaning solutions (the sensors and connections are not waterproof). Do not use irradiation, steam, or ethylene oxide for sterilization. WARNING The SpO2 in the Spot LXi device is not intended for use as an apnea monitor. WARNING Consider the SpO2 an early warning device. As a trend toward patient deoxygenation is indicated, use laboratory instruments to analyze blood samples to completely understand the patient’s condition.
Directions for Use
Introduction
9
WARNING Tissue damage can be caused by incorrect application or duration of use of a Nellcor OxiMax sensor. Inspect the sensor site as directed in the sensor Directions for Use. WARNING Do not use the sensors during magnetic resonance imaging (MRI) scanning. Induced current could potentially cause burns. The MS board pulse oximeter may affect the MRI image, and the MRI unit may affect the accuracy of the oximetry measurements. WARNING Carefully route patient cabling to reduce the possibility of patient entanglement or strangulation. WARNING Failure to cover the Nellcor OxiMax sensor site with opaque material in high ambient light conditions may result in inaccurate measurements.
General Cautions A caution statement in this manual identifies a condition or practice, which if not corrected or discontinued immediately, could lead to equipment failure, equipment damage, or data loss. These cautions pertain to the entire Spot Vital Signs LXi device. Caution If the accuracy of any measurement is in question, check the patient’s vital sign(s) with an alternate method and then check to verify the device is functioning properly. Caution Place the device on a secure surface or use one of the optional mounting accessories. Caution Do not place fluids on or near the device. Caution It is recommended that the device is used within stated operating temperature ranges (see “Environmental” on page 59). The device will not meet its performance specifications if used outside these temperatures ranges. Caution For proper patient electrical isolation, use only a Welch Allyn power supply (4500-101A) to charge Spot LXi and its attached peripheral devices. Do not use an external charger while the printer is attached to Spot LXi. Caution Always unplug the AC power transformer from the outlet before moving the mobile stand to a new location. Caution The basket has a three-pound weight limit. Take care not to exceed this limit. Caution Only use a 9V battery with the Healthometer scale. Remove and discard the wall mounted power supply. Caution The Healthometer scale, REF 349KLX, is not sold with an RS-232 cable to connect the scale to the Spot LXi. The customer must purchase a standard RS-232 cable. Caution When using the Healthometer scale, remove the two hex nuts on the RS-232 cable, as supplied, before screwing the cable into the scale base.
10
Introduction
Welch Allyn Spot Vital Signs LXi
Blood Pressure Cautions These cautions pertain to the Spot LXi blood pressure feature. Caution Minimize extremity and cuff motion during blood pressure readings. Caution If the blood pressure cuff is not at heart level, note the difference in reading due to the hydrostatic effect. Add the value of 1.80 mmHg (.2 kPa) to the displayed reading for every inch (2.5 cm) above heart level. Subtract the value of 1.80 mmHg (.2 kPa) from the displayed reading for every inch (2.5 cm) below heart level. Caution Proper blood pressure cuff size and placement is essential to the accuracy of the blood pressure determination. See Reusable Two-Piece Cuff Measurements (Table 10) or Durable One-Piece Cuff Measurements (Table 11) on page 35 for sizing information. Caution The position and physiologic condition of the subject can affect a blood pressure reading.
Temperature Cautions These cautions pertain to the Spot LXi temperature feature. Caution The SureTemp Plus feature only operates with the probe well in place. Caution Do not use alkaline batteries in the Braun ThermoScan PRO 4000. Caution Biting the probe tip may result in damage to the probe.
SpO2 Cautions These cautions pertain to the Spot LXi SpO2 feature. Caution The pulse oximeter is calibrated to determine the percentage of arterial oxygen saturation of functional hemoglobin. Significant levels of dysfunctional hemoglobin such as carboxyhemoglobin or methemoglobin may affect the accuracy of the measurement. Caution Some sensors may not be appropriate for a particular patient. If at least 10 seconds of perfusion pulses cannot be observed for a given sensor, change sensor location or sensor type for perfusion to resume. Caution Physiological conditions, medical procedures, or external agents that may interfere with the pulse oximeter’s ability to detect and display measurements include dysfunctional hemoglobin, arterial dyes, low perfusion, dark pigment, and externally applied coloring agents such as nail polish, dye, or pigmented cream. Caution When selecting a sensor, consider the patient’s weight and activity level, the adequacy of perfusion, the available sensor sites, the need for sterility, and the anticipated duration of monitoring.
Directions for Use
Introduction
19
Contents Checklist Unpack the Spot LXi and any applicable accessories and then inspect for missing items. Retain the shipping materials in the event of shipping damage or for return, if necessary, to Welch Allyn for repair or warranty service. Report any signs of shipping damage to the carrier. Report any missing or damaged items to the Welch Allyn Service Center near you. All Spot LXi devices include the following components: Spot LXi Device. This device automatically measures and displays blood pressure, pulse rate, and temperature. Directions for Use Manual. Read this manual thoroughly before using Spot LXi. Save this manual for reference. Warranty Card. This card validates the Spot LXi warranty. Fill out the warranty card and mail it today. Blood Pressure Cuff. One cuff with connectors. Other size cuffs are available separately. Blood Pressure Hose. Latex-free pressure hose with connectors to attach various sizes of blood pressure cuffs to the Spot LXi. AC Power Transformer and Cord Assembly. Provides power to the Spot LXi and charges the internal battery. Quick Reference Card. Attach this quick operating guide to the device handle, mobile stand, or wall mount.
Possible Attachments Spot LXi may include the following items based on the model and accessories purchased: SureTemp Plus Temperature Probe, Well, and Covers. One oral temperature probe (blue ejection button and well) and one box of 25 single-use, disposable probe covers. Braun ThermoScan PRO 4000 Thermometer and Covers. One ear thermometer; one box of 20 single-use, disposable probe covers; one rechargeable battery pack; and one lock release pin. Wireless Radio and Mounting Bracket. Attach these items to the back of the mobile stand or wall mount basket. Barcode Scanner and Mounting Bracket. Attach these items on the basket of the mobile stand or wall mount. Pulse Oximetry (SpO2). The finger clip SpO2 sensor and extension cable are for use with adult and pediatric patients. Other sensors are available separately. Printer. Attach to the basket of the mobile stand or wall mount. The printer comes with paper. Rolls of paper and labels are available separately.
20
Introduction
Welch Allyn Spot Vital Signs LXi
21
2
Controls, Display Window, and Connections Drawings and text are representative of Spot Vital Signs LXi with all available options. Your device may not include all functions based on the model purchased.
Controls Figure 1. Spot LXi Front Panel with SureTemp Plus Thermometer SureTemp Plus thermometer: remove the probe to take a temperature. BP SYS
127 73
4 WEST
mmHg
DIA
mmHg
TEMP
SpO2
985 %
AM Reading
01
F PULSE
/min
62
Removable probe well: store the temperature probe here when not in use; remove to clean or replace to eliminate cross-contamination.
Patient ID 0123456789 Weight Pain Clear Height Respiration Send/Next Reading
Enter
Navigation button: move throughout the options in the Navigation Window or to increase/decrease parameter units.
Probe cover storage area: storage space for one box of probe covers.
Select button: accepts the current option.
Power button: controls power to the device. Charging LED: indicates when device is plugged in and charging the battery.
Spot Vital Signs LXi
Blood Pressure Start/Stop button: initiates a new blood pressure cycle. Pressing again aborts an active blood pressure measurement.
Memory button: recalls up to 50 most recent readings.