Quick Reference Guide
2 Pages
Preview
Page 1
Ear Temperatures
with the Braun ThermoScan PRO 4000 ®
1. Remove the thermometer from the case and attach a probe cover. The device will turn on. 2. Wait for the “ready” beep and place the probe snugly into the ear canal; correct probe placement is essential for accurate readings. 3. Press and release the Start button. The thermometer will beep once. The green ExacTemp™ light above the Start button will flash, indicating stable probe placement. 4. A long beep and a solid green ExacTemp light will signal the end of the measurement. 5. If the probe is not placed in a stable position while taking a temperature, the device beeps, the green ExacTemp light goes out and POS (Position Error) will display.
Ear Canal
98.4 94.1 Probe Tip
C10126_BraunTipsCard.indd 1
97.8
99.1
6. Press the button to recall the last temperature. 7. To change default temperature scale: With the device off, press and hold the button for about 5 seconds. Release the button when desired scale is shown.
Tympanic Membrane
©2002 Welch Allyn, Inc. All rights reserved
1/6/13 9:05 P
Tips for Taking Ear Temperatures with the Braun ThermoScan PRO 4000 ®
1. Always apply a probe cover. If the device is placed in the ear canal without a probe cover, the lens must be cleaned before the next use. 2. Clean the lens with an alcohol swab and allow five minutes for complete drying. 3. Use only Braun ThermoScan probe covers. 4. The following will interfere with accurate temperature readings: > Poor probe placement
> Ambient temperature
> Dirty, wet or damaged probe lens
> Lying on pillow
> No or used probe cover
> Hearing aid
5. The following will not interfere with accurate temperature readings: > Cerumen (ear wax) > Otitis Media (ear infection) > Tympanostomy tubes
CUSTOMER SERVICE For technical or clinical support, please contact Welch Allyn at 800.535.6663. www.welchallyn.com © 2013 Welch Allyn SM2678 Rev F
C10126_BraunTipsCard.indd 2