Quick Reference Guide
2 Pages
Preview
Page 1
Welch Allyn Spot Vital Signs 4400 ®
Quick reference
Power button
Start an NIBP averaging program
• Located on the device housing • Powers up the device • When the device is on, opens pop-up dialog with controls to power down, enter sleep mode, or cancel out of the power menu
Power down 1. Touch the power button. 2. Touch Power down.
Results of an NIBP averaging program
Battery status Charging
Battery removed or not holding a charge
Approximate operating time remaining
START
STOP
STOP START AVERAGE
111/62 SYS/DIA mmHg ( 76 MAP )
1. 120/80 2. 123/82 3. 110/64 4. 111/62 5. 108/61 6. 105/59
Displays the average of the readings. Start average/Cancel/Skip
START
NIBP AVERAGED
TOTAL READINGS AVERAGED: 4
NIBP measurement Start/Stop
The averaging program takes consecutive NIBP measurements within a configurable amount of time. When the program is complete, the average measurement is displayed in the Home tab. 1. Place the proper cuff around the patient’s bare upper arm. 2. On the Home tab, touch Start Average. 3. If necessary, touch Skip to stop the current measurement. The measurement restarts at the end of the countdown. 4. When averaging completes, touch Save to save the patient data, or touch Clear to discard the data.
START SKIP CANCEL AVERAGE
CANCEL
SKIP
Displays “NIBP AVERAGED.” Displays the readings taken by the program. A reading with a line through it is excluded from the average. Note To retain an NIBP average, touch Save.
Manually enter vitals measurements NIBP
Clean the equipment CAUTION
PULSE RATE
START
SYS/DIA mmHg ( MAP )
START AVERAGE
SpO2
/MIN TEMPERATURE
°F (°C )
1. 2. 3. 4. 5.
Press and hold a frame, such as NIBP. A pop-up dialog appears. On the left, touch an empty field and enter the vitals measurement. Touch OK. When you are finished entering measurements, touch OK. Touch Save to save the patient data.
Enter additional parameters Height
Weight
in
lbs
Home
Pain
RR
BMI
Adult
bpm
Review
Clear
Save
Settings
1. Touch the Additional parameters frame. 2. Modify the parameters as desired. If height and weight are entered, BMI will be automatically calculated. 3. Touch OK. 4. When you are finished entering measurements, touch OK. 5. Touch Save to save the patient data.
Do not use unapproved cleaning agents. Use of unapproved cleaning agents may cause damage to components. See “Prepare to clean the equipment” in Directions for use for a list of approved cleaning agents.
Follow the cleaning agent manufacturer’s instructions to prepare solution, if applicable, and clean all exposed surfaces of the device, basket, cords and cables, and stand. Wipe all surfaces until no visible soil remains. Change the wipe or cloth throughout the cleaning procedure as needed. 1. Disconnect the AC power cord from the mains outlet. 2. Wipe the top of the device. 3. Wipe the sides, front, and rear of the device. 4. Clean the LCD screen with an agent that is approved for cleaning the display. To avoid residual film buildup after cleaning, wipe the LCD screen with a clean cloth dampened with water and then wipe the screen dry with a dry cloth. 5. Wipe the bottom of device. 6. Wipe the AC power cord. 7. Wipe the stand from top to bottom.
Pull vitals data into an EMR Refer to hillrom.com to learn how to integrate the Spot 4400 with your EMR provider. Once the device is integrated, you can pull patient vitals measurements to your EMR host computer using the USB cable provided.