Guide
18 Pages
Preview
Page 1
Stethoscope [the stethoscope)
A
Since 1915, Welch Allyn has been a company dedicated to bringing you the best in diagnostic instrumentation for the front line health care professional. The new Welch Allyn Electronic Stethoscope continues this tradition. The Welch Allyn Electronic Stethoscope based upon proprietary technology from Meditron AS - provides superior sound reproduction due to its sensor-based methodology. Other electronic stethoscopes depend upon a microphone for sound reproduction. Unfortunately, no matter the fidelity, microphones pick up ambient noise in the environment transmitted through the stethoscope diaphragm and chestpiece.
[thedistributor) is designed for use when there is a need for several people to listen to a patient simultaneously. By using [thedistributor), up to six people can participate. The "Master" listener controls the auscultation process with his or her stethoscope. Others are co-listeners who will hear the same sounds by connecting their stethoscopes or headsets to [thedistributor). Up to three groups of listeners may be linked together with additional [thedistributor)’s as needed for larger groups.
The Welch Allyn Electronic Stethoscope utilizes a highly directional, recessed pressure sensor which eliminates much of the ambient noise in the environment while faithfully reproducing body sounds in the same "color" as a traditional Welch Allyn stethoscope. Amplification can be varied as desired by the practitioner. Recognizing the need for improved auscultation skills, the Welch Allyn Electronic Stethoscope is just one part of a complete teaching and diagnostic system. When combined with [thedistributor) - a product for sending sounds from the stethoscope to others and [theanalyzer) - a software accessory for recording sounds and phonocardiograms, the Welch Allyn Electronic Stethoscope System is a powerful addition to any hospital, clinic, or health care professional's diagnostic toolbox.
2
Table of contents Overview
English
4-5
French
6-7
Spanish
8-9
German
10 - 11
Italian
12 - 13
Warranty
Product Diagram
3
14 - 15
16
English On/Off
Master
1-2-3-4-5
[thedistributor) has an automatic On and Off function. The unit turns itself on automatically when the master stethoscope is on and a signal is detected. This takes place shortly (0.5 seconds) after the On button of the stethoscope is pressed. Please note that the unit can only be turned on from the master stethoscope. As long as signals are being emitted by the master stethoscope or the PC, the unit will remain on. When [thedistributor) is on, the diode will light up. [thedistributor) will remain on for one minute after all signals have ceased.
The stethoscope that carries out the active listening must be connected to the external cable port marked "Master". Use one of the accompanying cables and make sure the plugs are properly pushed in. When the ON/OFF button is pressed, the audio signal will go to [thedistributor). The unit will then turn itself on and the diode will light up. The "Master" is then controlling the unfiltered auscultation signal that is being fed into [thedistributor).
Co-listeners link their stethoscopes or headphones to one of the external cable ports marked from 1 to 5. If you are using a Stethoscope, use the accompanying cables and ensure that the plugs are properly pushed in. Users will now experience the same sound as the "Master". For listeners using Headsets, the volume and frequency control on [thedistributor) can be used in a similar way to that of the stethoscope. For listeners using Stethoscopes and not Headsets, it is important that the frequency selector should be in the middle position.
4
PC
Light-emitting diode
Batteries
This external cable port should be used when the user wants to send an audio signal to a PC for visualisation, storage or further processing in the software program.
The green light-emitting diode located in the top cover of the unit has two funtions: 1: ON/OFF indicator. 2: Battery indicator. The unit will turn itself on and the diode will light up when a signal is received from the stethoscope. If no signal is received, the unit will turn itself off after one minute. A blinking diode indicates that the batteries are weak. No light indicates that there is no signal or that the batteries are completely drained.
[thedistributor) runs on two easily replacable highenergy TL-5902 lithium batteries with an international reference size of 1/2 AA, ER14252. The batteries are estimated to provide an effective listening time of 200 hours. Change batteries by pushing the cover back. Check that the batteries are installed with the correct polarity.
Choosing the frequency [thedistributor) has filters for three predetermined frequency ranges. The filters are controlled with a small selector switch. 1. Pulmonary (H: High-frequency sounds) 2. Wide listening range (Mid-position) 3. Cardiology (L:Low frequency sounds)
Volume control [thedistributor) is equipped with a variable volume control. Turning in the direction of the arrow reduces the volume.
5
English Guarantee Welch Allyn will remedy, through repair or replacement, all defects in the product that are due to errors in manufacture, materials, or construction and which become apparent within two years of sale. All claims must be made to your supplier without unreasonable delay. Liability Neither Welch Allyn nor Meditron AS is responsible for diagnoses, analyses, or outcomes resulting form the use of this equipment. Service Should any technical problems arise, please contact your dealer representative. Do not try to disassemble the product to carry out repairs. Service may also be obtained by contacting the authorized Welch Allyn Service Centers below: In the United States: Welch Allyn Repair Dept. 95 Old Shoals Rd. Arden, NC 28704 USA In Latin America: MDI, Repair Dept. 11300 NW 41st. Street, Miami, Florida 33178 USA In Europe: Welch Allyn GmbH. Postfach 31, Zollerstrasse 2-4 D-72417, Jungingen, Deutschland In Canada: Welch Allyn 160 Matheson Blvd - East, Unit 2, Mississauga, Ontario, Canada L4Z1V4 In Asia: Welch Allyn 21-09 Golden Mile 6001 Beach Road, Singapore 199589, Rep of Singapore In Australia: Welch Allyn Ground Floor, 18-20 Orion Road, Lane Cove, NSW 2066, Australia
Pour l’entretien de l’équipement, il est aussi possible de d’adresser aux centres d’entretien Welch Allyn agréés ci-dessous: Aux USA: Welch Allyn Repair Dept. 95 Old Shoals Rd. Arden, NC 28704 USA En Amérique latine: MDI, Repair Dept. 11300 NW 41st. Street, Miami, Florida 33178 USA En Europe: Welch Allyn GmbH. Postfach 31, Zollerstrasse 2-4 D-72417, Jungingen, Deutschland Au Canada: Welch Allyn 160 Matheson Blvd - East, Unit 2, Mississauga, Ontario, Canada L4Z1V4 En Asie: Welch Allyn 21-09 Golden Mile 6001 Beach Road, Singapore 199589, Rep of Singapore En Australie : Welch Allyn Ground Floor, 18-20 Orion Road, Lane Cove, NSW 2066, Australia
Spanish Garantía Welch Allyn reparará o sustituirá cualquier fallo del producto debido a errores de fabricación, materiales o construcción y que aparezcan antes de transcurridos dos años de la venta. Todas las reclamaciones deberán dirigirse al proveedor en un plazo razonable. Responsabilidades Ni Welch Allyn ni Meditron AS podrán ser considerados responsables de diagnósticos, análisis u otros resultados derivados del empleo de este equipo.
French Garantie Welch Allyn corrigera, par la réparation ou le remplacement, tout défaut du produit causé par une anomalie de fabrication, de matériaux ou de construction et apparu dans un délai de deux ans suivant la vente. Toute réclamation doit être adressée au concessionnaire sans délai excessif. Responsabilité Ni Welch Allyn ni Meditron AS ne sont responsables des diagnostics, analyses ou résultats découlant de l’usage du présent équipement.
Service En cas de problème technique, veuillez prendre contact avec votre concessionnaire. N’essayez pas de démonter le produit pour effectuer vous-même une réparation.
Mantenimiento Si surge algún problema técnico, póngase en contacto con el representante de su distribuidor. No desmonte el producto para hacer ninguna reparación. También puede recibir atención poniéndose en contacto con los centros de mantenimiento autorizados de Welch Allyn que se indican a continuación: En Estados Unidos: Welch Allyn Repair Dept. 95 Old Shoals Rd. Arden, NC 28704 EE.UU. En América Latina: MDI, Repair Dept. 11300 NW 41st. Street, Miami, Florida 33178 EE.UU.
14
English Declaration of Conformity [thedistributor), a part of the stethoscope system, conforms to the following standards or other normative documents: EN 1441: 1997 IEC 601-1-2: 1993-04, with exception of all claims related to immunity (3V/m) ISO 10993-1: 1997-12-15. CISPR 11: 1990-10 + amendment #1  Following the provisions of Medical Device Directive 93/42/EØF from 14th of june 1993. FDA 510(k) 991367
French Déclaration de conformité [thedistributor), un élément du système de stéthoscope, satisfait aux normes et autres documents normatifs suivants: EN 1441: 1997 IEC 601-1-2: 1993-04, à l’exception de toute réclamation relative à l’immunité (3V/m) ISO 10993-1: 1997-12-15 CISPR 11: 1990-10 + amendement #1  Satisfait aux provisions de la directive sur les instruments médicaux 93/42/EØF datée du 14 juin 1993. FDA 510(k) 991367
Spanish
95P 501 Rev A
Declaración de conformidad [thedistributor), parte del sistema del estetoscopio, cumple las siguientes normas y lo establecido en otros documentos normativos: EN 1441: 1997 IEC 601-1-2: 1993-04, con excepción de lo relativo a inmunidad (3V/m) ISO 10993-1: 1997-12-15 CISPR 11: 1990-10 + enmiendas 1 y 2
Conforme a las disposiciones de la Directiva sobre dispositivos médicos 93/42/EØF de 14 de junio de 1993. FDA 510(k) 991367
German Komformitätserkärung [thedistributor), entspricht als Teil des Stethoskopsystems den Anforderungen folgender Normen und Richtlinien: EN 1441: 1997 IEC 601-1-2: 1993-04, mit Ausnahmen der Anforderungen an die Immunität (3V/m) ISO 10993-1: 1997-12-15 CISPR 11: 1990-10 + Zusatz #1  Es entspricht den Anforderungen der Medizingeräte-Richtlinie 93/42/EWG vom 14. Juni 1993. FDA 510(k) 991367
Italian Dichiarazione di conformità [thedistributor), un componente del sistema stetoscopio, è conforme ai seguenti standard o documenti normativi: EN 1441: 1997 IEC 601-1-2: 1993-04, ad eccezione di tutti reclami relativi all'immunità (3V/m) ISO 10993-1: 1997-12-15 CISPR 11: 1990-10 + modifica #1  Conforme alle disposizioni della direttiva riguardante i dispositivi medicali 93/42/EØF del 14 giugno 1993. FDA 510(k) 991367
[thedistributor) enables the tutor and the students to listen simultaneously to the same patient.