XLTEK
XLTEK NeuroMax User Manual Rev E
User Manual
122 Pages
Preview
Page 1
XLTEK NEUROMAX
Page 2
User Manual
XLTEK NeuroMax
Publisher’s Notice 105352 Rev E XLTEK NeuroMax User Manual Natus Medical Incorporated Excel-Tech Ltd. (XLTEK) 2568 Bristol Circle
Oakville, Ontario, L6H 5S1 Canada Tel: 905-829-5300 or Fax: Free (US & Canada): Technical Support Email: Customer Service Email:
EUROPEAN AUTHORIZED REPRESENTATIVE: Natus Europe GmbH Robert-Koch-Str. 1 82152 Planegg Germany
Copyright © 2012 by Excel-Tech Ltd. (XLTEK) The information in this publication is provided for reference only. All information contained in this publication is believed to be correct and complete. Excel-Tech Ltd. shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing, performance, or use of this material. All product specifications, as well as the information contained in this publication, are subject to change without notice. This publication may contain or reference information and products protected by copyrights or patents and does not convey any license under the patent rights of Excel-Tech Ltd., nor the rights of others. Excel-Tech Ltd. does not assume any liability arising out of any infringements of patents or other rights of third parties. All rights reserved. This document contains confidential or proprietary information of Excel-Tech Ltd. No part of this document may be reproduced or transmitted in any form or by any means without the written permission of Excel-Tech Ltd. Excel-Tech Ltd. makes no warranty of any kind with regard to this material, including but not limited to the implied warranties of merchantability and fitness of a particular purpose.
Page 1
User Manual
XLTEK NeuroMax
TABLE OF CONTENTS NEUROMAX SAFETY AND STANDARDS CONFORMITY
6
1. WELCOME TO THE NEUROMAX
12
1.1. INTRODUCTION
13
1.1.1. INTENDED USE
13
1.2. CUSTOMER SERVICE 1.3. USING THE MANUAL
13 14
2. THE XLTEK NEUROMAX
15
2.1. OPERATING CONDITIONS
15
2.1.1. ENVIRONMENT PARAMETERS 2.1.2. TRANSPORT AND STORAGE PARAMETERS
15 15
2.2. SPECIFICATIONS MATERIALS EXTERNAL CONNECTORS CHANNELS SAMPLING RATE SIZE AND TYPE RESOLUTION CMRR NOISE MAXIMUM SIMULATOR VOLTAGE MAXIMUM STIMULATOR CURRENT STIMULATOR CURRENT ACCURACY STIMULATOR DURATION ACCURACY VOLTAGE ACCURACY ON SCREEN 2.3. WARNINGS AND CAUTIONS
16 16 16 16 16 16 16 16 16 17 17 17 17 17 17
2.3.1. W ARNINGS 2.3.2. CAUTIONS
18 20
2.4. EXPLANATION OF LABELING SYMBOLS 2.5. MAIN MENU 2.6. KEY PAD 2.7. HOT KEYS
22 23 24 27
2.7.1. SENSORY NERVE CONDUCTION HOT KEYS 2.7.2. ELECTROMYOGRAPHY HOT KEYS 2.7.3. EVOKED POTENTIAL HOT KEYS
27 27 28
2.8. CREATING A PATIENT FILE 2.9. GENERATING TEST REPORTS
29 30
2.9.1. REPORT FUNCTIONS
30
3. NERVE CONDUCTION TESTS
33
3.1. MOTOR AND SENSORY NERVE CONDUCTIONS
34
Page 2
User Manual
XLTEK NeuroMax
3.1.1. SETTING UP A NERVE CONDUCTION STUDY 3.1.2. CONDUCTING AN NCS 3.1.3. ABOUT THE REP STIM TEST 3.1.4. PERFORMING A REP STIM TEST 3.1.5. ABOUT THE F-W AVE TEST 3.1.6. PERFORMING AN F-W AVE TEST 3.1.7. SETTING UP AN H-REFLEX TEST 3.1.8. ACQUIRING H-REFLEX RESPONSES
34 36 37 38 40 41 44 44
4. EMG TESTS
45
4.1. SETTING UP THE EMG TEST MENU
45
4.1.1. SETTING UP AN EMG 4.1.2. ACQUIRING AN EMG
45 46
4.2. FREE RUN EMG FEATURES
48
4.2.1. REVIEWING FREE RUN EMG 4.2.2. ANALYZING TURNS AND AMPLITUDE
48 48
4.3. TRIGGERED EMG FEATURES
49
4.3.1. ACTIVATING TRIGGERS 4.3.2. SAVING A TEST 4.3.3. ANALYZING AND REVIEWING MOTOR UNITS
49 50 50
5. OTHER TESTS
50
5.1. OTHER TESTS
50
5.1.1. SETTING UP EVOKED POTENTIALS (EP) 5.1.2. CONDUCTING EVOKED POTENTIALS 5.1.3. SEP TESTS 5.1.4. SUGGESTED SEP PROTOCOLS 5.1.5. DERMATOMAL SEPS
51 51 53 54 56
5.2. BLINK REFLEX
57
5.2.1. SETTING UP A BLINK REFLEX TEST 5.2.2. CONDUCTING BLINK REFLEX TESTS
57 57
5.3. INCREMENTAL STIMULATION
58
5.3.1. SETTING UP INCREMENTAL STIM TESTS 5.3.2. CONDUCTING INCREMENTAL STIM TESTS 5.3.3. INCREMENTAL STIM TEST VALUES
58 60 60
5.4. HEART RATE VARIABILITY (HRV)
61
5.4.1. SETTING UP HRV STIMULATION TESTS 5.4.2. CONDUCTING HRV TESTS 5.4.3. SYMPATHETIC SKIN RESPONSE
61 61 63
5.5. MULTI-CHANNEL EMG/IOM 5.6. MULTI-CHANNEL NERVE CONDUCTIONS 5.7. THE P300 TEST
64 65 65
5.7.1. GETTING STARTED 5.7.2. RUNNING THE TEST 5.7.3. HINTS AND FEATURES 5.7.4. EXPLANATION OF OPTIONS
65 67 67 69
5.8. THE ERG (ELECTRORETINOGRAM) TEST
71
5.8.1. ABOUT THE ERG TEST 5.8.2. CONFIGURING THE ERG PROTOCOL
71 71
5.9. THE EOG (ELECTRO-OCULOGRAM) TEST
72
5.9.1. ABOUT THE EOG TEST
72
Page 3
User Manual
XLTEK NeuroMax
5.9.2. CONFIGURING THE E0G PROTOCOL
72
6. AV STIM 1000
73
6.1. AV STIM 1000 FRONT PANEL 6.2. AV STIM 1000 REAR PANEL 6.3. CONNECTING THE AV STIM 1000 6.4. WARNINGS AND CAUTIONS
73 73 74 75
6.4.1. W ARNINGS 6.4.2. CAUTIONS
75 75
6.5. CALIBRATION AND MAINTENANCE 6.6. SUGGESTED AEP PROTOCOLS
76 77
6.6.1. AEP IMPEDANCE CHECK
77
6.7. SUGGESTED VEP PROTOCOLS
77
6.7.1. VEP IMPEDANCE CHECK
78
6.8. ACQUIRING A FULLFIELD VEP RESPONSE 6.9. ACQUIRING A HEMIFIELD VEP RESPONSE
78 79
7. SETTING THE DEFAULTS
80
7.1. TEST MENU PARAMETERS
80
7.1.1. CREATING AN ELECTROMYOGRAPHY SUITE
81
7.2. EDITING TEST DEFAULTS
81
8. ADMINISTRATIVE FUNCTIONS
83
8.1. ADMINISTRATIVE FUNCTIONS
83
8.1.1. PATIENT DIRECTORY 8.1.2. MEMORY MANAGEMENT 8.1.3. BATCH PRINT 8.1.4. SYSTEM OPTIONS 8.1.5. EDIT REPORT FORMAT 8.1.6. EDIT SITE NAME LIST 8.1.7. EDIT EMG NOTEPAD 8.1.8. EDIT PATIENT INFORMATION FIELDS 8.1.9. CHANGING THE DATE ON STORED DATA
84 85 86 86 87 87 87 87 87
8.2. MANAGING THE REPORTS
88
8.2.1. RICH TEXT FILES (RTF)
88
8.3. EXTENDED WARRANTY SERVICES INFORMATION
89
9. APPENDIX 1: IN-SERVICE CHECKLIST
90
10. APPENDIX 2: TROUBLESHOOTING/MAINTENANCE
93
10.1. SOFTWARE
93
10.1.1. CHECKING SOFTWARE VERSION ON NEUROMAX 10.1.2. UPDATING SOFTWARE
93 93
10.2. NERVE CONDUCTIONS
93
10.2.1. CREATING A NCS
94
Page 4
User Manual
XLTEK NeuroMax
10.2.2. EDITING AN NCS 10.2.3. SETTING UP STIMULUS AND RECORDING SITES
94 94
10.3. ELECTROMYOGRAPHY
95
10.3.1. CREATING AN EMG SUITE 10.3.2. USING AN EMG SUITE
95 95
10.4. ADMINISTRATIVE FUNCTIONS
96
10.4.1. EDITING STUDIES 10.4.2. INTERPRETATION MACROS
96 96
10.5. TROUBLESHOOTING
97
10.5.1. SIGNAL CLIPPING IN NCS AND EMG 10.5.2. STIMULUS ARTIFACT 10.5.3. ELECTRODE IMPEDANCE 10.5.4. SKIN PREPARATION 10.5.5. ELECTRODE TYPE AND PLACEMENT 10.5.6. STIMULATOR
97 97 98 99 99 100
10.6. AV STIM 10.7. PRINTING
103 103
10.7.1. NEUROMAX 10.7.2. PRINTER
103 103
10.8. RECOMMENDED USER PERFORMED MAINTENANCE
104
10.8.1. NEUROMAX ENCLOSURE 10.8.2. KEYBOARD 10.8.3. SCREEN 10.8.4. BACK PANEL/CONNECTORS 10.8.5. HEADBOX AND CABLE 10.8.6. STIMULUS PROBE AND CABLE 10.8.7. PRINTER AND CABLE 10.8.8. ELECTRODES AND ACCESSORIES
105 106 106 106 106 106 107 107
10.9. USER ADJUSTMENTS
107
10.9.1. NO ELECTRICAL STIMULUS 10.9.2. NO RESPONSE FROM ELECTRODES 10.9.3. LARGE STIMULUS ARTIFACT 10.9.4. NOISY DATA 10.9.5. UNIT DOES NOT POWER ON 10.9.6. ERROR MESSAGES
108 108 109 110 111 112
10.10. CHECKING CALIBRATION OF NEUROMAX
115
10.10.1. CHECKING THE STIMULATOR CALIBRATION: 10.10.2. CHECKING THE W AVEFORM
115 116
INDEX
117
Page 5
User Manual
XLTEK NeuroMax
NEUROMAX SAFETY AND STANDARDS CONFORMITY STANDARDS COMPLIANCE AND NORMATIVE REFERENCES EMG System, model "NeuroMax 1002" and “NeuroMax 1004"; rated 120/240V, 50/60Hz, 1.4A/0.7A; detachable cord connected; Class I;Type BF applied parts. 1. Type of protection against electric shock: Class I 2. Degree of protection against electric shock: Type BF The NeuroMax and its accessories have been designed to comply with the following national and international standards.
Table 1 – Safety Standard of Compliance and Normative References CAN/CSA C22.2 No 601.1-M90 CSA 601.1 Supplement 1:1994
CSA 601.1 Amendment 2:1998
UL Std No 2601-1 (2nd Edition)
Medical Electrical Equipment part 1: General requirements for Safety adopted IEC 601-1 2ed (90) Supplement No 1-94 to CAN/CSA C22.2 601.1-M90 Medical Electrical Equipment Part 1: General Requirements for Safety Amendment 2 to CAN/CSA C22.2 601.1-M90 Medical Electrical Equipment Part 1: General Requirements for Safety Medical Electrical Equipment part 1: General requirements for Safety
IEC 601-1:1988 + A1:1991 + A2:1995
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-2-40 (1998-02)
Medical Electrical Equipment part 2-40: Particular requirements for the Safety of Electromyographs and Evoked Response Equipment
Table 2 – EMC Standard of Compliance and Normative References EN 60601-1-2 :2001 (2nd Edition)
Medical Electrical Equipment, Part 1-2:General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
IEC 61000-4-2:1995 / EN 61000-42:2001
Electromagnetic compatibility (EMC) - Part 4-2: Testing and measurement techniques - Electrostatic discharge immunity test
IEC 61000-4-3:2002 / EN 61000-43:2006
Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test
IEC 61000-4-4:2004 / EN 61000-44:2004
Electromagnetic compatibility (EMC) - Part 4-4: Testing and measurement techniques - Electrical fast transient/burst immunity test
Page 6
User Manual
XLTEK NeuroMax
IEC 61000-4-5:1995 / EN 61000-45:2006
Electromagnetic compatibility (EMC) - Part 4-5: Testing and measurement techniques - Surge immunity test
IEC 61000-4-6:1996 / EN 61000-46:2007
Electromagnetic compatibility (EMC) - Part 4-6: Testing and measurement techniques - Immunity to conducted disturbances, induced by radio-frequency fields
IEC 61000-4-8 / EN 61000-4-8
Electromagnetic compatibility (EMC) - Part 4-8: Testing and measurement techniques - Power frequency magnetic field immunity test
IEC 61000-4-11:2004 / EN 61000-411:2004
Electromagnetic compatibility (EMC) - Part 4-11: Testing and measurement techniques - Voltage dips, short interruptions and voltage variations immunity tests
IEC 61000-3-2:2005 / EN 61000-32:2006
Electromagnetic compatibility (EMC) - Part 3-2: Limits Limits for harmonic current emissions
IEC 61000-3-3:1994 / EN 61000-33:1995 +A1:2001 +A2:2005
Electromagnetic compatibility (EMC) - Part 3-3: Limits Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems
CISPR 11:2004 / EN 55011:1998+A1:1999 & A2:2002
Industrial, Scientific and Medical (ISM) Radio-Frequency Equipment - Electromagnetic Disturbance Characteristics Limits and Methods of Measurement
ANSI C63.4:2003
American National Standard for Methods of Measurement of Radio-Noise Emissions from Low-Voltage Electrical and Electronic Equipment in the Range of 9 KHz to 40 GHz
CISPR 16-1-1: 2003
Specification for radio disturbance and immunity measuring apparatus and methods. Part 1-1: Measuring Apparatus
CISPR 16-2-1:2004
Specification for radio disturbance and immunity measuring apparatus and methods. Part 1-2: Conducted disturbances
FCC CRF47 Part 15, Subpart B Class A
Federal Communications Commission (FCC) - Unintentional Radiators
Page 7
User Manual
XLTEK NeuroMax
DECLARATION OF COMPLIANCE FOR IEC 60601-1-2 Table 1 - Electromagnetic Emissions Guidance and manufacturer’s declaration – electromagnetic emissions The NeuroMax 1004/1002 is intended for use in the electromagnetic environment specified below. The customer or the user of the NeuroMax 1004/1002 should assure that it is used in such an environment.
Emissions test RF emissions
Compliance
Electromagnetic environment - guidance
Group 1
The NeuroMax 1004/1002 uses RF energy only for its internal function. Therefore, its RF emissions are very low and not likely to cause any interference in nearby electronic equipment
Class A
The NeuroMax 1004/1002 is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
CISPR 11 RF emissions CISPR 11
Harmonic Class A emissions IEC 61000-3-2 Voltage Complies fluctuations/ flicker emissions IEC 61000-3-3
Page 8
User Manual
XLTEK NeuroMax
Table 2 - Electromagnetic Immunity
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The NeuroMax 1004/1002 is intended for use in the electromagnetic environment specified below. The customer or the user of the NeuroMax 1004/1002 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic Discharge (ESD)
±6 kV contact
Compliance level Complies
±8 kV air
IEC 61000-4-2 Electrostatic fast transient/burst
±2 kV for power supply lines
IEC 61000-4-4
±1 kV for input/output lines
Surge
±1 kV differential mode
IEC 61000-4-5
Complies
IEC 61000-4-11
<5% UT (>95% dip in UT) for 0.5 cycle
Mains power quality should be that of a typical commercial or hospital environment.
Complies
Mains power quality should be that of a typical commercial or hospital environment. If the user of the NeuroMax 1004/1002 requires continued operation during power mains interruption, it is recommended that the NeuroMax 1004/1002 be powered from an uninterruptible power supply or a battery.
Complies
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
40% UT (60% dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 cycles
Power frequency (50/60 Hz) magnetic field
<5% UT (>95% dip in UT) for 5 sec 3 A/m
IEC 61000-4-8 NOTE: UT is the AC supply voltage prior to application of the test level.
Page 9
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. Mains power quality should be that of a typical commercial or hospital environment.
Complies
±2 kV common mode Voltage dips, short interruptions and voltage variations on power supply input lines
Electromagnetic environment - guidance
User Manual
XLTEK NeuroMax
Table 3 - Electromagnetic Immunity – for EQUIPMENT and SYSTEMS that are not LIFESUPPORTING
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The NeuroMax 1004/1002 is intended for use in the electromagnetic environment specified below. The customer or the user of the NeuroMax 1004/1002 should assure that it is used in such an environment
Immunity test
Compliance Electromagnetic environment - guidance level
IEC 60601 test level
Portable and mobile RF communications equipment should be used no closer to any part of the NeuroMax 1004/1002, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz
150kHz to 80MHz
Complies 80MHz to 800MHz
Radiated RF
3 V/m 80 MHz to 2.5 GHz
IEC 61000-4-3
800MHz to 2.5GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic sitea should be less than the compliance level in each frequencyb. Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the NeuroMax 1004/1002 is used exceeds the applicable RF compliance level above, the NeuroMax 1004/1002 should be observed to verify normal operation. If abnormal operation is observed, additional measures may be necessary, such as re-orienting or relocating the NeuroMax 1004/1002.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Page 10
User Manual
XLTEK NeuroMax
Table 4 - Recommended Separation Distances
Recommended separation distances between portable and mobile RF communications equipment and the NeuroMax 1004/1002 The NeuroMax 1004/1002 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the NeuroMax 1004/1002 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the NeuroMax 1004/1002 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter (W)
Separation distance according to frequency of transmitter (m) 150 kHz to80MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
DECLARATION OF COMPLIANCE FOR FCC Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. Warning: Changes or modifications not expressly approved by the manufacturer could void the user's authority to operate the equipment.
Page 11
User Manual
XLTEK NeuroMax
1. WELCOME TO THE NEUROMAX Congratulations, you have purchased the NeuroMax from XLTEK, one of the world's top manufacturers of neurodiagnostic equipment and software. The NeuroMax offers full-featured EMG in a simple, easy to use, and affordable line of instruments. XLTEK NeuroMax is available in two and fourchannel format and provides you with the highest performance and greatest reliability of any EMG instrument on the market. From simple NCS studies through to a complex quantitative EMG, the XLTEK NeuroMax is the system you can rely on.
XLTEK NeuroMax 1002 CE Two-channel neurodiagnostic for basic and advanced NCS, EMG, and EP. Incorporates a large 12.1" XGA (1024 x 768) screen. Advanced engineering and the latest in technology allow for improved performance and speed, advanced NCS capabilities, advanced EMG capabilities, and advanced MultiChannel Studies.
XLTEK NeuroMax 1004 CE Four-channel neurodiagnostic for basic and advanced NCS, EMG, and EP. Incorporates a large 12.1" XGA (1024 x 768) screen. Advanced engineering and the latest in technology allow for improved performance and speed, advanced NCS capabilities, advanced EMG capabilities, and advanced MultiChannel Studies.
XLTEK Headboxes and Accessories XLTEK provides two and four-channel amplifiers, AV
Stimulators, and a full range of accessories.
Page 12
User Manual
XLTEK NeuroMax
1.1. INTRODUCTION XLTEK is commitment to continual product
improvements and quality design to meet the needs of our clients. We thus encourage all feedback and any suggestions you have regarding any aspect of the EMG system, the manual, our line of accessories, and our support services.
1.1.1. INTENDED USE The NeuroMax EMG is intended to be used as a clinical electromyograph to acquire, display, store and archive neurophysiological signals.
1.2. CUSTOMER SERVICE XLTEK is committed to providing you with support so you can operate the NeuroMax with ease and confidence. If you need help, follow these steps to find a solution: Step 1: Document the Incident Carefully document the incident step by step. If possible, note error messages, the type of test and what you did before the problem occurred, including the sequence of keystrokes. Step 2: Shut Down the NeuroMax Sometimes you need to shutdown completely in order to solve a problem. Press the MAIN MENU key to return to the Main Menu. Turn the power off to the unit. Make sure all cables are connected and intact. Turn the power back on. Step 3: Call or Email Technical Support First, write down the serial number (located on the back of the computer). Then phone XLTEK‟s Customer Support at 1-800-303-0306 or contact XLTEK Technical Support at [email protected].
Page 13
User Manual
XLTEK NeuroMax
1.3. USING THE MANUAL XLTEK is committed to providing you with clear
instructions and unqualified support so you can operate our equipment with ease and confidence. The manual presents step-by-step instructions which take you through the testing, customizing, and operation of the equipment and software so that our system meets your specific needs. It will guide you through the acquisition of a patient record and its review, storage, and recall. You will learn how to develop a report and to archive studies for future reference. For your convenience, a thorough Table of Contents is provided which details the topics covered in each chapter. An Index is also available at the back of the manual. The in-depth procedures describe the detailed operation and customization of the EMG system. The procedures, which are accompanied by detailed graphics, are designed to tailor the system to your specific circumstances. We encourage all users to explore the manual and to take advantage of everything that XLTEK has designed the EMG system to do. When going through the procedures, we recommend that you read the whole section before starting the sequence. Please follow the instructions carefully. We have also placed TIPS and NOTES alongside the instructions. These will list Hot Keys, operation tips, shortcuts, and testing information. TIPS
NOTES
Page 14
User Manual
XLTEK NeuroMax
2. THE XLTEK NEUROMAX The NeuroMax is designed to conduct a range of tests, including Nerve Conduction Studies (NCS), Electromyography Studies (EMG), and other tests such as Evoked Potentials, Blink Reflex, Incremental Stimulation, and Heart Rate Variability. This chapter takes you through the Warnings and Cautions you need to observe while operating the NeuroMax and introduces you to its basic functions.
2.1. OPERATING CONDITIONS The NeuroMax is designed for optimum performance under safe conditions. To ensure the safety of the operator and of the patient, please read the following sections carefully.
2.1.1. ENVIRONMENT PARAMETERS Temperature Range: +10 to + 40 degrees Celsius Relative Humidity Range: 30 % to 75 % Atmospheric Pressure Range: 700 hPa to 1,060 hPa Altitude: To a maximum of 4600 meters above sea level
2.1.2. TRANSPORT AND STORAGE PARAMETERS Ambient Temperature Range: – 40 to +70 degrees Celsius Relative Humidity Range: 10% to 100%, including condensation Atmospheric Pressure Range: 500hPa to 1,060hPa
Page 15
User Manual
XLTEK NeuroMax
2.2. SPECIFICATIONS Specification
Value General Description Dimensions 4” (H) x 13” (W) x 13” (D) Weight 11.4lbs (5.2kg) Colour White Injection moulded PC ABS Materials chassis External Connectors Headbox, Headphones, Printer, Footswitch (pneumatic) 4 (NeuroMaxCE 1004) Channels 2 (NeuroMaxCE 1002) 60 kHz Sampling Rate Size and Type Resolution
Display 12.1” active matrix colour LCD screen 1024 x 768
Electrical Specifications Maximum Rated 92 VA Input Power Heat Loss 92 Watts Maximum Supply ± 10% for either 50 Hz or 60 Current Tolerance Hz transformer* Electrical Supply ± 10% for either 50 Hz or 60 Frequency Tolerance Hz transformer* Insulation Class and Class 1, Type BF Type Power Input Voltage, 120 VAC, 60 Hz, 1A Frequency, and 230 VAC, 50 Hz, 0.4 A Rated Current Mains Connection Protectively grounded detachable power supply cord ~105 dB CMRR Noise
~2.9 nV / √Hz
Page 16
User Manual
XLTEK NeuroMax
Stimulator Specifications Maximum Simulator ~450 V Voltage Maximum Stimulator 100 mA Current Stimulator Current Accuracy
± 1%
Stimulator Duration Accuracy
± 1%
Voltage Accuracy on ± 1% Screen * A power supervisor circuit shuts off power if insufficient power is present.
2.3. WARNINGS AND CAUTIONS The following Warnings and Cautions are marked with a and must be followed very closely to ensure the safety of both the patient and the user of the NeuroMax and the AV Stim. It is therefore important to read and observe ALL of the Warnings and Cautions before attempting to use the NeuroMax and the AV Stim. If there is any malfunction or perceived malfunction of the system, please call an authorized XLTEK service representative immediately at 1-800-3030306. All internal system checks and/or service must only be conducted by an authorized XLTEK service representative. IMPORTANT: 'SYSTEM' REFERS TO THE NEUROMAX, AV STIM, AND ALL ACCESSORIES ATTACHED TO IT.
The NeuroMax and AV Stim carry the ordinary equipment classification (as per IEC 529) for the level of protection against ingress of liquids. They are not drip or splash proof. Regarding protection against electrical shock, the NeuroMax and AV Stim are classified as Class I devices (as per EN 60601-1). The NeuroMax
Page 17
User Manual
XLTEK NeuroMax
requires a properly grounded electrical outlet. The mode of operation for both the NeuroMax and AV Stim is continuous operation.
2.3.1. WARNINGS Warnings MUST be followed when using the equipment. Warnings apply to conditions which can injure the patient and/or the operator.
WARNING: Care must be taken in the delivery of any level of stimulus. If used improperly, injury may be caused to the patient. The delivery of stimulus to the patient is done through the Start/Stop key on the NeuroMax and the stimulus probe.
WARNING: The level of stimulator intensity is controlled by the Up/Down arrows on the NeuroMax and the stimulation probe and is displayed on the active test screen in mA. Close attention must be paid to the level of stimulus intensity and the duration of the impulse. This unit has enough electrical power to harm a patient if used improperly.
WARNING: This system must only be plugged into a properly grounded electrical outlet. The internal isolation transformer of the system must not be bypassed, under any circumstances.
WARNING: Hazardous voltages are exposed when the lid of the NeuroMax is removed.
WARNING: This system is not suitable for use in the presence of flammable mixtures. The system is not AP or APG rated.
WARNING: Do not turn the system on until all cable connections are made and their integrity is checked.
Page 18