ZOLL

AEDs

AED PLUS Quick Guide Rev J

Quick Guide

2 Pages

Warning!  Use the AED Plus unit only as described in this manual. Improper use of the device can cause death or injury. DO NOT use or place the AED Plus unit in service until you have read the AED Plus Operator’s and Administrator’s Guides. DO NOT use or place the AED Plus unit in service if the unit’s status indicator window (located on the left side of the handle) displays a red “X”. DO NOT use or place the AED Plus unit in service if the unit emits a beeping tone. Connect the electrode cable to the AED Plus unit after installing batteries. Keep the electrode cable connected to the AED Plus unit at all times. This device should only be used by properly trained individuals. Only use electrodes labeled “Infant/Child” on children less than 8 years old or weighing less than 55 lbs (25 kg). Use CPR-D padz if patient is older than 8 years or weighs more than 55 lbs (25 kg). Always stand clear of patient when delivering treatment. Defibrillation energy delivered to the patient may be conducted through the patient’s body and cause a lethal shock to those touching the patient. DO NOT touch the electrode surfaces, the patient, or any conductive material touching the patient during ECG analysis or defibrillation. Move patient away from electrically conductive surfaces prior to use of equipment. DO NOT use the unit near or within puddles of water. Keep the patient as motionless as possible during ECG analysis. DO NOT use the unit near flammable agents, such as gasoline, oxygen-rich atmospheres, or flammable anesthetics. Avoid radio frequency interference from high-power sources that might cause the defibrillator to interpret cardiac rhythms incorrectly by turning off cell phones and 2-way radios. Disconnect non-defibrillation protected electronic devices or equipment from patient before defibrillation. Dry victim’s chest, if wet, before attaching electrodes.  Set-up and Check-out Procedure: 1. Insert 10 new batteries into AED Plus unit. 2. Connect electrode cable to AED Plus unit and pack sealed electrodes inside unit cover. Close cover. 3. Turn unit on and wait for “Unit OK” audio message. Verify that unit issues appropriate “Adult Pads” or “Pediatric Pads” audio message. 4. Turn unit off. 5. Wait 2 minutes. Verify that green check symbol ( ) appears in status indicator window (located on left side of handle) and that unit does not emit a beeping tone. 6. Place AED Plus unit in service. 7. Check AED Plus unit periodically to ensure that green check symbol ( ) appears in status indicator window.  Battery Replacement Replace batteries before expiration date or if unit prompts. Use only type 123A lithium manganese dioxide batteries from recommended manufacturers. Remove all batteries from battery compartment and discard before installing any new batteries. Insert 10 new batteries into battery well. Do not use old batteries. Press button in battery well only after installation of new batteries.  Cleaning Clean and disinfect unit with soft, damp cloth using 90% isopropyl alcohol or soap and water, or chlorine bleach (30 ml/liter water). Do not immerse any part of the unit in water. Do not use ketones (MEK, acetone, etc.). Avoid using abrasives (e.g., paper towels) on the LCD display, if so equipped. Do not sterilize the unit.  Automated External Defibrillator  Operator’s Guide AHA/ERC 2005  Apply freshly opened and undamaged electrodes, within the electrode expiration date, to clean and dry skin to minimize burning. DO NOT place electrodes directly over the patient’s implanted pacemaker. Pacemaker stimuli may degrade the accuracy of ECG rhythm analyses or the pacemaker may be damaged by defibrillator discharges. Check labeling inside the ZOLL AED Plus cover before using the cover as a Passive Airway Support System (PASS) device to ensure it is intended for this use. DO NOT use the Passive Airway Support System (PASS) if there is a suspected head or neck injury. Place the patient on a firm surface before performing CPR. DO NOT recharge, disassemble, or dispose of batteries in fire. Batteries may explode if mistreated.  TROUBLESHOOTING Problem  Recommended Action  Self-test failed.  Manually test by pressing and holding the ON/OFF button for more than 5 seconds. If unit fails test again, remove from service.  “Change batteries” prompt.  Replace all batteries at the same time.  Red “X” in Status Indicator window OR beeping noise when unit is OFF.  Perform manual test. Check to see if cable is attached properly to unit. Replace batteries. If unit still does not operate correctly, remove from service.  DO NOT use or stack the AED PLUS unit with other equipment. If the unit is used or stacked with other equipment, verify proper operation prior to use.  Caution! DO NOT disassemble the unit. A shock hazard exists. Refer all servicing to qualified personnel. Use only commercially available type 123A lithium manganese dioxide batteries. Discard batteries properly after removal from unit. Use only batteries from recommended manufacturers. See the AED Plus Administrator’s Guide (P/N 9650-0301-01) for a list of recommended battery manufacturers. If the device is stored outside the recommended environmental conditions, the electrode pads and/or batteries may be damaged or their useful life reduced. The CPR-D Padz Electrode can be connected to other ZOLL defibrillators with Multifunction Cables. Defibrillation can be administered when connected to other ZOLL defibrillators.The CPR function does not operate with any device other than the AED Plus defibrillator.  Red “X” in Status Indicator window when unit is ON.  Power cycle the unit. If Red “X” is still present in Status Indicator window, remove unit from service.  Technical Service Department ZOLL Medical Corporation  ZOLL Medical Europe  269 Mill Rd. Chelmsford, MA 01824-4105 978-421-9655 • 800-348-9011 Fax: 978-421-0010  Edisonring 3a 6669 NA Dodewaard Netherlands Tel: +31 488-411-183 Fax: +31 488-411-187 International Service: Contact your local distributor.  Important! This symbol indicates that an AED Plus unit is equipped for treating adult and pediatric patients. An AED Plus unit without this symbol is not equipped to treat pediatric patients and will NOT work with pedi•padz II ™ pediatric electrodes. To upgrade an AED Plus unit for use with ZOLL pedi•padz II pediatric electrodes, contact ZOLL Medical Corporation or an authorized ZOLL distributor for information on the ZOLL AED Plus Pediatric Upgrade Kit.  For Technical support or repair:  0197 Federal (U.S.A.) law restricts this device for sale to or on the order of a physician.  © ZOLL Medical Corp. Inc. 2008 REF 9650-0300-01 Rev. J
File Type: PDF
File Size: 274 KB
File Name: ZOLL - AED PLUS Quick Guide Rev J.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions