ZOLL
AED 3 Service Manual Rev F
Service Manual
32 Pages
Preview
Page 1
Service Manual
9650-000751-01 Rev. F
The issue date for the ZOLL AED 3 Service Manual (REF 9650-000751-01 Rev. F) is October 2023. Copyright © 2023 ZOLL Medical Corporation. All rights reserved. CPR-D-padz, Pedi-padz, PlusTrac, Program Management Onboard, RescueNet, Stat-padz, Uni-padz, ZOLL, and ZOLL AED 3 are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and/or other countries. Revision F of this manual applies to software revision 04.03.500.6452 or higher.
ZOLL Medical Corporation 269 Mill Road Chelmsford, MA USA 01824-4105
0123
ZOLL International Holding B.V. Einsteinweg 8A 6662 PW Elst Netherlands
Table of Contents Preface ... 1 Conventions ... 2 Safety Considerations ... 2
Chapter 1 Maintenance ... 3
Self Tests ... 3 Self Test Functions ... 4 Automatic Self-Test Transfer ... 4 Cleaning the AED ... 5 Manual Maintenance Tests ... 6 Physical Inspection Test ... 6 Test Equipment ... 7 Power Supply Test (Optional) ... 8 Equipment Needed ... 8 Set Up ... 8 Pediatric Mode Test ...10 Equipment Needed ...10 Set Up ...10 Energy Test ...11 Equipment Needed ...11 Set Up ...11 Wi-Fi Test ...14 Equipment Needed ...14 Set Up ...14 USB Test ...16 Audio Recording Test (ZOLL AED 3 BLS models only) ...17 Equipment Needed ...17 Set Up ...17
Chapter 2 Troubleshooting ... 19 Troubleshooting the AED ...19
Chapter 3 Functional Description ... 21 High Voltage Circuit Section ...21 Low Voltage Analog Circuit Section ...21 Digital Section ...22 Isolation 22 Readiness Indicator Board ...23 USB Connection ...23 Other External Connections ...23
9650-000751-01 Rev. F
ZOLL AED 3 Service Manual
i
Safety Function ...23 Safety on High Voltage Capacitor Charge ...23 Safety on Discharge ...24 Safety on Digital Circuits ...24 Power Supply ...25
Appendix Hardware/System Architecture ... 27 Overview ...27 ZOLL AED 3 Hardware/System Architecture ...28
Checklist ... 29
ii
ZOLL AED 3 Service Manual
9650-000751-01 Rev. F
Preface The ZOLL AED 3TM defibrillator is an Automated External Defibrillator (AED) that can be used for both adult and child victims of sudden cardiac arrest. The defibrillator incorporates a sequence of audio and visual prompts to help rescuers follow established AHA/ERC/ILCOR Guidelines for use of AEDs. The ZOLL AED 3 Service Manual is to be used primarily by the ZOLL® Technical Service Department in conjunction with the ZOLL AED 3 Operator’s Manual and the ZOLL AED 3 Administrator’s Guide. The ZOLL AED 3 defibrillator does not require preventative maintenance due to its automatic self tests. However, this manual provides optional maintenance tests you can perform periodically according to local protocols. This manual contains the following sections: Preface-Contains safety warnings and an overview of the manual’s contents. Review this section thoroughly before attempting to use or service the ZOLL AED 3. Chapter 1-Maintenance provides the recommended procedures for evaluating the condition and performance of the ZOLL AED 3 defibrillator. Chapter 2-Troubleshooting describes possible problems and solutions and lists the error log messages that indicate the AED requires technical service. Chapter 3-Functional Description provides technical descriptions of the ZOLL AED 3 defibrillator modules. Appendix-Hardware/System Architecture contains a diagram of the ZOLL AED 3 hardware/system. Checklist-Located on the back page of this manual and used to record the results of the maintenance tests in Chapter 1.
9650-000751-01 Rev. F
ZOLL AED 3 Service Manual
1
Conventions Throughout this document, text prompts and voice prompts are indicated by capital italicized letters, such as CALL FOR HELP. WARNING! Warning statements describe conditions or actions that can result in personal injury or death. CAUTION Caution statements describe conditions or actions that can result in damage to the defibrillator.
NOTE Notes contain important additional information on servicing the defibrillator.
Safety Considerations Only qualified personnel should service a ZOLL AED 3 defibrillator. Before using this defibrillator, read the ZOLL AED 3 Operator’s Manual. Before servicing any equipment, review these safety considerations and read this manual carefully. WARNING! The ZOLL AED 3 defibrillator can generate as much as 2250 volts with sufficient current to cause lethal shocks.
Before discharging the defibrillator, warn everyone near the equipment to STAND CLEAR. Do not recharge, disassemble, or dispose of the battery in fire. The battery might explode if mishandled.
2
ZOLL AED 3 Service Manual
9650-000751-01 Rev. F
Chapter 1 Maintenance The ZOLL AED 3 defibrillator automatically performs functional testing during periodic self tests. However, this chapter provides optional maintenance tests you can perform periodically according to local protocols. The following tests and maintenance instructions are included in this chapter: • Self Tests
• Cleaning • Manual Maintenance Tests
Self Tests The ZOLL AED 3 defibrillator performs the following self tests to verify AED integrity and its readiness for emergency use: • Battery Installation
• Power On • Manual • Automatic (Daily or Weekly, based on configuration) • Automatic Monthly Following successful completion of all self tests, the status indicator displays a
AED Passed Self Test green check () to show that all tests passed and that the AED is ready to use.
If the status indicator is blank following the completion of any self test, the ZOLL AED 3 defibrillator is not ready for use and may be defective. Remove the AED Failed Self Test
9650-000751-01 Rev. F
ZOLL AED 3 Service Manual
3
AED from service and consult the section “Chapter 2 Troubleshooting” on page 17 to help determine the problem.
Self Test Functions The following functions are verified during self tests. Table 1 on page 5 lists the functions that are included in each self test. • Battery Capacity: Verifies that the battery usage indicator shows adequate battery capacity remaining.
• Defibrillation Pads Connection: Verifies that the defibrillation pads are properly preconnected to the device.
• Defibrillation Pads (CPR Uni-padzTM)/Battery Expiration
Verification: Verifies that the defibrillation pads are within their expiration date and the battery is within its install-by date. Note: CPR Uni-padz are the only electrodes that have an expiration date that is recognized by the ZOLL AED 3 device.
• ECG Circuitry: Tests that the ECG signal acquisition and processing electronics are functional.
• Defibrillator Charge and Discharge Circuitry: Verifies that the device’s defibrillator electronics are functional and can charge and discharge at 2 joules. The following tests also include a separate 200 joule charge/ discharge test: Battery Installation Self Test, Automatic Monthly Test.
• Microprocessor Hardware/Software: Verifies proper function of the AED microprocessor electronics and the integrity of its software.
• CPR Circuitry and Sensor: Determines that CPR monitoring and
compression depth detection are functional (if defibrillation pads with CPR functionality are connected).
• Audio Circuitry: Verifies that voice prompts are working.
Automatic Self-Test Transfer All ZOLL AED 3 defibrillators come standard with Program Management OnboardTM intelligent monitoring technology to transfer self-test data. If configured, the AED can automatically transfer self-test information over Wi-Fi to the ZOLL PlusTracTM AED program management system or to another AED program management provider.
4
ZOLL AED 3 Service Manual
9650-000751-01 Rev. F
Table 1: Self Test Functions Battery Installation Self Test
Power On Self Test
Manual Self Test
Automatic Self Test
Automatic Monthly Test
Battery Capacity
Defibrillation Pads Connection
Defibrillation Pads (CPR Uni-padz)/ Battery Expiration
ECG Circuitry
Defibrillator Charge and Discharge Circuitry (2 Joule Charge/ Discharge Test)
Microprocessor Hardware/Software
CPR Circuitry and Sensor (if defibrillation pads with CPR functionality are connected)
Audio Circuitry
200 Joule Charge/ Discharge Test
Cleaning the AED After each use, clean and disinfect the defibrillator with a soft, damp cloth using either 90% isopropyl alcohol, or soap and water. You can also use a chlorine bleach and water mixture (30 ml/liter water) to clean the defibrillator (except on the contacts and connectors). NOTE Wipe down the defibrillator with water after using any cleaning solution. Chlorine residue that remains on the LCD screen may cause damage. DO NOT:
9650-000751-01 Rev. F
ZOLL AED 3 Service Manual
5
• Immerse any part of the defibrillator in water. • Use chlorine mixture on contacts or connectors; this will degrade the contacts over time. • Use ketones (MEK, acetone, etc.) to clean the defibrillator. • Use abrasives (e.g., paper towel) on the display window or graphic screen. • Sterilize the defibrillator.
Manual Maintenance Tests This section provides instructions for performing a physical inspection of and maintenance tests on the ZOLL AED 3. The maintenance tests verify that the AED is working properly and is ready for use. Use the checklist on the back page of this manual to record the results (pass/fail rating) of these maintenance tests.
Physical Inspection Test Check the following:
Pass/Fail
HOUSING 1.
Is the AED clean, undamaged, and free of excessive wear?
o o
2.
Does the handle work properly?
o o
3.
Are input connectors clean and undamaged?
o o
4.
Are there any cracks or loose parts in the housing?
o o
5.
Does the front panel/display have any damage or cracks?
o o
ELECTRODES 6.
Is the cable free of cracks, cuts, or exposed or broken wires?
o o
7.
Are the defibrillation pads connected to the AED and sealed in their package? Replace defibrillation pads if expired.
o o
BATTERY 8.
Is the battery correctly seated in the battery well?
o o
9.
Is the battery housing free of cracks and damage?
o o
10.
Are the connectors free of damage and excessive wear?
o o
11.
Is the battery pack within the install by date? Replace if the install by date has passed.
o o
GENERAL 12.
6
Turn on the AED and verify the green check () indicates ready for use; then turn it off.
ZOLL AED 3 Service Manual
o o
9650-000751-01 Rev. F
Test Equipment The following is a list of equipment that you need to perform the maintenance tests in this chapter. You can substitute an equivalent device for a listed device; however, not all simulators and analyzers will produce the same results. Be sure to follow the manufacturer's recommendation for conducting the maintenance tests. • Fluke Impulse 7000 Defib Analyzer (or equivalent)
• Stopwatch • ZOLL AED 3 Uni-padz Defib Test Cable Kit (REF 8900-000268) • Blank USB Flash Drive (minimum 512 MB, USB 1.1 or 2.0 compliant) • Computer (Wi-Fi capable with ZOLL RescueNet® software loaded and speakers if testing audio recording)
• ZOLL AED 3 Battery Pack • Battery Test Fixture (REF 9100-000369-TF) (optional) • Power Supply 10A • Digital Multi-Meter (Agilent 34401A or equivalent) • Wi-Fi Access Point NOTE If the Battery Test Fixture is not available, the tests in this section can be conducted with a ZOLL AED 3 Battery Pack.
9650-000751-01 Rev. F
ZOLL AED 3 Service Manual
7
Power Supply Test (Optional) This test is comprised of two tests: the Off Current test and the On Current test. In order to perform the Off Current test, two measurements are required: one with the fixture installed in the device and one for the test fixture. The difference of the two readings is the Off Current Value. NOTE If a Power Supply is not available for this test, record N/A for this test on the Maintenance checklist on the back page of this manual.
Equipment Needed Battery Test Fixture, Power Supply, Digital Multi-Meter.
Set Up 1. On the Digital Multi-Meter, set the function to DC Amps. 2. Connect the positive lead from the I jack to the positive current post (Amp meter) on the Battery Test Fixture. 3. Connect the negative lead from the LO jack to the negative current post on the Battery Test Fixture. 4. Set the Power Supply to 12.0V, then turn it off. 5. Connect the leads from the Battery Test Fixture to the proper polarity of the Power Supply and install the Battery Test Fixture into the battery well.
1.
Action
Expected Result(s)
Pass/Fail
Turn on the Power Supply.
Verify that the AED runs a self test (approximately one minute) and then powers off.
o o
Note: AED may issue the messages CHANGE BATTERY and UNIT FAIL. Disregard the messages and continue the test. 2.
Verify that the AED is turned off to perform the Off Current test. On the Battery Test Fixture, turn the I Measure switch to the sleep mode position.
Write down the current measurement on the Digital Multi-Meter. Value #1______________mA
After the measurement is taken, turn the I Measure switch back to the off position. 3.
Remove the Battery Test Fixture from the unit. Turn the I Measure switch to the sleep mode position. After the measurement is taken, turn the I Measure switch back to the off position.
8
Write down the current measurement on the Digital Multi-Meter. Value #2______________mA
ZOLL AED 3 Service Manual
9650-000751-01 Rev. F
Action
Expected Result(s)
4.
Subtract Value #2 from Value #1. This is the device Off Current Value.
Verify that the value is between 004mA and 0.015mA.
o o
5.
Press the On/Off button on the AED to turn it on to perform the On Current test.
Verify that the AED powers on in Rescue Mode.
o o
On the Battery Test Fixture, turn the I Measure switch to the sleep mode position.
Observe that the current measurement on the Digital Multi-Meter is between 200mA to 440mA.
o o
Verify that the AED issues the prompt UNIT POWERING OFF.
o o
6.
Pass/Fail
Note: The AED may issue the CHANGE BATTERY message. Disregard the message and continue the test.
After the measurement is taken, turn the I Measure switch back to the off position. 7.
Press the On/Off button on the AED to turn it off.
9650-000751-01 Rev. F
ZOLL AED 3 Service Manual
9
Pediatric Mode Test Equipment Needed Power Supply, Battery Test Fixture, Defib Test Cable, Fluke Impulse 7000 Defib Analyzer (or equivalent). NOTE The test values in the following procedure are based of the default pediatric energy settings (50, 70, and 85 joules) into a 50 ohm load. If the Analyzer you are using has a different impedance value, refer to the table “Delivered Energy at Each Defibrillator Setting into a Range of Loads” on page 13.
Set Up Connect the Defib Test Cable to the Defib Analyzer, and set the waveform to VFIB. Action
Expected Result(s)
1.
Connect the ZOLL AED 3 Uni-padz Defib Test Cable Kit to the AED, then press the On/Off button to turn on the AED.
Verify that the AED issues the prompts UNIT OK, ADULT PATIENT SELECTED followed by IF PATIENT IS A CHILD, PRESS CHILD BUTTON.
o o
2.
Press the Child button.
Verify that the Child button LED illuminates, and that the AED issues the prompt CHILD PATIENT SELECTED.
o o
3.
(First shock energy test. First pediatric shock energy is 50 joules.)
Verify that the AED issues the prompt DON'T TOUCH PATIENT ANALYZING, followed by, SHOCK ADVISED, DON'T TOUCH PATIENT, and PRESS FLASHING SHOCK BUTTON.
o o
4.
Press the flashing shock button.
Verify that AED prompts SHOCK DELIVERED.
o o
5.
Verify that the energy delivered into analyzer is 43 joules to 58 joules.
o o
6.
Verify that the AED issues the prompt START CPR.
o o
Verify that the AED issues the prompt UNIT POWERING OFF.
o o
7.
10
Press the On/Off button to turn off the AED.
ZOLL AED 3 Service Manual
Pass/Fail
9650-000751-01 Rev. F
Energy Test Equipment Needed Power Supply, Battery Test Fixture, ZOLL AED 3 Uni-padz Defib Test Cable Kit, Fluke Impulse 7000 Defib Analyzer (or equivalent). NOTE The test values in the following procedure are based of the default adult energy settings (120, 150, and 200 joules) into a 50 ohm load. If the Analyzer you are using has a different impedance value, refer to the table “Delivered Energy at Each Defibrillator Setting into a Range of Loads” on page 13.
Set Up This test continues from the previous test. The ZOLL AED 3 Uni-padz Defib Test Cable Kit is connected to the Defib Analyzer, and the waveform is set to VFIB. Action
Expected Result(s)
Pass/Fail
1.
Disconnect the ZOLL AED 3 Unipadz Defib Test Cable Kit from the AED, then press the On/Off button to turn on the AED.
Verify that the AED issues the prompts UNIT OK and PLUG IN PADS CABLE.
o o
2.
Connect the ZOLL AED 3 Uni-padz Defib Test Cable Kit to the AED.
Verify that the AED issues the voice prompt ADULT PATIENT SELECTED, followed by IF PATIENT IS A CHILD, PRESS CHILD BUTTON.
o o
3.
(First shock energy test. First adult shock energy is 120 joules.)
Verify that the AED issues the voice prompts DON’T TOUCH PATIENT, ANALYZING, followed by SHOCK ADVISED, DON’T TOUCH PATIENT, and PRESS FLASHING SHOCK BUTTON.
o o
Note: For Automatic models, instead of PRESS FLASHING SHOCK BUTTON the AED issues the prompt SHOCK WILL BE DELIVERED IN THREE, TWO, ONE. 4.
Press the flashing shock button. Note: For Automatic models, the AED automatically shocks the patient.
Verify that the AED issues the prompt SHOCK DELIVERED.
o o
Note: For Automatic models, the AED discharges the charge.
5.
Verify that the energy delivered into the analyzer is 100 joules to 135 joules.
o o
6.
Verify that the AED issues the prompt START CPR.
o o
9650-000751-01 Rev. F
ZOLL AED 3 Service Manual
11
7.
Action
Expected Result(s)
Hold the CPR sensor (attached to the ZOLL AED 3 Uni-padz Defib Test Cable Kit) in the palm of your hand. Move your hand in front of you in an up and down motion of approximately 2 inches at a rate of 100bpm. (Conduct test for about 20 seconds.)
Verify that the AED displays the CPR depth compression indicator.
8.
9.
(Second shock energy test. Second adult shock energy is 150 joules.)
Pass/Fail
o o
Note: The ZOLL AED 3 BLS model only displays the CPR depth compression indicator when the display is configured for lay user.
Verify that the AED issues the prompt STOP CPR after 2 minutes (default setting).
o o
Verify that the AED issues the voice prompts DON’T TOUCH PATIENT, ANALYZING, followed by SHOCK ADVISED, DON’T TOUCH PATIENT, and PRESS FLASHING SHOCK BUTTON.
o o
Note: For Automatic models, instead of PRESS FLASHING SHOCK BUTTON, the AED issues the prompt SHOCK WILL BE DELIVERED IN THREE, TWO, ONE. 10.
Press the flashing shock button. Note: For Automatic models, the AED automatically shocks the patient.
Verify that the AED issues the prompt SHOCK DELIVERED.
o o
Note: For Automatic models, the AED discharges the charge.
11.
Verify that the energy delivered into analyzer is 123 joules to 167 joules.
o o
12.
Verify that the AED issues the prompt START CPR; after 2 minutes (default setting), verify that the AED issues the prompt STOP CPR.
o o
Verify that the AED issues the voice prompts DON’T TOUCH PATIENT, ANALYZING, followed by SHOCK ADVISED, DON’T TOUCH PATIENT, and PRESS FLASHING SHOCK BUTTON.
o o
13.
(Third shock energy test. Third adult shock energy is 200 joules.)
Note: For Automatic models, instead of PRESS FLASHING SHOCK BUTTON, the AED issues the prompt SHOCK WILL BE DELIVERED IN THREE, TWO, ONE.
12
ZOLL AED 3 Service Manual
9650-000751-01 Rev. F
14.
Action
Expected Result(s)
Press the flashing shock button.
Verify that the AED issues the prompt SHOCK DELIVERED.
Note: For Automatic models, the AED automatically shocks the patient. 15. 16.
Press the On/Off button to turn off the AED.
Pass/Fail
o o
Note: For Automatic models, the AED discharges the charge. Verify that the energy delivered into analyzer is 178 joules to 240 joules.
o o
Verify that the AED issues the prompt UNIT POWERING OFF.
o o
Table 2: Delivered Energy at Each Defibrillator Setting into a Range of Loads Load
Selected Energy 50 J
70 J
85 J
120 J
150 J
200 J
25
37 J
54 J
57 J
86 J
109 J
139 J
50
50 J
69 J
80 J
118 J
145 J
209 J
75
61 J
82 J
97 J
134 J
166 J
196 J
100
60 J
84 J
95 J
142 J
165 J
194 J
125
57 J
80 J
91 J
133 J
155 J
178 J
150
65 J
91 J
103 J
124 J
145 J
192 J
175
60 J
84 J
95 J
116 J
135 J
177 J
Accuracy
±15%
±15%
±15%
±15%
±15%
±15%
9650-000751-01 Rev. F
ZOLL AED 3 Service Manual
13
Wi-Fi Test Equipment Needed Battery Test Fixture, Power Supply, Wi-Fi Access Point. NOTE To test the server connectivity, the AED must be connected to an access point. For more information, refer to “Setting up your Wireless Configuration” in the ZOLL AED 3 Administrator's Guide.
Set Up • Set the Power Supply to 12 volts. • Connect the Battery Test Fixture to the Power Supply. Action
Expected Result(s)
1.
Press and hold the On/Off button for more than 5 seconds to turn on the AED in AED Management (NonRescue) Mode.
Verify that the AED displays the Select Operating Mode window.
o o
2.
Press the Device Configuration icon
Verify that the AED displays the Configuration window.
o o
Verify that the AED displays the Advanced Configuration window.
o o
Verify that the AED displays the Wi-Fi Settings window.
o o
Verify that the AED displays the Device History Settings window.
o o
Verify that the AED displays the TESTING SERVER CONNECTIVITY message along with the IN PROGRESS message.
o o
Verify that the AED displays the message TESTING SERVICE CONNECTIVITY, COMPLETE.
o o
Verify that the AED displays the Advanced Configuration window.
o o
( 3.
4.
).
) and Press the Supervisor icon ( enter the six-digit passcode (default passcode is 123456) and press OK. Press the Wi-Fi icon (
).
5.
In the Device History Settings field, select Edit.
6.
In the Selected Access Point field, choose an access point.
7.
Press the Test Wi-Fi button.
8.
9.
14
Press the OK button, press Cancel, and then press OK.
ZOLL AED 3 Service Manual
Pass/Fail
9650-000751-01 Rev. F
USB Test Equipment Needed USB Flash Drive (Blank), Power Supply, Battery Test Fixture, computer
Set Up This test continues from the previous test. The AED should display the Advanced Configuration window.
1.
2.
Action
Expected Result(s)
Insert the USB Flash Drive into the USB connector on the back of the AED.
Verify that the blinking USB device icon
Press the Export Files icon ( select Clinical Archives (all).
)and
(
) appears in the status bar at the bottom of the display screen. (The blinking icon switches to steady status.)
Pass/Fail
o o
Verify that the green check appears next to the Clinical Archives (all) option.
o o
3.
Press Save (USB).
When the transfer is complete, verify that the AED displays the message DATA TRANSFER COMPLETE.
o o
4.
Press OK, then press Cancel. Remove the USB Flash Drive.
) no longer Verify that the USB icon ( appears in the status bar at the bottom of the display screen.
o o
5.
Press the On/Off button to turn off the AED.
Verify that the AED issues the prompt UNIT POWERING OFF.
o o
6.
Insert USB Flash Drive in the computer USB port. Open File Explorer, and select the USB drive.
Verify that there is at least 1 CRD file present. (The AED can store up to 2 CRD files.)
o o
9650-000751-01 Rev. F
ZOLL AED 3 Service Manual
15
Audio Recording Test (ZOLL AED 3 BLS models only) If the defibrillator does not have audio recording functionality enabled, record N/A for this test on the Maintenance checklist on the back page of this manual.
Equipment Needed USB Flash drive (containing Clinical Event data from the previous USB test), Power Supply, Battery Test Fixture, PC (with RescueNet Code Review version =>5.72).
Set Up Audio recording must be set to ON; the default for this configuration setting is ON. Refer to the Configuration Settings in the ZOLL AED 3 Administrator’s Guide if you need to change this setting to ON. NOTE The ZOLL AED 3 BLS defibrillator (with the Audio Recording option) is capable of recording up to 120 minutes of continuous ECG/Audio and clinical event data during the rescue of a patient case.
16
Action
Expected Result(s)
1.
Insert USB Flash Drive (containing the Clinical Event data from the previous USB test) in the computer USB port. Open File Explorer, and select the USB drive.
Verify that CRD files are present.
o o
2.
Double-click the desired file and then click Yes in the message box.
Verify that the RescueNet Code Review Standard Edition window appears.
o o
3.
Select 3-Magnified ECG at the top of the screen and click Play.
Verify that you can hear all voice prompts and events (from the computer speaker) that were recorded by the AED.
o o
4.
Press the On/Off button to turn off the AED.
Verify that the AED displays the message UNIT POWERING OFF.
o o
ZOLL AED 3 Service Manual
Pass/Fail
9650-000751-01 Rev. F