ZOLL
Propaq MD Operators Guide Rev D March 2019
Operators Guide
338 Pages
Preview
Page 1
Propaq® MD Operator’s Guide
Includes Real CPR Help ® and See-Thru CPR ®
REF: 9650-1806-05 Rev. D SW VER: 02.32.06.00
The issue date for the Propaq MD Operator's Guide (REF 9650-1806-05 Rev. D) is March, 2019. Copyright © 2019 ZOLL Medical Corporation. All rights reserved. AutoPulse, Shock Sync, ResQCPR, AED Plus, AED Pro, RapidShock, CPR-D-padz, Pedi-padz, OneStep, Real CPR Help, Rectilinear Biphasic, RescueNet, See-Thru CPR, Stat-padz, SurePower, and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and/or other countries. All other trademarks are the property of their respective owners. Masimo, Rainbow, SET, SpCO, SpMet, SpHb, SpOC, and PVI are trademarks or registered trademarks of Masimo Corporation in the United States and/or other countries. Propaq is a registered trademark of Welch Allyn Inc. Oridion Microstream FilterLine® and Smart CapnoLine® are registered trademarks of Medtronic plc.
ZOLL Medical Corporation 269 Mill Road Chelmsford, MA USA 01824-4105 ZOLL International Holding B.V. Newtonweg 18 6662 PV ELST The Netherlands
0123
Indications for Use Defibrillator Function The Propaq MD system is indicated for defibrillation on victims of cardiac arrest where there is apparent lack of circulation as indicated by: • Unconsciousness. • Absence of breathing. • Absence of pulse.
The Propaq MD system in the Manual mode is indicated for synchronized cardioversion of certain atrial or ventricular arrhythmias. A qualified physician must decide when synchronized cardioversion is appropriate. The Propaq MD system Semiautomatic and Manual mode is indicated for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol. The Propaq MD system Semiautomatic and Manual mode is indicated for adult and pediatric patients.
Electrocardiogram (ECG) Monitoring The Propaq MD system is indicated to monitor and/or record 3-, 5-, or 12-lead electrocardiogram (ECG) waveform and heart rate, and to alarm when heart rate is above or below limits set by the operator. ECG monitoring is indicated for patients from newborn (neonate) to adult, with and without heart dysfunction.
CPR Monitoring The Propaq MD system is indicated to provide visual and audio feedback via the CPR Monitoring function, designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended depth and rate of 2 inches (5 cm) and 100 compressions per minute.
External Transcutaneous Pacing The Propaq MD system is indicated for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation. External Pacing is indicated for pediatric and adult patients.
Non-Invasive Blood Pressure Monitoring The Propaq MD system is indicated for use to make non-invasive measurements of arterial pressure and heart rate, and to alarm if either parameter is outside of the limits set by the user. The non-invasive blood pressure monitoring feature is indicated for patients from newborn (neonate) to adult.
Temperature Monitoring The Propaq MD system is indicated for use to make continuous temperature measurements of rectal, esophageal, or surface temperatures, and to alarm if the temperature is outside of the limits set by the user. The temperature monitoring feature is indicated for use in patients from newborn (neonate) to adult.
SpO2 Monitoring The Propaq MD system is indicated for use for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and/or carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin (SpHb), oxygen content (SpOC), pleth variability index (PVI), and perfusion index (PI) via the pulse Co-oximeter and accessories. The pulse Co-oximeter and accessories are indicated for use on adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused, in hospitals, hospital-type facilities, or in mobile environments.
Respiration Monitoring The Propaq MD system is indicated for use to continuously monitor respiration rate and to alarm if the rate falls outside of the range set by the operator. Because the measurement method actually measures respiratory effort, apnea episodes with continued respiratory effort (such as obstructive apnea) may not be detected. It is not intended to be used as an apnea monitor. The respiration monitoring feature is indicated for use on patients from newborn (neonate) to adult.
CO2 Monitoring The Propaq MD system is indicated for use in continuous noninvasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and breath rate. The CO2 monitoring feature is indicated for use on patients from newborn (neonate) to adult.
Invasive Pressure Monitoring The Propaq MD system is indicated for use to display and make continuous invasive pressure measurements via a compatible pressure transducer. The invasive pressure monitoring feature is indicated for use on patients from newborn (neonate) to adult.
12-Lead Analysis The Propaq MD system is indicated for use in acquiring, analyzing and reporting physiological data via 12-lead ECG Analysis, and to provide interpretation of the data for consideration by caregivers. The 12-lead ECG Analysis feature is indicated for use on adults (> 18 years of age).
Web Console The Propaq MD system is indicated for the remote display of physiological data displayed on connected Propaq MD systems via the Web Console feature, including electrocardiogram (ECG), non-invasive blood pressure (NIBP), temperature, and heart rate.
Contraindications Semiautomatic Operation Contraindications for Use The rhythm analysis function may not reliably identify ventricular fibrillation in the presence of an implanted pacemaker. Inspection of the electrocardiogram and clinical evidence of cardiopulmonary arrest should be the basis for any treatment of patients with implanted pacemakers. Do not use the rhythm analysis function during patient movement on a stretcher. A patient must be motionless during ECG analysis. Do not touch the patient during analysis. Cease all movement of the stretcher prior to analyzing the ECG.
General Information Federal (U.S.A.) law restricts this defibrillator to sale by or on the order of a physician.
TABLE OF CONTENTS
Table of Contents Chapter 1
General Information
Product Description ... 1-1 Propaq MD Optional Features ... 1-3 How to Use This Manual... 1-3 Operator’s Guide Updates ... 1-3 Unpacking... 1-4 Symbols Used on the Equipment ... 1-4 Conventions ... 1-7 Intended Use ... 1-8 Propaq MD Product Functions... 1-8 Defibrillator Function ... 1-8 Defibrillator Output Energy ... 1-8 External Pacemaker ... 1-9 ECG Monitoring ... 1-9 Electrodes ... 1-9 Batteries ... 1-10 Ready For Use (RFU) Indicator ... 1-11 Safety Considerations... 1-12 Warnings... 1-12 General ... 1-12 ECG Monitoring ... 1-13 Defibrillation ... 1-14 Pacing ... 1-15 CPR ... 1-16 Pulse Oximeter ... 1-16 Noninvasive Blood Pressure ... 1-16 IBP ... 1-17 CO2 ... 1-17 Respiration ... 1-17 Ferromagnetic Equipment ... 1-18 Battery ... 1-18 Operator Safety ... 1-18 Patient Safety ... 1-19 Cautions... 1-20 Restarting the Defibrillator ... 1-21 FDA Tracking Requirements ... 1-21 Notification of Adverse Events ... 1-22 Software License ... 1-22 Service... 1-23 The ZOLL Serial Number... 1-24
9650-1806-05 Rev. D
Propaq MD Operator’s Guide
i
TABLE OF CONTENTS
Chapter 2
Product Overview
Defibrillator Controls and Indicators... 2-1 The Front Panel ... 2-2 Display Screen ... 2-5 Battery Status and Auxiliary Power Indicators ... 2-6 Patient Cables and Connectors ... 2-7 AutoPulse with Shock Sync™ ... 2-10 External Paddles ... 2-11 Auxiliary Power Adapter ... 2-14 Navigating the Display Screen... 2-15 Quick Access Keys ... 2-15 Navigation Keys ... 2-18 Display Brightness ... 2-18 Common Tasks ... 2-19 Setting the Date and Time ... 2-19 Changing the Display Brightness ... 2-20 Replacing a Battery Pack on the Propaq MD ... 2-21 Using Treatment Buttons ... 2-22
Chapter 3
Monitoring Overview
Propaq MD Monitoring Functions ... 3-1 ECG ... 3-2 Heart Rate ... 3-2 Respiration Rate ... 3-2 Temperature ... 3-2 Invasive Pressures (IBP) ... 3-2 Non-Invasive Blood Pressure (NIBP) ... 3-2 Capnography (CO2) ... 3-3 Pulse Oximetry (SpO2) ... 3-3 Monitoring Display Options ... 3-4 Configuring the Waveform Display ... 3-7
Chapter 4
Trends
Displaying the Trends Status Window ... 4-2 Printing Trend Information ... 4-3 Printing Trend Summary ... 4-3 Printing The 10 Most Recent Trend Snapshots ... 4-3 Printing an Individual Trend Snapshot ... 4-3 Printing Specific Trend Snapshots ... 4-3 Changing the Trends Status Window Display... 4-4 Continuous Waveform Recording ... 4-5
Chapter 5
Alarms
Visual Alarm Indicators ... 5-2
ii
www.zoll.com
9650-1806-05 Rev. D
TABLE OF CONTENTS
Audible Alarm Indicators ... 5-2 Alarm Indicator Self-Test... 5-2 Patient Alarm Display... 5-3 Life Threatening Rhythm Alarms... 5-3 Equipment Alert Display... 5-4 Responding to Active Alarms -- Silencing the Alarm ... 5-5 Re-enabling an Alarm ... 5-5 Latching Alarms ... 5-5 Pausing (Suspending) Alarms ... 5-6 Alarm Reminders ... 5-6 Alarm Options ... 5-7 Selecting Default Alarm Limits ... 5-7 Setting Alarm Limits Relative to the Patient -- Stat Set Option ... 5-8
Chapter 6
Monitoring ECG
ECG Monitoring Setup... 6-3 Preparing the Patient for Electrode Application ... 6-3 Applying Electrodes to the Patient ... 6-4 Connecting the ECG Cable To the Propaq MD Unit ... 6-6 Selecting ECG Waveforms for Display ... 6-6 Selecting the Waveform Trace Size ... 6-8 ECG Monitoring and Pacemakers ... 6-9 ECG System Messages... 6-10
Chapter 7
Monitoring Respiration (Resp) and Heart Rate (HR)
Respiration/Breath Rate Meter ... 7-2 Using Impedance Pneumography to Measure Respiration ... 7-2 Configuring Respiration (RR/BR) Alarms and Settings... 7-4 Enabling/Disabling RR/BR Alarms and Setting Alarm Limits ... 7-4 Using the Resp Parameter Control Panel ... 7-5 Heart Rate Meter ... 7-6 Configuring Heart Rate (HR) Meter Alarms ... 7-6 Enabling/Disabling HR Alarms and Setting Alarm Limits ... 7-6 Life Threatening Rhythm Alarms ... 7-8 Using the Heart Rate Parameter Control Panel ... 7-10 RESP System Message... 7-10
Chapter 8
Monitoring Non-Invasive Blood Pressure (NIBP)
How does NIBP Work? ... 8-3 The NIBP Numeric Display ... 8-4 NIBP Setup and Use... 8-4 Selecting the NIBP Cuff ... 8-5 Connecting the NIBP Cuff... 8-6 Applying the Cuff to the Patient ... 8-9 9650-1806-05 Rev. D
Propaq MD Operator’s Guide
iii
TABLE OF CONTENTS
Ensuring Correct Cuff Inflation Settings... 8-10 Configuring NIBP Alarms and Settings ... 8-11 Enabling/Disabling NIBP Alarms and Setting Alarm Limits ... 8-11 Using the NIBP Parameter Control Panel ... 8-14 NIBP System Messages ... 8-16
Chapter 9
Monitoring CO2
Overview... 9-1 CO2 Monitoring Setup and Use ... 9-3 Selecting the CO2 Sampling Line ... 9-3 Connecting the CO2 Sampling Lines ... 9-4 Applying a FilterLine Set ... 9-5 Applying a Smart CapnoLine Nasal or Nasal/Oral Cannula ... 9-6 Measuring CO2... 9-7 Setting CO2 and Respiration Rate Alarms ... 9-8 Enabling/Disabling Alarms and Setting CO2 Alarm Limits ... 9-8 Using the CO2 Parameter Control Panel ... 9-10 System Messages ... 9-11 Patents... 9-12
Chapter 10
Pulse CO-Oximetry (SpO2)
Warnings -- SpO2 General ... 10-3 Cautions... 10-5 SpO2 Setup and Use ... 10-6 SpO2 Accuracy Specifications ... 10-7 Selecting the SpO2 Sensor ... 10-11 Applying the SpO2 Sensor ... 10-12 Applying a Two-Piece Single-Use Sensor/Cable ... 10-12 Applying a Reusable Sensor/Cable ... 10-14 Cleaning and Reuse of Sensors ... 10-14 Connecting the SpO2 Sensor ... 10-15 Displaying Measurements ... 10-15 Enabling/Disabling SpO2 Alarms and Setting Alarm Limits ... 10-16 Setting Upper and Lower SpO2 Alarm Limits ... 10-16 Setting Upper and Lower SpCO and SpMet Alarm Limits ... 10-17 Setting Upper and Lower SpHb Alarm Limits ... 10-17 Setting Upper and Lower SpOC Alarm Limits ... 10-18 Setting Upper and Lower PVI Alarm Limits ... 10-18 Setting Upper and Lower PI Alarm Limits ... 10-18 Using the SpO2 Parameter Control Panel ... 10-19 SpCO and SpMet, and SpHb, SpOC, PVI and PI Monitoring ... 10-19 Specifying the SpO2 Averaging Time ... 10-19 Selecting the SpO2 Sensitivity ... 10-19 Selecting SpHb Venous Mode ... 10-20
iv
www.zoll.com
9650-1806-05 Rev. D
TABLE OF CONTENTS
Specifying the SpHb Averaging Time ... 10-20 Selecting the Heart Rate/ Pulse Rate (HR/PR) Tone ... 10-20 SpO2 System Messages ... 10-20 Functional Testers and Patient Simulators... 10-22 Patents... 10-22
Chapter 11
Monitoring Invasive Pressures (IBP)
Invasive Pressure Transducers ... 11-1 IBP Setup... 11-2 Attaching the Invasive Pressure Transducer ... 11-3 Zeroing the Transducer... 11-4 Rezeroing a Transducer ... 11-5 Displaying IBP Measurements... 11-5 Conditions Affecting IBP Measurements ... 11-5 Enabling/Disabling IBP Alarms and Setting Alarm Limits... 11-6 Setting Upper and Lower Systolic (SYS) Alarm Limits ... 11-6 Setting Upper and Lower Diastolic (DIA) Alarm Limits ... 11-7 Setting Upper and Lower Mean Arterial Pressure (MEAN) Alarm Limits ... 11-7 Setting IBP Source Label ... 11-8 IBP System Messages... 11-8
Chapter 12
Monitoring Temperature
Temperature Monitoring Setup ... 12-1 Selecting and Applying Temperature Probes ... 12-1 Connecting the Temperature Probe ... 12-2 Displaying Temperature ... 12-2 Enabling/Disabling Temperature Alarms and Setting Alarm Limits ... 12-3 Setting Upper and Lower Temperature Alarm Limits ... 12-3 Setting Upper and Lower Temperature Alarm Limits ... 12-4 Selecting the Temperature Label ... 12-4 Temperature System Messages ... 12-5
Chapter 13
Automated External Defibrillator (AED)
AED Operation... 13-2 Determine Patient Condition Following Medical Protocols ... 13-2 Begin CPR Following Medical Protocols ... 13-2 Prepare Patient ... 13-3 1 Turn on unit ... 13-4 2 Analyze ... 13-5 3 Press SHOCK ... 13-7 RapidShock ... 13-8 The Shock Conversion Estimator ... 13-8 Operating Messages ... 13-8 Audio and Display Messages ... 13-9 9650-1806-05 Rev. D
Propaq MD Operator’s Guide
v
TABLE OF CONTENTS
Switching to Manual Mode Operation ... 13-10
Chapter 14
12-Lead ECG Interpretive Analysis
Entering Patient Information ... 14-3 Entering the Patient Name and ID ... 14-3 Entering Patient Age and Gender ... 14-4 12-Lead ECG Monitoring Setup... 14-4 Preparing the Patient for Electrode Application ... 14-5 Applying Electrodes to the Patient ... 14-5 Connecting the 12-Lead Cable ... 14-7 Observing the 12-Lead Waveform Traces ... 14-7 12-Lead Interpretive Analysis ... 14-8 Fault Conditions Affecting 12-Lead Interpretive Analysis ... 14-11 Printing 12-Lead Waveform Traces ... 14-12 12-Lead Print and Display Options ... 14-13 Selecting 12-Lead Acquire ... 14-13 Specifying the Number of 12-Lead Print Copies ... 14-13 Specifying the 12-Lead Print Format ... 14-14 Printing 10 Seconds of Waveform Traces ... 14-16 Specifying the 12-Lead Frequency Response ... 14-16 Enabling 12-Lead Analysis ... 14-16 Enabling Interpretive Text ... 14-16
Chapter 15
Manual Defibrillation
Emergency Defibrillation Procedure with Paddles ... 15-1 Determine the Patient’s Condition Following Local Medical Protocols ... 15-3 Begin CPR Following Local Medical Protocols ... 15-3 Turn On Unit ... 15-3 1 Select Energy Level ... 15-3 2 Charge Defibrillator ... 15-5 3 Deliver Shock ... 15-7 Emergency Defibrillation Procedure with Hands-Free Therapy Electrodes... 15-8 Determine the Patient’s Condition Following Local Medical Protocols ... 15-8 Begin CPR Following Medical Protocols ... 15-8 Prepare Patient ... 15-8 Turn On Unit ... 15-9 1 Select Energy Level ... 15-9 2 Charge Defibrillator ... 15-10 3 Deliver Shock ... 15-11 Internal Paddles ... 15-11 Verification Prior to Use ... 15-12 Synchronized Cardioversion ... 15-13 Synchronized Cardioversion Procedure ... 15-14 Determine the Patient’s Condition and Provide Care Following Local Medical Protocols ... 15-14 Prepare Patient ... 15-14 Turn On Unit ... 15-14 vi
www.zoll.com
9650-1806-05 Rev. D
TABLE OF CONTENTS
Press the Sync Button ... 15-14 1 Select Energy Level ... 15-15 2 Charge Defibrillator ... 15-15 3 Deliver Shock ... 15-16
Chapter 16
Advisory Defibrillation
Advisory Defibrillation Procedure... 16-2 Determine the Patient’s Condition Following Local Medical Protocols ... 16-2 Begin CPR Following Local Medical Protocols ... 16-2 Prepare Patient ... 16-2 1 Turn on unit ... 16-3 2 Press ANALYZE ... 16-4 3 Press SHOCK button ... 16-5
Chapter 17
Analysis/CPR Protocol Defibrillation
Analysis/CPR Protocol Defibrillation Procedure ... 17-2 Determine the Patient’s Condition Following Local Medical Protocols ... 17-2 Begin CPR Following Local Medical Protocols ... 17-2 Prepare Patient ... 17-2 1 Turn on unit ... 17-3 2 Press ANALYZE ... 17-4 3 Press SHOCK button ... 17-5 RapidShock ... 17-6 The Shock Conversion Estimator ... 17-6
Chapter 18
External Pacing
External Pacing... 18-2 Pacer Modes ... 18-2 Pacing in Demand Mode ... 18-2 Determine Patient Condition and Provide Care Following Local Medical Protocols. 18-2 Prepare the Patient ... 18-2 1 Turn On Unit ... 18-2 2 Apply ECG Electrodes/Hands-Free Therapy Electrodes ... 18-3 3 Press Pacer button ... 18-3 4 Set Mode ... 18-4 5 Set Pacer Rate ... 18-4 6 Turn On Pacer ... 18-4 7 Set Pacer Output ... 18-4 8 Determine Capture ... 18-5 9 Determine Optimum Threshold ... 18-5 Pacing in Fixed Mode ... 18-6 1 Turn On Unit ... 18-6 2 Apply ECG Electrodes/Hands-Free Therapy Electrodes ... 18-6 3 Press Pacer button ... 18-7 4 Set Mode ... 18-7
9650-1806-05 Rev. D
Propaq MD Operator’s Guide
vii
TABLE OF CONTENTS
5 Set Pacer Rate ... 18-7 6 Turn On Pacer ... 18-7 7 Set Pacer Output ... 18-8 8 Determine Capture ... 18-8 9 Determine Optimum Threshold ... 18-8 Pediatric Pacing ... 18-9 Pace Fault ... 18-9
Chapter 19
Real CPR Help
CPR Voice Prompts (Adult Only) ... 19-3 CPR Metronome ... 19-3 Constant Metronome Enabled ... 19-3 Constant Metronome Disabled ... 19-4 FULLY RELEASE Prompt ... 19-5 CPR Dashboard ... 19-5 CPR Rate and Depth Measurements ... 19-5 CPR Release Indicator (Adult Only) ... 19-6 Chest Compression Indicator (Adult Only) ... 19-6 CPR Countdown Timer ... 19-6 CPR Idle Time Display ... 19-6 CPR Compression Bar Graph (Adult Only) ... 19-7 Real CPR with AutoPulse ... 19-8 Real CPR with the ResQCPR™ System ... 19-8
Chapter 20
See-Thru CPR
Using See-Thru CPR ... 20-2 Examples ... 20-3
Chapter 21
Patient Data
Storing Data... 21-1 Log Capacity Indicator ... 21-2 Capturing a Data Snapshot ... 21-2 Reviewing and printing snapshots ... 21-2 Treatment Summary Report ... 21-3 Printing Treatment Summary Report ... 21-3 Transferring Data to a USB Device ... 21-4 Clearing the Log ... 21-5
Chapter 22
Communications
The Wireless Icon ... 22-2 The Wireless Menu... 22-4 Selecting a Pre-Configured Access Point Profile ... 22-5 Creating a Temporary Access Point Profile ... 22-6 Bluetooth Device Pairing ... 22-10 viii
www.zoll.com
9650-1806-05 Rev. D
TABLE OF CONTENTS
Sending a 12-Lead Report... 22-12 Sending Disclosure Logs ... 22-13 Transmitting Readiness Test Logs ... 22-14 Communications System Messages... 22-14 Transmission Status Icons... 22-15
Chapter 23
Printing
Printing Patient Data ... 23-1 Printer Setup ... 23-2 Automatic Prints ... 23-2 Printing Waveforms ... 23-2 Printing Reports ... 23-3 Printing Trends ... 23-4 Printing Readiness Test Logs ... 23-5
Chapter 24
Maintenance
Daily/Shift Check Procedure ... 24-2 Inspection ... 24-2 Defibrillator/Pacing Test with Hands-Free Therapy Electrodes... 24-3 Defibrillator Testing with External Paddles... 24-5 Recommended Minimum Preventive Maintenance Schedule ... 24-6 Annually ... 24-6 Guidelines for Maintaining Peak Battery Performance ... 24-6 Cleaning instructions ... 24-7 Cleaning the NIBP Blood Pressure Cuff ... 24-7 Cleaning SpO2 Sensors ... 24-7 Cleaning Cables and Accessories ... 24-7 Loading Recorder Paper ... 24-8 Cleaning the Print Head ... 24-9
Appendix A
Specifications
Defibrillator...A-2 CPR Monitoring ...A-14 Monitor/Display ...A-14 Impedance Pneumography...A-15 Alarms...A-16 Recorder ...A-17 Battery ...A-17 General ...A-18 Pacer ...A-19 CO2 ...A-19 Pulse Oximeter ...A-20
9650-1806-05 Rev. D
Propaq MD Operator’s Guide
ix
TABLE OF CONTENTS
Non-Invasive Blood Pressure ...A-23 Invasive Pressures ...A-24 Temperature...A-25 Clinical Trial Results for the Biphasic Waveform ...A-26 Randomized Multicenter Clinical Trial for Defibrillation of Ventricular Fibrillation (VF) and Ventricular Tachycardia (VT) ...A-26 Randomized Multi-Center Clinical trial for Cardioversion of Atrial Fibrillation (AF) ...A-27 Pre-Clinical Study ...A-29 Published Clinical Data ...A-30 Synchronized Cardioversion of Atrial Fibrillation ...A-31 Electromagnetic Compatibility Guidance and Manufacturer’s Declaration ...A-32 Electromagnetic Immunity (IEC 60601-1-2) ...A-33 Electromagnetic Immunity ...A-34 ECG Analysis Algorithm Accuracy ...A-38 Clinical Performance Results ...A-38 Shock Conversion Estimator ...A-41 Wireless Output Guidance and Manufacturer’s Declaration ...A-43 RF Transmission Emitted (IEC 60601-1-2) ...A-43 FCC Notice ...A-43 Canada, Industry Canada (IC) Notices ...A-43
Appendix B Accessories
x
www.zoll.com
9650-1806-05 Rev. D
Chapter 1 General Information Product Description The ZOLL® Propaq MD® unit is an easy-to-use portable defibrillator that combines defibrillation and external pacing with the following monitoring capabilities: • ECG • CO-Oximeter • Non-invasive Blood Pressure • IBP • CO2 • Temperature • Respiration
The Propaq MD is a rugged, compact and lightweight unit that is designed for all resuscitation situations and is ideal for transport. The Propaq MD is powered by auxiliary power as well as an easily replaceable battery pack that quickly recharges when the Propaq MD is connected to auxiliary power. In addition, you can use a ZOLL SurePower™ Battery Charger Station to recharge and test the Propaq MD battery. Note:
The Propaq MD has defibrillation and pacing functionality, but some of the monitoring functions are optional features. See the complete list of options in Fig. 1-1. All features are included in this manual, but only purchased features will be available on your unit. The product is designed for use in both the hospital and the rugged EMS environment. The device is a versatile automated external defibrillator with manual capabilities and may be configured to operate in Manual, Advisory or Semiautomatic modes. It can be configured to start up in Semiautomatic (AED) mode or manual mode.
9650-1806-05 Rev. D
Propaq MD Operator’s Guide
1-1
Chapter 1 General Information
When operating in manual configuration, the device operates as a conventional defibrillator where the device’s charging and discharging is fully controlled by the operator. In Advisory and AED modes, some features of the device are automated and a sophisticated detection algorithm is used to identify ventricular fibrillation and determine the appropriateness of defibrillator shock delivery. Units may be configured to automatically charge, analyze, recharge, and prompt the operator to “PRESS SHOCK”, depending on local protocols. The unit is switched from AED mode to Manual mode for ACLS use by pressing the appropriate key on the front panel. The Propaq MD unit assists caregivers during cardiopulmonary resuscitation (CPR) by evaluating the rate and depth of chest compressions and providing feedback to the rescuer. Real CPR Help® requires the use of ZOLL authorized CPR electrodes. When using these pads, the displayed ECG waveforms can be adaptively filtered, using the See-Thru CPR® feature, to reduce the artifact caused by chest compressions. The unit has a large colorful LCD display of numerics and waveform data that provides easy visibility from across the room and at any angle. ECG, plethysmograph, and respiration waveform traces can be displayed simultaneously, giving easy access to all patient monitoring data at once. The display screen is configurable, so you can choose the best visual layout to fit your monitoring needs. The Propaq MD includes a transcutaneous pacemaker consisting of a pulse generator and ECG sensing circuitry. Pacing supports both demand and fixed noninvasive pacing for adult patients and adolescent, child, and infant pediatric patients. The Propaq MD has a patient data review and collection system that allows you to view, store, and transfer patient data. The Propaq MD unit contains a printer and USB port, which you can use to print the data and transfer it to a PC. The Propaq MD unit can send data through a wireless connection to remote locations. The Propaq MD unit can send 12-lead report snapshots (including trend data) or Disclosure Logs to a recipient via a ZOLL server. In addition full disclosure cases, which also contain trend data, can be automatically retrieved from the Propaq MD unit using ZOLL RescueNet or ePCR software.
1-2
www.zoll.com
9650-1806-05 Rev. D
How to Use This Manual
Propaq MD Optional Features The following features are optional in the Propaq MD unit. Note:
All features are included in this manual, but only purchased features will be available on your unit.
Figure 1-1 Propaq MD Optional Features Optional Feature 12-Lead ECG with Interpretation SpO2 (Masimo®) with SpCO® and SpMet® SpHb® (Masimo®) with SpOCTM, PVI® and PI NIBP (with Smartcuf® and SureBPTM) EtCO2 (Oridion® Microstream®) Temperature (2 Channels) Invasive Pressures (3 Channels) Advanced CPR Help Pacing Microphone
How to Use This Manual The Propaq MD Operator's Guide provides information operators need for the safe and effective use and care of the Propaq MD product. It is important that all persons using this device read and understand all the information contained within. Please thoroughly read the safety considerations and warnings section. Procedures for daily checkout and unit care are located in the Chapter 24, "Maintenance". This manual is supplemented by manual insert(s) for option(s) available on the Propaq MD. These insert(s) contain additional warnings, precautions, and safety-related information.
Operator’s Guide Updates An issue or revision date for this manual is shown on the front cover. If more than three years have elapsed since this date, contact ZOLL Medical Corporation to determine if additional product information updates are available. All users should carefully review each manual update to understand its significance and then file it in its appropriate section within this manual for subsequent reference. Product documentation is available through the ZOLL website at www.zoll.com. From the Products menu, choose Product Manuals.
9650-1806-05 Rev. D
Propaq MD Operator’s Guide
1-3
Chapter 1 General Information
Unpacking Carefully inspect each container for damage. If the shipping container or cushion material is damaged, keep it until the contents have been checked for completeness and the instrument has been checked for mechanical and electrical integrity. If the contents are incomplete, if there is mechanical damage, or if the defibrillator does not pass its electrical self-test, U.S.A. customers should call ZOLL Medical Corporation (1-800-348-9011). Customers outside of the U.S.A. should contact the nearest ZOLL authorized representative. If the shipping container is damaged, also notify the carrier.
Symbols Used on the Equipment Any or all of the following symbols may be used in this manual or on this equipment: Symbol
Description Class II equipment.
Dangerous voltage.
General warning: Observe and follow all safety signs.
Hot surface.
Fragile, handle with care.
Keep dry.
This end up.
Temperature limitation.
Conformité Européenne Complies with medical device directive 93/42/EEC.
Type B patient connection.
Type BF patient connection.
1-4
www.zoll.com
9650-1806-05 Rev. D
Symbols Used on the Equipment Symbol
Description Type CF patient connection.
Defibrillator-proof type BF patient connection.
Defibrillator-proof type CF patient connection.
Fusible link.
Equipotentiality.
Alternating current (ac).
Direct current (dc).
Auxiliary power adapter operation.
Caution, high voltage.
Earth (ground).
Negative input terminal.
Positive input terminal.
Power On/Off
Protective earth (ground).
9650-1806-05 Rev. D
Propaq MD Operator’s Guide
1-5
Chapter 1 General Information
2.
,I
Description 45 2%
)/.
Symbol
Contains lithium. Recycle or dispose of properly.
R EC Y C LE
,I )/.
Keep away from open flame and high heat.
Do not open, disassemble, or intentionally damage.
Do not crush.
Do not discard in trash. Recycle or dispose of properly.
Return to a collection site intended for waste electrical and electronic equipment (WEEE). Do not dispose of in unsorted trash.
Date of manufacture.
Use by.
Latex-free.
Do not reuse.
Do not fold.
Not sterile.
Manufacturer.
Authorized representative in the European Community.
1-6
www.zoll.com
9650-1806-05 Rev. D