4 Pages
Preview
Page 1
Ackermann
INTERCHANGEABLE SUCTION-IRRIGATION serious injury if not avoided.
INSTRUCTIONS FOR USE
Picture deviations are possible.
WARNING
SAFETY INSTRUCTIONS Before using the instrument Read and follow the instructions for use. Keep the instructions for use in a clearly visible place for later use. The following summarises the basic hazards identified with the product according to the type of hazard. Warnings that refer to a specific action can be found again directly with the corresponding action together with a classification of the hazard. / See “Explanation of symbols”. Risk of infection • Reprocess the instrument before use • Reprocess before returning to the manufacturer
Indicates a hazard that can lead to death or serious injury if not avoided.
invasive
device
for
INDICATION The instrument is designed for use in minimally invasive surgery. The instrument serves to irrigate the surgical site and/or to suction away any fluids and tissue fragments which may collect during the intervention.
again.
• Insert springs into the pressure pins.
CAUTION • The trumpet valve is pushed back upwards by the spring.
• Pull out the springs.
important Indicates measures to prevent damage to property. Note: For warnings of personal injury, the danger sign according to ANSI Z535.6 is used in these operating instructions.
Depending on the order, the instrument is supplied individually or in a set. • After unpacking, check all parts for transport damage. • In case of complaints, contact the manufacturer. / See manufacturer`s address.
PRODUCT DESCRIPTION INTENDED USE
Suction-irrigation system with luer lock
Suction-irrigation system with trumpet valve
Indicates a hazard that can lead to injury if not avoided.
SCOPE OF DELIVERY
Risk of injury • The instrument may only be used by qualified, medically and technically trained personnel. Operative use only by doctors. • Only use original accessories. • Do not use or repair damaged instruments. • Only combine with instruments that are compatible in their useful length and diameter
PURPOSE Reusable surgically temporary use.
* These instructions for use apply to several models.
The suction and irrigation tube is connected with the suction and irrigation handle and inserted via a trocar sleeve into the surgical site. Additional holes on the suction and irrigation tube enlarge the suction and irrigation range. Depending on the model, you can switch between the suction and irrigation function on the handle using either a control or a trumpet valve. Marks on the handle aid assignment. Suction-irrigation system with thread
Suction-irrigation system - connection via Luer,
APPLICATION
• Attach the valve to the handle as per the marking.
WARNING
illustration*
A Suction-Irrigation tube B Suction-Irrigation handle with control 1 Suction-Irrigation holes 2 LL connection 3 Combined Luer, tube connection 4 Control for suction-irrigation function 5 LL connection
Risk of infection from non-sterile instruments. • Reprocess the instrument before assembly. Suction and irrigation system with control
• Screw on the knurled rings.
• Unscrew the spring cap and remove the cock plug.
* These instructions for use apply to several models. Picture deviations are possible.
CAUTION
Note: The suction and irrigation tube cannot be disassembled. The instrument is inserted via a trocar sleeve. Select the trocar sleeve according to the diameter of the instrument. / More detailed information on this can be found in the „Compatibility“ section.
Tissue is punched when a trocar sleeve with an overly large diameter is used. • Only use trocar sleeves with a diameter which is slightly larger than that of the instrument. / See section „Compatibility“. Suction-irrigation system with control
TECHNICAL DATA
• Insert the cock plug and screw on the spring cap.
Product Life Designation
Value
Reprocessing
≤1000 cycles
Time
≤5 years
PERFORMING & FUNCTIONAL TEST The functional test shows whether perfect functioning of the instrument and its components can be ensured. Perform the functional test immediately after assembly. Note: Perform the functional test for the suction and irrigation handle.
Operating Conditions Disassembly
This instrument is not designed for use on the central nervous and circulatory system.
Designation
Value
Irrigation pressure
max. 600 mbar
Note: The suction and irrigation tube cannot be disassembled.
Prerequisite: The instrument is assembled. • Push control to “Suction” and “Irrigation” position.
Suction-irrigation system - connection via thread,
• Material fatigue and loss of functionality if the product lifetime is exceeded. • Tissue is punched when a trocar sleeve with an overly large diameter is used.
EXPLANATION OF SYMBOLS DANGER Indicates a hazard that will result in death or Page:
2 // 28
Risk of injury from the use of incompatible instruments. • Only use original accessories.
Note: During suctioning tissue fragments may settle in the suction and irrigation holes. For this reason, rinse the suction and irrigation tube several times outside of the surgical site during the intervention. • Insert the suction and irrigation tube in the suction and irrigation handle.
Suction-irrigation handle with control
CONTRAINDICATIONS
RISKS FROM IMPROPER USE
WARNING
illustration*
A Suction-Irrigation tube B Suction-Irrigation handle with trumpet valve C Suction-Irrigation handle with control 1 Suction-Irrigation holes 2 Thread 3 Irrigation valve 4 Suction valve 5 Tube connection 6 Knurled ring 7 Tube connection 8 Control for Suction-Irrigation function
ASSEMBLE & DISASSEMBLE Assembly
Suction-irrigation system with trumpet valve • Unscrew the knurled rings. • The control can be moved easily.
WARNING Suction-irrigation handle & ELectrode with trumpet valve
Risk of infection from non-sterile instruments. • Reprocess the instrument before assembly.
• Remove the valve.
• Connect the connections to a suction and irrigation device: LL connection -> Irrigation, Tube connection -> Suction. • Insert the instrument via a trocar sleeve. Note: With irrigation and suction handles with a control, you are free to choose which connection should be used for irrigation and which for suction • Perform the intervention. • Remove the instrument. • Reprocess the instrument.
Prerequisite: The instrument is assembled. • Push the trumpet valve together and let go IFU_00007-3_11_Interchangeable_Suction-Irrigation-Revision _04/21
0483
Ackermann
INTERCHANGEABLE SUCTION-IRRIGATION REPROCESSING WARNING Risk of infection due to non-sterile instruments. Reprocess reusable components before use. Note:The instrument must be disassembled before cleaning./ See chapter “Disassembly” / Detailed information on reprocessing and care can be found in the reprocessing instructions. The instrument is assigned to group “Reprocessing group 1” in chapter “Overview of reprocessing procedures”.
In addition, the warranty becomes null and void.
WARNING Risk of infection due to non-sterile instruments. • Prepare the instrument before returning it to the manufacturer. • Return the reconditioned instrument in its original packaging to the manufacturer./ See manufacturer’s address
SPARE PARTS STORAGE Only transport and store the instrument with adequate protection. / Detailed information on storage and care can be found in the reprocessing instructions.
REPAIRS Risk of injury due to improper repair. • Repairs may only be carried out by the manufacturer or by persons authorised by the manufacturer.
All product codes covered by these instructions are listed below:
GUIDELINES FOR CARE, MAINTENANCE & CLEANING
interchangeable suction-irrigation line
SAFETY INSTRUCTIONS
11-1236 11-1237-400 11-1237 11-1237-500 11-1237-100 11-1237-600 11-1237-100-45 11-1237-700 11-1237-200 11-1237-800 11-1237-300 11-1238
SYMBOLS USED ON LABELLING (ACC. DIN EN ISO 15223-1)
WARNING
Legal manufacturer Risk of injury from the use of incompatible spare parts. • Only use original spare parts.
Manufacturing date Consult instructions for use
suction valve T11-1236-500
Batch code
Irrigation valve
Product number
T11-1236-700
Quantity
SPring
Caution, consult accompanying documents
11-1236-100
WARNING Risk of infection due to non-sterile instruments. • Prepare the instrument before returning it to the manufacturer. • Return the reconditioned instrument in its original packaging to the manufacturer. / See manufacturer’s address
DISPOSAL Environmentally sound disposal enables valuable raw materials to be recycled. Dispose of the product in an environmentally friendly manner in accordance with the valid hospital guidelines.
WARRANTY A guarantee of two years from handover to the end customer is given for material and manufacturing defects. Transport costs and shipping risk are not assumed. In addition, the warranty stated in the General Terms and Conditions applies. This warranty does not apply to instruments that have been repaired, modified or improperly handled by unauthorised persons. In this case, no liability is assumed for the operational safety of the instrument either.
Page:
4 // 28
11-1269 11-1271L 11-1271L-45 11-1272L 11-1236COM
Do not use if package is opened or damaged
o-ring, 8X1MM 11-1236-200
Non-sterile
COMPATIBILITY
Keep out of direct sunlight
WARNING
Protect against moisture
Risk of injury from the use of incompatible instruments. • Only use original accessories. Suction-Irrigation tube
Compatible trocar
Diameter 3 mm
Diameter 3.5 mm
Diameter 5 mm
Diameter 5.5 mm
Diameter 10 mm
Diameter 11 mm
ADDIATIONAL INFORMATION Do not exceed maximum loading capacity of the sterilizer when processing multiple instruments in one sterilization cycle.
APPENDIX
ONLY
0483
Prescription use only mdc medical device certification GmbH, Kriegerstraße 6, 70191 Stuttgart, Germany
CONTACT DETAILS
These Guidelines for Care, Maintenance and Cleaning are not a substitute for the Instructions for Use. Before using the instrument Read and follow the Guidelines for Care, Maintenance and Cleaning. Keep these instructions in a place where they can be easily seen for reference at a later date.
OVERVIEW OF REPROCESSING METHODS VALIDATED REPROCESSING METHODS In the following synoptic table, the validated reprocessing methods are shown itemised according to the instrument groups. The individual instruments are assigned to the different reprocessing groups in the respective Instructions for Use. This synoptic table is not a substitute for the following sections of these Guidelines for Care, Maintenance and Cleaning. Reprocessing step
Reprocessing methods
The manufacturer does not assume any liability for damage as a result of incorrect reprocessing and care. The methods described for manual and mechanical reprocessing have been validated by the manufacturer. These are recommendations from the manufacturer. If using different and/or modified reprocessing methods, the user must provide proof of the effectiveness of the method The basic risks identified for the reprocessing process are summarised below based on the nature of the hazard. Warnings which relate to specific actions can be found directly next to the relevant action together with a classification of the risk. / See section “Explanation of symbols”. Risk of infection • Reprocess the instrument before initial use. • Reprocess the instrument before each use. • Reprocess the instrument before returning it to the manufacturer. • Follow the instructions for use of the cleaning agents and disinfectants used as well as of the cleaning and sterilisation devices used. • Wear personal protective equipment during reprocessing. • Discard disposable components after initial use. • Note the special reprocessing requirements if there is a suspicion of prions and Creutzfeldt-Jakob disease.
Reprocessing group 1
Reprocessing group 2
++ Validated reprocessing method 0 Process not validated by the manufacturer - Incompatibility
MATERIAL COMPATIBILITY The following synoptic table shows for which instrument groups and which reprocessing methods materials are compatible. The user must validate these methods! The individual instruments are assigned to the different reprocessing groups in the respective Instructions for Use. This synoptic table is not a substitute for the following sections of these Guidelines for Care, Maintenance and Cleaning. Reprocessing step
Reprocessing methods
Reproc. group 1
Reproc. group 2
Risk of injury • Do not use damaged instruments and do not repair. • Only use original accessories.
Ackermann Instrumente GmbH Eisenbahnstrasse 65-67 78604 Rietheim-Weilheim Germany Phone: Fax: E-Mail: Web:
+49 (0)7461 966 17 - 0 +49 (0)7461 966 17 - 70 [email protected] www.ackermannsurgical.com
+ Method with verified material compatibility 0 Process not validated by the manufacturer - Incompatibility
Note:More detailed information on reprocessing is available from the Robert Koch Institute (RKI) and the Instrument Reprocessing Working Group (AKI): RKI: Hygiene Requirements for Reprocessing Medical Devices (www.rki.de, as last amended) AKI: Proper Maintenance of Instruments (www.a-k-i.org, as last amended)
IFU_00007-3_11_Interchangeable_Suction-Irrigation-Revision _04/21
Ackermann
INTERCHANGEABLE SUCTION-IRRIGATION SCOPE These Guidelines for Care, Maintenance and Cleaning are valid for all products where explicit reference is made to the Guidelines for Care, Maintenance and Cleaning in the Instructions for Use.
EXPLANATION OF SYMBOLS DANGER Indicates a hazard that will result in death or serious injury if not avoided.
WARNING Indicates a hazard that can lead to death or serious injury if not avoided.
CAUTION Indicates a hazard that can lead to injury if not avoided.
IMPORTANT Indicates measures to prevent damage to property. Note: For warnings of personal injury, the danger sign according to ANSI Z535.6 is used in these operating instructions.
PREPARING DECONTAMINATION To prevent surgical residue from drying on, the following steps must be performed directly after surgery. • Rinse the instrument with cold water. • Remove coarse dirt with cold water. • Rinse out cavities with cold water. Note: If it is not possible to rinse with cold water, the instrument must be wrapped in a moist cloth to prevent any residues from drying on. • Always transport the instrument to the reprocessing site in a closed container to prevent product damage and contamination of the environment.
PRE-CLEANING Pre-cleaning prevents surgical residue from drying on. It must therefore be carried out directly after surgery. Pre-cleaning was validated with the cleaning agent Cidezyme from Johnson & Johnson:
Cleaning Enzymatic
Dosage 0.8%
pH value 7.8-8.8
6 // 28
Risk of infection due to insufficient reprocessing. • Remove protective caps (if relevant).
WARNING Risk of infection and pyrogenicity from residues if unsuitable cleaning agents are used. • Do not use fixing agents. • Do not rinse with hot water
IMPORTANT Avoid damaging product. • Do not use abrasive brushes or scourers. • Only use the cleaning agents which are listed in this section. • Use disinfectant with corrosion protection. • Tip: Remove caked-on tissue residues with a plastic brush. • Immerse the instrument in a cold water bath with 0.8% cleaning solution: >5 min. • Brush the instrument under cold water until all visible signs of soiling have been removed. • Dismantle the instrument (if possible). / See instrument’s Instructions for Use. • Open stop cocks (if relevant). Note: Rinse the instrument under the surface of the water. This prevents contamination of the surrounding area. • Brush the outside and inside under cold water with a round brush until no more residue is visible. • Rinse out cavities, drill holes and threads (if relevant) with a cleaning gun: >10 s at 3-5 bar. • Remove from the water bath and rinse off with cold water. • Immerse in combined cleaning and disinfectant solution until subsequent cleaning to prevent any residue from drying on.
CLEANING & DISINFECTION General Information Note: Unless stated otherwise in the instrument’s Instructions for Use, cleaning is performed with the instrument disassembled. Unless specified otherwise in the description of the reprocessing method, use water of the following quality: • Pre-rinsing: Tap water • Cleaning and rinsing: Demineralised water • Final rinsing: Demineralised water
Page:
WARNING
IMPORTANT Avoid damaging product. • Do not use abrasive brushes or scourers. • Only use the cleaning agents which are listed in the individual sections. • With plastic instruments avoid contact with hydrogen peroxide (H2O 2). • Choose between manual and mechanical cleaning.
WARNING Risk of infection due to insufficient reprocessing. • Special reprocessing requirements must be observed if there is a suspicion of prions and Creutzfeldt-Jakob disease. / To this end, note Annex 7 of the recommendation “Hygiene Requirements for Reprocessing Medical Devices” from RKI and BfArM.
a cleaning gun with cold water: >20 s at 3-5 bar. • Unless specified otherwise in the instrument’s Instructions for Use, clean the components in the ultrasonic bath with 0.8% cleaning agent with the following settings: Temperature
Frequency
Duration
40-45°C
35 kHz
10 - 15 min.
Chemical Disinfection Manual cleaning was validated with the cleaning agents Cidezyme/Enzol from Johnson & Johnson and Mucadont Zymaktiv from Merz Hygiene GmbH: Cleaning agent
Cleaning
Dosage
Cidezyme/Enzol
Enzymatic
0.8%
Mucadont Zymaktiv
Enzymatic
0.8%
Note: Rinse the instrument under the surface of the water. This prevents contamination of the surrounding area.
Preparation Dismantle the instrument (if possible). / See instrument’s Instructions for Use.
• Place the instrument in cold water: >10 min. • Brush the instrument under cold water until all visible signs of soiling have been removed. • Rinse out cavities, drill holes and threads with
Cleaning
CHECKING & CARE
• Unless specified otherwise in the instrument’s Instructions for Use, additionally clean the components prior to or in combination with mechanical cleaning in an ultrasonic bath with the following settings:
The right care of instruments will lengthen their service life and should therefore be carried out after every cleaning process.
Temperature
Frequency
Duration
40-45°C
35-45 kHz
10 - 15 min.
Note: Also clean the inner chambers of the instrument below water using a cleaning brush.
• Turn and move the components several times during cleaning in the ultrasonic bath.
Drying • Dry on the inside and outside: >10 min. at 50-100°C and/or blow through with sterile compressed air.
Disinfection was validated with the disinfectant Mucocit-T from Merz Hygiene GmbH: • Disinfect with disinfectant: >10 min. / See manufacturer’s Instructions for Use Disinfect
Dosage
pH value
Mucocit-T
4%
10.5 (diluted)
Proper cleaning depends on the right cleaning program. Compare the cleaning program with that of your cleaning system and save where necessary: Start the cleaning program. 1. Pre-rinsing with cold water: 1 min. 2. Emptying 3. Repeated pre-rinsing with cold water: 3 min. 4. Emptying 5. Cleaning with 0.5% alkaline cleaning agent: at 55°C for 5 min. 6. Emptying 7. Neutralising with deionised water: 3 min. 8. Emptying 9. Rinsing with deionised water: 2 min. 10. Emptying
Drying
Manual Cleaning (ALKALINE) & THERMAL Disinfection Mechanical cleaning was validated with the cleaning device G 7735 CD from Miele. Mechanical alkaline cleaning was validated with the cleaning agent Neodisher FA from Dr. Weigert: Cleaning
Dosage
Alkaline
0.5%
pH value 12.2-14 (diluted)
• Open stop cocks (if relevant).
Cleaning
/ Observe the national requirements as regards the A0 value (see ISO 15883)
• Turn and move the components several times during cleaning in the ultrasonic bath. • Rinse out cavities, drill holes and threads with a cleaning gun with cold water: >20 s at 3-5 bar. • Immerse the instrument in deionised water and rinse through the cavities several times with deionised water.
Disinfection Manual Cleaning (enzymatic) &
cleaning agent can reach all inner and outer surfaces. • Connect flush opening (if any) to MIS mobile unit.
Preparation • Dismantle the instrument (if possible). / See instrument’s Instructions for Use. • Open stop cocks (if relevant). • Place instruments in a sieve tray on the cleaning device MIS mobile unit so that the
* Deionised/demineralised water
WARNING Risk of injury from faulty or damaged components Do not use damaged instruments and do not repair. • Check to ensure they are clean and, if necessary, repeat cleaning. • Check for damage (e.g., sharp edges, rough surfaces). • Replace brittle and cracked seals (if relevant). • Lubricate moving parts (e.g., joints, rotating stop cocks) with medical oil. • Remove any excess oil. • Assemble instruments (if possible) and check to ensure they are in perfect working order. / See instrument’s Instructions for Use.
PACKAGING The instrument must be packed appropriately prior to sterilisation to ensure that the sterile barrier remains intact after removal from the steriliser. • Package the instrument to comply with ISO 11607 and EN 868.
• Dry inside and outside in the cleaning and disinfection device: 15-25 min. at 90-110°C. The cleaning and disinfection device cycle must include a drying phase. • Remove the product from the cleaning and disinfection device immediately at the end of the cycle. • If necessary, also blow through the product with sterile compressed air until it is completely dry.
STERILIZATION Unless stated otherwise in the instrument’s Instructions for Use, sterilisation is performed with the instrument assembled. / See instrument’s Instructions for Use.
Disinfection
Sterilisation was validated with the sterilisation device Selectomat S 3000 from MMM Group and Varioclav 400 E from Fisher Scientific.
Disinfection was validated with the device Getinge 88 series:
IMPORTANT
Disinfect
Avoid damaging product. • Observe the device’s maximum load. / See manufacturer’s Instructions for Use.
Demineralised water*
Temperature
Dwell time
90+3°C
≥ 5 min
IFU_00007-3_11_Interchangeable_Suction-Irrigation-Revision _04/21
Ackermann
INTERCHANGEABLE SUCTION-IRRIGATION WARNING
REPAIR
Risk of infection reprocessing.
due
to
insufficient
• Special reprocessing requirements must be observed if there is a suspicion of prions and Creutzfeldt-Jakob disease.
WARNING Risk of injury from improper repairs. • Only allow repairs to be performed by the manufacturer or by persons authorised by the manufacturer.
WARNING
Sterilisation was validated at 134°C and 4 minutes. The sterilisation time must therefore be at least 4 minutes. The following countries make different stipulations which must be observed: Country
Sterilization time
Germany
≥ 4 - 30 min.
France
≥ 18 - 30 min.
Switzerland
≥ 18 - 30 min.
Note: Dwell times of 4-30 min. do not have a negative influence on the instrument/material. • Open stop cocks (if relevant). • Place in the sterilisation device so that the components are not touching each other and the steam can circulate freely. • Set the sterilisation parameters (3-fold fractionated pre-vacuum): Temperature 134 -137°C
Pressure
Duration
3 bar
/ See details in
44 psi
this section.
• Start the sterilisation process.
STORAGE OF STERILE DEVICES To avoid reducing durability and forfeiting any resistance to bacteria the following storage conditions must be observed: • Store the sterile device sealed in a clean, dust-free and dry sterile container. • Protect from direct light. • Store the sterile container in a clean and dry environment with controlled humidity at room temperature. • Do not store the sterile container in the vicinity of aggressive substances (e.g., alcohols, acids, bases, solvents and disinfectants).
Risk of infection from non-sterile instruments. • Reprocess the instrument before returning it to the manufacturer. • Send the instrument back to the manufacturer in a reprocessed state and in its original packaging. / See manufacturer’s address
DISPOSAL Environmentally sound disposal enables valuable raw materials to be recycled. Dispose of the device in an environmentally friendly manner in accordance with the valid hospital guidelines.
INFORMATION ON VALIDATION The following materials and machines were used for validation: Cleaning agent: • Cidezyme/Enzol from ASP • Mucadont Zymaktiv from Merz Hygiene GmbH • Neodisher FA from Dr. Weigert • deconex 23 Neutrazym from Borer Disinfectant: • Mucocit-T from Merz Hygiene GmbH Cleaning and disinfection devices: • G 7735 CD from Miele Sterilisation devices: • Selectomat S 3000 from MMM Group • Varioclav 400 E from Fisher Scientific Sterilising agent: • Moist heat End of document
Note: Also observe your internal storage standards for sterile devices.
Page:
8 // 28
IFU_00007-3_11_Interchangeable_Suction-Irrigation-Revision _04/21