ACTEON
Diamond-coated ball tip for smoothing the vestibular bone window and wwprecise osteoplasty
Pre-disinfection, cleaning and sterilisation instructions
23 Pages
Preview
Page 1
Pre-disinfection, cleaning and sterilisation instructions
Tips and files
This document is an English translation of the original French version. Reference J02000 version V3 and drawing number RB01FR010C
Contents 1 Applicability 2 Documentation 2.1 Associated documentation 2.2 Electronic documentation 2.3 Latest document update 3 Warnings 3.1 Precautions to be taken 3.2 Cleaning cycle limits 3.3 Containment and transportation 3.4 Preparation for pre-disinfection 4 Manual instructions 4.1 Pre-disinfection and cleaning – Manual method 4.2 Sterilisation 4.3 Inspection 4.4 Packaging 4.5 Storage 5 Automatic Instructions 5.1 Pre-disinfection and cleaning – Manual method 5.2 Cleaning, automated method 5.3 Thermal disinfection 5.4 Sterilisation 5.5 Inspection 5.6 Packaging 5.7 Storage 6 Regulations and standards 6.1 Manufacturer identification 7 Index 8 Glossary
5 7 7 7 7 9 9 9 9 11 13 13 13 14 14 14 15 15 15 15 15 16 16 16 17 17 19 21
1 Applicability These instructions do not apply to tips and files for veterinary use. Tips and files IRR 25-45, SPR 30, SPR 60, SPR 80, PLU 60, PLU 90 and PLU 110 are not dealt with in this document.
Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C - Page 5/23
Page 6/23 - Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C
2 Documentation This document contains the following information: l l l
preparation for cleaning; medical device disinfection; medical device sterilisation.
2.1 Associated documentation Document title
References
Consulting electronic user instructions
J00000
Cleaning, disinfection and sterilisation instructions for handpieces
J12911
2.2 Electronic documentation
The user instructions for your device are available in electronic format on the specified website and not in printed format. If the website is unavailable, try again later. You can also request a free printed copy of the user instructions within seven days via our website, by telephone or in writing. The electronic user instructions are available in PDF format (Portable Document Format). You will need to have a PDF file reader installed to read the electronic user instructions. It is important for you to have read and understood the content of the user instructions relating to the use of your device and its accessories. Never use your device without first reading the user instructions. The device user instructions can be consulted at the following addresses:www.ultradent.com and www.satelec.com. When you receive your device, you are asked to print and/or to download all documents or sections of documents that you may need to refer to in the event of an emergency, if you are unable to connect to the internet or if your electronic display tool is not working (computer, tablet, etc.). We recommend that you visit the website regularly to consult and/or to download the latest version of your device's user instructions. Users are asked to keep documentation on hand so that it can be consulted when necessary.
2.3 Latest document update 07/2016
Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C - Page 7/23
Page 8/23 - Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C
3 Warnings Do not clean the medical device with steel wool or abrasive cleaning products. Do not use solutions containing iodine or with a high chlorine content. The pH of the detergents and disinfectants must be between 7 and 11. Contaminated or used devices cannot be placed in sterilisation box or cleaned in a machine. The cleaning method for tips and files recommended by SATELEC, a company of Acteon group is manual or automatic. All devices must be carefully cleaned and then undergo a final sterilisation before use. The sterilization parameters are only valid for correctly cleaned devices. Tips and files require special attention during cleaning. During automatic cleaning, tips or files must be placed on suitable instrument holders or in small baskets to prevent them from being damaged during washing. It is the responsibility of the end user to ensure that all equipment used to recondition SATELEC, a company of Acteon group products is properly installed, validated, maintained and calibrated. When possible, use a washer-disinfector. Prevent the overloading of wash baskets during ultrasonic cleaning or in a washer-disinfector. Throughout the procedure, wipe away blood and debris to prevent it from drying on the surfaces. After the procedure, soiled devices must be covered with a damp cloth to prevent residue from drying. Soiled devices must also be separated from non-contaminated devices to avoid contamination of personnel or surroundings.
3.1 Precautions to be taken At the end of the procedure, rinse, clean and disinfect the irrigation line. l l l
l l l l l
remove the perforator from the irrigation solution; immerse it in a container filled with an enzymatic or alkaline liquid; run the purge cycle on your medical device and allow the liquid to pass through the irrigation line for at least one minute; stop the purge cycle, then immerse the perforator in distilled or purified water; run the purge cycle for at least one minute to rinse the irrigation line; remove the perforator from the liquid and purge the line until all liquid has been removed ; wipe blood and debris from the tips or files to prevent them from drying on the surfaces; cover the tip or the file with a soft lint-free cloth dampened with purified water to prevent blood and debris from drying.
3.2 Cleaning cycle limits Repeated packaging cycles involving manual washing have little effect on tips and files. End of service life is normally determined by wear and damage due to use.
3.3 Containment and transportation Soiled devices must be transported separately from non-contaminated devices to avoid any contamination.
Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C - Page 9/23
Page 10/23 - Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C
3.4 Preparation for pre-disinfection It is advisable to recondition devices as soon as possible after use. SATELEC, a company of Acteon group devices must be reconditioned within two hours of use. Prior to cleaning and after each use, remove the tip or the file from the handpiece. Prior to cleaning and after each use, remove the tip from the front of the handpiece and place it in the appropriate container. Its cleaning is subject to a different set of instructions. l
l l l l
Remove the single-use or autoclavable irrigation line and the clips from the handpiece-cord assembly, if applicable. Their cleaning is subject to a different set of instructions. Always dispose of the single-use perforator and/or irrigation line. Disconnect the handpiece from the cord. Unscrew the nosepiece. Depending on the model, remove the optical light guide. Depending on the model, disconnect the LED ring from the handpiece by pulling on it gently.
Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C - Page 11/23
Page 12/23 - Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C
4 Manual instructions 4.1 Pre-disinfection and cleaning – Manual method Equipment: soft brush, soft lint-free swab, lint-free cloth, syringe, pipette or water spray, alkaline cleaner, ultrasonic cleaner. Minimum duration of step
Cleaning instructions
1 minute
Rinse the soiled device under cold running water. Use a soft-bristled brush, a swab or a lint-free cloth to remove most of the contamination. Use a syringe, a pipette or a water spray with an alkaline or enzymatic cleaning solution to rinse the tip cannulation.
10 minutes
Immerse the medical device in a freshly prepared alkaline cleaning solution in an ultrasonic cleaner for at least ten minutes. Adhere to the manufacturer's exposure time, concentration, water quality and temperature recommendations.
1 minute
Rinse the device under cold running water. Use a syringe, a pipette or a water spray to rinse the tip cannulation.
4 minutes
Wash and disinfect the medical device using an alkaline cleaning solution. Remove surface contamination using a soft brush or a swab. Wash the medical device in water to prevent contaminants from spreading into the air
1 minute 30 seconds
Rinse the medical device in deionised or purified water. Use a syringe, a pipette or a water spray to rinse the tip cannulation. Visually inspect the medical device. Repeat this procedure until the medical device is visibly clean. Perform a final rinse of the device using distilled or purified water. Dry the medical device using a soft lint-free cloth or medical grade clean compressed air
4.2 Sterilisation Unless otherwise specified, non-sterile products can be resterilised using validated steam sterilisation methods (ISO 17665 or national standards). SATELEC, a company of Acteon group recommends the following: Sterilisation exposure time
Sterilisation exposure temperature
Drying time
4 minutes
132 °C
15 minutes minimum and 20 minutes
18 minutes
134 °C
15 minutes minimum and 20 minutes
4 minutes
134 °C
15 minutes minimum and 20 minutes
3 minutes
134 °C
15 minutes minimum and 20 minutes
Saturated steam sterilisation with pre-vacuum The drying times vary from 15 to 60 minutes according to the following criteria: l l l l l l l
the type of packaging material, such as a sterile barrier system or rigid reusable containers; steam quality; device materials; total mass; steriliser performance; usual practices for the geographical area; varying cool-down times. The distributor and manufacturer accept no responsibility for sterilisation procedures performed by the end user or the customer that are not performed according to the manufacturer’s recommendations.
Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C - Page 13/23
4.3 Inspection l
l l
Devices must be inspected to check that no contamination remains, that they are not corroded, dulled, discoloured or damaged. Before conditioning and sterilising the cleaned devices, check they are clean, undamaged and function properly. Damaged devices must be discarded, they must not be lubricated.
4.4 Packaging Use suitable packaging or a rigid reusable container for sterilisation; the sterile barrier system must comply with ISO standard 11607. Prevent any contact between devices and other objects that could damage their surface or the sterile barrier system. Use suitable packaging; the sterile barrier system must comply with ISO standard 11607. Prevent any contact between devices and other objects that could damage their surface or the sterile barrier system.
4.5 Storage Storage conditions are printed on the packaging label. Packaged products should be stored in a clean, dry environment, protected from direct sunlight, pests, humidity and extreme temperatures. Use products in the order in which they are received First in, First out, taking into account the expiry date indicated on the label.
Page 14/23 - Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C
5 Automatic Instructions 5.1 Pre-disinfection and cleaning – Manual method The manual pre-disinfection/pre-cleaning method must be performed prior to the automatic cleaning. Equipment: soft brush, soft lint-free swab, lint-free cloth, syringe, pipette or water spray, alkaline cleaner, ultrasonic cleaner. Minimum duration of step
Cleaning instructions
1 minute
Rinse the soiled device under cold running water. Use a soft-bristled brush, a swab or a lint-free cloth to remove most of the contamination. Use a syringe, a pipette or a water spray with an alkaline or enzymatic cleaning solution to rinse the tip cannulation.
1 minute 30 seconds
Wash and disinfect the medical device using an alkaline cleaning solution. Remove surface contamination using a soft brush or a swab. Wash the medical device in water to prevent contaminants from spreading into the air
1 minute
Rinse the device under cold running water. Use a syringe, a pipette or a water spray to rinse the tip cannulation.
10 minutes
Immerse the medical device in a freshly prepared alkaline cleaning solution in an ultrasonic cleaner for at least ten minutes. Adhere to the manufacturer's exposure time, concentration, water quality and temperature recommendations.
1 minute 30 seconds
Rinse the medical device in deionised or purified water. Use a syringe, a pipette or a water spray to rinse the tip cannulation. Visually inspect the medical device. Repeat this procedure until the medical device is visibly clean. Perform a final rinse of the device using distilled or purified water. Dry the medical device using a soft lint-free cloth or medical grade clean compressed air
5.2 Cleaning, automated method Step
Minimum duration
Cleaning instructions
Pre-washing
2 minutes
Cold tap water
Washing
10 minutes
Warm tap water, hotter than 40°C. Use an alkaline cleaning solution
Neutralisation
2 minutes
Warm tap water (40°C), with neutraliser if necessary.
Rinsing
2 minutes
Distilled or purified water, hotter than 40°C
Drying
40 minutes
At a temperature of 90°C.
5.3 Thermal disinfection Thermal disinfection at 90°C for at least five minutes.
5.4 Sterilisation Unless otherwise specified, non-sterile products can be resterilised using validated steam sterilisation methods (ISO 17665 or national standards). SATELEC, a company of Acteon group recommends the following: Sterilisation exposure time
Sterilisation exposure temperature
Drying time
4 minutes
132 °C
15 minutes minimum and 20 minutes
18 minutes
134 °C
15 minutes minimum and 20 minutes
Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C - Page 15/23
Sterilisation exposure time
Sterilisation exposure temperature
Drying time
4 minutes
134 °C
15 minutes minimum and 20 minutes
3 minutes
134 °C
15 minutes minimum and 20 minutes
Saturated steam sterilisation with pre-vacuum The drying times vary from 15 to 60 minutes according to the following criteria: l l l l l l l
the type of packaging material, such as a sterile barrier system or rigid reusable containers; steam quality; device materials; total mass; steriliser performance; usual practices for the geographical area; varying cool-down times. The distributor and manufacturer accept no responsibility for sterilisation procedures performed by the end user or the customer that are not performed according to the manufacturer’s recommendations.
5.5 Inspection l
l l
Devices must be inspected to check that no contamination remains, that they are not corroded, dulled, discoloured or damaged. Before conditioning and sterilising the cleaned devices, check they are clean, undamaged and function properly. Damaged devices must be discarded, they must not be lubricated.
5.6 Packaging Use suitable packaging or a rigid reusable container for sterilisation; the sterile barrier system must comply with ISO standard 11607. Prevent any contact between devices and other objects that could damage their surface or the sterile barrier system. Use suitable packaging; the sterile barrier system must comply with ISO standard 11607. Prevent any contact between devices and other objects that could damage their surface or the sterile barrier system.
5.7 Storage Storage conditions are printed on the packaging label. Packaged products should be stored in a clean, dry environment, protected from direct sunlight, pests, humidity and extreme temperatures. Use products in the order in which they are received First in, First out, taking into account the expiry date indicated on the label.
Page 16/23 - Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C
6 Regulations and standards 6.1 Manufacturer identification SATELEC A Company of ACTEON Group 17, avenue Gustave Eiffel BP 30216 33708 MERIGNAC cedex FRANCE Tel. +33 (0) 556.34.06.07 Fax: +33 (0) 556.34.92.92 E.mail: [email protected]. www.acteongroup.com
Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C - Page 17/23
Page 18/23 - Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C
Index: drying – user responsibility
7 Index D drying 13, 16
E electronic 7 electronic user instructions 7 equipment 13, 15
F file 9 First in, First out 14, 16
H handpiece 7, 11
N neutralisation 15
P pH 9
S saturated steam with pre-vacuum 13, 16 storage conditions 14, 16
T Thermal disinfection 15 tip 9
U update 7 user responsibility 9
Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C - Page 19/23
Page 20/23 - Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C
Glossary: Active implantable device – Medical device
8 Glossary
D disinfection voluntary and temporary removal of some germs to stop or prevent an infection, risk of infection or secondary infection by unwanted or pathogenic viruses or micro-organisms
A Active implantable device Active implantable medical devices are devices which are designed to be wholly or partially implanted in the human body or located in a natural orifice, and that require the supply of electricity or another energy source to operate correctly other than energy that is directly
E expiry date date up to which the medical device can be used. After this date, the medical device will need to be resterilised
generated by the human body or gravity. (Public Health Code L.5211-1)
alcohol wipe disposable wipe soaked in an alcoholic solution designed to disinfect medical devices
I Implantable device Any device designed to be completely implanted in the human body or to replace an epithelial surface or the surface of the eye, during a
autoclave container with thick walls and hermetic seal designed to steam sterilise under a pressure of
surgical procedure and to remain in place after surgery. An implantable device also incorporates any device designed to be partially implanted in the human body during surgery and that is intended to remain in place after the surgical procedure for a period of at least thirty days. (directive 93/42/EEC)
several bar. For an item to be considered sterile, the theoretical probability of isolating a germ must be less than 1 in a million. This is the sterility assurance level (SAL) set forth in standard EN 556
C M
cleaning essential pre-conditioning step to remove contamination through the physical-chemical action of a suitable product such as a detergent, combined with a mechanical action to ensure that the medical device is fully operational and clean. After cleaning, the cleanliness of the medical device components should be checked in addition to the cleanliness of the reassembled medical device. It is also important to make sure there is no damage likely to impact the safety, integrity or correct operation of the device
Medical device A medical device is any instrument, device, item of equipment, material or product, excluding products of human origin, or any other item used alone or with another item, including accessories and software required to ensure correct operation of the medical device, designed by the manufacturer to be used on humans for medical purposes and of which the main desired action is not obtained using pharmacological or immunological means or metabolism, but whose
Instructions | tips and files | J02001 | V3 | (13) | 07/2016 | RB01EN010C - Page 21/23