2-8-245HAN/2-8-245CAS HOOKUP CONNECTION GUIDE
READ ALL INSTRUCTIONS PRIOR TO CONNECTION Locate the endoscope port and follow the instructions for attaching the corresponding Endoscope Connector. All connectors must be attached.
Using Medivators® ADVANTAGE® Automated Endoscope Reprocessors
ADVANTAGE HOOKUP (Figure 1)
Connector Suction Align silicone tube with suction
barb on Karl Storz Cleaning Adapter and push onto barb until fully engaged.
P/N: X-X-XXXHAN
LOT: XXXXXX ID: XXXX
Test Connector Leak Align the slot on connector
Each ADVANTAGE Hookup is comprised of multiple ports and connector assemblies. Each port is labeled with a Port ID number (circled). This same number also identifies the corresponding Connector Assembly. Each Connector Assembly is fitted with a specific endoscope connector. (see Fig. 2).
Karl Storz Cleaning Adapter PN 11301CD (available from Karl Storz) Attach Karl Storz 11301CD adapter to the suction valve port of the endoscope following Karl Storz instructions.
EXAMPLE CONNECTOR ASSEMBLY (Figure 2)
Ferrule
Connector Coupling
Label
Tubing
Endoscope Connector example
Inlet Connector Biopsy Align stainless steel fitting
with biopsy port, push and turn clockwise to connect.
CONNECTOR ASSEMBLY IDENTIFICATION CHART Connector Assembly Description
with the pin on the scope. Push on and turn clockwise to engage.
Port ID
Port Name
1
LEAK TEST
Leak Test Connector Assembly
H28245-1
Order Number
3
BIOPSY
Biopsy Inlet Connector Assembly
H28245-3
7
SUCTION
Suction Connector Assembly
H28245-7
If a Connector Assembly needs replacement, locate the port number/name and use the corresponding order number found in the Connector Assembly Identification Chart. Connector Assemblies are specific to individual hookups. Check each Hookup Connection Guide for relevant information.
DESIGNATED HOOKUP FOR THE FOLLOWING KARL STORZ® ENDOSCOPES*: 11001BN1 11001BN4 11002BD1
11004BC1 11009BC1 11009BII
11301AA1 11301AAN1 11301AAP1
11301BN1 11301BND1 11301BNN1
11301BNP1 11301BNX 11302BD1
11302BD2 11302BDD1 11302BDD2
11302BDP1 11302BDX 11304BC1
11340BC1
* This list of corresponding endoscope models is subject to change without notice. For a full list, please go to:
www.medivators.com/hookuplookup * Representative endoscope for illustrative purposes only. For exact port locations, refer to endoscope manufacturer’s instruction guide.
Ensure proper alignment of all connectors with the endoscope mating parts prior to attachment.
If you have any questions regarding this hookup, in the U.S., please contact MEDIVATORS Technical Support at 1-800-444-4729. Outside the U.S., please contact your MEDIVATORS Distributor.
HOOKUP CONNECTOR BLOCK LABEL
CASSETTE-STYLE HOOKUPS (not available in US)
The Hookup Connector Block Label contains useful information for quick reference.
An ADVANTAGE Hookup Cassette combines an endoscope tray with the hookup connector block. Specific endoscope connections for this hookup are identical to non-cassette versions. Cassette-style hookup part numbers are designated with a CAS suffix. ADVANTAGE Hookup Cassettes are available outside the United States only. Connector Assemblies are listed in the Connector Assembly Identification Chart.
Alternative Channel Connector Names
Handle Locking Instructions
WARNINGS & CAUTIONS This hookup is intended for connection only with the specific endoscope models identified in the appropriate MEDIVATORS online hookup guide. Use of this hookup with endoscopes other than those specified by the hookup application guide may result in inadequate HLD. Users operating MEDIVATORS AERs and hookups must be trained and competent in the understanding of endoscope channel systems. Prior to reprocessing, users must verify that the hookup used contains connections for all appropriate channels that require a separate reprocessing connection according to the hookup instructions All connectors and adapters must remain firmly attached and unrestricted for the entire disinfection cycle to ensure adequate HLD. Users must inspect the endoscope and hookup combination to verify proper connection and flow.
Port ID Number
Modifications or repairs to hookups which do not correspond to the manufacturer’s specifications may result in inadequate HLD and/or damage to endoscopes.
Port Name
Failure to properly clean and prepare endoscope for immersion prior to hookup connection may result in inadequate HLD and/or damage to the endoscope. Endoscopes must be inspected for damage and verified to be in proper working order prior to hookup connection. Connection of hookups to damaged endoscopes or endoscopes with obstructed/restricted channels could result in inadequate HLD. Manual leak testing must be carried out in accordance with manufacturer’s recommendations prior to reprocessing. Install all necessary waterproof caps, plugs and cleaning adapters prior to immersion.
P/N: X-X-XXXHAN
Lot Number
LOT: XXXXXX ID: XXXX
Hookup Barcode
Do NOT autoclave MEDIVATORS hookups. MEDIVATORS makes no claim on high-level disinfection efficacy when these reprocessing instructions are not followed, or when this hookup kit is applied to endoscopes other than those specified in the online interactive hookup guide. Avoid biological contamination. Always wear personal protective equipment when handling endoscopes.
Hookup Part Number
Handle Locking Instructions - insert block onto basin channel manifold and move handle to the left to lock. Move to the right to unlock. Lot Number - identifies date of manufacture. Hookup Part Number - identifies hookup. Hookup Barcode - barcode of part number. Port Name - identifier of port location in ADVANTAGE system. Error logs identify issues using these names. Also identifier of Connector Assembly replacements for older model hookups. Port ID Number - identifier for Connector Assembly replacements. Not found on older model hookups. Alternative Channel Connector Names - connector names vary depending on specific endoscope channel configuration.
Manufactured in the USA by: Medivators Inc. 14605 28th Avenue North Minneapolis, MN 55447 USA Toll Free: +1.800.444.4729
Medivators BV Sourethweg 11 6422PC Heerlen The Netherlands Tel: +31.45.5.471.471
www.medivators.com
Cantel Medical Asia/Pacific Pte. Ltd. 1A International Business Park #05-01 Singapore 609933 Tel: +65.6227.9698
Medivators Inc. Beijing Representative Office Room 1801, Floor 18th, Tower A, Beijing Marriott Hotel, Office Building No. 7, Jianguomen South Avenue, Dongcheng District, Beijing 100005 China Tel: +86 10 65204039
© 2015 Medivators Inc. 50097-548/B
SLB.BY.BALLOON
Bal Component Nbr Item Number
Component Item Description
40100-102|01 PAPER, COMP/COPY 8.5X14
14:06 Mon Feb 02 2015
Medivators
Sys hpm2k Acct lad:MINN.MAIN
Single Level BOM for Part 50097-548|01 IFU/DFU 2-8-245 IFU/DFU FOR 2-8-245 HAN/CAS Revision B, Effective as of 02-02-15 CPN CPN Quantity Make Phant Bin PS Rev Ref UM Type Per Assy Purch Code Code Type Lvl Desig EA
OS
0.000
P
A