Setup Guide
4 Pages
Preview
Page 1
Hysterolux™ canister setup
Setup Guide Four canisters
SIDE LEVER set to split suction
2
3
4
REG position
2
Rear port
1
2
5
3
4
3
4
Front port
Vacuum tube from vacuum source
5
Yellow outflow tube dropped into patient drape†
7
2
3
4
1
6
6 Blue outflow tube from patient drape
6
†No connection to canister. 1. TruClear™ shavers [instructions for use]. Mansfield, MA: Medtronic; 2017. Always refer to the Instructions For Use included with the product for complete indications, instructions, warnings, precautions, and contraindications.
7
Suction Regulator Settings1 Soft Tissue Shaver Plus:
150
Soft Tissue Shaver Mini:
200-300
Dense Tissue Shaver Plus:
150-250
Dense Tissue Shaver Mini:
250-300
mmHg mmHg mmHg mmHg
White outflow tube from handpiece
6
5
• Attach the tubes as indicated, in the order and locations specified by the numbered steps. 5
• Use the comparison chart on page 4 as a visual reference to cross-check the finished assembly. • Use the suction regulator settings chart below to determine the correct suction setting for each shaver.
Hysterolux™ canister setup
Setup Guide Two canisters
2
SIDE LEVER set to split suction
3
Yellow outflow tube dropped into patient drape†
5
Rear port
4
Blue outflow tube from patient drape
4
†No connection to canister. 1. TruClear™ shavers [instructions for use]. Mansfield, MA: Medtronic; 2017. Always refer to the Instructions For Use included with the product for complete indications, instructions, warnings, precautions, and contraindications.
• Use the suction regulator settings chart below to determine the correct suction setting for each shaver.
REG position
2
3
• Attach the tubes as indicated, in the order and locations specified by the numbered steps.
3
2 1
Front port
Vacuum tube from vacuum source
1
5 White outflow tube from handpiece
4
4
Suction Regulator Settings1 Soft Tissue Shaver Plus:
150
Soft Tissue Shaver Mini:
200-300
Dense Tissue Shaver Plus:
150-250
Dense Tissue Shaver Mini:
250-300
mmHg mmHg mmHg mmHg
Hysterolux™ canister setup
Setup Guide Common errors
Rear port
Front port
When connected this way, the front canisters - when full - will "block" the suction. The back canisters will not fill.
Do not connect the regulator to the front canisters. You should always connect the regulator to the back canisters.
Hysterolux™ canister setup
Setup Guide Comparison Chart
OFF LINE
REG.
After assembling your hoses, use this illustration to compare your system to the correct layout:
VACUUM TUBE From vacuum source
• Double-check all tubes are well attached to canisters. • The White OutflowTube is placed onto the TissueTrap. • The Blue Outflow Tube is connected to patient port on the canister. • The Yellow Outflow Tube is dropped into patient drape - no connection needed to canister.
BLUE OUTFLOW TUBE From patient drape
VACUUM
VACUUM
TANDEM
TANDEM PATIENT
PATIENT
VACUUM
TANDEM PATIENT
©2022 Medtronic. Medtronic, Medtronic logo, and Engineering the extraordinary are trademarks of Medtronic. All other brands are trademarks of a Medtronic company. 07/2022 – US-UW-2200033 – [WF#3940942]
WHITE OUTFLOW TUBE From handpiece
VACUUM
TANDEM PATIENT