8 Pages
Preview
Page 1
Recommended Transducer Cleaner and Disinfectant [a]: Cleaners
[b]: Wipes
[c]: Sprays
b32: Super b1: CLEANISEPT® WIPES b16: CaviWipes MetriZyme b17: Dispatch Towels b2: mikrozid® AF SaniTristel Preb18: Accel TB Wipes Cloth® Wipes Jumbo Clean Wipes b33: SANIb19: CaviWipes 1 b3: PROTEXTM DISINa3: Liquinox b20: Tuffle 5 a4: Revital-OX CLOTH® FECTANT Wipes b21: Sani-Cloth Active b4: Sani-Cloth® Plus PRIME Enzymatic b22: Septiwipes b5: SONOTM ULTRAWIPES Detergent b23: Mikorbac Tissues a5: MetriSponge SOUND WIPES a6: Prolystica 2X b7: Tristel Sporicidal Wipes b24: Sani-Cloth Germicidal Wipes Concentrate b8: Tristel Rinse Wipes b9: Clinell Universal Wipes/ b25: WIP'ANIOS Enzymatic CLEAN'UP Clinell® Surface Wipes Cleaner b26: CaviWipes XL b10: mikrozid® Sena7: Endozime b27: OXIVIR™ WIPES sitive Wipes and Endozb28: Oxivir 1 Wipes ime Sponge b11: Wip’Anios premium b29: PDI SANI-CLOTH b12: ultrasound probe a8: klenzyme a9: ANIOSYME 5 BLEACH WIPES cleaning wipes b30: wip anios excel b13: Sani-Cloth AF3 (gray) a10: DDN9 b31: Virusolve® + Pronb14: Protex Ultra Wipes b15: Sani-Cloth HB te all’uso Wipes a1: a2:
Probe
OxivirTM/MC Tb PI-SPRAY II Surfa’safe TRANSEPTIC PROTEXTM DISINFECTANT SPRAY c7: Tristel Duo c8: IODOCLEAN c9: Protex Spray c10: CaviCide c11: T-Spray c12: Indican Form c13: Transeptic Spray c14: CAVICIDE 1 c15: OXIVIR 1 c16: WIP'ANIOS SPOR'ACTIV c17: Accel TB Liquid c1: c2: c3: c4: c5:
[d]: Solutions
c18: Sani-HyPerCide GERMICIDAL SPRAY c19: Sani-24 GERMICIDAL SPRAY
[e]: Devices
e1: d27: Minncare d12: Cavicide d1: CIDEX OPA d2: Cidex Activated Diliquid Liquid disinfectant aldehyde Solution d13: Metricide d28: Virusolve® d14: Metricide 28 d3: Minncare® Cold d15: Metricide + ConSterilant d4: Ster-Bac e2: centrate OPA Plus d29: Virex II 256 d16: Cidex Plus d5: Triacid-N d17: Gigasept AF d6: Revital-Ox® e3: d18: Osvan Resert® High Level Disinfectant/ d19: Neojodin UltrOx™ High-Lev- d20: Milton d21: hibitane el disinfectant d22: Sterihyde d7: Gigasept® PAA d23: Metricide 14 concentrate d8: DESCOTON extra d24: Sekusept plus d9: Gigasept® FF(neu) d25: Wavicide-01 e4: d10: ANIOXYDE 1000 d26: SALVAd11: SALVANIOS pH10 NIOS pH7
[a]: Cleaners
[b]: Wipes
[c]: Sprays
[d]: Solutions
C5-1s/C5-1E/C5-1U/C5-1/C5-2/C5-2s/C5-2E/C6-2E/ C6-2/C6-2s/C6-2P/C11-3E/C11-3s/C11-3U/C11-3
a1, a2, a3, a4, a5, a6, a7, a9, a10
b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, b18, b19, b24, b25, b26, b27, b28, b29, b30, b32, b33
c1, c2, c3, c5, c7, c8, c14, c15, c16, c17, c18, c19
d1, d4, d5, d6, d7, d8, d9, d10, d11, d12, d26, d29
C6-2Gs/C6-2GE/C6-2GU/SC7-1U/SC9-2U/SL103U/SCM7-1U/SLM10-3U/SC6-1GU/C6-1
a1, a2, a3, a4, a5, a6, a7, a9, a10
b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c5, c7, c8, c14, c15, c16, d1, d3, d4, d5, d6, d7, d8, d9, d10, d11, d12, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c17, c18, c19 d15, d26, d27, d28, d29
e1, e2
C7-3E/3C1/3C1s/6C2/6C2P/6C2s/35C20EA/35C 50EA/35C50EB/35C50P/65C15EAV/65C15EA
a1, a2, a3, a4, a5, a6, a7, a9, a10
b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c4, c5, c7, c8, c14, c15, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c16, c17, c18, c19
e1, e2
d1, d2, d3, d4, d5, d6, d7, d8, d9, d10, d11, d12, d13, d14, d15, d26, d27, d28, d29
a1, a2, a3, a4, a5, b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c4, c5, c7, c8, c14, c15, d1, d2, d3, d4, d5, d6, d7, d8, d9, d10, d11, a6, a7, a9, a10 b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c16, c17, c18, c19 d12, d13, d14, d15, d26, d27, d28, d29 a1, a2, a3, a4, a5, b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b16, b18, b19, c1, c2, c3, c5, c7, c14, c15, c16, SC5-1U/SC5-1E d1, d4, d6, d12, d15 a6, a7 b24, b25, b26, b27, b28, b29 c17 a1, a2, a3, a4, a5, b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c5, c7, c8, c14, c15, c16, d1, d4, d5, d6, d7, d10, d11, d12, d15, d26, SC6-1U/SC6-1E/SC6-1s/SC5-1NE/SC5-1Ns/SC5-1N a6, a7, a9, a10 b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c17, c18, c19 d28, d29 SC8-2U/SC8-2E/SC8-2s a1, a8 b15, b25 c4, c11, c16 d1, d9, d16, d20, d25 a1, a2, a3, a4, a5, b1, b2, b3, b4, b5, b7, b8, b10, b11, b12, b16, b18, b19, c1, c2, c3, c4, c5, c7, c14, c15, c16, d1, d2, d3, d4, d5, d6, d12, d13, d14, d15, 65EC10EA/65EC10EB/65EC10EC a6, a7, a9 b24, b25, b26, b27, b28, b29, b30, b31 c17, c18, c19 d27, d28 a1, a2, a3, a4, a5, b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c4, c5, c7, c8, c14, c15, d1, d2, d4, d5, d6, d7, d8, d9, d10, d11, d12, 65EC10ED a6, a7, a9, a10 b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c16, c17, c18, c19 d13, d14, d15, d26, d28, d29 C4-1/C4-1U/C4-1s a1, a3, a4, a5, a6, a7 b4, b5, b13, b14, b15, b16, b17, b18, b26, b32, b33 c2, c9 d1, d2, d6, d12, d13, d14, d15 C9-3Ts a1, a3, a4, a5, a6, a7 b16, b26 c18, c19 d1, d6, d12, d13, d14, d15 3C5/3C5s/3C5A/3C5P
Rely+On PeraSafe
[e]: Devices [f]: Powder e1, e2
e1, e2
Convex
e1, e2 e1, e2 e1 e1, e2, e3 e1, e2, e3 e2 e1, e4
Linear
L7-3/L7-3s/L7-3E/7L4/7L4s/7L4A/7L4P/7L4B/7L4 Bs/7L4BP/7L5/7L5P/7L5s/7L6/7L6s/75L38EA/75 L38EB/75L53EA/75L60EA/75L38P/10L4/10L4s
a1, a2, a3, a4, a5, a6, a7, a9, a10
b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c4, c5, c7, c8, c14, c15, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c16, c17, c18, c19
d1, d2, d3, d4, d5, d6, d7, d8, d9, d10, d11, d12, d13, d14, d15, d26, d27, d28, d29
e1, e2
65EL60EA/6LE7/6LE7s/6LE7P/6LE5V/ 6LE5Vs/6LE5VP/7LT4/7LT4s/7LT4P/7LT4E/ 50L60EAV/65L50HAV/75L50EAV/75LT38EA
a1, a2, a3, a4, a5, a6, a7, a9
b1, b2, b3, b4, b5, b7, b8, b10, b11, b12, b16, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31
c1, c2, c3, c4, c5, c7, c14, c15, c16, c17, c18, c19
d1, d2, d3, d4, d5, d6, d12, d13, d14, d15, d27, d28
e1, e2, e3
a1, a2, a3, a4, a5, a6, a7, a9, a10
b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, b18, b19, b24, b25, b26, b27, b28, b29, b30, b32, b33
c1, c2, c3, c5, c7, c8, c14, c15, c16, c17, c18, c19
d1, d4, d5, d6, d7, d8, d9, d10, d11, d12, d26, d29
e4
7LT4s (only for the socket with black cover) L9-3/L9-3U/L9-3E/L9-3s/L12-3/L12-3E/L133s/L13-3/L12-4/L12-4s/L14-6N/L14-6Ns/L146NP/L14-6NE/L14-6WE/L14-6Ws/L14-6WU
[f]: Powder
f1: TrophonSonex-HL (Used with Trophon/ Trophon2) Germitec UV-C Vaporized Hydrogen Peroxide (Used with V-PRO Low Temperature Sterilization System) STERRAD® system
P/N: 046-008412-00 (36.0)
e1, e2
Release date: 2023-05 1
[a]: Cleaners
[b]: Wipes
[c]: Sprays
b32: Super b1: CLEANISEPT® WIPES b16: CaviWipes MetriZyme b17: Dispatch Towels b2: mikrozid® AF SaniTristel Preb18: Accel TB Wipes Cloth® Wipes Jumbo Clean Wipes b33: SANIb19: CaviWipes 1 b3: PROTEXTM DISINa3: Liquinox b20: Tuffle 5 a4: Revital-OX CLOTH® FECTANT Wipes b21: Sani-Cloth Active b4: Sani-Cloth® Plus PRIME Enzymatic b22: Septiwipes b5: SONOTM ULTRAWIPES Detergent b23: Mikorbac Tissues a5: MetriSponge SOUND WIPES a6: Prolystica 2X b7: Tristel Sporicidal Wipes b24: Sani-Cloth Germicidal Wipes Concentrate b8: Tristel Rinse Wipes b9: Clinell Universal Wipes/ b25: WIP'ANIOS Enzymatic CLEAN'UP Clinell® Surface Wipes Cleaner b10: mikrozid® Senb26: CaviWipes XL a7: Endozime b27: OXIVIR™ WIPES sitive Wipes and Endozb28: Oxivir 1 Wipes ime Sponge b11: Wip’Anios premium b12: ultrasound probe b29: PDI SANI-CLOTH a8: klenzyme a9: ANIOSYME 5 cleaning wipes BLEACH WIPES b13: Sani-Cloth AF3 (gray) b30: wip anios excel a10: DDN9 b14: Protex Ultra Wipes b31: Virusolve® + Pronb15: Sani-Cloth HB te all’uso Wipes a1: a2:
Probe L10-3E/L10-3s L11-3U/L11-3VNs/L13-3N/L13-3Ns/L14-3s/ L14-3WU/L14-3WE/L14-3Ws/L14-3W L11-4/L11-4s/L14-6/L14-6s/L14-6P/10L24EA L16-4HE/L16-4Hs/L16-4HU
[a]: Cleaners a1 a1, a2, a3, a4, a5, a6, a7, a9, a10 a1, a2, a3, a4, a5, a6, a7, a9, a10 a8
OxivirTM/MC Tb PI-SPRAY II Surfa’safe TRANSEPTIC PROTEXTM DISINFECTANT SPRAY c7: Tristel Duo c8: IODOCLEAN c9: Protex Spray c10: CaviCide c11: T-Spray c12: Indican Form c13: Transeptic Spray c14: CAVICIDE 1 c15: OXIVIR 1 c16: WIP'ANIOS SPOR'ACTIV c17: Accel TB Liquid c1: c2: c3: c4: c5:
[d]: Solutions
c18: Sani-HyPerCide GERMICIDAL SPRAY c19: Sani-24 GERMICIDAL SPRAY
[b]: Wipes
[e]: Devices
e1: d27: Minncare d12: Cavicide d1: CIDEX OPA d2: Cidex Activated Diliquid Liquid disinfectant aldehyde Solution d13: Metricide d28: Virusolve® d14: Metricide 28 d3: Minncare® Cold d15: Metricide + ConSterilant e2: d4: Ster-Bac centrate OPA Plus d29: Virex II 256 d16: Cidex Plus d5: Triacid-N e3: d17: Gigasept AF d6: Revital-Ox® d18: Osvan Resert® High Level Disinfectant/ d19: Neojodin UltrOx™ High-Lev- d20: Milton d21: hibitane el disinfectant d22: Sterihyde d7: Gigasept® PAA d23: Metricide 14 concentrate d8: DESCOTON extra d24: Sekusept plus d9: Gigasept® FF(neu) d25: Wavicide-01 e4: d10: ANIOXYDE 1000 d26: SALVAd11: SALVANIOS pH10 NIOS pH7
[c]: Sprays
[d]: Solutions
b25 c5 d1, d25 b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c5, c7, c8, c14, c15, c16, d1, d3, d4, d5, d6, d7, d8, d9, d10, d11, d12, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c17, c18, c19 d15, d26, d27, d28, d29 b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c5, c7, c8, c14, c15, c16, d1, d4, d5, d6, d7, d10, d11, d12, d15, d26, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c17, c18, c19 d28, d29 d1, d2
Linear
a1, a2, a3, a4, a5, a6, a7, a9, a10 a1, a3, a4, a5, a6, a7, a9, a10
b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c4, c5, c7, c8, c14, c15, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c16, c17, c18, c19
d1, d2, d3, d4, d5, d6, d7, d8, d9, d10, d11, d12, d13, d14, d15, d26, d27, d28, d29
e1 e1, e2 e1, e2
e1, e2
b4, b5, b13, b14, b15, b16, b17, b18, b25, b26, b31
c2, c9
d1, d2, d6, d12, d13, d14, d15, d28, d29
a9
b1, b4, b5, b11, b13, b16, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33
c1, c3, c14, c15, c16, c17, c18, c19
d12, d28
a9, a10
b13, b25, b29, b30, b31, b32, b33
c7, c16, c18, c19
d28, d29
L14-5sp
a1, a3, a4, a5, a6, a7
b4, b5, b13, b14, b15, b16, b17, b18, b26, b32, b33
c2, c9
d1, d2, d6, d12, d13, d14, d15
e1, e2
L14-5WU/L14-5WE/L14-5Ws
a1, a3, a4, a5, a6, a7
b4, b5, b13, b14, b15, b16, b17, b18, b26, b32, b33
c2, c9, c11
d1, d2, d6, d12, d13, d14, d15
e2
L20-5U/L20-5s/L20-5E L12-3RCs/L12-3VNs L13-3WE/L13-3WU/L13-3Ws/L15-3WE/L15-3WU/ L15-3Ws/L18-5WU/LM18-5WU/LM24-6WU
L30-8U/L33-8U/L33-8s
a8
P7-3/P7-3s/P7-3E/P7-3U/P7-3P P12-4/P12-4s
Phased
SP5-1U/SP5-1s/SP5-1E/SP5-1 SP5-1Ns/SP5-1N LFP5-1s/LFP5-1U/LFC5-1s/LFC5-1U
P8-2s/P8-2/P8-2U/P8-2P/SP9-2U/SPM6-1U/XP5-1U
e1, e2
d2 c18, c19
L9-3PAU P4-2/P4-2s/P4-2E/P4-2NE/P4-2Ns/P10-4E/ P10-4s/P10-4U/P10-4/2P2/2P2s/2P2P
a1, a2, a3, a4, a5, a6, a7, a9, a10 a1, a8 a1 a1, a2, a3, a4, a5, a6, a7, a9, a10 a1, a2, a3, a4, a5, a6, a7, a9, a10 a10 a1, a2, a3, a4, a5, a6, a7, a9, a10
b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c4, c5, c7, c8, c14, c15, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c16, c17, c18, c19 b3, b4, b15, b25, b29
c2, c4, c5, c16
b5, b25, b29
c1, c2, c5, c16
b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c5, c7, c8, c14, c15, c16, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c17, c18, c19 b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c5, c7, c8, c14, c15, c16, b18, b19, b24, b25, b26, b27, b28, b29, b30, b32, b33 c17, c18, c19 b1, b4, b5, b11, b13, b16, b18, b19, b24, b26, b27, b28, c1, c3, c14, c15, c17, c18, c19 b32, b33
d1, d2, d3, d4, d5, d6, d7, d8, d9, d10, d11, d12, d13, d14, d15, d26, d27, d28, d29
e1, e2
d1, d2, d11, d13, d16, d20, d25
e1
d1, d2, d3, d4, d5, d27
e1
d1, d4, d5, d6, d7, d10, d11, d12, d15, d26, d28, d29 d1, d4, d5, d6, d7, d8, d9, d10, d11, d12, d26, d29
2
e1, e2 e1, e2
d12, d29
b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c5, c7, c8, c14, c15, c16, d1, d3, d4, d5, d6, d7, d8, d9, d10, d11, d12, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c17, c18, c19 d15, d26, d27, d28, d29
Rely+On PeraSafe
[e]: Devices [f]: Powder
e4
L16-4Hs (only for the socket with black cover) LM14-6E/LM14-6s/LM16-4U
[f]: Powder
f1: TrophonSonex-HL (Used with Trophon/ Trophon2) Germitec UV-C Vaporized Hydrogen Peroxide (Used with V-PRO Low Temperature Sterilization System) STERRAD® system
e1, e2
[a]: Cleaners
[b]: Wipes
[c]: Sprays
b32: Super b1: CLEANISEPT® WIPES b16: CaviWipes MetriZyme b17: Dispatch Towels b2: mikrozid® AF SaniTristel Preb18: Accel TB Wipes Cloth® Wipes Jumbo Clean Wipes b33: SANIb19: CaviWipes 1 b3: PROTEXTM DISINa3: Liquinox b20: Tuffle 5 a4: Revital-OX CLOTH® FECTANT Wipes b21: Sani-Cloth Active b4: Sani-Cloth® Plus PRIME Enzymatic b22: Septiwipes b5: SONOTM ULTRAWIPES Detergent b23: Mikorbac Tissues a5: MetriSponge SOUND WIPES a6: Prolystica 2X b7: Tristel Sporicidal Wipes b24: Sani-Cloth Germicidal Wipes Concentrate b8: Tristel Rinse Wipes b9: Clinell Universal Wipes/ b25: WIP'ANIOS Enzymatic CLEAN'UP Clinell® Surface Wipes Cleaner b10: mikrozid® Senb26: CaviWipes XL a7: Endozime b27: OXIVIR™ WIPES sitive Wipes and Endozb28: Oxivir 1 Wipes ime Sponge b11: Wip’Anios premium b12: ultrasound probe b29: PDI SANI-CLOTH a8: klenzyme a9: ANIOSYME 5 cleaning wipes BLEACH WIPES b13: Sani-Cloth AF3 (gray) b30: wip anios excel a10: DDN9 b14: Protex Ultra Wipes b31: Virusolve® + Pronb15: Sani-Cloth HB te all’uso Wipes a1: a2:
Probe Endo-cavity
V11-3/V11-3BE/V11-3B/V11-3WE/ V11-3Ws/V11-3E/V11-3s V11-3HU/V11-3HE/V11-3Hs/V11-3H/V113HB/V11-3HBs/V11-3HBE/SV10-2U
[a]: Cleaners
OxivirTM/MC Tb PI-SPRAY II Surfa’safe TRANSEPTIC PROTEXTM DISINFECTANT SPRAY c7: Tristel Duo c8: IODOCLEAN c9: Protex Spray c10: CaviCide c11: T-Spray c12: Indican Form c13: Transeptic Spray c14: CAVICIDE 1 c15: OXIVIR 1 c16: WIP'ANIOS SPOR'ACTIV c17: Accel TB Liquid c1: c2: c3: c4: c5:
[d]: Solutions
c18: Sani-HyPerCide GERMICIDAL SPRAY c19: Sani-24 GERMICIDAL SPRAY
[b]: Wipes
[e]: Devices
e1: d27: Minncare d12: Cavicide d1: CIDEX OPA d2: Cidex Activated Diliquid Liquid disinfectant aldehyde Solution d13: Metricide d28: Virusolve® d14: Metricide 28 d3: Minncare® Cold d15: Metricide + ConSterilant e2: d4: Ster-Bac centrate OPA Plus d29: Virex II 256 d16: Cidex Plus d5: Triacid-N e3: d17: Gigasept AF d6: Revital-Ox® d18: Osvan Resert® High Level Disinfectant/ d19: Neojodin UltrOx™ High-Lev- d20: Milton d21: hibitane el disinfectant d22: Sterihyde d7: Gigasept® PAA d23: Metricide 14 concentrate d8: DESCOTON extra d24: Sekusept plus d9: Gigasept® FF(neu) d25: Wavicide-01 e4: d10: ANIOXYDE 1000 d26: SALVAd11: SALVANIOS pH10 NIOS pH7
[c]: Sprays
[d]: Solutions
a1, a2, a3, a4, a5, a6, a7, a9, a10 a1, a2, a3, a4, a5, a6, a7, a9, a10
b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c4, c5, c7, c8, c14, c15, d1, d2, d3, d4, d5, d6, d10, d11, d12, d13, b18, b19, b24, b25, b26, b27, b28, b29, b30, b32, b33 c16, c17, c18, c19 d14, d15, d26, d27, d29 b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c5, c7, c8, c14, c15, c16, d1, d3, d4, d5, d6, d7, d8, d9, d10, d11, d12, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c17, c18, c19 d15, d26, d27, d28, d29
a1, a2, a3, a4, a5, a6, a7, a9, a10
b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c4, c5, c7, c8, c14, c15, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c16, c17, c18, c19
V11-3HBs V11-3Hs (only for the socket with black cover) V10-4/V10-4s/V10-4B/V10-4Bs/V10-4BP/ 6CV1/6CV1s/6CV1P
[e]: Devices [f]: Powder e1, e2, e3 e1, e2
d1, d2, d4, d5, d6, d7, d8, d9, d10, d11, d12, d13, d14 d15, d26, d28, d29
e1, e2, e3
Pencil
d1, d2, d4, d5
Biplane
CB10-4/CB10-4P/CB10-4E
a1, a2, a3, a4, a5, a6, a7, a9
b1, b2, b3, b4, b5, b7, b8, b10, b11, b12, b16, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31
c1, c2, c3, c4, c5, c7, c14, c15, c16, c17, c18, c19
d1, d2, d3, d4, d5, d6, d12, d13, d14, d15, d27, d28
e1, e2, e3
6LB7/6LB7s/6LB7P/6LB7E/65EB10EA
a1, a2, a3, a4, a5, a6, a7, a9
b1, b2, b3, b4, b5, b7, b8, b10, b11, b12, b16, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31 b1, b4, b5, b11, b16, b18, b19, b24, b26, b27, b28, b30, b31
c1, c2, c3, c4, c5, c7, c14, c15, c16, c17, c18, c19
d1, d2, d3, d4, d5, d6, d12, d13, d14, d15, d27, d28
e2, e3
c1, c3, c7, c14, c15, c17
d1, d12, d15, d28
a9
d1, d2, d4
4CD4/4CD4s/D6-2E D7-2/D7-2E/D7-2s/SD8-1E/SD8-1s/SD8-1/ SD8-1U/D6-2/D6-2P/D6-2EA/D6-2B/D6-2A D6-2NE/DL14-3U
a1, a2, a3, a4, a5, a6, a7, a9
b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, c1, c2, c3, c4, c5, c7, c8, c14, c15, b18, b19, b24, b25, b26, b27, b28, b29, b30, b31, b32, b33 c16, c17, c18, c19
a1
d1, d2, d3, d4, d5, d6, d10, d11, d12, d13, d14, d15, d26, d27, d28
e1
d1, d18, d19, d20, d21, d22
e1 e1
4D
D8-2E/D8-2U
b10, b16, b19, b20, b21
c2, c10
d1, d2, d9
D8-4U
b4, b10, b16, b19, b20
c2, c10, c11
d1, d2, d16, d17
c2, c10, c11
d1, d2, d16, d17
b10, b16, b19, b20, b21, b32, b33
DE10-3/DE10-3E/DE10-3U/DE10-3s DE11-3U/DE11-3s/DE11-3E/ DE11-3/DE11-3WE/DE11-3Ws/DE11-3WU DE10-3WU/DE10-3WE (only for the probe with gray strain relief)
Rely+On PeraSafe
e4
CW2s/CW5s/CW5
ELC13-4U/ELC13-4s/ELC10-4
[f]: Powder
f1: TrophonSonex-HL (Used with Trophon/ Trophon2) Germitec UV-C Vaporized Hydrogen Peroxide (Used with V-PRO Low Temperature Sterilization System) STERRAD® system
a1, a2, a3, a4, a5, a6, a7
b1, b2, b3, b4, b5, b7, b8, b9, b10, b11, b12, b13, b16, b18, b19, b24, b25, b26, b27, b28, b32, b33
a8
b1, b12, b15, b16, b20, b21, b22, b23, b24
DE10-3WU/DE10-3WE/DE10-3Ws (only for the probe with white strain relief)
3
c1, c2, c3, c5, c7, c8, c14, c15, c17, d1, d3, d4, d5, d6, d7, d8, d9, d10, d11, d12, c18, c19 d15, d26, d27 c2, c11, c12, c13
d1, d2, d15, d16, d17, d23, d24, d25
c18
d1
f1 f1
e1, e2
f1
e1, e3 e1, e2
f1
Active Ingredients of the Cleaner Cleaner MetriZyme/MetriSponge: Revital-OX Enzymatic Detergent:
Endozime and Endozime Sponge: Prolystica 2X Concentrate Enzymatic Cleaner/klenzyme/ Liquinox/Tristel Pre-Clean Wipes: ANIOSYME 5: DDN9
Disinfectant Active Ingredients
Proteinase substilisin Citric acid, Triethanolamine, Ethanolamine, Ethoxylated coconut oil alkyl amine, Subtilisins (proteolytic enzymes), Glycerine Subtilisins (proteolytic enzymes) enzymatic detergent
ionic surfactants, sequestering agent, stabilising agent, enzymatic complex, Excipients N PROPIONATE, N-DIDECYL-N-METHYLPOLY{OXYETHYL)AMMONIUM TETRAPOTASSIUM ETHYLENEDIAMINETETRAACETATE EDETIC ACID
Active Ingredients of the Disinfectant Disinfectant
Active Ingredients
Tristel Trio Wipes System/Tristel Jet/ Tristel Duo/Tristel Sporicidal Wipes:
chlorine dioxide
IODOCLEAN
sodium thiosulfate and excipients
Ster-Bac/PI-SPRAY II/mikrozid® Quaternary Ammoniums Sensitive Wipes/Clinell Universal Wipes/Sani-Cloth HB/Sani-Cloth Active/T-Spray/Mikorbac Tissues/SaniCloth Germicidal Wipes/SALVANIOS pH7/SALVANIOS pH10/Protex Ultra Wipes/CaviWipes XL/CAVICIDE 1: Tristel Rinse Wipes: deionized water Cidex OPA: 0.55% Ortho-phthladehyde DESCOTON extra/Wavicide-01: glutaraldehyde OxivirTM/MC Tb: 0.5% hydrogen peroxide Sani-Cloth® Plus: n-Alkyl dimethyl benzyl ammonium chloride n-Alkyl ethylbenzyl ammonium chloride CaviWipes: isopropanol, ethylene glycol monobutyl ether (2-butoxyethanol), diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride, water CaviWipes 1: isopropanol, ethanol, ethylene glycol monobutyl ether (2-butoxyethanol), didecyldime-thylammonium chloride, water Dispatch Towels: Sodium hydroxide, Sodium metasilicate, Sodium hypochlorite TM SONO ULTRASOUND WIPES: octyl decyl dimethyl ammonium chloride; dioctyl dimethyl ammonium chloride; didecyl dimethyl ammonium chloride; dimethyl benzyl ammonium chloride Minncare® Cold Sterilant/ 22% Hydrogen Peroxide, 4.5% Minncare liquid disinfectant: Peroxyacetic Acid mikrozid® AF Wipes Jumbo: 25% ethanol; 35% propan-1-ol CLEANISEPT® WIPES: 0.25g didecyldimethylammoniumchloride 0.5g quaternary ammonium compounds, benzyl-C12-16-alkyldimethyl-, chlorides Wip’Anios premium: didecyldimethylammonium chloride 1.4mg/g, polyhexamethylene biguanide hydrochloride 0.96mg/g Cidex Activated Dialdehyde Solution: glutaraldehyde TRANSEPTIC: isopropyl alcohol, chlorhexidine gluconate Protex Spray: Quaternary ammonium compounds diC8-10-alkyldimethyl, chlorides Water
Active Ingredients
Sani-HyPerCide GERMICIDAL SPRAY hydrogen Sani-24 GERMICIDAL SPRAY Quaternary Ammoniums, Ethanol SANI-CLOTH ® PRIME WIPES Quaternary Ammoniums ® Hydrogen peroxide plasma STERRAD system gigasept® FF(neu):
®
gigasept PAA concentrate: ProtexTM Disinfectant Spray/ ProtexTM Disinfectant Wipes:
Triacid-N: Surfa’safe:
®
®
Revital-Ox Resert High Level Disinfectant/TrophonSonex-HL: ANIOXYDE 1000: Sani-Cloth AF3: Metricide OPA Plus: Metricide/Metricide 14/Metricide 28: CaviCide:
Gigasept AF: Rely+On PeraSafe: Cidex Plus: Sonogel: Sterihyde: Osvan: Neojodin: Milton: Hibitane: WIP’ANIOS CLEAN’UP: OXIVIR 1/OXIVIR™ WIPES/ Oxivir 1 Wipes/Accel TB Liquid/Accel TB Wipes: WIP’ANIOS SPOR’ACTIV: PDI SANI-CLOTH BLEACH WIPES: wip anios excel: ®
Virusolve + Pronte all’uso Wipes: ®
Virusolve + Concentrate:
Super Sani-Cloth Virex II 256
®
0.11g succindialdehyde; 0.3g dimethoxytetrahydrofuran; < 5% anionic surfactant; non-ionic surfactants; anticorrosion compounds; fragrance peracetic acid (5%); hydrogen peroxide: acetic acid; potassium hydroxide: corrosion inhibitor octyl decyl dimethyl ammonium chloride; dioctyl dimethyl ammonium chloride; didecyl dimethyl ammonium chloride; dimethyl benzyl ammonium chloride N-Dodecylpropan-1,3-diamin; propan-2-ol; isotridecanol, ethoxylated; non-ionic detergent didecyldimethylammonium chloride 1.4mg/g, polyhexamethylene biguanide hydrochloride 0.96mg/g Hydrogen peroxide 3% Hydrogen Peroxide Quaternary ammonium chlorides. Ortho-phthladehyde Glutaraldehyde isopropanol, ethanol, ethylene glycol monobutyl ether (2-butoxyethanol), diisobutylphe-noxyethoxyethyl dimethyl benzyl ammonium chloride, water didecyldimethylammonium chloride, glycine, aminoalkyl derivs tridecylpolyethylenglycoether. disodium carbonate, citric acid glutaraldehyde carbomer, polyacrylat Glutaraldehyde Ammonia benzalkonium povidone-iodine sodium hypochlorite Hlorhexidine Association of surfactants, excipients Hydrogen peroxide
Registered disinfectants in Canada Drug Identification NO./ License NO.
Name
74736
V-PRO Low Temperature Sterilization System
2981
STERRAD system
®
®
®
02286467
Revital-Ox Resert High Level Disinfectant
100935
Trophon/Trophon2
01963996
Metricide
01964461
Metricide 28
02197170
Wavicide-01
02277484
Minncare liquid disinfectant
02161656
Cavicide Liquid
02247354
Super Sani-Cloth
02517000
SANI-CLOTH ® PRIME WIPES
®
Registered disinfectants in FDA region Registration NO.
Name
FDA K190103 510(k)
V-PRO Low Temperature Sterilization System
Sterilization
K924434
Cidex Activated Dialdehyde Solution
High-level disinfectant and Sterilization
K030004
CIDEX OPA
High-level disinfectant
K923744
Cidex Plus
High-level disinfectant and Sterilization
K080420
Revital-Ox Resert High Level Disinfectant
High-level disinfectant
K103059
Trophon
High-level disinfectant
K173865
Trophon2
High-level disinfectant
K930284
Metricide
High-level disinfectant and Sterilization
K931052
Metricide 28
High-level disinfectant and Sterilization
K140703
Metricide OPA Plus
High-level disinfectant
K914749
Wavicide-01
High-level disinfectant and Sterilization
K954142
STERRAD system
EPA
Microbicidal Method
®
®
®
Sterilization
®
9480-6
Sani-Cloth Plus
Low-level disinfectant
70627-56
Oxivir
TM/MC
Low-level disinfectant
Hydrogen peroxide and peroxyacetic acid mixture stabilized Sodium hypochlorite Didecyldimethylammonium chloride, non-ionic surfactants, sequestering agent, excipients Polycarboxylate, non ionic emulsifier, alcohol, Didecyldimethyl ammonium Chloride, Bis(3-aminopropyl)dodecylamine 2-Aminoethanol, Didecyldimethyl ammonium Chloride, Propan-2-ol, Potassium Carbonate
9480-9
Sani-Cloth AF3 (gray)
46781-8
CAVIWIPES
Low-level disinfectant
46781-6
CAVICIDE
Low-level disinfectant
70627-60
OXIVIR WIPES
46781-12
CAVICIDE 1
9480-4
Super Sani-Cloth
9480-12
SANI-CLOTH PRIME WIPES
Low-level disinfectant
Quaternary Ammoniums
9480-14
Sani-HyPerCide GERMICIDAL SPRAY
Low-level disinfectant
421829-9480
Sani-24 GERMICIDAL SPRAY
Low-level disinfectant
9480-8
PDI SANI-CLOTH BLEACH WIPES
Low-level disinfectant
70627-24
Virex II 256
Low-level disinfectant
Quaternary Ammoniums
4
Tb
™
Low-level disinfectant
Low-level disinfectant Low-level disinfectant ®
®
Low-level disinfectant
•
Cleaning, Disinfecting (Sterilizing) Procedures After completing each examination, clean, disinfect or sterilize the probes as required. If necessary, repeat the cleaning, disinfection (or sterilization) process before next use. When biopsy procedures have been performed, be sure to sterilize the needle-guided bracket. Fail to do so may result in the probe and the needle-guided bracket to becoming sources of infection. Please follow the instructions in the manual for cleaning.
•
WARNING Never immerse the probe connector into liquids such as water or disinfectant, for the connector is not waterproof. Immersion may cause electric shock or malfunction.
–
CAUTION • •
•
No cleaning and disinfecting may result in the probe becoming a source of infection. Please follow the disinfectant manufacturer’s manual for performing cleaning and disinfection, including preparing sterile water and cleaning and disinfection time.
• • • • • • •
• •
Sterilization describes a process that destroys or eliminates all forms of microbial life and is carried out in healthcare facilities by physical or chemical methods.
Probes can be divided into three categories based on their intended use according to the standard ISO 17664-1:2021. Some probes may fall into more than one category (e.g. probes use for biopsy procedures). When selecting a disinfectant, determine the required level of disinfection based on intended use and possibility of cross-contamination. • Non-critical items: come into contact with intact skin only or are devices not intended for direct patient contact. Probes that only come into contact with clean, intact skin are considered noncritical devices and require cleaning after every use. Cleaning may be followed by a lowlevel disinfectant spray or wipe. For details, see “ Processing Non-Critical Probes”. • Semi-critical items: come into contact with mucous membranes. This category includes all endocavity probes - transvaginal, transrectal, and transesophageal (TEE). These semicritical probes must be cleaned with an appropriate cleaner after use followed by high-level disinfection. For details, see “ Processing Semi-Critical Probes”. • Critical items: enter normally sterile parts of the human body. These probes are considered critical and include all intraoperative probes. These probes must be cleaned with an appropriate cleaner after each use, followed by a sterilization process. For details, see “ Processing Critical Probes”.
After the examination, wipe off the ultrasound gel thoroughly. Otherwise, the ultrasound gel may solidify and degrade the image quality of the probe. DO NOT make the probe to become overheated (more than 55 °C) during cleaning and disinfections. High temperature may cause the probe to become deformed or damaged. Observe the illustration graph carefully to immerse the probe. Only soak parts of the probe below the strain relief. Repeated disinfection or sterilization will eventually damage the probe, please check the probe performance periodically. Clean the probe thoroughly in accordance with the cleaning procedure before disinfection or sterilization. For details about probe types, refer the operator’s manuals of the ultrasound system. For details about recommended disinfectants for probes, see the “Cleaner and Disinfectant” table. For use of each disinfectant, please refer to the manufacturer’s instructions. For V-PRO Low Temperature Sterilization System: – The probe should be placed into a STERIS Sterilization Tray and wrapped with sterilization wrap. In Canada/FDA region, the STERIS Sterilization Tray and sterilization wrap should be cleared by the FDA/HC, such as H600 OneStep® sterilization wrap. –
High-Level Disinfection (HLD): Destruction/removal of all microorganisms except bacterial spores.
Selecting a Cleaning, Disinfection/Sterilization Method
NOTE: •
Cleaning is the removal of visible soil (e.g. organic and inorganic material) from objects and surfaces and normally is accomplished manually or mechanically using water with detergents or enzymatic products. Thorough cleaning is essential before high-level disinfection and sterilization because inorganic and organic material that remains on the surfaces of instruments interfere with the effectiveness of these processes. Disinfection describes a process that eliminates many or all pathogenic microorganisms, except bacterial spores. – Low-Level Disinfection (LLD): Destruction of most bacteria, some viruses, and some fungi.
How to Select a Processing Method
Semi-critical
Non-critical
Start the sterilization system using the Non Lumen Cycle according to the instructions provided by the manufacturer.
Percutaneous procedure
Body surface
Please select the proper disinfectant for the probe of the Diagnostic Ultrasound System. The content of this document shall prevail in case of other new editions. Refer to local regulations for the use of each disinfectant.
Clean intact skin
Cleaning, Disinfection, and Sterilization Overview
Contaminated intact skin
No cover
Cleaning and disinfection refer to two distinct processes. According to the Centers for Disease Control and Prevention (CDC) “Guideline for Disinfection and Sterilization in Healthcare Facilities” (2008):
Single-use cover
Non-sterile gel
• Cleaning • LLD
5
Non-intact skin
• Cleaning • LLD*
Sterile gel
Needle placement
Percutaneous biopsies
• Single-use cover • Sterility dictated by procedure sterility
Sterile cover
Sterile gel
Critical
Other endocavity use
Transesophageal
Intra-operative
Single-use cover
Sterility dictated by procedure sterility
Sterility dictated by procedure sterility
Sterile or bacteriostatic gel
Sterile gel
Sterile gel
• Cleaning • HLD
• Cleaning • Sterilization
NOTE •
•
This step is to remove the ultrasound gel or other visible dirt. 1. Wear a pair of gloves to prevent infection through the whole processing. 2. Disconnect the probe from the system. If the sheath is used, take off the sheath and discard it. Wipe off the ultrasound gel or other visible dirt on the surface of the probe by using a damp piece of disposable lint-free soft cloth or tissue.
LLD marked with * indicates that those categories must undergo low-level disinfectants that are effective against the mycobacteria and bloodborne pathogens. For details, consider referencing the position statement of the American Institute of Ultrasound in Medicine “Guidelines for Cleaning and Preparing External- and Internal-Use Ultrasound Transducers and Equipment Between Patients as well as Safe Handling and Use of Ultrasound Coupling Gel” at https://www.aium.org/officialstatements/57. For non-critical probes that require use of cover, if there is a suspected protective cover failure, then they are considered as semi-critical items, and HLD is recommended.
Cleaning Select wipes or detergent to clean the probe. For details about recommended cleaners or disinfectants for probes, see the “Cleaner and Disinfectant” table.
Processing Non-Critical Probes
Cleaning with Wipes Perform the following procedure: 1. Use an approved cleaning or disinfectant wipe, cleaning sponge, or a soft cloth soaked in approved cleaner or disinfectant to clean all surfaces of the probe thoroughly. 2. Dry the probe with a piece of disposable lint-free soft cloth or tissue. Do not dry the probe by heating. 3. Inspect the probe. If visible dirt still exists, repeat the preceding steps to wash the probe until it is all clean.
Processing of non-critical probes requires a two-step process: Cleaning of the probe followed by low-level disinfection.
WARNING Use protective eyewear when disinfecting the probe using sprays. Perform the following procedure: 1. Wear a pair of gloves to prevent infection through the whole processing. 2. Disconnect the probe from the system. If the sheath is used, take off the sheath and discard it. 3. Clean the probe. a. Select an appropriate low-level disinfectant wipe or a piece of disposable lint-free soft cloth soaked with a disinfectant spray. For details about recommended disinfectants for probes, see the “Cleaner and Disinfectant” table. b. Wipe all the surface of the probe according to the wiping duration specified in the operator’s manual provided by the manufacturer. When necessary, clean and disinfect the seams or biopsy guide features by using disposable cotton swabs. 4. Disinfect the probe. Prepare a new low-level disinfectant wipe or a piece of disposable lint-free soft cloth soaked with a disinfectant spray to wipe the probe again. 5. Inspect the probe. If visible dirt still exists, repeat the preceding steps to wipe the probe until it is all clean. 6. Allow the probe to air dry in a clean and well-ventilated place or dry the probe with a piece of disposable lint-free soft cloth or tissue. Do not dry the probe by heating. 7. Check whether the probe has defects such as peeling, rifts, bumps, cracks, or liquid spill. If such defects exist, the probe has reached the end of its service life. In this case, stop using it and contact the Mindray service department. 8. Store the probe in a cool, clean and dry environment.
Cleaning with Detergent Perform the following procedure: 1. Choose an appropriate cleaning agent including mild detergents, and enzymatic. 2. Immerse the probe fully in the cleaning fluid for at least 1 minute or according to manufacturer’s instructions. Lightly clean the probe with a piece of lint-free soft cloth or soft sponge until no dirt is visible. When necessary, clean the seams or biopsy guide features by using disposable cotton swabs. Avoid using a brush to wash the lens because it may damage the probe. Observe the graph here carefully to immerse the probe. Only soak parts of the probe below the strain relief.
Strain relief Probe handle Liquid level Approx.10 mm
Processing Semi-Critical Probes Processing of semi-critical probes requires a two-step process: Cleaning of the probe followed by high-level disinfection. For detailed information on the TEE cleaning and disinfection, see the accompanying TEE manual.
3. Rinse the probe thoroughly by using a large amount of clean water (about 7.5 L/2 gallons) at room temperature for about 1 minute to remove the residual dirt and cleaning solvent. 4. Dry the probe with a piece of disposable lint-free soft cloth or tissue. Do not dry the probe by heating.
Before Processing 6
5. Inspect the probe. If visible dirt still exists, repeat the preceding steps to wash the probe until it is all clean.
Before Processing This step is to remove the ultrasound gel or other visible dirt. 1. Wear a pair of gloves to prevent infection through the whole processing. 2. Disconnect the probe from the system. If the sheath is used, take off the sheath and discard it. Wipe off the ultrasound gel or other visible dirt on the surface of the probe by using a damp piece of disposable lint-free soft cloth or tissue.
High-Level Disinfection Perform the following procedure: 1. Disinfect the probe by using an appropriate high-level disinfectant solution or device. For how to use a high-level disinfectant or device, see the operator’s manual provided by the manufacturer. Prepare a disinfectant by using sterile distilled or softened water when necessary. For details about recommended disinfectants or devices for probes, see the “Cleaner and Disinfectant” table. – Soaking: Immerse the probe head in the disinfectant and shake the probe appropriately to remove any bubbles on the probe surface. For details about the probe immersion duration, see the operator’s manual provided by the manufacturer.
Cleaning Select wipes or detergent to clean the probe. For details about recommended cleaners or disinfectants for probes, see the “Cleaner and Disinfectant” table.
Cleaning with Wipes Perform the following procedure: 1. Use an approved cleaning or disinfectant wipe, cleaning sponge, or a soft cloth soaked in approved cleaner or disinfectant to clean all surfaces of the probe thoroughly. 2. Dry the probe with a piece of disposable lint-free soft cloth or tissue. Do not dry the probe by heating. 3. Inspect the probe. If visible dirt still exists, repeat the preceding steps to wash the probe until it is all clean.
Observe the graph here carefully to immerse the probe. Only soak parts of the probe below the strain relief.
Strain relief Probe handle
Cleaning with Detergent
Liquid level Approx.10 mm
–
Perform the following procedure: 1. Choose an appropriate cleaning agent including mild detergents, and enzymatic. 2. Immerse the probe fully in the cleaning fluid for at least 1 minute or according to manufacturer’s instructions. Lightly clean the probe with a piece of lint-free soft cloth or soft sponge until no dirt is visible. When necessary, clean the seams or biopsy guide features by using disposable cotton swabs. Avoid using a brush to wash the lens because it may damage the probe. Observe the graph here carefully to immerse the probe. Only soak parts of the probe below the strain relief.
Wiping: Use a market disinfection wipe product or sterile disposable lint-free soft cloth wetted with disinfection spray and wipe all surfaces of the probe for a duration according to the manufacturer instructions.
Strain relief
2. Rinse the probe thoroughly by using a large amount of clean water (about 7.5 L/2 gallons) at room temperature for about 1 minute to remove the residual disinfectant. Or follow the disinfectant manufacturer’s instructions regarding rinsing. Dry the probe with a piece of disposable lint-free soft cloth or tissue. Do not dry the probe by heating. 3. Check whether the probe has defects such as peeling, rifts, bumps, cracks, or liquid spill. If such defects exist, the probe has reached the end of its service life. In this case, stop using it and contact the Mindray service department. 4. Store the probe in a cool, clean and dry environment.
Probe handle Liquid level Approx.10 mm
Processing Critical Probes Processing of critical probes requires a two-step process: Cleaning of the probe followed by sterilization. For detailed information on the laparoscopic cleaning and disinfection, see the accompanying laparoscopic manual.
3. Rinse the probe thoroughly by using a large amount of clean water (about 7.5 L/2 gallons) at room temperature for about 1 minute to remove the residual dirt and cleaning solvent. 4. Dry the probe with a piece of disposable lint-free soft cloth or tissue. 7
Do not dry the probe by heating. 5. Inspect the probe. If visible dirt still exists, repeat the preceding steps to wash the probe until it is all clean.
Sterilization For intra-operative probes, they have to be thoroughly cleaned and sterilized after completing each examination. Perform the following procedure: 1. Sterilize the probe by using an appropriate sterilant or device. For details about recommended sterilants or devices for probes, see the “Cleaner and Disinfectant” table. For how to use a device, see the operator’s manual provided by the manufacturer. When using a sterilant, follow the following steps: a. Prepare a sterilant by using sterile distilled or softened water when necessary. b. Immerse the probe head in the sterilant and shake the probe appropriately to remove any bubbles on the probe surface. For details about the probe immersion duration, see the operator’s manual provided by the manufacturer.
Liquid level
Immersion Sterilization (7LT4s as example)
c. Rinse the probe thoroughly by using a large amount of sterile distilled or softened water (about 7.5 L/2 gallons) at room temperature for about 1 minute to remove the residual disinfectant. Or follow the sterilant manufacturer’s instructions regarding rinsing. Dry the probe with a piece of sterile disposable lint-free soft cloth. Do not dry the probe by heating. 2. Check whether the probe has defects such as peeling, rifts, bumps, cracks, or liquid spill. If such defects exist, the probe has reached the end of its service life. In this case, stop using it and contact the Mindray service department. 3. Store the probe in a cool, clean and dry environment.
8