NSK

SCL-LED Coupling

Coupling Reprocessing Instructions

2 Pages

Reprocessing Instruction in accordance with ISO17664  Coupling 1 Preparation 20sec.  Cleaning wipes  2 Cleaning, Disinfection and Drying Manual 38°C  3.5L/min  500lx  15sec.  5sec.  15sec. Coupling  Coupling Wrench / Wrench  *PTL-CL-LED and KCL-LED are shown as typical models.  2sec.  M4  B2  3times 1min  WL-Adapter 01 (ALPRO)  Minuten Wipes (ALPRO)  WL-clean (ALPRO)  WL-Adapter 01 (ALPRO)  3sec. 1time  3sec.  M4  B2  WL-Adapter 01 (ALPRO)  1time  WL-cid (ALPRO)  WL-dry (ALPRO) 2min  M4  B2  WL-Adapter 01 (ALPRO)  3sec. 1time  WL-Blow (ALPRO)  ・After cleaning remove any remaining moisture using a dry cloth or compressed air ensuring it is dry internally and externally. ・Cover the product with a cloth when drying with compressed air to prevent water aerosol from being produced. ・Use a nationally approved disinfectant with proven bactericidal, fungicidal, and virucidal properties.
File Type: PDF
File Size: 586 KB
File Name: NSK - NAKANISHI - OM-DT0926EN - Coupling Reprocessing Instructions - 2020-07.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions