CUSTOMER NOTIFICATION | MARCH, 2023
Summary of Validated Cleaning & Sterilization Methods for St. Jude Medical™ Brand EP Catheter Cables
BACKGROUND Abbott (formerly St. Jude Medical) published cleaning and sterilization methods for many of our St. Jude Medical™ brand electrophysiology catheter cables in 2015. This letter replaces the past communication (SJM-AFGEN-0914-0007.) Please be aware that the exact same information on cleaning and sterilization of our cables is in each Instructions for Use (IFU) that is packaged with each cable. Please refer to the IFU accompanying the product when in doubt. Please take note that for all of the listed cables, the maximum number of uses, regardless of the sterilization and cleaning method combination used, is 10. You will be able to determine which methods are applicable to each individual cable based on the information provided in this document. The appendix in this document lists all St. Jude Medical brand cables affected by this change, as well as detailed cleaning and sterilization instructions.
QUESTIONS AND ADDITIONAL INFORMATION: If you have any questions regarding this change, please contact your local Abbott Representative or local Abbott Customer Service.
2 | Summary of Validated Cleaning and Sterilization Methods for St. Jude Medical™ Brand EP Catheter Cables
APPENDIX
EtO
STERRAD
Auto-clave
EP EXT CBL 20-pole, 150 cm (for Reflexion
Yes
Yes
Yes
No
EP EXT CBL 2/3/4-pole, 150 cm, black
Yes
Yes
Yes
No
EP EXT CBL 6-pole, 150 cm, black
Yes
Yes
Yes
No
401972
EP EXT CBL 6/8/10-pole, 150 cm, black
Yes
Yes
Yes
No
401973
EP EXT CBL 2/3/4-pole, 150 cm, blue
Yes
Yes
Yes
No
401974
EP EXT CBL 2/3/4-pole, 150 cm, grey
Yes
Yes
Yes
No
401975
EP EXT CBL 2/3/4-pole, 150 cm, red
Yes
Yes
Yes
No
401976
EP EXT CBL 6/8/10-pole, 210 cm, black
Yes
Yes
Yes
No
401977
EP EXT CBL 6/8/10-pole, 150 cm, red
Yes
Yes
Yes
No
401980
EPS EXT CBL 2/3/4-pole, 150 cm, black
Yes
Yes
Yes
No
401981
EPS EXT CBL 2/3/4-pole, 150 cm, blue
Yes
Yes
Yes
No
401982
EPS EXT CBL 2/3/4-pole, 150 cm, grey
Yes
Yes
Yes
No
401983
EPS EXT CBL B2/3/4-pole, 150 cm, red
Yes
Yes
Yes
No
401984
EPS EXT CBL 6-pole, 150 cm, black
Yes
Yes
Yes
No
401985
EPS EXT CBL 8/10-pole, 150 cm, black
Yes
Yes
Yes
No
401986
EPS EXT CBL 8/10-pole, 210 cm, black
Yes
Yes
Yes
No
402558
SAF EXT CBL to EPT-1000‡ XP, 150 cm
Yes
Yes
Yes
No
402560 SAF EXT CBL to IBI-1500T™ & Ampere™
Yes
Yes
Yes
No
402561
SAF EXT CBL to IBI-1500T & Ampere gen.,
Yes
Yes
Yes
No
402566
SAF EXT CBL to Stockert EP-Shuttle™, 150
Yes
Yes
Yes
No
402567
SAF EXT CBL to Stockert EP-Shuttle, 300
Yes
Yes
Yes
No
402505 EPTC EXT CBL to Stockert EP-Shuttle, 300
Yes
Yes
Yes
No
402543
EPTC EXT CBL to IBI-1500T & Ampere
Yes
Yes
Yes
No
402544 EPTC EXT CBL to IBI-1500T & Ampere
Yes
Yes
Yes
No
85735
1735-D EXT CBL to IBI-1500T & Ampere
No
Yes
Yes
No
85930
1910-S EXT CBL 10-pole, 150 cm
Yes
Yes
Yes
Yes
85931
1924-S EXT CBL 20/24-pole, 150 cm
Yes
Yes
Yes
Yes
85942
1910-S EXT CBL 10-pole, 250 cm
Yes
Yes
Yes
Yes
85945
1914-SA EXT CBL 14-pole, 210 cm
Yes
Yes
Yes
Yes
85953
1904-SA EXT CBL 4-pole, 150 cm
Yes
Yes
Yes
Yes
85954
1910-SA EXT CBL 10-pole, 150 cm
Yes
Yes
Yes
Yes
85955
1914-SA EXT CBL 14-pole, 250 cm
Yes
Yes
Yes
Yes
85201
2101-C EXT CBL for 1400-CV, 240 cm
No
Yes
Yes
No
85205
2102-C EXT CBL for 1400-CV II, 240 cm
No
Yes
Yes
No
Reorder number
Auto-wash
The section below provides updated cleaning and sterilization methods for the following cable model numbers. Manual cleaning and the sterilization with EtO and STERRAD‡ systems are applicable to all cables as described below. Please observe the limited applicability of the auto wash cleaning method and the autoclave sterilization method for individual cables.
Cables for Response™, 401661 Reflexion™ and Livewire™ 401970 diagnostic catheters 401971
Cables for Supreme™ diagnostic catheters
Cables for Safire™ dry tip ablation catheters
Cables for Livewire™ TC ablation catheters
Cables for Inquiry™ diagnostic catheters
Cables for Inquiry™ cardioversion catheters
Abbott Medical Australia Pty Ltd | 3
Auto-wash
EtO
STERRAD
Auto-clave
85641
1641 EXT CBL to IBI-1500T & Ampere™
No
Yes
Yes
No
85644
1641-A EXT CBL to IBI-1500T & Ampere gen., 250 cm
Yes
Yes
Yes
Yes
85711
1711-M EXT CBL to Medtronic ATAKR‡, 250 cm
No
Yes
Yes
No
85713
1713-W EXT CBL to Stockert EP-Shuttle™, single thermocouple, 250 cm
No
Yes
Yes
No
85739
1739-W EXT CBL to Stockert EP-Shuttle, dual thermocouple 250 cm
No
Yes
Yes
No
85763
1763-E EXT CBL to EPT-1000‡ XP, 4mm-tip thermistor, 250 cm
No
Yes
Yes
No
85778
1778-E EXT CBL to EPT-1000‡ XP, 8mm-tip thermistor, 250 cm
No
Yes
Yes
No
85643
1689 EXT CBL to Osypka, dual thermocouple, 250 cm
No
Yes
Yes
No
85708
1711-MA EXT CBL to Medtronic ATAKR‡, 250 cm
Yes
Yes
Yes
Yes
85709
1713-WA EXT CBL to Stockert EP-Shuttle, single thermocouple, 250 cm
Yes
Yes
Yes
Yes
85761
1761-W EXT CBL to Stockert EP-Shuttle, 4mm-tip thermistor, 250 cm
No
Yes
Yes
No
85809
1804-S St. Jude Medical™ Generators to Recording System
Yes
Yes
Yes
Yes
85684
1684-TH EPT Thermistor EXT CBL to IBI- 1500T, 250 cm
No
Yes
Yes
No
85719
1719-W Biosense Webster Thermocouple EXT CBL to IBI-1500T, 250 cm
No
Yes
Yes
No
85760
1760-M Medtronic Thermocouple EXT CB to IBI-1500T, 250 cm
No
Yes
Yes
No
85762
1762-W Biosense Webster Thermistor EXT CBL to IBI-1500T™, 250 cm
No
Yes
Yes
No
85764
1764-W Biosense Webster Dual Thermocouple EXT CBL to IBI-1500T, 250
No
Yes
Yes
No
85769
1769-BS Bard Thermistor EXT CBL to IBI-1500T, 250 cm
No
Yes
Yes
No
A402892
1611 EXT CBL to EnSite™ Contact™ Hardware Kit, 250 cm
No
Yes
Yes
No
Reorder number Cables for Therapy™ & Safire™ irrigated ablation catheters Cables for Therapy™ ablation catheters
EP Monitor cable Cable for non-St. Jude Medical ablation catheters
Cable for St. Jude Medical Contact Enabled™ catheters
4 | Summary of Validated Cleaning and Sterilization Methods for St. Jude Medical™ Brand EP Catheter Cables
Auto-wash
EtO
STERRAD
Auto-clave
D-AVSECBL12
Sensor Enabled™ Diagnostic Catheter Cable 12-Pin
Yes
Yes
Yes
Yes
Cable for Advisor HD Grid, Sensor Enabled™ Diagnostic Catheter Cables
D-AVSECBL22
Sensor Enabled™ Diagnostic Catheter Cable 22-Pin
Yes
Yes
Yes
Yes
Cable for FlexAbility Sensor Enabled™ Ablation catheters
A-FASECBL4
Sensor Enabled™ Ablation Connection Cable
Yes
Yes
Yes
Yes
Reorder number Cable for Advisor Fixed Loop Sensor Enabled™ catheters
Abbott Medical Australia Pty Ltd | 5
CABLE REUSE »P roperly clean and re-sterilize cables within twentyfour (24) hours after use. »D o not submerge cables in an enzymatic detergent or water bath during transport. Avoid contact with contaminated equipment or instruments during cable transport. »C ables are intended to be reused a maximum of ten (10) times. Discard cables that have reached this number of uses. » I nspect cables prior to reuse. Do not use cables if there is visible damage. Discard damaged cables. »E nsure that the connection is secure between the cable and connecting equipment. Note: To ensure cables are properly cleaned and externally sterile, use the specified cleaning and sterilization methods/parameters. Deviations from the specified methods/parameters may result in reduced instrument life or inadequate cleaning or sterilization. If alternate methods/parameters are used, it is the customer’s responsibility to ensure cables are effectively cleaned and sterilized. The sterilization and cleaning methods listed in this manual are only valid when using properly installed, maintained, calibrated and compliant cleaning and sterilization equipment.
*Only applicable to select cables as indicated in the cable listing above in Autowash column. 6 | Summary of Validated Cleaning and Sterilization Methods for St. Jude Medical™ Brand EP Catheter Cables
CABLE CLEANING METHODS
MANUAL METHOD
1. S oak the cable for five (5) minutes, fully submerged in ENZOL‡ enzymatic detergent (or equivalent), diluted according to the manufacturer’s recommendations with warm 31°–43° C (88°–110° F) tap water. Remove visible soil with a soft-bristled brush. 2. Rinse the cable with clean, warm, running tap water for a minimum of one (1) minute. 3. D ry with a clean, lint-free cloth. 4. A fter cleaning, visually inspect cables for any remaining soil or moisture. Repeat cleaning and/or drying if necessary. Note: Cable must be thoroughly dried before sterilization. AUTOMATED METHOD* Preparation for Automated Cleaning: Prior to putting cables in the washer/disinfector, perform the following steps to ensure that any organic matter on the cables is rehydrated. 1. S oak the cable for one (1) minute, fully submerged in ENZOL‡ enzymatic detergent (or equivalent), diluted according to the manufacturer’s recommendations with warm 31°–43° C (88°–110° F) tap water. 2. R emove visible soil with a soft-bristled brush. 3. Rinse the cable with clean, warm 31°–43° C (88°–110° F), running tap water for a minimum of one (1) minute. Automated Cleaning: 1. Load cables into a tray in a manner that is consistent with the manufacturer’s instructions. Do not place heavy objects on top of cables or place cables next to sharp objects during an automated cleaning cycle. Doing so may damage the cables. 2. S et the automated washer/disinfector RELIANCE‡ 444 (or legally marketed equivalent) to the following settings prior to running the cleaning cycle: Note: No thermal disinfection is recommended. Cycle
Time
Water Temperature (Tap or Purified)
Detergent & Concentration
Pre-Wash
1 min
Cold/Warm ≈ 16°-43°C (61°- 110°F)
N/A
Enzymatic Wash
1 min 30 sec
Warm/Hot ≈ 43°-60°C (110°- 140°F)
PROLYSTICA‡ 2X concentrate enzymatic presoak and cleaner (or equivalent), diluted per manufacturer’s instructions
Wash
2 min 30 sec
Heated 60°-70°C (140°-158°F)
PROLYSTICA‡ 2X concentrate neutral detergent (or equivalent), diluted per manufacturer’s instructions
Final Rinse
1 min 30 sec
Heated 60°-70°C (140°-158°F)
N/A
Drying
1 hour
Air Temperature 70°-80°C (158°-176°F)
N/A
3. A fter cleaning, visually inspect cables for any remaining soil or moisture. Repeat cleaning and/or drying if necessary.
Abbott Medical Australia Pty Ltd | 7
STERILIZATION METHODS ETO STERILIZATION
STERRAD‡ STERILIZATION SYSTEMS
Place cable into a breathable package suitable for EtO sterilization (e.g. TYVEK‡ pouch or equivalent). Load cables into chamber and run equipment in a manner that is consistent with the manufacturer’s instructions.
Ensure cables are thoroughly dried before loading into the STERRAD system to avoid a potential cycle cancellation. Package the cable per facility practice in an appropriately sized single pouch (e.g., TYVEK pouch or equivalent), tray, or container labeled as appropriate for the STERRAD hydrogen peroxide gas plasma systems. Do not overlap cables. Load cables into STERRAD system and run equipment in a manner that is consistent with the manufacturer’s instructions.
St. Jude Medical has validated the following cycle parameters: Cycle Parameters
55°C Temperature Cycle
Pre-Conditioning Chamber Pre-Conditioning Time
30-45 min
Relative Humidity
40%-80%
Ethylene Oxide (EtO) Process
The following STERRAD‡ systems and cycles are recommended for reprocessing: STERRAD™ System and Cycle within the United States
STERRAD™ System and Cycle outside United States
STERRAD™ 50
STERRAD™ 50
STERRAD™ 200
STERRAD™ 200 (Short Cycle)
Exposure Time
1 hour
Vacuum Set
1.3 psia
Eto Gas
100% EtO
Eto Concentration
725 mg/L-880 mg/L
STERRAD™ 100S
STERRAD™ 100S (Short Cycle)
12 hours
STERRAD™ NX™ (Standard cycle)
STERRAD™ NX™ (Standard cycle)
STERRAD™ 100NX™ (Standard cycle)
STERRAD™ 100NX™ (Standard cycle)
Aeration Time
8 | Summary of Validated Cleaning and Sterilization Methods for St. Jude Medical™ Brand EP Catheter Cables
STEAM AUTOCLAVE Only applicable to select cables as indicated in the cable listing above in Autoclave column. The cables must be handled in the same manner as other reusable instruments that require packaging or wrapping during sterilization. All cables must be sterilized in packaging that is permeable to steam. Load cables and run equipment in a manner that is consistent with the manufacturer’s instructions. St. Jude Medical has validated the following cycle parameters: Temperature Cycle Cycle Parameters
Low
High
Dynamic/Pre-Vac Temperature
121°C (250°F)
132°C-135°C (270°F-275°F)
Time
40 min
4 min
Gravity Displacement Temperature
121°C (250°F)
132°C-135°C (270°F-275°F)
Time
40 min
10 min
Note: Sterilization temperatures must not exceed 135°C (275°F).
Abbott Medical Australia Pty Ltd | 9
Brief Summary: Prior to using these devices, please review the Instructions for Use for a complete listing of indications, contraindications, warnings, precautions, potential adverse events and directions for use. Abbott Abbott Medical Australia Pty Ltd 299 Lane Cove Road, Macquarie Park NSW 2113 Australia Tel: 1800 839 259 Abbott Medical New Zealand Limited 4 Pacific Rise, Mt Wellington Auckland 1060 New Zealand Tel: 0800 756 269 Abbott.com ™Indicates a trademark of the Abbott group of companies. ‡ Indicates a third party trademark, which is property of its respective owner. © 2023 Abbott. All Rights Reserved. MAT-2302365 v2.0 | Item approved for Australia and New Zealand use only.