Instructions for Use
124 Pages
Preview
Page 1
PressureWireTM Receiver
Instructions for Use
20844 IFU PressureWire Receiver R05.indd 1
2010-10-18 13:51:18
20844 IFU PressureWire Receiver R05.indd 2
2010-10-18 13:51:18
ENGLISH CONTENTS
4
DESCRIPTION (SEE DIAGRAM 1)
4
1. INTENDED USE/INDICATION FOR USE
4
2. CONTRAINDICATIONS
4
3. GENERAL WARNINGS
4
4. GENERAL PRECAUTIONS
5
5. INSTALLATION
5
6. PREPARATION FOR USE
6
7. DIRECTIONS FOR USE
6
8. CLEANING AND MAINTENANCE
8
9. PRESSURE SIGNAL PERFORMANCE
9
10. TECHNICAL SPECIFICATIONS
9
11. COMPLIANCE WITH REGULATORY REQUIREMENTS
10
WARRANTY AND LIMITATIONS
13
SYMBOLS WITH EXPLANATION
14
3
20844 IFU PressureWire Receiver R05.indd 3
2010-10-18 13:51:18
CONTENTS ▪▪
PressureWire Receiver
▪▪
PressureWire Receiver mounting bracket
NOTE: A monitor cable is required to connect PressureWire Receiver to a compatible cathlab recording system. (Not part of contents but part of the delivery according to customer order).
DESCRIPTION (SEE DIAGRAM 1) ▪▪
PressureWire Receiver is a receiver to the PressureWire transmitter and is connected to a cathlab hemodynamic recording system.
▪▪
PressureWire Receiver receives PressureWire signal via radio and converts it to a pressure signal compatible with standard invasive blood pressure transducers.
▪▪
PressureWire Receiver is connected to a recording system with a standard invasive blood pressure connection (ANSI/AAMI BP22-1994, DC excitation).
▪▪
Refer to section Technical Specifications for compatibility with hemodynamic recording systems.
▪▪
PressureWire Receiver user interface consists of three buttons with different light indicators and buzzer sound.
Diagram 1: PressureWire
C
A B C D E F G H
A
B PressureWire Receiver Transmitter (Not a part of this package) PressureWire Not a part of this package) CONNECT button Plus/Minus buttons Plus/Minus light indicators Function light indicators Monitor cable connector
F E
D G
F
H
1. INTENDED USE/INDICATION FOR USE INTENDED USE: PressureWire Receiver is intended for use as an interface between PressureWire and a compatible cathlab recording system. INDICATION FOR USE: PressureWire Receiver together with PressureWire is indicated for use in catheterization and related cardiovascular specialty laboratories for measurement of physiological parameters on patients requiring such information for further treatment.
2. CONTRAINDICATIONS
No identified contraindications.
3. GENERAL WARNINGS
No identified general warnings. 4
20844 IFU PressureWire Receiver R05.indd 4
2010-10-18 13:51:19
4. GENERAL PRECAUTIONS
Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. PressureWire Receiver is only intended for use as an interface between PressureWire and a compatible cathlab recording system. PressureWire Receiver shall only be used with monitor cable approved by St. Jude Medical Systems. Before using PressureWire Receiver refer to instructions for use supplied with PressureWire or warnings, precautions, contraindications and instructions for use. Make sure PressureWire Receiver is connected to intended PressureWire and not to adjacent or previously used PressureWire units. The PressureWire Receiver housing is protected against the effects of a discharge of a cardiac defibrillator. Handle and dispose of PressureWire Receiver in accordance with medical practice and applicable local, state and federal laws and regulations.
5. INSTALLATION
5.1 Open the PressureWire Receiver package and use the enclosed mounting bracket to fasten PressureWire Receiver on or near cathlab table. At delivery the mounting bracket is assembled for pole mounting. If mounting on the table rail is desirable, disassemble the mounting bracke and reassemble according to the picture below.
If PressureWire Receiver is used in a sterile environment it is recommended to cover it in a sterile plastic bag or under a plastic sheet. 5.2 Connect PressureWire Receiver to one available invasive blood pressure channel on the cathlab hemodynamic recording system using the monitor cable supplied by St. Jude Medical Systems. NOTE: Refer to section Technical Specification for compatibility with hemodynamic recording systems. NOTE: Radio range is reduced by objects and walls, keep transmitter and receiver in line of sight wherever possible. CAUTION: PressureWire Receiver is powered by the excitation voltage from the invasive blood pressure channel. Therefore the cathlab hemodynamic recording system must be powered in order for PressureWire Receiver to be activated. On some cathlab systems the pressure channels must also be activated and zeroed before the excitation is available.
5
20844 IFU PressureWire Receiver R05.indd 5
2010-10-18 13:51:20
Refer to instructions for use for the cathlab hemodynamic recording system for correct handling instructions. 5.3 A steady yellow light and a short beep on PressureWire Receiver indicate that it is successfully connected and powered. 5.4 Zero pressure channel on the cathlab hemodynamic recording system.
6. PREPARATION FOR USE
6.1 If necessary, verify balance between PressureWire Receiver and cathlab recording system, see chapter “Balance PressureWire Receiver and cathlab recording system” below. 6.2 Activate the receiver by pressing the CONNECT button. The receiver is now ready to connect to PressureWire transmitter (during 60s), indicated by blinking green light on the receiver. 6.3 Turn on the PressureWire transmitter. PressureWire will now calibrate and connect to the receiver. Refer to Instructions for use for PressureWire Aeris for correct handling before calibration. 6.4 When the receiver is successfully connected to the transmitter the CONNECT button indicators changes to steady green light and a double beep is heard. -- If a connection is not established and the green light on the receiver is still blinking, restart the transmitter by turning it off and then on again. -- If a connection is not established within 60s the receiver returns to standby mode, indicated by a yellow light and a long beep. Reactivate PressureWire Receiver by pressing the CONNECT button, then restart the transmitter by turning it off and then on again. -- Refer to the table below for other light indicators. 6.5 After calibration, confirm that the pressure reading on the cathlab monitor is zero; re-zero the cathlab pressure channel if necessary. PressureWire is now ready for use.
7. DIRECTIONS FOR USE NORMAL OPERATION After successful connection to the PressureWire transmitter, pressure is received by PressureWire Receiver and converted for the cathlab hemodynamic recording system. ADJUST PRESSURE OFFSET (EQUALIZATION) To remove any residual pressure difference between aortic and PressureWire pressure use the plus (+) and minus (–) buttons on the PressureWire Receiver. To increase PressureWire offset, in steps of 1 mmHg, press the plus (+) button. The plus (+) indicator (F) is lit as long as there is a positive pressure offset. To decrease PressureWire offset, in steps of 1 mmHg, press the minus (-) button. The minus (-) indicator (F) is lit as long as there is a negative pressure offset. Alternatively, the equalization function on the cathlab recording system can be used to remove pressure offset. Refer to the Instructions for use for the cathlab recording system. NOTE: Before equalization, the height of the aortic pressure transducer should be aligned to heart level. If the ratio PressureWire Aortic pressure after height alignment is higher than 1.1 or below 0.9 it is recommended to take out PressureWire place it in the packaging coil, flush and then recalibrate according to PressureWire Aeris Instructions for use, before adjusting pressure offset.
6
20844 IFU PressureWire Receiver R05.indd 6
2010-10-18 13:51:20
BALANCE PRESSUREWIRE RECEIVER AND CATHLAB RECORDING SYSTEM It is recommended to daily verify that the pressure output from PressureWire Receiver is in balance with the pressure displayed on the cathlab recording system. To provide reference pressure output levels to the cathlab recording system, enter Refout mode by pressing and holding the CONNECT button >3s. When Refout mode is activated, the plus (+) and minus (-) indicators (F) are lit.
40 20
100
0
200
To increase/decrease pressure reference output level press plus (+) / minus (-) buttons. Pressure output level changes according to 0, 20, 40, 100, 200 mmHg, indicated by the function light indicators, see picture. NOTE: Reference output levels are unaffected by any previous pressure offset. To exit Refout mode shortly press the CONNECT button. For complete information about light indicators see table below.
7
20844 IFU PressureWire Receiver R05.indd 7
2010-10-18 13:51:20
RECEIVER LIGHT INDICATORS Light indicators
Explanation
Yellow steady light
At startup: Unit power on and self test passed. During measurement: Lost connection to PressureWire or PressureWire is connected from transmitter.
Green blinking light
Connect mode – searching for connection to PressureWire Connect mode is automatically deactivated if no connection has been found after 60s, or if the CONNECT button is pressed again.
Green steady light
Established connection to PressureWire receiving pressure data.
Incremental green lights
Reference pressure level output. See chapter “Balance PressureWire Receiver and cathlab recording system”.
Plus indicator steady blue
Positive pressure offset added to PressureWire pressure.
Minus indicator steady blue
Negative pressure offset added to PressureWire pressure.
Green steady light & yellow blinking light
PressureWire battery low. It is recommended to finish case or to exchange PressureWire.
Yellow blinking light
Internal error. Attempt restart by dis- and re-connecting PressureWire Receiver from cathlab recording system.
CHANGE OF PRESSUREWIRE When changing PressureWire be sure to re-activate PressureWire Receiver by pressing the CONNECT button before turning on a new PressureWire transmitterto secure correct connection. Refer to Instructions for use for PressureWire Aeris.
8. CLEANING AND MAINTENANCE
PressureWire Receiver shall only be cleaned with regular cleaning agents. No maintenance other than normal functional control needed. See section “Balance PressureWire Receiver and cathlab recording system”. If PressureWire Receiver performance is outside specification contact a St. Jude Medical Systems representative.
8
20844 IFU PressureWire Receiver R05.indd 8
2010-10-18 13:51:21
9. PRESSURE SIGNAL PERFORMANCE (All values together with PressureWire Aeris) Operating pressure: Zero Thermal Effect: Sensitivity Thermal Effect: Zero drift: Total Accuracy for the combination of PressureWire and PressureWire Receiver:
Frequency response: Delay time:
-30 to +300 mmHg 0.3 mmHg/°C 0.3%/°C <7 mmHg/h ±1 mmHg plus ±1% of reading (over the pressure range – 30 mmHg to 50 mmHg) ±3% of reading (over the range 50 mmHg to▪ 300 mmHg) DC to 25 Hz <10 ms
10. TECHNICAL SPECIFICATIONS Physical Weight: Dimensions: Material:
60 g 100 x 55 x 27mm PC/ABS
Radio Frequency range: Type: Range:
2.4000-2.4835 GHz (ISM-band) Frequency Hopping Spread Spectrum 0-2m
NOTE: Radio range is reduced by objects and walls, keep transmitter and receiver in line of sight wherever possible. WARNING: Changes or modifications not expressly approved by St. Jude Medical Systems could void user´s authority to operate the equipment. Electrical interface compatibility Compatible with invasive blood pressure channels on cathlab hemodynamic recording systems according to: Signal: ANSI/AAMI BP 22-1994 Excitation voltage: 4-8 V DC Excitation impedance: 200 to 500 Ω Signal impedance with tolerance: 56±10 Ω Nominal sensitivity: 5 μV/V/mmHg Environmental conditions Temperature: Air humidity: Ambient pressure:
15-35°C 30-75% 425-850 mmHg
Transport and storage Transportation temperature: Air humidity: Store at room temperature in a dry place.
-25-70°C 10-95%
9
20844 IFU PressureWire Receiver R05.indd 9
2010-10-18 13:51:21
11. COMPLIANCE WITH REGULATORY REQUIREMENTS
Hereby, St Jude Medical Systems AB, declares that this PressureWire Receiver is in compliance with the essential requirements and other relevant provisions of Medical Device Directive (93/42/EEC), Radio and Telecommunications Terminal Directive (1999/5/EC) and Quality System Regulation (US). GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC EMISSIONS The PressureWire system is intended for use in the electromagnetic environment specified below.The customer or the user of PressureWire system should assure that it is used in such an environment. Emission test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The PressureWire system uses RF energy only for its internal function.Therefore, its RFemissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
Harmonic emissions IEC 61000-3-2
Class A
The PressureWire system is suitable for use in all establishments including domestic establishments and those directly connected to the public low-voltage power supply network that supplies building for domestic purposes.
Voltage fluctuations /flicker emissions IEC 61000-3-3
Complies
10
20844 IFU PressureWire Receiver R05.indd 10
2010-10-18 13:51:21
GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY The PressureWire system is intended for use in the electromagnetic environment specified below. The customer or the user of the PressureWire system should assure that it is used in such an environment. Immunity test
IEC 60601 Test level
Compliance level
Electromagnetic environment – guidance
Electrostatic discharge (ESD) ▪ IEC 61000-4-2
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst ▪ IEC 61000-4-4
±2 kV for power supply lines ±1 kV for input/ output lines
N/A 1) 2)
N/A
Surge ▪ IEC 61000-4-5
±1 kV differential mode ±2 kV common mode
N/A 2)
N/A
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
<5% UT (>95% dip in UT) for 0,5 cycle
N/A 1)
N/A
3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
40% UT (60% dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 cycles <5% UT (>95% dip in UT) for 5 sec
Power frequency (50/60 Hz) magnetic filed IEC 61000-4-8
3 A/m
NOTE: UT is the a.c. mains voltage prior to application of the test level. 1) No mains power input 2) Patient connection
11
20844 IFU PressureWire Receiver R05.indd 11
2010-10-18 13:51:21
GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY The PressureWire system is intended for use in the electromagnetic environment specified below. The customer or the user of the PressureWire system should assure that it is used in such an environment. Immunity test
IEC 60601 Test level
Compliance level
Electromagnetic environment – guidance Portable and mobile RF communication equipment should be used no closer to any part of the PressureWire system, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz
N/A 1)
d = 1,2√ P
Radiated RF IEC 61000-4-3
3 V/m 80 MHz to 2,5 GHz
3 V/m
d = 1,2√ P 80 MHz to 800 MHz d = 2,3√ P 800 MHz to 2,5 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site surveya, should be less than the compliance level in each frequency rangeb. Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the PressureWire system is used exceeds the applicable RF compliance level above, the PressureWire system should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the PressureWire system. b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 1) No mains power input
12
20844 IFU PressureWire Receiver R05.indd 12
2010-10-18 13:51:21
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND THE PRESSUREWIRE SYSTEM The PressureWire system is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the PressureWire system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the PressureWire system as recommended below, according to the maximum output powerof the communications equipment. Rated maximum output power of transmitter (W)
Separation distance to frequency of transmitter (m) 150 kHz to 80 MHz d = 1,2√ P
80 MHz to 800 MHz d = 1,2√ P
800 Mhz to 2,5 GHz d = 2,3√ P
0,01
0,12
0,12
0,24
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
STANDARDS PressureWire Receiver is classified as follows in accordance with relevant parts in: IEC 60601-1 ed.02 Amd 1 & Amd 2 IEC 60601-1-4 IEC 60601-1-2 (2001) Amd 1 (2004) EN 300328 v.1.7.1 ANSI/AAMI BP 22-1994 Conforms to UL 60601-1 Certified to CAN/CSA C22.2 No. 601.1 -M90
WARRANTY AND LIMITATIONS
Although PressureWire Receiver, hereafter referred to as “product”, has been manufactured under carefully controlled conditions, St Jude Medical Systems AB, hereafter called St Jude, has no control over the conditions under which the product is used. St Jude, therefore disclaims all warranties, both expressed and implied, with respect to the product, including, but not limited to, any implied warranty of merchantability or fitness for a particular purpose. St Jude shall not be liable to any person or entity for any medical expenses or any direct, incidental or consequential damages caused by any use, defect, failure or malfunction of the product, whether a claim for such damages is based upon warranty, contract, tort or otherwise. No person has any authority to bind St Jude to any representation or warranty with respect to the product. The exclusions and limitations set out above are not intended to, and should not be construed so as to contravene mandatory provisions of applicable law. If any part or term of this Disclaimer of Warranty is held to be illegal, unenforceable or in conflict with applicable law by court of competent jurisdiction, the validity of the remaining portions of this Disclaimer of Warranty shall not be affected, and all rights and obligations shall be construed and enforced as if this Disclaimer of Warranty did not contain the particular part or term held to be invalid. PressureWire is developed and manufactured according to Medical Device Directive 93/42/EEC, Radio and Telecommunication Terminal Equipment Directive 1999/5/EC and Quality System Regulation (US). 13
20844 IFU PressureWire Receiver R05.indd 13
2010-10-18 13:51:21
SYMBOLS WITH EXPLANATION Caution, (Attention consult accompanying documents). See Instructions for Use. Type CF equipment. Contents of the package. Caution: Federal law restricts this device to sale by or on the order of a physician. Catalogue number.
REF
Electronic waste bin. Disposal according WEEE.
Classified
cETLus valid for Canada and US Classified. Conforms to UL 60601-1. Certified to CAN/CSA STD C22.2 NO 601.1 –M90.
Receiver
Receiver.
Receiver mounting bracket
Receiver mounting bracket.
SN
Serial number. Do not use if package is damaged. Manufacturer. Complies with Medical Device Directive 93/42/EEC and Radio and Telecommunication Terminal Equipment Directive 1195/5/EC. Manufacturing Date.
Made in Sweden
Made in Sweden.
Instructions for Use
Instructions for Use.
FCC ID: U4L01080410 IC: 8466A-PW
Compliance ID for FCC, IC and C-tick.
N 17554
LOT
Lot number.
St. Jude Medical Systems AB Palmbladsgatan 10 Box 6350, SE-751 35 Uppsala, Sweden + 46 (0)18 16 10 00 www.sjm.com 14
20844 IFU PressureWire Receiver R05.indd 14
2010-10-18 13:51:21
SVENSKA INNEHÅLL
16
BESKRIVNING (SE DIAGRAM 1)
16
1. INTENDED USE/INDICATION FOR USE
16
2. KONTRAINDIKATIONER
16
3. GENERELLA VARNINGAR
16
4. GENERELLA FÖRSIKTIGHETSÅTGÄRDER
17
5. INSTALLATION
17
6. FÖRBEREDELSER INNAN ANVÄNDNING
18
7. BRUKSANVISNINGAR
18
8. RENGÖRING OCH UNDERHÅLL
20
9. TRYCKSIGNALPRESTANDA
21
10. TEKNISKA SPECIFIKATIONER
21
11. UPPFYLLANDE AV MYNDIGHETSKRAV
22
SYMBOLER MED FÖRKLARINGAR
26
15
20844 IFU PressureWire Receiver R05.indd 15
2010-10-18 13:51:21
DEUTSCH INHALT
28
BESCHREIBUNG (SIEHE ABBILDUNG 1)
28
1. GEBIETE UND INDIKATION FÜR DIE ANWENDUNG
28
2. KONTRAINDIKATIONEN
28
3. ALLGEMEINE WARNHINWEISE
28
4. ALLGEMEINE VORSICHTSMASSNAHMEN
29
5. INSTALLATION
29
6. VORBEREITUNGEN FÜR DEN GEBRAUCH
30
7. GEBRAUCHSANWEISUNG
30
8. REINIGUNG UND WARTUNG
33
9. DRUCKSIGNALPERFORMANCE
33
10. TECHNISCHE SPEZIFIKATIONEN
33
11. ERFÜLLUNG GENEHMIGUNGSBEHÖRDLICHER ANFORDERUNGEN
34
SYMBOLE MIT ERLÄUTERUNGEN
38
27
20844 IFU PressureWire Receiver R05.indd 27
2010-10-18 13:51:22
FRANÇAIS CONTENU
40
DESCRIPTION (VOIR DIAGRAMME 1)
40
1. UTILISATION PREVUE/INDICATIONS D’UTILISATION
40
2. CONTRE-INDICATIONS
40
3. AVERTISSEMENTS D’ORDRE GENERAL
40
4. PRECAUTIONS D’ORDRE GENERAL
41
5. INSTALLATION
41
6. PREPARATION POUR L’UTILISATION
42
7. INDICATIONS D’USAGE
42
8. NETTOYAGE ET MAINTENANCE
45
9. EMISSION DU SIGNAL DE PRESSION
45
10. CARACTERISTIQUES TECHNIQUES
45
11. COMPLIANCE WITH REGULATORY REQUIREMENTS
46
SYMBOLES AVEC EXPLICATIONS
50
39
20844 IFU PressureWire Receiver R05.indd 39
2010-10-18 13:51:23
ESPANÕL CONTENIDO
52
DESCRIPCIÓN (VÉASE EL DIAGRAMA 1)
52
1. USO PREVISTO/INDICADO
52
2. CONTRAINDICACIONES
52
3. ADVERTENCIAS GENERALES
52
4. PRECAUCIONES GENERALES
53
5. INSTALACIÓN
53
6. PREPARACIÓN PARA EL USO
54
7. INSTRUCCIONES DE USO
54
8. LIMPIEZA Y MANTENIMIENTO
56
9. FUNCIONAMIENTO DE LA SEÑAL DE PRESIÓN
57
10. ESPECIFICACIONES TÉCNICAS
57
11. CUMPLIMIENTO DE LOS REQUISITOS DE LAS NORMATIVAS
58
SÍMBOLOS CON EXPLICACIONES
62
51
20844 IFU PressureWire Receiver R05.indd 51
2010-10-18 13:51:25
ITALIANO CONTENUTO
64
DESCRIZIONE (CONSULTARE LO SCHEMA 1)
64
1. USO PREVISTO/INDICAZIONI PER L’USO
64
2. CONTROINDICAZIONI
64
3. AVVERTENZE GENERALI
64
4. PRECAUZIONI GENERALI
65
5. INSTALLAZIONE
65
6. PREPARAZIONE PER L’USO
66
7. ISTRUZIONI PER L’USO
66
8. PULIZIA E MANUTENZIONE
68
9. ESECUZIONE DEL SEGNALE DI PRESSIONE
69
10. SPECIFICHE TECNICHE
69
11. CONFORMITÀ AI REQUISITI NORMATIVI
70
SIMBOLI CON LA RELATIVA SPIEGAZIONE
74
63
20844 IFU PressureWire Receiver R05.indd 63
2010-10-18 13:51:26
NEDERLANDS INHOUD
76
BESCHRIJVING (ZIE AFBEELDING 1)
76
1. BEOOGD GEBRUIK/GEBRUIKSINDICATIES
76
2. CONTRA-INDICATIES
76
3. ALGEMENE WAARSCHUWINGEN
77
4. ALGEMENE VOORZORGSMAATREGELEN
77
5. INSTALLATIE
77
6. VOORBEREIDINGEN VOOR GEBRUIK
78
7. GEBRUIKSAANWIJZING
78
8. REINIGING EN ONDERHOUD
80
9. PRESTATIES VAN HET DRUKSIGNAAL
81
10. TECHNISCHE SPECIFICATIES
81
11. VOLDOEN AAN WETTELIJKE VOORSCHRIFTEN
82
SYMBOLEN EN HUN VERKLARINGEN
86
75
20844 IFU PressureWire Receiver R05.indd 75
2010-10-18 13:51:27